Portable pH Meter Instruction Manual: SECTION 1 INTRODUCTION

Pages 13 to 18 of 40, from Portable pH Meter Instruction Manual

The original document is the PDF (597 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Hach SensION1

Page 13

SECTION 1

INTRODUCTION This manual describes the operation and use of the Hach sension™1 Portable pH/mV/Temperature Meter (see Figure 1). This meter features a custom digital LCD display which simultaneously shows temperature and measurement results. This meter has all the features of a simple pH meter plus a millivolt mode, IP67 waterproof design, electrode holder, ergonomic design, and automatic buffer recognition. The electrode holder on the back of the instrument allows the user to store the electrode and meter in one convenient package when not in use. The electrode holder is designed so the user can slip a vinyl electrode cover over the electrode tip, creating a humid environment that allows the electrode membrane to remain hydrated when not in use (see Figure 2 Electrode Holder). The meter is designed to be maintenance-free. If the meter gets dirty, wipe the surface with a damp cloth. Use a cotton-tipped applicator to clean or dry the connectors if they get wet.

1.1 Unpacking the Instrument Remove the instrument and accessories from the shipping container and inspect each item for any damage that may have occurred during shipping. Verify that all items listed on the packing slip are included. If any items are missing or damaged, contact Hach Customer Service, Loveland, Colorado for instructions. Hach’s toll free phone number for customers within the United States is 800-227-4224. For customers outside the United States, contact the Hach office or distributor serving you.

1.1.1

Standard Accessories

•

Batteries - 4 alkaline AA (not rechargeable)

•

Instrument Manual

•

May include electrode and related accessories (covered in the electrode manual).

13

Page 14 read from a scan by OCR; it may hold reading errors

SECTION 1, continued

Figure 1 Hach sens/ON1 Meter ey sension1

Figure 2 Electrode Holder

Page 15

SECTION 1, continued 1.2 Keypad Description Figure 3 illustrates the meter’s keypad. The description and function of each key is given in Table 1. Figure 3 sension1 Meter Keypad

Table 1 Keys and Description Key

Description

Exit/Power On-Off

Turns the instrument on; turns it off from the Reading mode. Acts as a NO or CANCEL key when the question mark icon is flashing. Performs the following and returns to the Reading mode: • Exits the Setup mode • Aborts a calibration • Exits a calibration review

Arrow Keys

Scroll between options in Setup mode. Change the default temperature when a temperature probe is not in use. In Setup mode, back up one step toward the Reading mode.

READ/ENTER Key

Acts as a YES key when the question mark is flashing. Allows user to edit a setup when the setup number is flashing. Initiates a new measurement when the meter has stabilized in the Display Lock Enabled mode.

pH/mV Key

Toggles between pH value and mV value in Reading, Calibration, and Cal Review modes.

Setup Key

Enters Setup mode (from Reading mode only).

Cal Key

Enters Calibration mode (from Reading mode only).

Review Key

Enters Calibration Review mode (from Reading mode only).

15

Page 16

SECTION 1, continued 1.3 Display Fields and Icons The display has two screens. The upper screen displays measurements or standard values, the operation mode in use, slope, sample/default temperature, pH or mV units, error codes, and indicates if the meter reading is stable. The lower screen displays the acrive keys. Figure 4 shows the icons and screens displayed by the meter and Table 2 describes each element. Several icons on the display are not used by the portable pH sension1 meter (indicated by NA in the table), but will be displayed if the power key is held down for several seconds. Figure 4

Display Elements

16

Page 17

SECTION 1, continued Table 2 Main Display Elements Item No.

Description

1

Indicates meter is in Calibration mode. If the ? is flashing, calibration is necessary.

2

Indicates meter is in Calibration Review mode.

3 (NA)

Indicates data is being sent to a printer/computer.

4 (NA)

Indicates recalled data that is currently displayed is being erased.

5

Indicates meter is in Setup mode.

6 (NA)

Indicates all data points are being printed or erased.

7

Numerical field that displays Setup, Sample, and Standard numbers when those words are displayed with the number. If Standard and 1 are displayed, the meter is measuring Standard 1.

8

Flashing ? and CAL indicate calibration is necessary. Also a prompt to press the ENTER or EXIT key.

9

Indicates the meter is measuring a sample (sample number is displayed to the right).

10

Indicates the meter is measuring a standard (standard number is displayed above).

11

Indicates the displayed number is the electrode slope.

12

Numerical field that displays the slope and pH or mV values of standards and samples.

13

Indicates measurement units (pH or mV).

14

When Default is displayed, the meter is using the default temperature value to calculate the temperature correction for the pH value.

15

Temperature units (choice of °C or °F).

16

Indicates the value displayed in small numerical field (item 17) is in millivolts.

17

Numerical field that displays temperature value.

18 (NA)

Indicates meter is using AC power (only displayed when in the docking station).

19

Indicates an inactive key has been pressed and that function is not allowed.

20

Indicates ENTER key is active.

21

Indicates arrow keys are active.

22 (NA)

In Setup mode, it indicates the date is being set.

23

Indicates EXIT key is active.

24

Indicates temperature compensation is being used.

25 (NA)

Indicates numeric part of the keys is active.

26

Display Lock icon. Displayed with item 27.

27

Indicates whether Display Lock setting is On or Off.

28

Faulty probe connection or incorrect probe attached. May also indicate the calibration is questionable. Usually displayed with an error code.

29

Low battery icon. Change batteries as soon as possible.

30

Indicates a meter function problem.

17

Page 18

SECTION 1, continued Table 2 Main Display Elements (Continued) Item No.

Description

31

When on or flashing, Stabilizing... indicates signal from sample is not yet stable. When it disappears, the reading is stable and may be recorded.

32

Used with ? icon. Asks if user wants to store the displayed sample data or the calibration that has been just completed.

33 (NA)

Used with large display to indicate the time is being set.

34 (NA)

Indicates meter is in recall mode and the data displayed is stored data.

1.4 Audible Signals The meter will beep under certain conditions:

• when a non-functional key press is made (one beep)

• when display lock is enabled and measurement stability is reached in reading mode (three beeps)

• any time measurement stability is reached during calibration mode, regardless of the Display Lock setting

• to signal an error condition.

18