The text of a manual · part 6 of 7
Portable pH Meter Instruction Manual: SECTION 3 INSTRUMENT OPERATION
Pages 25 to 30 of 40, from Portable pH Meter Instruction Manual
The original document is the PDF (597 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Hach SensION1
Page 25
SECTION 3
INSTRUMENT OPERATION
3.1 Setup Menu The sension™1 Portable pH meter has a setup menu that allows the analyst to choose options for connector choice, display lock, temperature units, automatic buffer recognition of pH 6.86 or 7.0, and the display resolution. To access the Setup menu, press the SETUP key. The arrow icons will be displayed, indicating that additional options are available within the menu. Press the up or down arrow key to scroll to the desired option, then press ENTER. Table 3 describes these options. Table 3 Setup Options Setup Number
3.1.1
Option Description
1
Use BNC or 5-pin connector
2
Display lock (On or Off)
3
Temperature units (°C or °F)
4
Measurement resolution (0.0, 0.00, or 0.000)
5
Automatic buffer recognition (pH 7.00 or 6.86)
Choosing the Probe Connector This setup ensures the potential from the appropriate electrode is detected by the meter. Connect only one electrode to the meter. 1. From the reading mode, press SETUP. 2. The Setup icon and the number 1 (flashing) will be displayed. Change the connector choice by pressing ENTER; this toggles the selection between BNC and 5 pin. 3. When the desired option is selected, press EXIT to return to the reading mode.
25
Page 26
SECTION 3, continued 3.1.2
Turning Display Lock Off and On Setup 2 is Display Lock option. This feature stops measurement reading fluctuation on the display once a stable reading is reached. The default setting is Off. When this feature is not used, the measurement value may continue to fluctuate and Stabilizing... may be displayed. 1. From the reading mode, press SETUP. 2. The Setup icon and the number 1 (flashing) will be displayed. Press the up arrow once so the Setup number is 2. 3. Change the display lock status by pressing ENTER; this toggles the display lock between off and on. When the display lock is disabled, the Display Lock icon and Off are displayed. When this feature is enabled, the Display Lock icon is displayed without Off. When this feature is enabled, the Display Lock icon is displayed when the stabilization criteria are met. 4. When the desired option is selected, press EXIT to return to the reading mode.
3.1.3
Selecting Temperature Units 1. From the reading mode, press SETUP. 2. The Setup icon and the number 1 (flashing) will be displayed. Press the up arrow until the Setup number is 3. 3. Change the temperature unit by pressing ENTER; this key toggles the temperature units between °C and °F. The default is °C. 4. When the desired option is selected, press EXIT to return to the reading mode.
26
Page 27
SECTION 3, continued 3.1.4
Selecting Measurement Resolution The meter can display measurement values to tenths (0.0), hundredths (0.00) or thousandths (0.000). The default is hundredths. 1. From the reading mode, press SETUP. 2. The Setup icon and the number 1 (flashing) will be displayed. Press the up arrow three times so the Setup number is 4. 3. Change the resolution by pressing ENTER; this toggles between the three resolution options. 4. When the desired option is selected, press EXIT to return to the reading mode.
3.1.5
Selecting Auto Buffer Recognition The sension1 Portable pH Meter is designed to auto-recognize and calibrate on 4.01, 6.86 or 7.00, and 10.01 pH buffers. Buffers with pH values other than these cannot be used to calibrate the sension1 meter. The only selection option for pH buffer auto recognition is 6.86 or 7.00. The default is pH 7.00. To change this option: 1. From the reading mode, press SETUP. 2. The Setup icon and the number 1 (flashing) will be displayed. Press the up arrow until the Setup number is 5. 3. Change the buffer value by pressing ENTER; this toggles the between the choices 6.86 and 7.00. 4. When the desired option is selected, press EXIT to return to the reading mode.
3.2 Calibrating the Meter Hach recommends a daily two- or three-point calibration using buffers that bracket the sample pH. This will verify the electrode is working properly and allow the slope value to be stored.
27
Page 28
SECTION 3, continued 3.2.1
Performing a Calibration Using pH 4, 6.86, 7, and 10 Buffers 1. If using a probe without a temperature sensor, see Section 2.3.1 on page 22 and Section 2.5 on page 22 for information about obtaining and using a default temperature. 2. Prepare two pH buffers, either 4.01 and 7.00 (or 6.86), or 7.00 (or 6.86) and 10.01, according to the electrode instruction manual. Note: Use a 6.86 or 7.0 pH buffer for the mid-range buffer. To view or change the setting for the mid-range buffer see Section 3.1.5. Note: The pH values for the buffers are given for 25 °C. If the sample temperature is not 25 °C, the pH values displayed for the buffers will reflect the correct pH value for the sample temperature.
3. Press I/O/EXIT to turn the instrument on. From the Reading mode, press CAL. CAL and flashing ? will appear in the upper display area, along with Standard and 1. 4. Press READ/ENTER. The temperature and pH values will be updated until a stable reading is reached. Note: If the meter is measuring in pH mode, it automatically moves to the next calibration step when stabilization is reached (indicated by three beeps). If measuring in mV mode, the three beeps will still sound when the stabilization occurs, but ENTER must be pressed to accept the reading. This lets the operator control the acceptance point of the buffer.
5. When the reading has stabilized or been accepted, the standard number will change to 2. 6. Remove the probe from the first buffer and rinse with deionized water. Place the probe in the second buffer. 7. Repeat steps 5 and 6 for the third buffer and press EXIT. 8. Press READ/ENTER. The temperature and pH values will be updated until a stable reading is reached. 9. When the reading has stabilized or been accepted, the slope value and the Store and ? icons will appear. Verify the slope value is within the ranges specified in the electrode manual.
28
Page 29
SECTION 3, continued 10. To save the calibration and return to the Reading mode, press ENTER. To exit the calibration without saving it and return to the reading mode, press EXIT.
3.3 Reviewing the Calibration 1. From the pH Reading mode, press the REVIEW key. 2. The display will show the standard number, standard pH, and temperature (view mV value by pressing pH/mV). Press the up arrow once. 3. The meter will continue to scroll through the standard information with each press of the up arrow key. When all the standards have been displayed, press the up arrow key again. 4. The meter will display the slope and offset of the calibration curve. 5. To exit Cal Review mode, press EXIT. To review any standard or slope information again, press the down arrow.
29
Page 30
SECTION 3, continued 3.4 Measuring Samples After successful calibration, the meter is ready to measure samples. See instructions in the electrode manual for more information and specific steps for using the electrode. 1. Place the electrode in the sample. Press READ/ENTER. Stabilizing... will appear, along with the sample temperature and the pH or mV reading. These values may fluctuate until the system is stable. 2. When the reading is stable Stabilizing... will disappear. If the Display Lock is enabled, the display will “lock in” on the pH or mV and sample temperature. If the Display Lock is off, the display will show the current reading and temperature, but the values may fluctuate. 3. Record the pH or mV value. 4. Remove the electrode from the sample, rinse with deionized water and place the electrode in the next sample. Repeat Steps 1-3 for each sample. 5. When measurements are complete, press the I/O/EXIT key to turn the meter off. Rinse the electrode with deionized water and blot dry. Replace the protective cap on the electrode and put the electrode in the electrode holder. To store the electrode for more than 30 days, see the electrode manual for specific instructions.
30