Page 31

5.0 Symbol Testing & Scan Results The Quick Check 200 Series acquires data during a measurement scan utilizing the new software-intensive “ScanProfile” methodology as specified in ANSI’s “Bar Code Print Quality Guideline” (ANSI x3.182). (NOTE: In all cases, the more smooth and slow the scan, the more accurate the results.) the user may be prompted with a series of beeps (one long and three short) and one of the following displays: [Scan Slower!] or [Scan Evenly!] or [Scan Faster!] indicating that the scan was either too fast, too uneven/erratic or too slow for proper data collection. A practiced, more deliberate and smooth rescan should be made to insure accurate results. Both pass/fail testing and detailed measurements on a given symbol are obtained simply by scanning across it with the QC200. The most accurate results will be obtained by starting just outside the quiet zone and scanning in a single, smooth motion with the optics tips and reflectance guide “foot” lightly touching the symbol surface. Practice helps! A successful measurement will be signaled by beeping, a blinking green (pass) or red (fail) LED and the display showing results such as:

If in Traditional P/F Test Mode. When in ANSI Grading Test Mode, the PCS value shown is replaced with an ANSI letter grade or its numerical equivalent.

23

Page 32

A symbol that consistently produces no beeping is either fatally out-of-spec or of an unknown or unprogrammed bar code symbology. The QC200’s customer LCD display shows the user a complete synopsis of the scan results. The information shown includes the decode of the symbol’s encoded characters (dot matrix characters in the middle of the LCD), the symbology type (across the LCD top), the PCS or ANSI Grade (on the right when in the Right Handed viewing orientation), an indication of bar growth/shrinkage (small arrows to the extreme left of the LCD when in the Right Handed viewing orientation) and in what parameter any error may have occurred (the key words DIM, REF and/or FMT will light just to the left of the symbol decode, when in the Right Handed viewing orientation). If the encoded message in a bar code symbol exceeds 12 characters, the right most character of the display will be a “right arrow” (>). In that case, pressing the Scroll[<>] Button will “scroll” the encoded message and allow the remainder of the message, in 11 character portion, to be viewed. As mentioned above, uniform bar growth or shrinkage is displayed by a vertical series of “arrows” on the LCD. One “arrow” will remain lit to indicate the average bar width error in relation to traditional print tolerance allowances. When centered between the “+” and “-“ on the label, the bar growth/shrinkage is at a minimum. As the arrow position moves toward the “-”, increasing amount of shrinkage is seen. Conversely, a move toward the “+” indicates bar growth. When the arrow directly across from either the “+” or “-“ is lit, traditional bar width tolerances have been exceeded. There are seven arrows in the scale. With each test scan of a symbol, the QC200 also indicates the sample’s pass/fail status with a characteristic “beep” pattern. One long beep with ONE short indicates that the symbol passes the appropriate battery of tests. One long beep with THREE short beeps indicates that it fails in some regard. Thus pass/fail testing proceeds by simply scanning symbols and listening for a triple beep. Upon any failure, the LCD indicates the parameter area where the failure occurred. These areas are indicated by a three-letter abbreviation. 24

Page 33

DIM = a DIMENSIONAL parameter error REF = a REFLECTANCE (contrast) parameter error FMT = a FORMAT parameter error In a symbol with multiple errors, you will have more than one of the above indicators lit. Exact details of the failure can be obtained by following the procedures in the section that follows. When in the ANSI Grading mode, every tenth measurement of a given symbol (or every eightieth measurement if also in Extended Accuracy mode) will cause the results message for the last scan to be replaced with: [**g/aa/lll**] x.x reflecting the ANSI Symbol Grade (average of the ten Scan Grades per ANSI procedures). In this display, “g” is the ANSI Grade (A,B,C,D or F), “aa” is the verifier’s aperture size, “lll” is the verifier’s spectral response and “x.x” is the numerical value the ANSI letter grade was derived from. 5.1 Test Result Details 5.1.1 Dimensional Parameters After a successful scan is made, detailed results in the Dimensional Parameters area are accessed by first pressing the Select [0] Button once. This will produce the display: [Dim’l Data>] P (or F) Either the “P” or “F” will be shown to indicate if the symbol under test passed or failed the Dimensional Parameters. Pressing the Scroll [<>] Button will move you into the Dimensional Parameters details. The order the displays are shown changes dependent upon whether the QC200 is set for “Traditional” or “ ANSI” testing criteria (though in each case all the same displays are shown). For this document, the order used with “Traditional” test criteria is illustrated, with the first display showing: [AvgBar x.xxX] P (or F) 25

Page 34

AvgBar is the average bar print error. It is the value or amount of average bar growth or shrinkage in the measured symbol relative to its X dimension. This value is compared against Traditional print tolerance criteria with the “P” of “F” showing compliance or non-compliance. Pressing the Scroll [<>] Button again changes the display to: [W/N Ratio x.xx] P (or F) W/N Ratio is the wide-to-narrow ratio and is measured for those symbologies where it is meaningful, namely Code 39, Interleaved 2/5 and USS-Codabar; otherwise, it is not displayed. This value is compared against either generic limit (1.8 to 3.2) or a target fixed value (e.g., 2.5 + 0.2) to determine if it passes or fails. An out-of-range W/N ratio is considered a dimensional error under BOTH Traditional and Graded testing criteria. Another press of the Scroll [<>] Button will display (ONLY IF applicable errors have been found) certain Global Threshold, Reference Decode or Reference Algorithm failure messages such as: [GloThrsFAILS] F or [RefDcod FAIL] F or [RefAlgoFAILS] F Global Threshold ANSI’s Symbol Grading methodology requires foremost that a global reflectance threshold set simply at the halfway point between R(min) and R(max) should “find” every bar and space in the symbol. The Quick Check 200 Series uses the ANSI prescribed methodology for precise edge location and indicates when the global threshold requirement has not been met. “Reference Decode or Algorithm” ANSI’s Graded symbol testing criteria requires that a given scan be decodable using certain “reference” algorithms, else the scan fails. Similarly, certain application specifications call for specific formats, checks or characters that must be in the symbol. The Quick Check 200 Series employs advanced decoding schemes to test the necessary reference decoding and algorithms. If any errors are encountered, a failure is indicated appropriately. If errors were not found in these areas, the next display would read: [D Margin xx%] Grade or [Decod’ty xx%]Grade 26

Page 35

Decodability/Decodability “Margin” Decodability” is a Graded measure of how close a given scan comes to a reference decode failure due to variations in bar and space widths. The Quick Check 200’s “Avg DMargin” is the character-averaged value of Decodability. Decodability is what percentage of the symbology’s safe decoding margin is left for the scanner after printing errors. Lower margin values and corresponding lower grades indicate an increased susceptibility to decoding failure due to scanning errors such as, for example, acceleration or bar or space size deviation. Decodability “Margin” is stated as “D Margin” in the QC200’s LCD Dimensional Details display and ANSI decodability is listed as “Decod’ty”. If the Scroll [<>] Button is again pressed, various other Dimensional Parameter errors might be displayed next. Examples of these follow: [BAD QuietZon] F -- Indicates insufficient Quiet Zone width [WIDE IC Gap] F -- Indicates the InterCharacter Gap is too wide [BAD StrtStop] F -- Indicates either the Start or Stop Characters are bad [BAD GuardPat] F – Indicates that the Guard Bar Pattern in a UPC/EAN/JAN symbol is bad [BAD Separatr] F – Indicates a bad Separator Pattern Pressing the Scroll [<>] Button again produces the display: [Total xxx”X”] Total xxx”X” indicates the total symbol width expressed in “X” dimensions. “Formally” a bar code symbol’s width includes the quiet zones, however the value presented here is a calculated count of the number of “X” dimensions from the starting bar through the ending bar of the scanned symbol. The operator can use this value and the symbol’s actual measured width to calculate X. By pressing Scroll [<>] again, the display becomes: [Calc “X”? 0] The “X” DIMENSION can now be calculated by first pressing the Select [0] Button (pressing Scroll [<>] will by-pass this operation) giving the following display: 27

Page 36

[SymLen=0.--“] By then entering the symbol’s width as measured with the supplied ruler in decimal inches (e.g., 1.24), the operator provides the key parameter for calculating X. To change the value of “0.--“, press the Scroll [<>] Button. This increments the first digit from “0” to “9”. When the desired digit is shown, press Select [0] to accept and move to the next digit. After all three are filled in, the Quick Check® 200 series then calculates the X Dimension and displays it in the following manner: [X is x.xxxx”] If 0.00 is entered, the X remains uncalculated. The “total” width of any EAN/UPC symbol is that of the main symbol, ignoring a possible addendum. 5.1.2 Reflectance Parameters After a successful scan is made, detailed results in the Reflectance Parameters area are accessed either by first stepping through the Dimensional Details as shown above or by pressing the Select [0] Button twice. This will produce the display: [Refl Data >] P (or F) Again, either the “P” or “F” will be shown to indicate if the symbol under test passed or failed the Reflectance Parameters. Pressing the Scroll [<>] Button will move you into the Reflectance Parameters details. The order of the displays are shown changes dependent upon whether the QC200 is set for “Traditional” or “ANSI” testing criteria (though in each case all the same displays are shown). For this document, the order used with “Traditional” test criteria is illustrated, with the first display showing: [PCS xx%] P (or F) and then pressing the Scroll [<>] Button again will give: [Rl:Rd xx:xx%] P (or F) Print Contrast Signal (PCS), Light Reflectance (Rl) and Dark Reflectance (Rd) are all “Traditional” measures of symbol contrast. PCS is based on the reflectance of light areas, 28

Page 37

Rl and dark areas, Rd. These values are determined using a definition of Rl and Rd as the Peak Light and Peak Dark points seen during the scan. Thus Rl = R(max) and Rd = R(min) in current terminology. The Traditional pass/fail criteria are PCS > 75% with Rl > 25% and Rd <30%. NOTE: IF DURING THE MEASUREMENT OF A SAMPLE SYMBOL AN “XX” IS DISPLAYED OR PRINTED AS A Rl, Rd or PCS REFLECTANCE VALUE, AND “OUT OF RANGE” READING HAS BEEN MADE. THE QC200 SHOULD BE RE-CALIBRATED AND THE SAMPLE SHOULD BE RESCANNED. Again pressing the Scroll [<>] Button will change the display to: [SymContr xx%] Grade Symbol Contrast (SC) is a Graded measure of the difference between the highest reflectance value R(max) and lowest reflectance value R (min) seen during a measurement scan. Another press of the Scroll [<>] Button changes the display to show: [Rmin/Rmx xx%] Grade Rmin/Rmx has an additional Pass/Fail (Grade “A” or “F”) requirement that it be at or below 50%. Press the Scroll [<>] Button again to get: [Modul’n xx%] Grade Modulation (MOD) is an ANSI graded criterion equal to Edge Contrast (min) divided by Symbol Contrast. [Edge(mn) xx%] Grade Edge Contrast Minimum (Ecmin or Edge(mn)) is the minimum value of Edge Contrast (EC) seen during a measurement scan. Edge Contrast (EC) is the reflectance difference between an adjacent bar and space. The Pass/Fail requirement (Grade “A” or “F” here is that the EC(min) be >15%. Another press of the Scroll [<>] Button brings us to the final Reflectance parameter detail of: [Defects xx%] Grade Defects is the Graded measure of the maximum Element Reflectance Non-uniformity (noise WITHIN any element due to spots, voids, etc.) normalized to symbol Contrast. 29

Page 38

5.1.3 Format Parameters After a successful scan is made, detailed results in the Format Parameters area are accessed either by first stepping through the Dimensional and Reflectance Details as shown above or by pressing the Select [0] Button three (3) times. [Msg Format >] P (or F) Either the “P” or “F” will be shown to indicate if the symbol under test passed or failed the Format Parameters. Pressing the Scroll [<>] Button will move you into the Format Parameters details. The order the displays are shown changes dependent upon whether the QC200 is set for “Traditional” or “ANSI” testing criteria (though in each case all the same displays are shown). For this document, the order used with “Traditional” test criteria is illustrated. If an Application Specification has been activated, the normal sequence will be superseded by the following display: [APPL *Type*] where APPL will be the appropriate Application and Type will be any possible sub-identifier within that application. Repeated pressing of the Scroll [<>] Button will then show any other application specific errors such as: [BAD Source] and will finally come to: [MsgLength xx] P (or F) Message Length is the length of an encoded message. The length may need to be checked against an application-specified length restriction. Another press of the Scroll [<>] Button displays: [No Check Chr] P or [ChkChr Checks] P or [ChkChr Fails] F Check Characters may be mandatory under a bar code symbology specification, required by the user or by an application specification: either, neither, or both! The display screen will reflect the requirement and the result. Repeated 30

Page 39

pressing of the Scroll [<>] Button will reveal any additional error situations such as: [BAD Symbology] F if a wrong code type is found. At this point (or any other point during the scrolling of results), pressing the Select [0] Button will cycle the QC200 into its “Store” (if not hooked to a printer) or “Print” (if hooked to a printer) sequence. The print sequence is explained in Section 6.

31

Page 40

32