Page 41

6.0 Printer Print Out Samples [Store Data?>] When a printer is not attached or is turned off, you are asked if you wish to store the measured results in a 20-scan print buffer. Pressing the Scroll [<>] Button will store the scan. The buffer’s contents can later be dumped to the printer. When a QC200, that has stored data, is hooked to a printer and the QC200 is then turned on, the first display will be a query on whether or not to dump the stored information. NOTE: The QC200 will also “see” the QCAC Charger as a “printer” when it is connected. [Print Data?>] When a printer is attached to the Quick Check® 200, you are asked if you wish to print the most recent scan. Pressing the Scroll [<>] Button will cause the current testing results to be printed and the following display will show: [ - Busy ! - ] NOTE: Printing can be halted after starting by momentarily pressing the POWER button on the QC200. Printing will stop as soon as the printer’s internal buffer is clear.

33

Page 42

6.1 Print Out Samples All of the measurements obtained during a measurement scan are saved in a temporary buffer. When a printer is not connected those measurements can be transferred to a 20-scan print buffer as described above. When a printer is attached and powered, it produces a listing of full measurement details. From a QCP or QCHSP, the print out would look similar to the following: \\\Quick Check 200\\\

660 nm, 06 mil Scanner UPC-A: 100% Mag. Factor 021200074424 Avg Bar Err = +.04X OK! DecodeMargin = 92% <A> Symbol Total = 095 “X” Prnt Contr Sig = 94% OK! Reflect (Light) = 75% OK! Reflect (Dark) = 03% OK! Symbol Contrst = 72% <A> R(min)/R(max) = 04% <A> Modulation = 79% <A> EdgeContr(min) = 57% <A> “Defects” = 05% <A> Message Length = 11 OK! Check Charctr Passes OK! Traditional Tests <PASS> Bar Growth: IN TOL (+) “P C S” – OK Traditional Tests - PASS Format Tests - PASS ANSI Parameters <A> Dimension Checks <P> Format Checks <P> Profile QualityGrade <A>

34

Page 43

From a QCSSP, a typical printout would look more like the following:

-

Quick Check 200 -

660 nm, 06 mil Scanner Code 39 *+A123B4C5D6E711* AvgBar = +.05 X OK! W/N Ratio = 2.9 OK! DecodMargn = 88% <A> Symbol Totals 264X PCS = 90% OK! R(L) = 79% OK! R(D) = 07% OK! SymbolContr = 72% <A> R(mn)/R(mx) = 09% <A> Modulation = 61% <B> EdgeCtr(mn) = 44% <A> “Defects” = 04% <A> Msg Length = 15 OK! -No-Check Character Bar Growth: IN TOL “P C S” : OK Traditional Tests -PASSFormat Tests -PASSANSI Parameters: <B> Dimension Checks: <P> Format Checks: <P> Profile Quality Grade is <B>

35

Page 44

For all printers, on the tenth scan of a given symbol (or eightieth scan if in “Extended Accuracy” mode), a message similar to the following will be appended to the normal printout. For the QCP and QCHSP:

*********************** **** * Symbol Quality Grade * * Avg 4.0 = > A/06/660* *********************** ****

And from the QCSSP:

*************************************************** ****** Symbol Quality Grade: Avg 3.0 = > B/06/660 *************************************************** ******

Each result shown in the printout is information also available on the display as previously described. Having the wrong printer setting, if encountered, may cause printing difficulties. This is set within the “Config.” Menu as explained previously. Printing can also be initiated in either a single (“on demand”) or continuous (“auto-print”) print mode by use of the Printer Control Command Codes found later in this manual. NOTE: Printing can be halted after starting by momentarily pressing the POWER button on the QC200. Printing will stop as soon as the printer’s internal buffer is clear. 36

Page 45

7.0 Special Functions SPECIAL FUNCTIONS The Quick Check 200 series can be set to a special “static” reflectance measurement function (Reflectometer Mode). This mode is especially useful in investigating the underlying optical characteristics of bar code symbols. The Reflectometer Mode, along with the Command Code mode is covered below. 7.1 Reflectometer Mode This menu selectable Special Function is entered by pressing the Scroll [<>] Button when [Reflectom’r>] is present on the LCD. Then in this special reflectometer mode, the reflectance “seen” by the QC200’s optics is continuously displays as both a two digit reflectance percentage and on a graphic bar graph as soon as the mode is invoked. Here the xx% and the graphic bar graph (or “gas gauge”) represent a reflectance value of 00% to 100% from the static reflectometer. This mode is not “memorized” when the unit is shut off. When turning the unit back on after power down, it will be necessary to re-establish the Reflectometer Mode if needed. For these values to be meaningful and consistent, it is critical first that reflectance calibration have been performed recently and second that the QC200 be held with its “reflectance foot” and optics tip both touching the sample to maintain the proper angle. 7.2 Command Code Programming The Quick Check 200 series has the unique ability to be “preprogrammed” by the user by scanning a specially encoded code 128 symbol. This symbol will automatically make all the selections normally made by the user stepping through the normal menu sequence. This is particularly useful if the user has one or two particular setups that they are constantly switching between. Samples of Command Codes, such as the

37

Page 46

Command Code to “reset” the QC200 back to all generic default settings, are included with each unit. NOTE: Display orientation (Right Handed vs. Left Handed) and Printer Interface selection ARE NOT set using the Command Code. Factory default settings for these are “Right handed” and “QCHSP”. Users with one of the accessory Quick Check Printers can produce a Command Code of their QC200’s present set-up by following the next steps; First be sure your Quick Check 200 is connected to a Quick Check Printer, that the proper interface selection has been made and both are turned on. Next, locate the Special Features Bar Code Menu in the back of this manual. On this Menu sheet are two codes, Command Code “Unlocked” and Command Code “Locked”. Scanning either will result in the Quick Check

Printer printing the specialized code 128 you can use for rapid one step programming of your QC200. The Command Code “Locked” code allows the user to add a feature to the Command Code that will “lock” the menu programming after that code is scanned the first time. This prevents access to reprogramming of the unit by the casual user. The “lock” can not be removed unless the user scans another, previously prepared, Command Code that does not have the “lock” function in it, such as the “reset” Command Code supplied in this manual. (NOTE: Before actually using this code, label it along the line provided with a description and make a copy of it on your office copier. This helps improve the bar code quality. Making a “reduced size” copy will help also.) These codes are also produced by HHP. Contact Customer Administration at HHP about this service. It will be necessary to make the following selections;

38

Page 47

OPERATING MODE Extended Accuracy (8-Scan/Scan path average) ON(ExtendAccur) or OFF(NormalAccur)

Decodability (choose one) Hand Held Products’ Decodability Margin (Avg DMargin) ANSI’s (Decodability)

Pass/Fail Criteria (choose one) “Traditional” P/F (Trad’l P/F) or ANSI Grades – D, C, B, or A

ANSI Grading by LETTER or NUMBER (choose one) Letter Grades (Ltr.Grades) Number Grades (Num. Grades)

Program Menu Lock On or Off Reflectometer Mode On or Off

SPECIFICATION Generic Codes vs. Application Spec (choose one) Generic Codes (if selected an Application cannot also be selected) Code 39 Yes or No and Mod Check? Yes or No Interleaved 2 of 5 Yes or No and Mod Check? Yes or No UPC/EAN/JAN Yes or No Magnification Factor?_ _ _ (Default is 100) Choices – 080/085/090/095/100/120/140/160/180/200

Codabar Yes or No Traditional Codabar? Yes or No USS Codabar? Yes or No Code 128 Yes or No and Modes or Messages (choose one) MSI Code Yes or No User Check Digit? Yes or No Mod 10 or Mod 11? (choose one) Code 16K Yes or No and Modes or Messages? (choose one)

39

Page 48

Application Specification CCBBA Yes or No HIBCC Yes or No AIAG Yes or No Bookland Yes or No SISAC Yes or No Logmars Yes or No CompTia/CITA(ABCD)Yes or No SCC Yes or No I 2/5 mag factor? _ _ _ (default 075)

40

Page 49

8.0 Quick Check Maintenance 8.1 Battery Charging Charging the battery pack > > when the batteries require recharging, the LOW BATTERY!! Prompt will flash on the LCD. The unit will not allow further scans until it has been turned off and the batteries recharge to a level above the operational minimum. The batteries will only recharge when the Quick Check is not in use. To charge the batteries in the unit: 1.

Plug the QC charger into the power jack on the end of the Quick Check.

2.

Plug the AC charger power cube into an appropriate VAC wall outlet.

3.

Charge the batteries for a minimum of six (6) hours.

After following the proper battery charging time, your Quick Check will be ready for operation. 8.2 Scanner Tip Cleaning Occasionally the scanner tip and its optically clear window may become partially obstructed by ink, paper fibers or other deposits. To insure reliable scanning, you should check the tip periodically and clean it if dirty. To clean the scanner tip: 1. With the Quick Check turned off, lightly dust the tip with a soft camel’s hairbrush or equally soft device. 2. If ink or deposits are still evident, using a mild cleaning solution and soft applicator (cotton swab), wipe the tip to clean. NOTE: Damaged wand tips may be replaced. Contact the Customer Administration or Service Department for details.

41

Page 50

8.3 Reflectance Calibration Page Replacement The Reflectance Calibration Page that comes with your Quick Check 200 is used for setting the unit’s reflection calibration point. This page is a precision made standard. If it should become physically damaged or dirty, it should be replaced in order to insure proper operation of the unit. Contact Customer Administration at HHP to order replacements. 8.4 Troubleshooting

If your Quick Check fails to operate properly after following all operation instructions carefully, do not open the case or attempt to repair. This will void the warranty. Contact HHP for assistance.

42