Page 51

Note: If you set the Test Criteria Passing parameter to an ISO/IEC grade and you set the #Scans/Symbol parameter to 1, the scan grade is the same as the ISO/IEC symbol grade. •

The first line of the symbol grade display provides the symbol or application name, the same as the first line of the scan grade display. • The second line of this display indicates this data is a final symbol grade, that is, it is an average of the scan grades you have obtained. • The third line provides a variety of information. The first of the three terms refers to the current average grade the symbol received. The subsequent terms, for example 06/660, refer to the wand aperture and the wavelength. These values are used to determine the symbol grade. Refer to "If the unit was set to use the database feature before being turned off, the bottom line shows the term DATABASE." on page 1-9 for additional information. • The last line lets you check your status in relation to the symbol grade. The first value in this line tells you how many scan grades were used to calculate the displayed symbol grade. The second value tells you the number of scan grades required to calculate the final symbol grade. • The large letter or number on the right of the display is the symbol grade. The verifier can display the symbol grade either as letter (A, B, C, D, F) or as a number (4, 3, 2, 1, 0), depending on how you set the Test Criteria Grades parameter. 5. To obtain the final symbol grade, you must satisfy the #Scans/Symbol setting. Repeat steps 2 through 4, scanning the same bar symbol at different paths within the symbol. •

When you satisfy the #Scans/Symbol setting, the final symbol grade display appears. To review the scan grade display, press ↓. Press ↑ to return to the final symbol grade display. The types of data provided in the final symbol grade display are the same as those in the symbol grade display.

♦ To review data from the last scan: 1. If the Quick Check Verifier is turned off, turn it on. The opening display appears. 2. Press ←. The results from the last scan appear.

LED Display The light-emitting diodes (LEDs) can indicate either bar growth and shrinkage based on traditional test criteria or a pass/fail grade result based on ISO/IEC grade criteria. You set the type of test criteria. Refer to "Changing the Test Criteria" on page 3-8 for additional information.

Quick Check® 600/800 Series User’s Guide

4-7

Page 52

For your convenience, the display option feature called “LEDs” allows you to select ISO/IEC test criteria yet have the LEDs display bar growth rather than ISO/ IEC pass/fail grade. Refer to "Changing the Output Options" on page 3-12 for additional information. If you set the verifier to use traditional pass/fail test criteria, the color of each LEDs is always the same (first is red, second is yellow, third is green, fourth is yellow, fifth is red). After a scan, one or two of the LEDs flash to indicate the amount of growth or shrinkage. The following table correlates the LED color and the percent tolerance used.

LED lit

% of bar growth tolerance used

LCD bar growth indication

Green

0% to25%

OK!

Green and yellow

25% to 50%

OK!

Yellow

50% to 75%

OK-,OK+

Yellow and red

75% to 100%

OK-, OK+

Red

>100%

LO!, HI!

For example, if you scan a UPC symbol at 100% magnification and the yellow LED on the left center flashes, it indicates that the tested symbol has used approximately 2 mil (2.0 to 3.0 mil [0.002” to 0.003”]) of the total tolerance (± 4 mil [0.004”]) and that the symbol’s bars are too narrow. If you set the verifier to use ISO/IEC methods to calculate the grade, the color of the LEDs will vary. After a scan, one of the LEDs flashes to indicate the scan grade (ISO/IEC Grade A, B, C or D). For example, if you selected ISO/IEC Grade C as passing and obtained a scan grade of A, the A LED flashes to indicate the scan grade the symbol received. In this case, the A, B and C LEDs will be green to indicate that all three grades are acceptable.

Audio Sounds The audio signal sounds to indicate the bar code symbol’s status to a battery of tests: • One long beep followed by one short beep indicates a passing grade. • One long beep followed by three short beeps indicates a failing grade. You can easily and quickly perform pass/fail testing by scanning symbols and listening for a set of triple beeps to indicate failure and then review the display for specific information.

4-8

Quick Check® 600/800 Series User’s Guide

Page 53

Reviewing the Dimensional Parameters The following table lists the dimensional parameters and summarizes their results. Descriptions of the parameters follow the tables. Parameter

Trad’l

Average Bar Error (Avg Bar Err) Wide-to-narrow ratio (Wide/Nar Ratio)

ISO/IEC

√ √ √

Decodability*

* The method you select (Avg Decode or ISO/IECISO/IEC) determines the scan and symbol grades.

Parameter

Result

Avg Bar Err

Compared against traditional print tolerance criteria set by symbology or application OK if within criteria Fail if outside criteria

Wide/Narrow Ratio

Compared against generic or user set limits for the symbology or application OK if within limits Fail if outside limits

Decodability

A if > .62 B if > .50 C if > .37 D if > .25 F if < .25

Errors in Dimensional Parameters If, after a scan, the abbreviation for dimensional appears as >DIM<, an error exists in one or more of the dimensional parameters. If there is an error, use the arrow buttons to review each parameter to identify the problem. If the abbreviation for dimensional appears as Dim, the symbol has no dimensional errors.

Average Bar The average bar print error (AvgBar) value is the amount of average bar growth or loss in the measured symbol relative to its “X” dimension. This value is compared against traditional print tolerance criteria and rated. If the AvgBar is within these limits, it is rated OK! or pass. If the AvgBar is outside these limits, this test fails.

Quick Check® 600/800 Series User’s Guide

4-9

Page 54

W/N Ratio The wide-to-narrow ratio (W/N) applies to symbologies with two element widths including Code 39, Interleaved 2 of 5 and USS-Codabar. W/N is not meaningful for the other symbologies and is not displayed. When measured, the W/N value is compared against generic or user-specified limits appropriate for the symbology. If the W/N Ratio is within these limits, it is rated OK! or pass. If the W/N Ratio is outside these limits (HI! or LO!) this test fails.

Decodability Decodability is the percentage of the symbology’s safe decoding margin available for the scanner after printing errors. Lower margin values and corresponding lower grades indicate an increased susceptibility to decoding failure due to scanning errors. Scanning variations such as acceleration can cause bar or space deviations due to uneven, too fast or too slow scanning and provide lower margin values. You can set the type of decodability in the test criteria options to either ISO/IEC Decodability or character averaged decodability (Avg). With either calculation method, the decodability is rated A if > .62, B if > .50, C if > .37, D if > .25 or F if < .25. • ISO/IEC Decodability is a graded measure of how close a given scan comes to reference decode failure due to variations in bar and/or space widths. • Character averaged decodability is also called “D margin.”

♦ To review the dimensional parameters: 1. Scan a symbol. 2. If the cursor is not next to Dim, use the arrows to place it there. The first two dimensional parameters appear in the results line. 3. To view other dimensional parameters, press ↓. The names of the dimensional parameters and their results scroll. 4. To review other types of data (Ref or Fmt), press →. OR 5. To resume scanning, start at the quiet zone and scan the symbol.

Calculating X The total shown as the dimensional parameter is the total symbol width in “X” dimension terms. Formally, a bar code symbol’s width includes the quiet zones. The value presented as total is a calculated count of the number of “X” dimensions from the starting bar through the ending bar of the scanned symbol. You can use this value and manually measure the symbol’s actual width to calculate “X.” 1. Press SELECT to display the menu with the term Print Results on top. 2. Press ↓ to move the cursor next to Calculate “X”.

4 - 10

Quick Check® 600/800 Series User’s Guide

Page 55

3. Press SELECT. The term Symbol Measures 00.00 Inches appears. The cursor appears in the whole number position. If you want to enter the symbol length in metric units (centimeters), enter the numeric value and then press → to move the cursor to Inches. Press ↑. The term changes to Centim. 4. Use the supplied ruler to measure the symbol length in decimal inches. Measure from the left edge of the first bar to the right edge of the last bar. 5. Press ↑ or ↓ repeatedly until you display the whole number of the symbol length. Press → to move the cursor to the tenths decimal place. Again, press ↑ or ↓ repeatedly until you display the value you want. In the same manner, enter the hundreths and thousandths values. 6. If you entered the symbol length in inches, continue with step 7. If you entered the symbol length in centimeters, press → to move the cursor to Inches. Press ↑. The term changes to Centim. Continue with step 7. 7. Press SELECT. The X value and the AvgBar value are calculated and displayed. If you entered 00.00, the X value is not calculated. 8. If you want to recalculate the X value using a different value for the symbol length, press ← to move the cursor to the left. The two lower lines of text disappear. Now press ↑ or ↓ repeatedly until you display the new value. Press SELECT. The X value and the AvgBar value are recalculated and displayed.

Reviewing the Reflectance Parameters The following table lists the reflectance parameters and summarizes their results. Descriptions of the parameters follow these tables. Parameter Print Contrast Signal(Prnt Contr Sig) Reflectance (Light) (Reflect (Light)) Reflectance (Dark)(Reflect (Dark))

Trad’l

√ √ √

Symbol Contrast (SymCntr) Reflectance min/max (Rmin/Rmax) Modulation Edge contrast min (EdgeContr (min)) Defects

Quick Check® 600/800 Series User’s Guide

ISO/IEC

√ √ √ √ √ √ √

4 - 11

Page 56

Parameter

Result

PCS

Pass if PCS > 75%,RL > 25% and RD < 30% Fail if not satisfied

RI/Rd

Pass if Rd/RI < 50% Fail if Rd/RI > 50%

SymCont

A if > 70% B if > 55% C if > 40% D if > 20% F if < 20%

Edge contrast (min)

A if > 15% F if < 15%

Modulation

A if > .70 B if > .60 C if > .50 D if > .40 F if < .40

Defects

A if < .15 B if < .20 C if < ..25 D if < .30 F if > .35

* For UPC, the minimum P.C.S. varies depending on the measured R-Max.

Errors in Reflectance Parameters If, after a scan, the abbreviation for Reflectance appears as >REF<, an error exists in one or more of the reflectance parameters. If so, you should use the arrow buttons to review each parameter. If the abbreviation for Reflectance appears as Ref, the symbol has no reflectance errors.

Print Contrast Signal, Light and Dark Reflectance During a scan, the Quick Check Verifier obtains the maximum light reflectance from spaces (Rmax or RI) and minimum dark reflectance from bars (Rmin or Rd). The print contrast signal, “P.C.S.,” is a traditional measure of print contrast. PCS= (R[L]-R[D]/R[L]. The traditional pass/fail criteria are PCS, 75% with RI > 25% and Rd < 30%. The traditional parameters are rated OK! or pass if the PCS, RI and Rd meet these criteria. If a value does not meet the criteria, the Ref parameters fail (LO!). If the PCS, RI or Rd values appears on the display or on a printout as “XX” or “ - - ,” the reading is out-of-range. You should recalibrate the unit and then rescan the symbol.

4 - 12

Quick Check® 600/800 Series User’s Guide

Page 57

Note: Certain high reflectance materials may always read “XX” or “ - - .” You should repeat the calibration. The ratio of Rd/RI (Rmin/Rmax) must be less than 50% to pass this test. Alternately, you can consider Rd < 50% RI. RI may be rated LO!, but it is never rated HI!, while Rd may be rated HI! but is never rated LO!.

Symbol Contrast, Minimum and Maximum Reflectance Symbol contrast, SymContr, is a graded measure of the difference between the highest reflectance value R (max) and the lowest reflectance value, R (min), obtained during a scan. SC = R(max)-R(min). Symbol contrast is rated A if > 70%, B if > 55%, C if > 40%, D if > 20% and F if < 20%.

Edge Contrast Edge Contrast minimum is the smallest edge contrast or “edge transition” in a scan reflectance profile. The edge contrast is the difference between the reflectance of an adjacent bar and space. EC (min) is the minimum value of the edge contrast obtained during a scan. The criteria that edge contrast must be > 15% must be met. The EC (min) is rated A if > than 15% or F < 15%.

Modulation Modulation is a ratio of the edge contrast minimum value divided by the symbol contrast (EC (min)/SC). Modulation is rated A if > .70, B > .60, C if > .50, D if > .40 and F if < .40.

Defects Defects is the graded measure of the maximum element reflectance non uniformity (noise within any element due to spots, voids, etc.), normalized to symbol contrast. High readings are caused by print defects found: • in the symbol • adjacent to the quiet zone • at the transition of the label’s edge when a label which may have a marginal or inadequate quiet zone is laid on top of another material. Defects are rated A if < .15, B if < .20, C if < .25, D if < .30 and F if > 30.

♦ To review the reflectance parameters: 1. Scan a symbol. 2. If the cursor is not next to Ref, use the arrows to place it there. The first two reflectance parameters appear in the results line. 3. To view other reflectance parameters, press ↑ or ↓. The names of the reflectance parameters and their results scroll. 4. To review other types of data (dimensional or format), press → or ←. OR

Quick Check® 600/800 Series User’s Guide

4 - 13

Page 58

5. To resume scanning, start at the quiet zone and scan the symbol.

Reviewing the Format Parameters The following tables list the format parameters and their results. Descriptions of the parameters follow the tables. Parameter

Trad’l

ISO/IEC

Symbol type

√ √ √

√ √ √

Parameter

Result

Message length

Compared against the appropriate symbology, user or application specification - OK if it meets the specification - Fail if it does not meet the specification Compared against the appropriate symbology or application specification or set by user - CkChr Checks if it meets the specification - Check Character Fails if it does not meet the specification - No Check Character if you do not set the unit to test check characters

Message Length Check Character

Check Character

Errors in Format Parameters If, after a scan, the abbreviation for format appears as >FMT<, an error exists in one or more of the format parameters. If so, use the arrow buttons to review each parameter. If the abbreviation for format appears as Fmt, the symbol has no format errors.

Message Length The length of an encoded message is the number of characters in the output less any of the following characters: • Start and stop characters (for Codabar and Code 39) • Any symbology-required check characters (for EAN/UPC and Code 128) • Any function characters (for Code 128, 16K and Code 49)

4 - 14

Quick Check® 600/800 Series User’s Guide

Page 59

Check Character A check character in a symbol may be mandated by a bar code symbology specification, by you, by both or not at all. The display reflects the requirement and the result, either CkChr Checks if it passes or Check Character Fails if it fails. If you do not set the unit to test for a check character, the message NO Check Character appears. If the check character is bad, a message indicating that appears.

Symbol Type The symbol type refers to differences in symbol formats used in various applications. That is, the same symbol used in different applications may have a different length or a unique data or application identifier. When you set the verifier to an application, the unit attempts to identify the symbol type. For example, if an AIAG Code 39 symbol is six or fewer numeric characters with a data identifier of Q, it is a “quantity” symbol. If the symbol is nine or fewer alphanumeric characters with a data identifier of S, it is a “serial number” symbol.

Chapter 8, Using a Quick Check 800 Verifier lists the symbol types and error messages. If an application does not have a unique symbol format, the unit will simply display the symbology.

♦ To review the format procedures: 1. Scan a symbol. 2. If the cursor is not next to Fmt, use the arrows to place it there. The first two format parameters appear in the results line. 3. To view other format parameters, press ↑ and ↓. The names of the format parameters and their results scroll. 4. To review other types of data (dimensional or reflectance), press → or ←. OR 5. To resume scanning, start at the quiet zone and scan the symbol.

Reviewing Messages Errors can result from a variety of causes. They can result from errors in dimensional, reflectance or format parameters generically or based on an application. An example of a generic (symbology) error message that may appear after scanning a symbol is “BAD QuietZon.” This message indicates that there is insufficient quiet zone width. Chapter 8, Using a Quick Check 800 Verifier lists error, application and type messages.

Quick Check® 600/800 Series User’s Guide

4 - 15

Page 60

Using SSTR Codes to Review Data A Single Scan to Result (SSTR) Code is a bar code symbol that tells the verifier to execute certain command functions. If you scan an SSTR Code, the unit displays the scan result parameter represented by that bar code symbol. After you scan a symbol, either the default results appear (listed below) or, if you changed the output display options, the results you specified appear. SSTR Codes

Type

Test Criteria and Notes

Parameter SSTR Codes Average Bar Error W/N Ratio Decodability R(I) : R(d) P.C.S. Symbol Contrast Rmin/Rmax Modulation Edge Contrast (Min) Defects Symbol Type Message Length

Dim Dim Dim Ref Ref Ref Ref Ref Ref Ref Fmt Fmt traditional traditional ISO/IEC traditional and ISO/IEC traditional ISO/IEC ISO/IEC ISO/IEC ISO/IEC ISO/IEC traditional and ISO/IEC traditional and ISO/IEC

Default

Dim/Ref

Install Upper Line Install Lower Line Save Settings, - - ISO/IEC Decodability & Modulation traditional Average Bar Error and P.C.S. installs upper result line installs lower result line saves settings

SSTR Code Type and Name

To review results that are not on the display, use the arrow buttons or simply scan one of the parameter SSTR codes for the type result you want. The result appears on the verifier’s lower display line. Use the arrow buttons to step through the verifier’s menus (Dim, Ref and Fmt) to view these results. Chapter 11, SSTR Codes includes a list of SSTR Codes. Use the other SSTR Codes (Default, Install Upper Line, Install Lower Line and Save Settings) to customize your verifier. If you want to review one or two parameter results after each scan, you can customize the verifier so these results appear as the default first screen. For example, if you are most interested in the message length and symbol contrast data, you can set your unit to display the message length and symbol contrast results automatically after each scan.

4 - 16

Quick Check® 600/800 Series User’s Guide