The text of a manual · part 2 of 13

Ultra High Accuracy Laser Displacement Meter LC-2400 Series Instruction Manual, pages 11 to 20

Pages 11 to 20 of 121

The original document is the PDF (805 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Keyence LC-2400A

Page 11

1-5 Safety Features Provided with the LC-2400 Series

• Protective enclosure We recommend that you install a protective enclosure around the sensor head to prevent any person from getting near the sensor head during operation. • Protective goggles We recommend that you wear protective goggles when using the LC2400 series. Note

The intense light from this laser can be harmful to the eyes during prolonged viewing. Normal reflex blinking is usually enough to prevent any eye damage. However, it is best to wear laser protective glasses whenever working around a sensor head.

1-5 Safety Features Provided with the LC-2400 Series The LC-2400 series comes with the following safety features: ■ Laser ON alarm LED Both the sensor head and the controller panel have a visible LED that lights when laser is ready to be and is being emitted. Note

LEDs can be checked to see if they are lit even when you are wearing laser protective glasses. Reference: Laser ON alarm LED (p.12, 14, 16)

■ Delay of laser beam emission To prevent an operator from being exposed to the laser beam, the laser beam is emitted three seconds after the laser ON alarm LED lights. ■ Laser emission remote control input connector The laser emission control connector is located on the rear panel of the control unit. The laser can be turned on or off by a remote control signal through this connector. Reference: Laser Remote Grounding Connector (p.15)

■ Key-operated power switch The controller power switch can be locked using the attached key. When the LC-2400 controller is OFF, the key can be removed. ■ Laser beam shield The sensor head transmitter comes with a laser beam shield. Precautions for parameter setup and measurements If there is a danger of an operator looking into the laser beam when working in front of the sensor head, close the laser beam shield as shown below before starting operations. Close or open the laser beam shield as shown below: • Regular-reflective sensor head (LC-2420/LC-2430)

Open

Close

• Diffuse-reflective sensor head (LC-2440/LC-2450)

Open Close

4

CHAPTER 1 Laser Safety Precautions

Page 12

CHAPTER 2

SYSTEM CONFIGURATION 2-1

Basic System Configuration ............................................ 6

2-2

Expansion System Configuration .................................... 7

2-3

Sensor Head Types and Functions ................................. 8

2-4

Quick Reference Table .................................................... 9

Page 13

2-1 Basic System Configuration

The LC-2400 series is equipped with I/O ports for connecting to external equipment and a slot for installing an optional expansion board. By connecting personal computers or other devices to the LC-2400 series, measurement data obtained with the LC-2400 series can be used for various applications.

2-1 Basic System Configuration The LC controller consists of a display unit and control unit. If required, the control unit can be used separately from the display unit. There are also two expansion slots; one for the optional LC-B3 board which, when installed, allows the use of a second sensor head and a second slot for either a DIGITAL I/O board or GP-IB board. However, the LC-2400W series does not incorporate these expansion slots.

Display unit (model: LC-D1A/LC-D1W)

Control unit (model: LC-C1A/LC-C1W/LC-C2)

Expansion slot for LC-B3 2CH sensor head board* 1., *2.

Expansion slot for DIGITAL I/O board* 1., *2. (model: LC-B1) or GP-IB board*(model: LC-B2)

1CH sensor head port: for LC-2420/LC-2430/LC-2440/LC-2450

*1. Optionally available from KEYENCE *2. Cannot be used with the LC-2400W series.

6

CHAPTER 2 System Configuration

Page 14

2-2 Expansion System Configuration

2-2 Expansion System Configuration For flexible expansion, a variety of external equipment can be connected to the LC-2400 series, as shown below.

With the optionally available 2CH board *1., *2. (model: LC-B3), you can connect the second sensor head (model: LC-2420 /-2430/-2440/-2450) to the controller.

Connect a PLC *1. , photoelectric sensor *1., or proximity sensor *1.. This port is used to switch memory banks, input auto-zero and hold signals, and output alarm signals.

Connect the first sensor head (model: LC-2420 /-2430/-2440/-2450).

2ch sensor head port

1ch sensor head port Control I/O port With a commercially available printer connected, you can record measurement results.

RS-232C port

Expansion slot

Analog voltage output

With the DIGITAL I/O board *1., *2. (model: LC-B1) or GP-IBboard *1., *2. (model: LC-B2), you can send data to and receive it from a personal computer.

With an oscilloscope or pen recorder connected, you can monitor changes in light intensity and target displacement through analog voltage signals.

You can send data to and receive it from a personal computer through one of the interface (RS-232C, DIGITAL I/O *1., GP-IB *1. ). *1. Optionally available from KEYENCE *2. Cannot be used with the LC-2400W series.

CHAPTER 2 System Configuration

7

Page 15

2-3 Sensor Head Types and Functions

2-3 Sensor Head Types and Functions The LC-2400 Series gives you a choice of 4 sensor head models. LC-2420/LC-2430: Regular-reflective sensor heads Regular-reflective sensor heads are used to measure targets having a highly reflective or mirror-like surface.

LC-2440/LC-2450: Diffuse-reflective sensor heads Diffuse-reflective sensor heads are used to measure targets having a lowreflective or opaque surface.

8

CHAPTER 2 System Configuration

Page 16

2-4 Quick Reference Table

2-4 Quick Reference Table Sensor setup, cable connections, and various parameters must be set before starting measurement. However, some of the parameters are not necessary, depending on system configuration and the type of measurements being done. The following table shows the procedures to follow, in sequential order, before starting measurement. Find the items pertaining to your setup in the table to determine which sections of the manual are required reading. Type of measurement Trial measurement Measuring height Measuring eccentricity Measuring thickness Measuring difference in height Section No. 5-1 Connecting Controller to Sensor Head 5-2 Connecting Power Supply Cable 5-3 Grounding Controller 5-4 Installing LC-B3 Board (2CH Sensor Head Board) 5-5 Installing GP-IB or DIGITAL I/O Board 5-6 Adjusting Sensor Head Position 6-1 Basic Operation of Controller 6-2 Selecting Measurement Mode 6-3 Calibrating Sensitivity (Gain Selection) 6-4 Selecting the Display/Output Mode 6-5 Correcting Measurement Error (Calibration) 6-6 Offset Values 6-7 Reducing Variation in Measurement Data (Setting Number of Averaging Measurements) 6-8 Reducing the Effect of Target Surface Irregularities (Setting Low-pass Filter Value) 6-9 Setting Range for Light Intensity (INT. LIMIT) 6-10 Setting Upper/Lower Limits for Measurements (Setting Tolerance) 6-12 Storing and Loading Settings (Programming) 7-1 Measuring with One Sensor Head 7-2 Measuring with Two Sensor Heads

Page No. p.22 p.22 p.23 p.24 p.24 p.25 p.28 p.30 p.30 p.31 p.32 p.35

R R R R

R R R R

R R R

R R R

R R R

R R R R R R O

R R R R R R O

R R R R

R R R R

R

R O

R O p.36

R

R

R

R p.36

O

O

O

O p.37

O

O

O

O p.39

O

O

O

O p.41 p.44 p.45

R

R

R R

R R

R

R

R

R: Required reading O: Optional reading

CHAPTER 2 System Configuration

9

Page 17

10 CHAPTER 2 System Configuration

Page 18 read from a scan by OCR; it may hold reading errors

CHAPTER 3

Part Names AND FUNCTIONS io oy a O10) 01116) |<] ee Ko PM =) 0 110) Ml (<1- (0

Page 19

3-1 Controller

3-1 Controller Front Panel

Display unit Display

Operation keys

LC-2400A OFF

LC-2400A

LASER DISPLACEMENT METER mm

NORMAL

PEAK

P–P

ON

BOTTOM

µm AUTO

1

2

3

4

GAIN

HIGH LIMIT OFFSET

CAL

GO LOW

1

2

3

INT. ANALOG BAUD LIMIT RATIO RATE

4

5

POWER GAIN

1CH

PRM

AVERAGE

2CH

ADD

SUB

FILTER

INTENSITY

ZERO

2–HEAD

DIGIT

PRM

LC-2400W

PROGRAM

LASER ON

LOCK

LOAD

STAND-BY

HOLD

OFF

SAVE ON

ON

PROGRAM

POWER

POWER

Reference: Refer to P.44.

Key-operated power switch

Display

1 mm µm 9

AUTO

1

2

3

4

GAIN

HIGH

8

LIMIT OFFSET

CAL

GO

INT. ANALOG BAUD LIMIT RATIO RATE

2 DIGIT

PRM

3

7 LOW

1

2

3

4

5

PROGRAM

4 LASER ON

6

No. 1

Name Display panel

2

GAIN LEDs

3

PARAMETER LEDs

4

PROGRAM LEDs

5 6

PANEL LOCK LED Laser ON alarm LED

7

LOW LED

8

GO LED

9

HIGH LED

LOCK

5

Function Displays measured values, preset values and error messages. Indicates the receiver sensitivity. One of the four LEDs corresponding to the current sensitivity will be lit. When the receiver sensitivity is set to AUTO, one of the four LEDs and the AUTO LED will be lit. Lights when setting parameters such as upper/lower tolerance limits, offset, calibration, intensity limit, analog ratio, baud rate and display digit. The LED corresponding to the parameter currently being changed lights. The LED corresponding to the selected program number will be lit. The LEDs blink during program selection. Lights when the operation keys are disabled. Lights when the laser beam is ready to be and is being emitted from the sensor head. Lights when a measured value falls below the preset lower limit in each measurement mode. Lights when a measured value is within the preset tolerance range of that measurement mode. Lights when a measured value exceeds the preset upper limit of that measurement mode.

12 CHAPTER 3 Part Names and Functions

Ref. page p.82 p.30 p.28 p.41 p.55 p.4., 26 p.39 p.39 p.39

Page 20

3-1 Controller

Operation Keys p.

0

A

NORMAL

P–P

B

C

D

PEAK

BOTTOM

E

R Q

GAIN

P O

1CH

2CH

ADD

SUB

G

2–HEAD

PRM

AVERAGE

F

FILTER

INTENSITY

ZERO

H I

N

LOAD

M

SAVE

HOLD

J

K

PROGRAM

L

No. 0 A B

Name NORMAL key P-P key 2CH key

C D E F

PEAK key BOTTOM key ADD key SUB key

G H I J K L

FILTER key ZERO key INTENSITY key HOLD key ENTER key. Numeral change keys

M N O P

SAVE key LOAD key PRM key AVERAGE key

Q R

GAIN key 1CH key

Function Press this key to measure targets in the NORMAL mode. Press this key to measure targets in the Peak-to-Peak mode. Press this key to measure targets with the second sensor head (2CH slot) when two sensor heads are connected. Press this key to measure target in the PEAK mode. Press this key to measure target in the BOTTOM mode. Press this key to measure thickness using two sensor heads. Press this key to measure height deviation using two sensor heads. Press this key to set a (Low-pass) FILTER value. Press this key to reset a measured value to "0". Press this key to display received laser beam intensity. Press this key to hold a measured value. Press this key to enter the value you set. Press this key to change values. Press or to shift digits and press or to increase or decrease values as well as change signs. Press this key to save a measured value. Press this key to load stored parameter settings. Press this key to change parameter settings. Press this key to set the number of averaging measurements. Press this key to change receiver sensitivity. Press this key to use the first sensor head (1CH port) when two sensor heads are connected.

Ref. page p.30 p.30 p.31, 45 p.30 p.30 p.31, 46 p.31, 47 p.36 p.29 p.29 p.29 p.28 p.28 p.41 p.41 p.28 p.36 p.30 p.31,44

CHAPTER 3 Part Names and Functions

13