Page 121

Configuring the 9560

6

SELECT CODES NO

3. Press Space. The 9560 displays this prompt: SELECT CODES YES

4. Press Enter to begin changing bar code settings. The 9560 displays this prompt: CODE 39 NO CK DIGIT

5. Press Enter until this prompt is displayed: CODE 93

ENABLE

6. Press Space. The prompt changes to: CODE 93

DISABLE

7. Press Enter to set CODE 93 to disabled. 8. Scan this label: Exit Configuration and Save

*$-* *$-*

The 9560 exits Prompting Configuration mode and saves your changes.

6-15

Page 122

9560 Transaction Manager User’s Manual

Standard Display Prompts The following table lists the configuration prompts and options for the standard 2x40 character display. The options are the same for each display, but the prompts are different. The prompts are listed in the order they appear.

Configuration Prompts for 2x40 Standard Display Prompt

Options

<?,?> FOR HELP <CR, FWD> START

Type ? to view help prompts, or press Enter to configure.

SELECT CODES NO

NO YES

CODE 39 NO CHK DIGIT

NO CHK DIGIT WITH CHK DIGIT HIBC DISABLED

FULL ASCII C39 DISABLED

DISABLED ENABLED

I 2/5 VARIABLE

VARIABLE CASECODE FIXED LENGTH DISABLED

UPC A / EAN-13 ENABLED

ENABLED UPC A ONLY DISABLED

UPC VERSION E ENABLED

ENABLED DISABLED

EAN-8 ENABLED

ENABLED DISABLED

SUPPLEMENTALS ENABLED

ENABLED DISABLED

CODABAR STANDARD

STANDARD CONCATENATED DISABLED ABC

START / STOP ABCD S/S

ABCD S/S DC1-DC4 S/S DISABLE S/S

CODE 11 1 CHECK DIGIT

1 CHECK DIGIT 2 CHECK DIGIT DISABLE

CODE 93 ENABLED

ENABLED DISABLED

6-16

Comments

Selecting No skips the following options and displays the OP PARAMETERS prompt.

Fixed length requires an even-numbered length.

Page 123

Configuring the 9560

6

Configuration Prompts for 2x40 Standard Display (continued) CODE 128 STANDARD

STANDARD UCC-128 DISABLED

AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA

OP PARAMETERS NO

NO YES

Selecting No skips the following options and displays the SELECT PROTOCOL prompt.

BEEP VOLUME X

0 THROUGH 7

Press Space to select a number from 0 through 7 for X.

DISPLAY MODE BUFFERED

BUFFERED TRANSPARENT

CHARACTER SET US-ASCII

US-ASCII FRENCH GERMAN UNITED KINGDOM DANISH SWEDISH ITALIAN SPANISH KATAKANA

CRRM DISABLED

DISABLED ENABLED

PREAMBLE A REQD NO

NO YES

SCANNER TIMEOUT 0 SECONDS

If you choose Katakana, you also specify Kana or Ro-maji entry style and select a function key to switch between US-ACII and Katakana.

Type a number from 0 through 60 msec.

SCANNER MODE 1 SHOT

1 SHOT AUTOMATIC

TRIGGER MODE LEVEL

LEVEL EDGE

APPEND TIME NO

NO YES

TIME IN SECONDS NO

NO YES

RESUME IRL PROG NO

NO YES

AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA

SELECT PROTOCOL NO (PT. TO PT.)

NO (PT. TO PT.) NO (USER DEFINED) NO (POLLING D) NO (MULTI-DROP) NO (U.D.M-DROP)

This prompt displays the currently selected protocol. Press Space to view the protocol options.

SELECT PROTOCOL USER DEFINED

USER DEFINED POINT TO POINT POLLING MODE D MULTI-DROP USER D. M-DROP

The following prompts and options are displayed if needed for the protocol you choose here.

6-17

Page 124

9560 Transaction Manager User’s Manual

Configuration Prompts for 2x40 Standard Display (continued) INTERFACE RS232

RS232 RS422/485

TERMINAL MODE NON BUFFERED FD

NON BUFFERED FD BUFFERED TRANSPARENT NON BUFFERED HD BLOCK

XON = NUL

NUL ANY ASCII CHARACTER

XOFF = NUL

NUL ANY ASCII CHARACTER

BAUD RATE 9600

110 300 600 1200 2400 4800 9600 19200

PARITY EVEN

EVEN ODD DISABLED

DATA BITS 7

7 8

STOP BITS 1

1 2

TIMEOUT DELAY 10 SEC

5 MSEC 100 MSEC 500 MSEC 2 SEC 10 SEC 20 SEC 40 SEC 60 SEC

TURNAROUND DELAY 0 MSEC

0 MSEC 1 MSEC 2 MSEC 5 MSEC 10 MSEC 20 MSEC 50 MSEC 100 MSEC

6-18

Page 125

Configuring the 9560

6

Configuration Prompts for 2x40 Standard Display (continued) INTERCHAR DELAY 0 MSEC

0 MSEC 1 MSEC 2 MSEC 5 MSEC 10 MSEC 20 MSEC 50 MSEC 100 MSEC

LRC DISABLED

DISABLED ENABLED

POL = NUL

NUL ANY ASCII CHARACTER

SEL = NUL

NUL ANY ASCII CHARACTER

RES = NUL

NUL ANY ASCII CHARACTER

REQ = NUL

NUL ANY ASCII CHARACTER

AFF = NUL

NUL ANY ASCII CHARACTER

NEG = NUL

NUL ANY ASCII CHARACTER

SOM = NUL

NUL ANY ASCII CHARACTER

#1 RX EOM = CR

CR ANY ASCII CHARACTER

#2 RX EOM LF

LF ANY ASCII CHARACTER

#1 TX EOM = CR

CR ANY ASCII CHARACTER

#2 TX EOM LF

LF ANY ASCII CHARACTER

#1 EOR = NUL

NUL ANY ASCII CHARACTER

#2 EOR = NUL

NUL ANY ASCII CHARACTER

RECORDS/BLOCK 0

Type a number from 0 through 99.

EOF = NUL

NUL ANY ASCII CHARACTER

SOP = SI

SI ANY ASCII CHARACTER

6-19

Page 126

9560 Transaction Manager User’s Manual

Configuration Prompts for 2x40 Standard Display (continued) EOP = SYN

SYN ANY ASCII CHARACTER

RUN = DC2

DC2 ANY ASCII CHARACTER

END = SO

SO ANY ASCII CHARACTER

PSS = CR

CR ANY ASCII CHARACTER

PAK = RS

RS ANY ASCII CHARACTER

BAK = BEL

BEL ANY ASCII CHARACTER

6-20

Page 127

Configuring the 9560

6

Large-Type Display Prompts This table lists the configuration prompts and options for the large-type 1x20 character display. The options are the same for the standard and large-type displays, but the prompts are shorter for the large-type display. The prompts are listed in the order they appear. For information on using Prompting Configuration, see “Entering Prompting Configuration Mode” or “Moving Around in Prompting Configuration Mode” earlier in this chapter.

Configuration Prompts for 1x20 Display Prompt

Options

<?>HELP <CR> START SELECT CODES

NO

Type ? to view help prompts, or press Enter to configure. NO YES

CODE 39 NO CK DIGIT

NO CK DIGIT CK DIGIT HIBC DISABLE

FL ASCI C39 DISABLE

DISABLE ENABLE

I 2/5

VARIABLE CASECODE FIXED LENGTH DISABLE

VARIABLE

UPCA/EAN13

ENABLE

ENABLE UPC A ONLY DISABLE

UPC VER E

ENABLE

ENABLE DISABLE

EAN-8

ENABLE

ENABLE DISABLE

SUPPLEMENTLS

ENABLE

ENABLE DISABLE

CODABAR

STANDARD

START / STOP

ABCD

Comments

Selecting No skips the following options and displays the OP PARAMETERS prompt.

Fixed length requires an even-numbered length.

STANDARD CONCATENATED DISABLE ABC ABCD DC1-DC4 DISABLE

6-21

Page 128

9560 Transaction Manager User’s Manual

Configuration Prompts for 1x20 Display (continued) CODE 11

1 CK DIGIT

1 CK DIGIT 2 CK DIGIT DISABLE

CODE 93

ENABLE

ENABLE DISABLE

CODE 128

STANDARD

STANDARD UCC-128 DISABLE

AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA

OP PARAMETERS

NO

NO YES

Selecting No skips the following options and displays the PROTOCOL prompt.

X

0 THROUGH 7

Press Space to select a number from 0 through 7 for X.

DISPLAY

BUFFERED

BUFFERED TRANSPARENT

CHAR SET

US-ASCII

US-ASCII FRENCH GERMAN UNITED KINGDOM DANISH SWEDISH ITALIAN SPANISH KATAKANA

CRRM

DISABLE

DISABLE ENABLE

BEEP VOLUME

PREAMBLE A REQD

NO

If you choose Katakana, you also specify Kana or Ro-maji entry style and select a function key to switch between US-ACII and Katakana.

NO YES

SCAN TIMEOUT XX SEC

Type a number from 0 through 60 msec.

SCANNER MODE 1 SHOT

1 SHOT AUTOMATIC

TRIGGER MODE

LEVEL

LEVEL EDGE

APPEND TIME

NO

NO YES

TIME IN SECONDS

NO

NO YES

RESUME IRL PROG

NO

NO YES

AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA

PROTOCOL (PT TO PT)

6-22

(PT TO PT) (USER DEF) (POLLING D) (MULTI-DROP) (U.D.MDROP)

This prompt displays the currently selected protocol. Press Space to view the protocol options.

Page 129

Configuring the 9560

6

Configuration Prompts for 1x20 Display (continued) PROTOCOL

USER DEF

USER DEF PT TO PT POLLING D MULTI-DROP USER MDROP

INTERFACE

RS232

RS232 RS422/485

TERM NON BUFFER FD

NON BUFFER FD BUFFERED TRANSPARENT NON BUFFER HD BLOCK

XON = NUL

NUL ANY ASCII CHARACTER

XOFF = NUL

NUL ANY ASCII CHARACTER

BAUD RATE

9600

110 300 600 1200 2400 4800 9600 19200

PARITY

EVEN

EVEN ODD DISABLE

DATA BITS

7

7 8

STOP BITS

1

1 2

TIMEOUT DLY

10 SEC

5 MSEC 100 MSEC 500 MSEC 2 SEC 10 SEC 20 SEC 40 SEC 60 SEC

TURNARND DLY

0 MS

0 MSEC 1 MSEC 2 MSEC 5 MSEC 10 MSEC 20 MSEC 50 MSEC 100 MSEC

The following options are displayed if needed for the protocol you choose here.

6-23

Page 130

9560 Transaction Manager User’s Manual

Configuration Prompts for 1x20 Display (continued) INTRCHAR DLY

LRC

0 MS

0 MSEC 1 MSEC 2 MSEC 5 MSEC 10 MSEC 20 MSEC 50 MSEC 100 MSEC

DISABLE

DISABLE ENABLE

POL = NUL

NUL ANY ASCII CHARACTER

SEL = NUL

NUL ANY ASCII CHARACTER

RES = NUL

NUL ANY ASCII CHARACTER

REQ = NUL

NUL ANY ASCII CHARACTER

AFF = NUL

NUL ANY ASCII CHARACTER

NEG = NUL

NUL ANY ASCII CHARACTER

SOM = NUL

NUL ANY ASCII CHARACTER

#1 RX EOM = CR

CR ANY ASCII CHARACTER

#2 RX EOM LF

LF ANY ASCII CHARACTER

#1 TX EOM = CR

CR ANY ASCII CHARACTER

#2 TX EOM LF

LF ANY ASCII CHARACTER

#1 EOR = NUL

NUL ANY ASCII CHARACTER

#2 EOR = NUL

NUL ANY ASCII CHARACTER

RECORDS/BLOCK =

0

Type a number from 0 through 99.

EOF = NUL

NUL ANY ASCII CHARACTER

SOP = SI

SI ANY ASCII CHARACTER

6-24