Chapter 2. Getting Started the BHT-6000 and System Mode
[4.5]
Setting the resume function
SET RESUME 1:ON
2:OFF
Selecting "5: RESUME" on the SET SYSTEM menu calls up the screen shown at left. Reversed is the current setting. 1 ON
: Activates the resume function which resumes whatever display was shown before the power was turned off, when the BHT-6000 is powered up.
2 OFF
: Deactivates the resume function.
Select a desired item by using the numerical keys or F5 and F6 keys. To return to the SET SYSTEM menu, press the C key.
47
Page 63
[4.6]
Setting the communications environments
After the BHT-6000 is initialized, the interface port and communications parameters are set as listed in the default table below. Do not access them unless necessary. Interface port
OPT (Optical interface port)
Communications protocol
BHT-protocol Defaults
Communications parameters for the optical interface port TRANSMIT SPEED
9600 bps
PULSE WIDTH (Output pulse width of IR beam)
1.63 µs
PROTOCOL (Protocol options)
SERIAL No. ON: Adds serial numbers to data blocks. H. PARITY ON:
Adds a horizontal parity.
LINKUP TIME:
30 seconds
FIELD SPACE:
Ignore
Listed below are the default communications parameters for the direct-connect interface Defaults
Communications parameters for the direct-connect interface port TRANSMIT SPEED
19200 bps
PARITY BIT
None
DATA BIT (Character length)
8 bits
STOP BIT
1 bit
PROTOCOL (Protocol options)
SERIAL No. ON: Adds serial numbers to data blocks. H. PARITY ON:
Adds a horizontal parity.
LINKUP TIME:
30 seconds
FIELD SPACE:
Ignore
48
Page 64
Chapter 2. Getting Started the BHT-6000 and System Mode
SET COM 1:OPTICAL 2:CONNECTOR 3:COM PORT 4:PROTOCOL TYPE
Selecting the "6: COM" on the SET SYSTEM menu calls up the screen shown at left. 1 OPTICAL:
Switches to the communications parameters setting screen for the optical interface.
2 CONNECTOR:
Switches to the communications parameters setting screen for the direct-connect interface.
3 COM PORT:
Switches to the interface port setting screen.
4 PROTOCOL TYPE:
Switches to the communications protocol type setting screen.
Select a desired screen by using the numerical keys or F5 and F6 keys, and then press the ENT key. To return to the SET SYSTEM menu, press the C key.
[4.6-1]
Setting the communications parameters for the optical interface
SET OPTICAL 1:TRANSMIT SPEED 2:PULSE WIDTH 3:PROTOCOL
Selecting "1:OPTICAL" on the SET COM menu calls up the screen shown at left. 1 TRANSMIT SPEED:
Switches to the transmission speed screen.
2 PULSE WIDTH:
Switches to the IR beam output pulse width screen.
3 PROTOCOL:
Switches to the communications protocol option menu.
Select a desired screen by using the numerical keys or F5 and F6 keys, and then press the ENT key. To return to the SET COM menu, press the C key.
49
Page 65
(1)
Transmission speed screen
SET SPEED < OPTICAL > 1: 2400 4: 38400 2: 9600 5: 57600 3:19200 6:115200
Selecting "1:TRANSMIT SPEED" on the SET OPTICAL screen calls up the screen shown at left. Select the desired transmission speed by using the numerical keys or F5 and F6 keys, and then press the ENT key. To return to the SET OPTICAL screen, press the C key.
(2)
IR beam output pulse width screen
SET PULSE WIDTH < OPTICAL >
Selecting the "2:PULSE WIDTH" on the SET OPTICAL screen calls up the screen shown at left.
1:1.63us 2:3/16 Bit Time
1 1.63us:
1.63 µs
2 3/16 Bit Time:
3/16 bit time
Select the desired pulse width by using the numerical keys or F5 and F6 keys, and then press the ENT key. To return to the SET OPTICAL screen, press the C key.
(3)
Communications protocol option menu
SET PROTOCOL < OPTICAL > 1:SERIAL No. 2:H.PARITY 3:LINKUP TIME 4:FIELD SPACE
Selecting "3: PROTOCOL" on the SET OPTICAL screen calls up the screen shown at left. 1 SERIAL No.:
Switches to the serial number ON/OFF screen.
2 H. PARITY:
Switches to the horizontal parity ON/OFF screen.
3 LINKUP TIME:
Switches to the timeout length screen.
4 FIELD SPACE:
Switches to the space code handling screen.
Select a desired screen by using the numerical keys or F5 and F6 keys, and then press the ENT key. To return to the SET OPTICAL screen, press the C key.
NOTE
If the BHT-Ir protocol has been selected, the serial number and horizontal parity settings will be ignored.
50
Page 66
Chapter 2. Getting Started the BHT-6000 and System Mode
SET SERIAL No. < OPTICAL > 1:ON
2:OFF
Selecting the "1:SERIAL No." on the SET PROTOCOL menu calls up the screen shown at left. Select whether or not the system should add serial numbers to data blocks, by using the numerical keys or F5 and F6 keys. Then press the ENT key. To return to the SET PROTOCOL menu, press the C key.
SET H.PARITY < OPTICAL > 1:ON
2:OFF
Selecting the "2:H.PARITY" on the SET PROTOCOL menu calls up the screen shown at left. Select whether or not the system should add a horizontal parity, by using the numerical keys or F5 and F6 keys. Then press the ENT key. To return to the SET PROTOCOL menu, press the C key.
SET LINKUP TIME < OPTICAL > 1:30 4:120 2:60 5:None 3:90
Selecting the "3:LINKUP TIME" on the SET PROTOCOL menu calls up the screen shown at left. Select the timeout length (in seconds) to be applied when a link is to be established, by using the numerical keys or F5 and F6 keys. Then press the ENT key. To return to the SET PROTOCOL menu, press the C key.
SET FIELD SPACE < OPTICAL > 1:Ignore 2:Data
Selecting the "4:FIELD SPACE" on the SET PROTOCOL menu calls up the screen shown at left. Select whether space codes in the tail of a data field should be ignored or handled as data, by using the numerical keys or F5 and F6 keys. Then press the ENT key. To return to the SET PROTOCOL menu, press the C key.
51
Page 67
[4.6-2]
Setting the communications parameters for the direct-connect interface
SET CONNECTOR 1:TRANSMIT SPEED 2:PARITY BIT 3:DATA BIT 4:STOP BIT 5:PROTOCOL
Selecting "2:CONNECTOR" on the SET COM menu calls up the screen shown at left. 1 TRANSMIT SPEED:
Switches to the transmission speed screen.
2 PARITY BIT:
Switches to the parity setting screen.
3 DATA BIT:
Switches to the character length screen.
4 STOP BIT:
Switches to the stop bit length screen.
5 PROTOCOL:
Switches to the communications protocol option menu.
Select a desired screen by using the numerical keys or F5 and F6 keys, and then press the ENT key. To return to the SET COM menu, press the C key.
NOTE
If the BHT-Ir protocol has been selected, the parity bit, character length, and stop bit length settings will be ignored.
(1) Transmission speed screen
SET SPEED < CONNECTOR > 1:1200 4: 9600 2:2400 5:19200 3:4800 6:38400
Selecting "1:TRANSMIT SPEED" on the SET CONNECTOR screen calls up the screen shown at left. Select the desired transmission speed by using the numerical keys or F5 and F6 keys, and then press the ENT key. To return to the SET CONNECTOR screen, press the C key.
(2) Parity setting screen
SET PARITY BIT < CONNECTOR > 1:None 2:Odd 3:Even
Selecting "2:PARITY BIT" on the SET CONNECTOR screen calls up the screen shown at left. Select the parity: none, odd or even, by using the numerical keys or F5 and F6 keys, and then press the ENT key. To return to the SET CONNECTOR screen, press the C key.
52
Page 68
Chapter 2. Getting Started the BHT-6000 and System Mode
(3)
Character length screen
SET DATA BIT < CONNECTOR >
Selecting "3:DATA BIT" on the SET CONNECTOR screen calls up the screen shown at left.
1:7 bits 2:8 bits
Select the desired character length by using the numerical keys or F5 and F6 keys, and then press the ENT key. To return to the SET CONNECTOR screen, press the C key.
(4)
Stop bit length screen
SET STOP BIT < CONNECTOR >
Selecting "4:STOP BIT" on the SET CONNECTOR screen calls up the screen shown at left.
1:1 bit 2:2 bits
Select the desired stop bit length by using the numerical keys or F5 and F6 keys, and then press the ENT key. To return to the SET CONNECTOR screen, press the C key.
(5)
Communications protocol option menu
SET PROTOCOL < CONNECTOR > 1:SERIAL No. 2:H.PARITY 3:LINKUP TIME 4:FIELD SPACE
Selecting "5:PROTOCOL" on the SET CONNECTOR screen calls up the screen shown at left. 1 SERIAL No.:
Switches to the serial number ON/OFF screen.
2 H. PARITY:
Switches to the horizontal parity ON/OFF screen.
3 LINKUP TIME:
Switches to the timeout length screen.
4 FIELD SPACE:
Switches to the space code handling screen.
Select a desired screen by using the numerical keys or F5 and F6 keys, and then press the ENT key. To return to the SET CONNECTOR screen, press the C key.
NOTE
If the BHT-Ir protocol has been selected, the serial number and horizontal parity settings will be ignored.
53
Page 69
SET SERIAL No. < CONNECTOR > 1:ON
2:OFF
Selecting the "1:SERIAL No." on the SET PROTOCOL menu calls up the screen shown at left. Select whether or not the system should add serial numbers to data blocks, by using the numerical keys or F5 and F6 keys. Then press the ENT key. To return to the SET PROTOCOL menu, press the C key.
SET H.PARITY < CONNECTOR > 1:ON
2:OFF
Selecting the "2:H.PARITY" on the SET PROTOCOL menu calls up the screen shown at left. Select whether or not the system should add a horizontal parity, by using the numerical keys or F5 and F6 keys. Then press the ENT key. To return to the SET PROTOCOL menu, press the C key.
SET LINKUP TIME < CONNECTOR > 1:30 4:120 2:60 5:None 3:90
Selecting the "3:LINKUP TIME" on the SET PROTOCOL menu calls up the screen shown at left. Select the timeout length (in seconds) to be applied when a link is to be established, by using the numerical keys or F5 and F6 keys. Then press the ENT key. To return to the SET PROTOCOL menu, press the C key.
SET FIELD SPACE < CONNECTOR > 1:Ignore 2:Data
Selecting the "4:FIELD SPACE" on the SET PROTOCOL menu calls up the screen shown at left. Select whether space codes in the tail of a data field should be ignored or handled as data, by using the numerical keys or F5 and F6 keys. Then press the ENT key. To return to the SET PROTOCOL menu, press the C key.
54
Page 70
Chapter 2. Getting Started the BHT-6000 and System Mode
[4.6-3]
Setting the interface port
SET COM PORT 1:BASIC OPT IFC 2:SYSTEM MODE OPT IFC
Selecting the "3:COM PORT" on the SET COM menu calls up the screen shown at left. Reversed is the current setting. 1 BASIC:
Selects the optical or directconnect interface port to be used for user programs written in BHT-BASIC 3.0 (OPEN "COM:" or XFILE statement).
2 SYSTEM MODE:
Selects the optical or directconnect interface port to be used for downloading or uploading files in System Mode.
Select a desired item by using the numerical keys or F5 and F6 keys, and then select a desired setting by using the F7 and F8 keys. To return to the SET COM menu, press the C key.
55
Page 71
[4.6-4]
Setting the communications protocol type
PROTOCOL TYPE 1:BHT Protocol 2:BHT IrProtocol
Selecting the "4:PROTOCOL TYPE" on the SET COM menu calls up the screen shown at left. Reversed is the current setting. 1 BHT Protocol:
Selects the BHT-protocol for downloading or uploading files in System Mode or for the execution of XFILE statement in BHT-BASIC 3.0. This protocol should be selected for file transmission with Transfer Utility.
2 BHT IrProtocol:
Selects the BHT-Ir protocol for downloading or uploading files in System Mode or for the execution of XFILE statement in BHT-BASIC 3.0. This protocol should be selected for file transmission with Ir-Transfer Utility C.
Select a desired item by using the numerical keys or F5 and F6 keys, and then press the ENT key. To return to the SET COM menu, press the C key.
SET ID
00001 >> _
Selecting the "2:BHT IrProtocol" on the PROTOCOL TYPE screen calls up the screen shown at left. Enter the ID number of the BHT-6000 by using the numerical keys, and then press the ENT key. If you do not need to modify the current setting, press the ENT key only.
SET ID
NOTE 00001 >> 65535_
An ID number should be five-digit decimal character string. The setting range is from 00001 to 65535. If the entered value is less than five digits, the ENT key does not become operable.
If you make a wrong entry, press the BS key to delete it and then enter the correct data. To return to the SET COM menu, press the C key.