Page 31
SECTION 4
TROUBLESHOOTING
4.1 Introduction Correcting problem conditions with the sension™ electrochemical meters is limited to responding to the error message displayed. Other problems must be handled by a Hach technician at a service center. Refer to REPAIR SERVICE on page 38. Do not attempt to service the meter as there are no fieldserviceable parts. Opening the meter case will void the warranty.
4.2 Shorting Test This test detects the meter offset. 1. Turn the meter on. Connect the shorting cap to the BNC connector (see Figure 7). Unplug any electrodes from the 5-pin connector. Select the BNC connector from Setup 1. 2. Change the output to display in mV (press pH/mV to toggle between pH and mV readings). After the meter has stabilized, simultaneously press CAL and ENTER. The display should show 0.0 mV. If it does not, contact Hach Service. Figure 7
Attaching Shorting Cap to The BNC Connector
31
Page 32
SECTION 4, continued 4.3 Error Codes Error codes indicate a functional problem with the meter and/or the electrode. Error code numbers will appear in the temperature field along with other icons. Table 4 describes the possible error codes and some possible solutions to eliminate the cause. Table 4 Error Codes Error Code
Error Name & Display Icons
Description
2
Cal slope error. ERROR and SLOPE will be displayed.
Calibration slope is too high or low. Ensure correct pH buffers are used. Be sure a pH probe is connected to the meter. Disconnect any probe connected to the BNC connector.
3
Cal Std Error.
The value read will not work in the calibration algorithm. Ensure one buffer was not measured twice and that the correct buffers were used. If not, repeat calibration with new buffers.
ERROR, ENTER, CAL and Standard will be displayed.
4.4 Meter Service Request Questionnaire 1. What is the complete lot code of the meter and electrode? 2. On what date was the meter purchased? 3. How long has the meter been in use? 4. What types of samples are being tested? 5. What is the temperature of the samples being tested? 6. How often is the meter being used? 7. How is the meter being stored between uses? 8. If the meter has been in use for a while, what maintenance has been performed? 9. Describe the suspected problem or failure of the meter. 10. Please have your meter, electrode, buffers/standards, and this completed questionnaire near the phone before calling technical support.
32
Page 33
GENERAL INFORMATION
At Hach Company, customer service is an important part of every product we make. With that in mind, we have compiled the following information for your convenience.
33
Page 34
34
Page 35
REPLACEMENT PARTS REQUIRED REAGENTS Quantity Required Per Test
Description
Unit
Cat. No.
Buffer, Powder Pillows pH 4.01, color-coded red ................................................. 1.......... 15/pkg..............22269-95 pH 7.00, color-coded yellow............................................ 1.......... 15/pkg..............22270-95 pH 10.00, color-coded blue ............................................. 1.......... 15/pkg..............22271-95 Buffer Solutions pH 4.01, color-coded red ............................................. 20 mL ... 500 mL..............22834-49 pH 7.00, color-coded yellow........................................ 20 mL ... 500 mL..............22835-49 pH 10.01, color-coded blue ......................................... 20 mL ... 500 mL..............22836-49 pH Electrode Storage Powder Pillows ................................................ 20/pkg..............26573-64 pH Electrode Storage Solution.......................................................... 475 mL..............50301-49
REQUIRED APPARATUS Beaker, poly, 50 mL ............................................................... 1.............. each................1080-41 Electrode, pH combination, w/temp, 5-pin connector ........... 1.............. each..............51910-00 sension™1 Portable pH Meter ............................................... 1............. each..............51700-00 Shorting Cap, BNC ................................................................ 1.............. each..............50071-00 Temperature Probe, 5-pin ....................................................... 1.............. each..............51980-00
OPTIONAL APPARATUS Demineralizer Bottle, 177 mL................................................................ each..............14299-00 Electrode, pH, gel-filled, w/temp, 5-pin connector................................. each..............51935-00 Electrode, pH combination, flat end, 5-pin ............................................. each..............51915-00 Electrode, reference, single junction ....................................................... each..............50220-00 Electrode, reference, double junction...................................................... each..............50225-00 Electrode Stand ....................................................................................... each..............45300-00 Electrode Stand with Electromagnetic Stirrer, 115 VAC ........................ each..............45300-01 Electrode Stand with Electromagnetic Stirrer, 230 VAC ........................ each..............45300-02 Electrode Washer..................................................................................... each..............27047-00 sension™2 Portable pH/ISE Meter ........................................................ each..............51725-11 sension™3 Benchtop pH Meter, 115 V, with electrode .......................... each..............51750-10 sension™3 Benchtop pH Meter, 220 V, with electrode .......................... each..............51750-11 sension™4 Benchtop pH/ISE Meter, 115 V, No. American style Plug . each..............51775-10 sension™4 Benchtop pH/ISE Meter, 230 V, European Style Plug ........ each..............51775-11 Stir Bar, 7/16 x 3/16 in. .............................................................................. each..............45315-00 Thermometer, mercury, -20 to 110 °C .................................................... each..............20959-11
35
Page 36
36
Page 37
HOW TO ORDER By Telephone: 6:30 a.m. to 5:00 p.m. MST Monday through Friday (800) 227-HACH (800-227-4224) By FAX: (970) 669-2932
By Mail: Hach Company P.O. Box 389 Loveland, Colorado 80539-0389 U.S.A. Ordering information by E-mail: orders@hach.com
Information Required •
Hach account number (if available)
•
Billing address
•
Your name and phone number
•
Shipping address
•
Purchase order number
•
Catalog number
•
Brief description or model number
•
Quantity
Technical and Customer Service (U.S.A. only) Hach Technical and Customer Service Department personnel are eager to answer questions about our products and their use. Specialists in analytical methods, they are happy to put their talents to work for you. Call 1-800-227-4224 or E-mail techhelp@hach.com.
International Customers Hach maintains a worldwide network of dealers and distributors. To locate the representative nearest you, send E-mail to intl@hach. com or call (970) 669-3050. In Canada: Hach Sales & Service Canada Ltd.; Manitoba, Canada Telephone: (204) 632-5598; FAX: (204) 694-5134
37
Page 38
REPAIR SERVICE
Authorization must be obtained from Hach Company before sending any items for repair. Please contact the Hach Service Center serving your location.
In the United States:
Hach Company
100 Dayton Ave.
Ames, Iowa 50010
(800) 227-4224 (U.S.A. only) Telephone: (515) 232-2533 FAX: (515) 232-1276
In Canada:
Hach Sales & Service Canada Ltd. 1313 Border Street, Unit 34 Winnipeg, Manitoba
R3H 0X4
(800) 665-7635 (Canada only) Telephone: (204) 632-5598
FAX: (204) 694-5134
E-mail: canada@hach.com
Other locations:
Hach Company World Headquarters, P.O. Box 389
Loveland, Colorado 80539-0389 U.S.A. Telephone: (970) 669-3050
FAX: (970) 669-2932
38
Page 39
WARRANTY
Hach warrants most products against defective materials or workmanship for at least one year from the date of shipment; longer warranties may apply to some items. HACH WARRANTS TO THE ORIGINAL BUYER THAT HACH PRODUCTS WILL CONFORM TO ANY EXPRESS WRITTEN WARRANTY GIVEN BY HACH TO THE BUYER. EXCEPT AS EXPRESSLY SET FORTH IN THE PRECEDING SENTENCE, HACH MAKES NO WARRANTY OF ANY KIND WHATSOEVER WITH RESPECT TO ANY PRODUCTS. HACH EXPRESSLY DISCLAIMS ANY WARRANTIES IMPLIED BY LAW, INCLUDING BUT NOT BINDING TO ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. LIMITATION OF REMEDIES: Hach shall, at its option, replace or repair nonconforming products or refund all amounts paid by the buyer. THIS IS THE EXCLUSIVE REMEDY FOR ANY BREACH OF WARRANTY. LIMITATION OF DAMAGES: IN NO EVENT SHALL HACH BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND FOR BREACH OF ANY WARRANTY, NEGLIGENCE, ON THE BASIS OF STRICT LIABILITY, OR OTHERWISE.
This warranty applies only to Hach products purchased and delivered in the United States. Catalog descriptions, pictures and specifications, although accurate to the best of our knowledge, are not a guarantee or warranty. Hach warrants the meter against defective materials or workmanship for three years from the date of shipment. The Docking station has a one year warranty from the date of shipment.
39
Page 40
HACH COMPANY WORLD HEADQUARTERS P.O. Box 389 Loveland, Colorado 80539-0389 Telephone: (970) 669-3050 FAX: (970) 669-2932
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. - Call toll-free 800-227-4224 Outside the U.S.A. - Contact the HACH office or distributor serving you. On the Worldwide Web - www.hach.com; E-mail - techhelp@hach.com