The text of a manual · part 5 of 7
Portable pH Meter Instruction Manual: SECTION 2 INSTRUMENT SETUP
Pages 19 to 24 of 40, from Portable pH Meter Instruction Manual
The original document is the PDF (597 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Hach SensION1
Page 19
SECTION 2
INSTRUMENT SETUP
2.1 Instrument Description This durable portable sension™1 pH/mV meter is designed for easy hand-held use when measuring samples in the field or laboratory. The meter operates on 4 alkaline batteries with an estimated battery life of at least 500 hours of continuous use. The meter measures from -2.0 to 19.99 pH units and the sample temperature. Displayed pH values are temperature corrected using the measured sample temperature or a default temperature setting. The meter also measures and displays mV.
2.2
Battery Installation
CAUTION Use only alkaline batteries in this product. Other types of batteries can result in safety hazards.
PRUDENCE Utiliser seulement des piles alcalines dans cet appareil. Les autres types de piles peuvent créer des risques pour la sécurité. ATENCIÓN Utilice solamente baterias alcalinas en este producto. El uso de otros tipos de baterias puede causar riesgos de seguridad. VORSICHT Verwenden Sie in diesem Produkt nur Alkali-Batterien. Die Verwendung anderer Batterien gefährdet die Betriebssicherheit. ATENÇÃO Use somente baterias alcalinas neste produto. Outros tipos de baterias podem resultar em risco a segurança.
The sension1 meter requires four alkaline AA batteries for portable operation. Other types of batteries should not be used in the meter. The battery compartment access panel is on the bottom of the instrument. To access the battery compartment, turn the instrument over and position the connectors away from you. Hold the instrument between your hands and use your thumbs to slide the panel towards you. See Figure 5.
19
Page 20
SECTION 2, continued When battery replacement is necessary, replace all four batteries. The position of the batteries in the instrument is important. Icons of the correct orientation are molded into the bottom of the battery compartment. With the connectors pointing away from you, insert a battery in left-most position so the positive end faces you. Insert the other three batteries so the positive and negative ends alternate (i.e, +, -, +, -). Insertion is easiest if the battery is pushed against the spring connector first, then pressed into place. Figure 5
Battery Replacement + + + +
Slid e to
Ope n
20
Page 21
SECTION 2, continued 2.3 Probe Connections Attach electrodes with 5-pin connectors to the sensor input by lining the pins up with the holes in the meter port (see Figure 6). Push the electrode connector toward the instrument. For probes with BNC connectors, slide the connector into the input. Push towards the instrument and turn clockwise to lock into position. Electrodes may be attached to both the 5-pin and BNC connectors at the same time as long as they are not in contact with the same solution. To select either of the connectors for measurement, go to the Setup 1 menu and select one. When using half-cells, connect reference electrodes with pin tip connectors by pushing the connector straight into the center reference input. If using a combination electrode with a BNC or 5-pin connector, the reference pin-tip jack is not used. Figure 6
Electrode Connectors
21
Page 22
SECTION 2, continued 2.3.1
Temperature Probe Connection Hach electrodes with the 5-pin connector have the temperature sensing unit included in the electrode probe. If using an electrode with BNC connector, connect a Hach Temperature Probe (Cat. No. 51980) to the 5-pin connector on the meter (see Figure 6). Or measure the temperature manually and enter the value as the default temperature using the arrow keys.
2.4 Turning the Meter On After the batteries are installed, turn the instrument on using the I/O/EXIT key (located on the upper left side of the keypad). Press the key once to power the instrument up. The display will show the software version number, then move to the Reading mode.
2.5 Temperature Measurement The meter displays temperature in the range of -10.0 to 110 °C simultaneously with sample results. If a temperature probe is properly connected, actual temperature measurements are displayed in the temperature/mV field. The meter requires a temperature to calculate temperaturecorrected pH readings. The meter uses temperature data from one of three sources:
•
The temperature sensor in the sample
•
The factory default setting (25 °C)
•
A user-entered setting (which becomes the default)
If a temperature probe is connected properly, the meter will display the current sample temperature and will not allow the temperature to be manually set. To manually set the temperature When a temperature probe is not used to supply the temperature for pH temperature compensation, the temperature must be set manually or the factory default temperature (25 °C) will be used. In either case, Default will be displayed above the temperature. To change the default temperature, first put the meter in Reading or Calibration mode. Press the arrow keys to increase or decrease 22
Page 23
SECTION 2, continued the displayed default temperature value to the desired temperature. The value adjusts in 0.1 °C (0.1 to 0.2 °F) increments. Note: Acceptable temperatures range from -10 to 110 °C.
To change the units of temperature measurement, see Section 3.1.3 on page 26.
2.6 Millivolt Measurements The meter can be used to measure absolute millivolts (mV). To display a current millivolt reading, press the pH/mV key. The mV value is displayed with mV in the units field. Absolute millivolts are displayed with 0.1 mV resolution in the range of -2000 to 2000. The millivolt mode is useful when measuring oxidation-reduction potential, performing potentiometric titrations, or preparing calibration curves. Detailed instructions for any Hach electrode are given in the electrode instruction manual. Titration instructions are included in the Hach ORP Electrode Instruction Manual, or in standard analytical chemistry texts.
2.7 Automatic Shut-off Function The meter is equipped with an automatic shut-off feature to prolong battery life. The meter will automatically shut off 15 minutes after the last key press unless the meter is in Calibration mode. If it is in Calibration mode, automatic shutoff will occur four hours after the last key press. Press the I/O/EXIT key after automatic shut-off to restore power to the instrument.
23