The text of a manual · part 3 of 13
Ultra High Accuracy Laser Displacement Meter LC-2400 Series Instruction Manual, pages 21 to 30
Pages 21 to 30 of 121
The original document is the PDF (805 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Keyence LC-2400A
Page 21
3-1 Controller
Rear Panel
T
Display unit
U
Control unit
V
S b
X
W
POWER ON CONTROL I/O
MONITOR
HEAD
No.
HEAD
No.
OFF a
100~240V AC INTENSITY
RS-232C
DISPLACEMENT
BATTERY
LASER ON
LASER REMOTE
LC–C1 FUSE 3.15A
`
No. S
Name CONTROL I/O port
T
INTENSITY output connector
U
DISPLACEMENT output connector Expansion I/O slot 1CH sensor head connector Ventilating fan Power supply connector F.G. terminal LASER REMOTE connector
V W X Y Z [
\
Laser ON alarm LED
] ^
BATTERY LED RS-232C I/O port a
LC-B3 board slot b
Power switch
* Optionally available
14 CHAPTER 3 Part Names and Functions
]
\
[
Z
Function Inputs signals (program, hold on/off, auto-zero on/off, etc.) from and outputs signals to external equipment. Outputs the laser beam intensity measured with the LC in analog voltage. Outputs the measured displacement value as an analog voltage. Install GP-IB board* or digital I/O board* in this slot. Connect the 1CH sensor head to this connector. Radiates heat from the controller. Connect the power supply cable to this connector. Ground the controller through this terminal. The LASER REMOTE connector is used to control laser emission by an external input signal. When this connector is not used, be sure to connect the attached laser remote grounding plug to this connector. Lights when a laser beam is ready to be and is being emitted from the sensor head. Lights when the internal backup battery is exhausted. Connect a personal computer or other device to this port to externally control the LC. Install the optional LC-B3 board in this slot. The LC-B3 cannot be used with the LC-2400W series. Use this switch to turn the main power supply on or off. (LC-2400W only)
Y
Ref. page p.71 p.74 p.73 p.24, 95 p.22 p.16 p.22 p.23 p.11 p.18, 26 p.15 p.50 p.24
Page 22
3-1 Controller
Laser Remote Grounding Plug
Follow the instructions below to insert this plug into the LASER REMOTE connector. 1. Check the connector’s orientation before insertion. LASER REMOTE connector
Laser remote grounding plug
2. Turn the stopper clockwise to lock it in place. Stopper
Note
1
Laser emission is enabled when pins 1 and 2 of the connector are shorted.
4
2
3
Control Unit Face Control unit
; ; ; ; ; ; ;;;;;; DC
DIGITAL I / O b
No. b
Name 96-pin DIGITAL I/O port c
Earth ground terminal d
Power supply terminals
+15V c
DC
0V
DC
–15V
+5V
0V d
Function The control unit is connected to the display unit via this connector. Digital signals are output through this connector when the control unit is used separately from the display unit. Ground the control unit through this terminal when using the control unit separately from the display unit. Connect the power supply cable to these terminals when using the control unit separately from the display unit.
Ref. page p.74 p.23 p.23
CHAPTER 3 Part Names and Functions
15
Page 23
3-2 Sensor Head
3-2 Sensor Head LC-2420/2430 (Regular-reflective) h i e f g
LC-2440/2450 (Diffuse-reflective)
i ef
No. e f g h i
Name Laser beam shield
Transmitter Receiver Cable connector Laser ON alarm LED
16 CHAPTER 3 Part Names and Functions g h
Function The laser beam shield is used to cover the laser beam transmitter. When operators are working near the sensor head, if there is a risk of looking directly at the laser beam, be sure to cover the transmitter with the shield. Emits laser beams. Receives the reflected laser beam. Connects the sensor head cable to the controller. Lights when a laser beam is ready to be and is being emitted from the sensor head.
Ref. page p.4 p.18 p.18 p.22 p.4, 26
Page 24
CHAPTER 4
OPERATING INSTRUCTIONS 4-1
Instructions for Using Controller and Sensor Head ............. 18
Page 25
4-1 Instructions for Using Controller and Sensor Head
Before operating the LC-2400 series, read the following instructions carefully. If you encounter any problems, please contact us.
4-1 Instructions for Using Controller and Sensor Head After Receiving the LC-2400 Series
• After receiving the LC, check the contents to see if any items were damaged during transportation. • Check that all the items were included. Reference: Contents of the Package (p.ii)
• Before using the LC, be sure to read this instruction manual for details on the correct use of this product.
Daily Maintenance Instructions
• Do not try to disassemble the LC or use a disassembled unit. Do not disassemble the sensor head in particular, since all of the sensor head parts have been factory-calibrated. • Please handle the controller and sensor head with care. The precision optical components in the sensor head are especially susceptible to shock. • The receiver and transmitter lenses are the critical components for accurate measurement. Any flaw or dust on the lenses may result in measurement errors. If dust has accumulated on the transmitter and receiver lenses, clean the lens surface by following the procedures given below. 1. Blow the dust off the receiver and transmitter lenses using a manual blower. 2. If dust persists, wipe the lens surface gently using a cotton swab dipped in alcohol.
18 CHAPTER 4 Operating Instructions
Page 26
4-1 Instructions for Using Controller and Sensor Head
Replacing Battery
WARNING
CAUTION
The LC controller has an internal backup battery for storing parameter settings. The expected battery life is 1 to 2.5 years. When the battery is exhausted, the BATTERY LED lights. When this LED is lit, replace the battery by following the procedure given below. Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent type recommended by the manufacturer. Dispose of used batteries according to the manufacturer's instructions.
Turn the power OFF and unplug the power cable from the wall outlet.
1. Remove the two screws from the back of the display unit and remove the control unit. Upper cover
Expansion slot
Control unit Screws
2. Turn the control unit upside down, remove the six cover screws from the cover and then remove the cover. Locate the battery holder in the control unit. Screws Cover
Battery holder
Screws
3. Remove the battery from the battery holder. Positive side Flat lithium battery: CR2032
Internal backup battery Battery holder
4. Install the new battery in the battery holder with the positive side facing up. When installing a new battery, be sure that the polarity of the battery is correct.
CHAPTER 4 Operating Instructions
19
Page 27
4-1 Instructions for Using Controller and Sensor Head
5. Put the cover back on the control unit and secure it with the six cover screws. 6. Reinsert the control unit into the display unit with the top of the control unit facing up, and secure the control unit with the two retaining screws.
Instructions for Sensor Setup CAUTION
Turn the power OFF when connecting or disconnecting the sensor head connector, power supply cables, or any of the optional boards. Otherwise, the laser diode or other electronic components may become degraded or damaged. • Place the LC in a location that meets the following requirements: Temperature: 0 to 40°C Humidity: 35 to 85% RH (No condensation) Placing the LC in a location that does not meet these requirements may cause the controller and sensor head to malfunction. Do not operate the LC in a location where extreme temperature variations are expected. • Do not operate the LC where volatile flammable objects or corrosive gases are present. • Do not expose the LC to extremely high humidity or water splashes. • When using the control unit separately from the display unit, be sure to provide a protective cover for the control unit during operation. • To prevent noise, keep the connection cables separate from the power cables. If this is not possible, shield or ground the connection cables. • Insulate the sensor head if noise is transferred through the sensor head. • Do not block the ventilation port or the ventilation fan of the controller. If they are blocked, the controller may overheat, resulting in a malfunction. Note also that a ventilation fan is not provided with the control unit. When using the control unit separately from the display unit, make sure that the room is sufficiently ventilated. Ventilation port
Ventilation fan
Hints on Correct Use
• Keep the ambient temperature at a constant level during measurement. • If the sensor head is operating in a location where there is thick fog or hot air, measurement errors may result. • Be sure to leave the power on for at least 60 minutes before starting measurements.
20 CHAPTER 4 Operating Instructions
Page 28
CHAPTER 5
CONNECTIONS 5-1
Connecting Controller to Sensor Head .......................... 22
5-2
Connecting Power Supply Cable ................................... 22
5-3
Grounding Controller ..................................................... 23
5-4
Installing LC-B3 Board (2CH Sensor Head Board) ....... 24
5-5
Installing GP-IB or DIGITAL I/O Board .......................... 24
5-6
Adjusting Sensor Head Position .................................... 25
Page 29
5-1 Connecting Controller to Sensor Head 5-2 Connecting Power Supply Cable
5-1 Connecting Controller to Sensor Head The 2CH sensor head cannot be used with the LC-2400W series. CAUTION
Turn the power OFF when connecting or disconnecting the sensor head connector, power supply cables, or any of the optional boards. Otherwise, the laser diode or other electronic components may become degraded or damaged. Connect the controller to the sensor head using the supplied connecting cable. At this time, make sure that the control unit and sensor head have the same serial number.
Options
Connecting cable Sensor heads 2CH sensor head
1CH sensor head
If you purchase the LC-B3 and sensor head later, you can connect them to the controller without adjustment because the sensor head and the LC-B3 have been factory-calibrated as a pair.
5-2 Connecting Power Supply Cable Connecting the Power Supply Cable When Using the Display Unit
CAUTION
Turn OFF the LC controller before starting the following procedure. Plug the power supply cable into the power supply socket located on the rear of the display unit, and then into a wall outlet. Power must be 100 to 240 VAC (50/60 Hz).
22 CHAPTER 5 Connections
Page 30
5-2 Connecting Power Supply Cable 5-3 Grounding Controller
Connecting the Power Supply Unit to a Separate Control Unit
CAUTION
Cannot be used with the LC-2400W series. Turn OFF the power supply unit before starting the following procedure.
;;;;; ;;;;;
Connect the power supply unit to the power supply terminals located on the front of the control unit.
DC
+15V
DC
0V
–15V +5V
DC
0V
Note that the control unit uses the following types of power: • ±15 VDC (±3%), 0.5A • ±5 VDC (±5%), 3A, ripple (p-p): 100 mV max.
CAUTION
Be sure to supply ±15 VDC and +5 VDC at the same time. Supplying only +5 VDC may cause a malfunction.
5-3 Grounding Controller The controller can be grounded either by connecting the earth ground terminal located on the rear of the display unit to a grounding electrode or by connecting the ground wire of the power supply cable to the ground screw of a wall outlet. To prevent any risk of malfunction, ground the LC both ways as shown below.
CAUTION
Be sure to ground the controller separately from other devices.
Grounding the Controller When Using the Display Unit
Earth-ground the controller through the earth ground terminal located on the rear of the display unit with the ground wire of the power supply cable.
Grounding the Control Unit
Cannot be used with the LC-2400W series. Earth-ground the control unit through the earth ground terminal located on the front of this unit.
;;;;; DC
+15V
DC
0V
–15V +5V
DC
0V
CHAPTER 5 Connections
23