The text of a manual · part 5 of 12
Quick Check 600/800 Series Bar Code Verifiers User's Guide, pages 41 to 50
Pages 41 to 50 of 120
The original document is the PDF (1.2 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the PSC QC-800, PSC QC-600, Datalogic QC-800 and HHP QC-800
Page 41
RI & RD, P.C.S., Symbol Contrast, Rmin/Rmax, Modulation, EC(min), Defects, Message Type, Message Length or Check Character Error Override: Yes or No LEDs: P/F Criteria or Bar Growth Output Options Printer Type: (Choose One) QCP, HSP, SSP Printout: Long form or short form AutoPrint/Store: On or Off With profile: Yes or No Generic Code Selections U.P.C./EAN/JAN Enable: On or Off Magnification Factor? (Choose One) 080/085/095/100/105/110/115/120/125/130/140/150/160/180/200 Expand UPC-EO: Yes or No Code 39 Enable: On or Off Character Check: None or Checked Fixed W/N: None or select value between 1.4 and 3.9 ± 0.2 Fixed message length: None or select value between 1 and 49 characters Interleaved 2 of 5 Enable: On or Off Character Digit: None or Checked Fixed W/N: None or select value between 1.4 and 3.9 ± 0.2 Fixed message length: None or select value between 2 and 78 digits Code 128 Enable: On or Off Modes Shown: Yes or No Fixed message length: None or select value between 1 and 70 characters Code 93 Enable: On or Off Fixed message length: None or select value between 1 and 80 characters (from Code 39 set) Codabar Enable: On or Off Character Font and Tolerances: (Choose one) Trad’l, “USS” or ISO/IEC ‘83 Fixed W/N: None or select value between 1.4 and 3.9 ± 0.2 Fixed message length: None or select value between 1 and 61 digits Code 11 Enable: On or Off
Quick Check® 600/800 Series User’s Guide
3 - 17
Page 42
Check digits: 1 or 2 Fixed message length: None or select value between 1 and 80 digits Code 16K Enable: On or Off Modes Shown: Yes or No Fixed message length: None or up to 154 characters, full symbol Code 49 Enable: On or Off Fixed message length: None or up to 49 alpha (81 numeric) characters, full symbol MSI code Enable: On or Off Character Digit: None or Mod10 or Mod 11 Fixed message length: None or select value between 1 and 50 digits IATA 2 of 5 Enable: Yes or No Regular 2 of 5 Enable: Yes or No Check Digit: Yes or None
Choosing a Command Code You may want to use a Command- Code rather than a Command+ Code if you often use the same setup on two units with different printers. With the Command- Code, you can easily change the setups for the units without changing the printer types, which may be more confusing because all the other settings on these two Command Codes would be identical.
Customized Command Codes Creating your own Command Code requires a Quick Check 600/800 Verifier printer. If you do not have a printer, you can request a Command Code from Hand Held Products. You can use your Quick Check 600/800 Verifier printer to produce your own Command Code using the Special Codes in Chapter 11. These four Special Codes are labeled “Command- Code (Locked),” “Command Code- (Unlocked),” “Command+ Code (Locked)” and Command+ Code (Unlocked)”. The locked codes have an additional setting which locks the changes into the unit after the Command Code is scanned. The lock prevents accidental reprogramming during routine use. The lock can only be removed by scanning another, previously prepared, Command Code that does not include the lock, such as an unlocked code. When shipped to you, the Quick Check 600/800 Verifier settings are unlocked. If you do not have a Quick Check 600/800 Verifier printer or if you want to have a sharper, more durable Command Code, Hand Held Products can prepare a customized code for you.
3 - 18
Quick Check® 600/800 Series User’s Guide
Page 43
♦ To use a customized Command Code: 1. Scan the Command Code for the type of work you will be doing. The settings stored in that Command Code are now set in your unit. ♦ To create a customized Command Code: 1. Connect the Quick Check 600/800 Verifier to a printer and turn on both units. Refer to Chapter 5, Saving and Printing for instructions on connecting the unit and the printer. If you locked your unit by scanning a Locked Command Code, you must scan an Unlocked Code before you can make any changes. 2. Make any settings changes on the unit that you want to save on your Command Code. If you want to prevent others or yourself from making accidental or intentional changes to the verifier settings, scan a locked Code. (Remember, if you later decide to make changes, you must scan an unlocked code.) Your Command Code begins printing. 3. Label your customized printed Command Code on the line provided with a description of the code. You may also want to indicate if it is a Command- or Command+ Code. 4. To improve the bar code quality, make a copy of it on an office copier. Making a “reduced size” copy may also improve the quality. 5. If you want to change any of the settings, make sure your unit is unlocked, and repeat steps 2 through 4.
♦ To request a Command Code from Hand Held Products: You can have Hand Held Products create your customized Command Code by mailing or faxing Hand Held Products a copy of the Customized Command Code Selections List with all of the options you want on the Code indicated. Be sure to keep the original list with your Command Code selections for your future reference. Mail your completed list to the Hand Held Products’ office nearest you (see Chapter 14, Customer Support).
Quick Check® 600/800 Series User’s Guide
3 - 19
Page 44
3 - 20
Quick Check® 600/800 Series User’s Guide
Page 45
4
Reviewing Scan Data
Reviewing Scan Results After you customize the verifier settings for your use, you are ready to scan bar code symbols. The verifier acquires and analyzes data using the ScanProfile Methodology specified in ISO/IEC’s Bar Code Print Quality Guideline (ANS X3, 182). After a scan, the audio signal and the LEDs provide auditory and visual information about the symbol. The verifier provides results using either traditional print quality parameters to make a pass/fail judgement about the symbol or the ISO/IEC methods to judge a symbol and report an ISO/IEC scan grade. You can also review specific traditional or ISO/IEC parameters (dimensional, reflectance and format parameters). Each parameter is described later in this section. The flow chart in Chapter 10, Flow Charts displays all the scan data you can review and how to access each item using the menus.
Scan Grades The scan grade is also known as the “Scan Profile Grade.” You can think of the scan grade as the grade resulting from a single scan across the symbol, for example, at exactly three millimeters from the top of the symbol. The scan grade is the lowest grade received by one of the following parameters after a specified number of scans (you specify the number of scans using the Extended Accuracy setting): • • • • • • •
Edge determination Rmin/Rmax Symbol contrast Minimum edge contrast Modulation Decodability Defects
The scan grade must also comply with any additional requirements imposed by the appropriate industry application standard, symbology specification or both for each scan. For example, if you set the passing option test criteria to traditional, the scan grade is a pass/fail judgement. However, if you set the passing option to ISO/IEC, the scan grade is a letter or number grade. Using the Extended Accuracy mode greatly reduces reading variability due to variations in scanning motion and minor scan path variations. The Extended Accuracy settings you can specify to calculate the scan grade include off (1), 2 through 10, 12, 15, 20, 30, 50 and 100.
Quick Check® 600/800 Series User’s Guide
4-1
Page 46
ISO/IEC Symbol Grade The ISO/IEC Symbol Grade is the average of the scan grades resulting from scans you take from different paths across the symbol. The ISO/IEC Symbol Grade is also referred to as the overall or final symbol grade. You can obtain a symbol grade only if the passing option test criteria is set to an ISO/IEC grade (not to traditional). You use the #Scans/Symbol setting to specify the number of individual scan grades that you want the verifier to use to calculate the symbol grade. The available settings are 1 through 10, 12, 15, 20, 30, 50, 100 and Var. The default setting for #Scans/Symbol is 10. The ISO/IEC standard requires you obtain ten scans of each symbol. For example, you might sample the symbol at ten different places in the symbol, four in the upper portion, two in the middle and four in the lower portion. Multiple scans of the same symbol across different scan paths take advantage of the vertical redundancy of bar codes and give you a more accurate value of the symbol’s overall print quality. The flow chart shown on page 4-3 explains how the ISO/IEC Symbol grade is determined.
4-2
Quick Check® 600/800 Series User’s Guide
Page 47
Quick Check® 600/800 Series User’s Guide
Edge determination Symbol contrast Edge contrast min Decodability Defects Modulation v
Edge determination rN y
¥
¥
¥
R min./R max
A< 50% F> 50%
Symbol contrast
A> 70% Bz 55% Cz 40% D> 20% F< 20%
Edge contrast min.
A> 15% F< 15%
Decodability A>.62 B>.50 C2.37 D>.25 F<.25
Defects
A<.15 B<.20 C<.25 D<.30 F > .30
Modulation
A>.70 B > .60 C>.50 D>.40 F< .40 y v y v grade of parameters A = highest F = lowest
Lowest alphabetic/numeric
Fail = F
[Decode?| me | Pass | le lowest scan grade or F
| hl
¥v
Determining the ISO/IEC Symbol Grade
Page 48
Quick Check® 600/800 Series User’s Guide
A 4
Conversion of scan grade to numeric grade tTAN-O [Te qmoou ma
Numeric average of X scan Compliance with any additional
Numeric grades from other > grades. X preset using 4 requirements imposed by the scan paths. #Scans / Symbol setting appropriate application, symbology specification or both. Graded A or F.
i
Conversion of symbol average to symbol grade
3.5< A<4.0 25< B<3.5 15< C<25 5< D<1.5
Fe 5
|
Symbol grade
Determining the ISO/IEC Symbol Grade (continued)
Page 49
Reviewing a Scan Example To clarify the difference between a scan grade and a symbol grade, this section presents an example of how you might set up your verifier, scan the symbols and review the results. In this example, we change the verifier settings and see how they affect the results. 1. To begin, we set up the verifier with the following settings: •
Test Criteria Passing Option set to ISO/IECISO/IEC; the letter grade selected does not matter • Scanning Options Extended Accuracy set to 3; note that this setting affects the scan grade. 2. Scan a bar code symbol. After the first scan, a message appears stating we have two more scans to go (of the same scan path or line across the symbol) to satisfy the preset Extended Accuracy setting of 3. 3. Continue to scan the same line of the symbol for a total of three scans. After the third scan, the scan grade appears on the display (the scan grade is the lowest of the grades of all of the parameters, in this case, averaged for the three scans). We now have the scan grade for a specific path across the symbol. 4. Scan another line or path of the symbol three times to obtain a second scan grade. After the third scan of the second set, a different display appears. This display is the value we set for the #Scans/Symbol (2) and the number of scan grades used to determine the symbol grade.
Note: If you set the Test Criteria Passing to traditional rather than to ISO/IEC grade, the scan grade is reported as Pass or Fail. The verifier does not report an ISO/IEC symbol grade when you select “traditional.”
♦ To obtain the scan grade and ISO/IEC Symbol grade: 1. Enter the Scanning Options Extended Accuracy setting you want. This setting determines the number of scans of the same path you want used to calculate the scan grade. If you set the Test Criteria Passing setting to ISO/IEC, you can obtain the ISO/IEC symbol grade: • •
Enter the Test Criteria Passing setting to an ISO/IEC grade Set the Test Criteria #Scans/Symbol setting to the value you want. This setting determines the number of scans grades you want to use to calculate the final symbol grade. Refer to Chapter 3, Changing the Settings for instructions on changing the settings. 2. Scan the bar code symbol. The first line of this display provides the symbol or application name and the encoded scan data. If the first line of scan data exceeds 16 characters in length, a right arrowhead appears as the rightmost character of the display. Press → on the verifier to review all the decoded data. A left arrowhead
Quick Check® 600/800 Series User’s Guide
4-5
Page 50
appears at the beginning of the decoded data to indicate there is more data to the left. Press ← to review the beginning of the message. The text on the other lines of the display depends on the Extended Accuracy option setting. - If the Extended Accuracy setting is off, the scan grade display appears after a single scan of a symbol. - If the Extended Accuracy setting is set to any value except off, a display appears with the term Extended Accuracy and the number of scans remaining to satisfy this setting. 3. To satisfy the Extended Accuracy setting, scan the same bar code symbol on the same path for the displayed number of times. - If you do not scan the same symbol within five seconds, the unit assumes you do not want to satisfy the Extended Accuracy setting you chose and displays the scan grade from that individual scan. If, after your first scan, you scan a different symbol type or a differently encoded symbol, the Extended Accuracy counter resets. - When you satisfy the Extended Accuracy setting, the scan grade display appears. - The second line of the scan grade display lists the three types of parameters, Dim (dimensional), Ref (reflectance) and Fmt (format). If an error exists in any of the parameters, that parameter name appears in all capital letters with a left and right caret pointing to it, for example, >DIM<. - The third and fourth lines of the scan grade display contain either ISO/IEC or traditional measurement results (parameters) depending on how you set the Test Criteria setting. Refer to later sections in this chapter for information about each parameter. You can review all the scan results using the arrow keys on the unit or use the SSTR (Single Scan to Result) codes described later in this section to review most of the scan results. For additional scan results, the LEDs light and/or flash, depending on how you set them up. - The large letter or number on the right of the display is the scan grade. If you set the verifier to use ISO/IEC criteria to test symbols, you can display the scan grade as a letter (A, B, C, D, or F) or as a number (4, 3, 2, 1, 0) using the Test Criteria Grades setting. If you set the unit to use traditional criteria to test symbols, a P or F appears for test pass or fail.
Note: You may want to print each scan grade result. The printout you obtain after the verifier calculates the final symbol grade shows the symbol grade and the results of the last scan grade only, but does not show each scan grade averaged to calculate the final symbol grade. The results of each scan grade may be helpful if you are trying to determine if particular paths (upper versus middle versus lower) consistently provide different results. 4. If you set the Test Criteria Passing setting to an ISO/IEC grade and you set the #Scans/Symbol setting to any value except 1, press ↑ to display the running symbol grade. To review the scan grade display, press ↓. Press ↑ to return to the symbol grade display.
4-6
Quick Check® 600/800 Series User’s Guide