RAPID CONFIGURATION
1
1 RAPID CONFIGURATION STEP 1 – ASSEMBLE THE READER The first step to perform is to assemble the accessories that make up the Matrix 400™ reader. The lens and either an internal or an external illuminator must be used. This procedure shows an internal illuminator.
CAUTION
Matrix 400™ must be disconnected from the power supply during this procedure.
1. In a dust-free environment, remove the Matrix 400™ Lens Cover by unscrewing it.
CAUTION
Do not touch the sensor aperture, lens glass or lens cover glass. These areas must be kept clean. Avoid any abrasive substances that might damage these surfaces during cleaning.
2. Remove the sensor protection label by pulling it off of the base. 3. Mount the lens by screwing it tightly onto the base. If the Locking Knobs on the lens are obstructed because they are aligned behind an illuminator spacer base and illuminator spacer, insert the Lens Spacer between the Matrix 400™ body and the C-Mount lens so that the Locking Knobs will be unobstructed. It is strongly recommended to apply a lens locking sticker (provided with the lens), for applications where Matrix 400™ is subjected to vibration. NOTE 4. If using an internal illuminator: a. Mount the four internal illuminator spacers onto the illuminator spacer bases provided on the Matrix 400™ body. b. Align and mount the Illuminator tightly onto the spacers using the four screws and washers provided in the illuminator package. The spacers are positioned asymmetrically to avoid incorrect alignment. 5. To keep dust and dirt off of the lens during mounting, temporarily replace the lens cover. Lens Spacer (optional)
Illuminator Spacer Base
Locking Knobs
Internal Illuminator
Lens Cover
Illuminator Spacers Protective Film (remove before calibration)
Figure 1 – Assembling Matrix 400™ Accessories
1
Page 12
MATRIX 400™ REFERENCE MANUAL
1
Required Accessories The following table shows the correct lens/illuminator combinations to be used for Matrix 400™ imager assembly. Lenses
Internal Illuminators
93ACC1793
LNS-1006
6 mm C-Mount Lens (only for Matrix 400 600-0x0 models)
93A401020 93A401022
LT-002 LT-004
Red Wide Angle White Wide Angle
93ACC1794
LNS-1109
9 mm C-Mount Lens
93ACC1795
LNS-1112
12.5 mm C-Mount Lens
93ACC1796
LNS-1116
16 mm C-Mount Lens
93ACC1797
LNS-1125
25 mm C-Mount Lens
93ACC1798 93ACC1799
LNS-1135 LNS-1150
35 mm C-Mount Lens 50 mm C-Mount Lens
93A401020 93A401022 93A401020 93A401022 93A401019 93A401021 93A401019 93A401021 93A401024 93A401024
LT-002 LT-004 LT-002 LT-004 LT-001 LT-003 LT-001 LT-003 LT-006 LT-006
Red Wide Angle White Wide Angle Red Wide Angle White Wide Angle Red Narrow Angle White Narrow Angle Red Narrow Angle White Narrow Angle Red Super Narrow Angle Red Super Narrow Angle
STEP 2 – CONNECT THE SYSTEM To connect the system in a Stand Alone configuration, you need the hardware indicated in Figure 2. In this layout the data is transmitted to the Host on the main serial interface. Data can also be transmitted on the RS232 auxiliary interface independently from the main interface selection. When One Shot or Phase Mode Operating mode is used, the reader is activated by an External Trigger (photoelectric sensor) when the object enters its reading zone.
PG 6000
CBX Main Interface
CAB-MS01
Matrix 400™ Host
P.S.*
I/O, AUX
* External Trigger or Presence Sensor (for One Shot or Phase Mode)
Figure 2 – Matrix 400™ in Stand Alone Layout
2
Page 13
RAPID CONFIGURATION
1
CBX100/CBX500 Pinout for Matrix 400™ The table below gives the pinout of the CBX100/CBX500 terminal block connectors. Use this pinout when the Matrix 400™ reader is connected by means of the CBX100/CBX500:
Vdc GND Earth
+V I1A I1B -V +V I2A I2B -V Shield
CBX100/500 Terminal Block Connectors Input Power Outputs Power Supply Input Voltage + +V Power Source - Outputs Power Supply Input Voltage -V Power Reference - Outputs Protection Earth Ground O1+ Output 1 + O1Output 1 O2+ Output 2 + Inputs Power Source – External Trigger O2Output 2 External Trigger A (polarity insensitive) Auxiliary Interface External Trigger B (polarity insensitive) TX Auxiliary Interface TX Power Reference – External Trigger RX Auxiliary Interface RX Power Source – Inputs SGND Auxiliary Interface Reference Input 2 A (polarity insensitive) ID-NET™ Input 2 B (polarity insensitive) REF Network Reference Power Reference – Inputs ID+ ID-NET™ network + IDID-NET™ network Shield Network Cable Shield Main Interface RS232 RS485 Full-Duplex RS485 Half-Duplex TX TX+ RTX+ RTS TXRTXRX *RX+ CTS *RXSGND SGND SGND
* Do not leave floating, see par. 4.2.2 for connection details.
CAUTION
Do not connect GND, SGND and REF to different (external) ground references. GND, SGND and REF are internally connected through filtering circuitry which can be permanently damaged if subjected to voltage drops over 0.8 Vdc.
3
Page 14
MATRIX 400™ REFERENCE MANUAL
1
19-pin Connector Pinout for Matrix 400™ The table below gives the pinout of the 19-pin M16 male connector for connection to the power supply and input/output signals. Use this pinout when the Matrix 400™ reader is connected by means of the 19-pin connector:
Figure 3 - 19-pin M16 Male Connector
19-pin M16 male connector pinout Pin A L
Name Vdc GND
K
CHASSIS
B C D E H F G I S O R P
I1A I1B I2A I2B O1+ O1O2+ O2RX TX ID+ ID-
Pin
Name
M U N T
MAIN INTERFACE (SW SELECTABLE)
Function Power supply input voltage + Power supply input voltage Cable shield internally connected by capacitor to the chassis External Trigger A (polarity insensitive) External Trigger B (polarity insensitive) Input 2 A (polarity insensitive) Input 2 B (polarity insensitive) Output 1 + Output 1 Output 2 + Output 2 Auxiliary RS232 RX Auxiliary RS232 TX ID-NET™ network + ID-NET™ network RS485 RS485 RS232 Full-Duplex Half-Duplex TX TX+ RTX+ RX *RX+ RTS TXRTXCTS *RX-
* Do not leave floating, see par. 5.4.2 for connection details.
4
Page 15
RAPID CONFIGURATION
1
STEP 3 – MOUNT AND POSITION THE READER 1. To mount the Matrix 400™, use the mounting brackets to obtain the most suitable position for the reader. Two of the most common mounting configurations are shown in the figures below. Other mounting solutions are provided in par. 3.3.
Pitch
Tilt
Figure 4 –Positioning with Mounting Bracket (Back)
Pitch
Skew
Figure 5 –Positioning with Mounting Bracket (Side)
2. When mounting the Matrix 400™ take into consideration these three ideal label position angles: Pitch or Skew 10° to 20° and Tilt 0°, although the reader can read a code at any tilt angle. P
T
S
Minimize
Assure at least 10°
Minimize
Figure 6 – Pitch, Skew and Tilt Angles
5
Page 16
MATRIX 400™ REFERENCE MANUAL
1
3. Refer to the Reading Features table in chp. 7 for FOV calculation and minimum distance requirements according to the base/lens combination used for your application.
NOTE
Rapid Configuration of the Matrix 400™ reader can be made either through the X-PRESS™ interface (steps 4-6) which requires no PC connection, or by using the VisiSet™ Setup Wizard (steps 7-8). Select the procedure according to your needs.
STEP 4 – FOCUS THE READER Matrix 400™ provides a built-in tool called Blue Diamonds™ to aid focusing the reader. The Blue Diamonds™ are accessed through the X-PRESS™ Interface. 1. Remove the lens cover in order to focus the reader. 2. Prepare the correct accessory lens for your application: a. Loosen the two Locking Knobs on the lens. b. Adjust the Focus ring to the "Far position" and the Diaphragm ring to the "F4" 1 number setting which is the preferred setting for installation. 3. Power the reader on. During the reader startup (reset or restart phase), all the LEDs blink for one second. On the connector side of the reader near the cable, the “POWER ON” LED (blue) indicates the reader is correctly powered. 4. Enter the Focus function by pressing and holding the X-PRESS™ push button until the Focus LED is on. 5. Release the button to enter the Focus function. The Blue Diamonds™ turn on. READY
The procedure is as follows: a. Adjust the Focus ring towards the "Near position" until the Blue Diamonds™ are perfectly in focus, see Figure 8. At long focal distances a "skew" angle may cause a noticeable difference in focus between the two diamonds, in this case select the best possible focus (both diamonds slightly out of focus). Tighten the Focus Locking Knob. b. Tighten the Diaphragm Locking Knob.
green
LEARN
GOOD green
SETUP
TRIGGER yellow
FOCUS
COM yellow
TEST
STATUS red Figure 7 – X-PRESS™ Interface: Focus Function
If necessary you can use the Fine Focusing Tool in the VisiSet™ Setup Wizard for fine focusing. See Step 8. NOTE
1
For far reading distances, the Diaphragm ring can be set to values between F2 and F4 to increase image lighting and Blue Diamond™ visibility.
6
Page 17
RAPID CONFIGURATION
1 Blue Diamond™ in focus FOV
Your reader can have one of the above types of Blue Diamonds™.
Figure 8 – Focus Function Using Blue Diamonds™
6. Exit the Focus function by pressing the X-PRESS™ push button once. The Blue Diamonds™ turn off. 7. Replace the lens cover, screwing it tightly to the base.
STEP 5 – CALIBRATE IMAGE DENSITY In order to function correctly to the fullest extent of its capabilities, Matrix 400™ must acquire information regarding image density or PPI (pixels per inch). This calibration takes place through the X-PRESS™ Interface and the Grade A Barcode Test Chart included in the package. This procedure is necessary for the first time installation, if the lens type is changed or if the focal distance is changed.
Locate 1. Enter the Focus function by pressing and holding the X-PRESS™ push button until the Focus LED is on. 2. Release the button to enter the Focus function. The Blue Diamonds™ turn on. 3. From the Grade A Barcode Test Chart, select the longest code whose length fits between the two Blue Diamonds™. Rotate the code 90 degrees and position the code at the center of the FOV (equidistant from the Blue Diamonds™).
READY green
LEARN
GOOD green
SETUP
TRIGGER yellow
FOCUS
COM yellow
TEST
STATUS red Figure 9 – X-PRESS™ Interface: Locate Function
4. Exit the Focus function by pressing the X-PRESS™ push button once. The Blue Diamonds™ turn off.
7
Page 18
MATRIX 400™ REFERENCE MANUAL
1
Setup 5. Enter the Setup function by pressing and holding the X-PRESS™ push button until the Setup LED is on. 6. Release the button to enter the Setup function. The Setup LED will blink until the procedure is completed. The Setup procedure ends when the Image Acquisition parameters are successfully saved in the reader memory, the Setup LED will remain on continuously and Matrix 400™ emits 3 high pitched beeps. If the calibration cannot be reached after a timeout of about 5 (five) seconds Matrix 400™ will exit without saving the parameters to memory, the Setup LED will not remain on continuously but it will just stop blinking. In this case Matrix 400™ emits a long low pitched beep.
READY green
LEARN
GOOD green
SETUP
TRIGGER yellow
FOCUS
COM yellow
TEST
STATUS red Figure 10 – X-PRESS™ Interface: Setup Function
7. Exit the Setup function by pressing the X-PRESS™ push button once.
Learn 8. Enter the Learn function by pressing and holding the X-PRESS™ push button until the Learn LED is on. 9. Release the button to enter the Learn function. The Learn LED will blink until the procedure is completed. The Learn procedure ends when the Image Density value is successfully saved in the reader memory, the Learn LED will remain on continuously, the Green Spot is activated and Matrix 400™ emits 3 high pitched beeps. If the calibration cannot be reached after a timeout of about 3 (three) minutes Matrix 400™ will exit without saving the parameters to memory, the Learn LED will not remain on continuously but it will just stop blinking. In this case Matrix 400™ emits a long low pitched beep. 10. Exit the Learn function by pressing the X-PRESS™ push button once.
8
READY green
LEARN
GOOD green
SETUP
TRIGGER yellow
FOCUS
COM yellow
TEST
STATUS red Figure 11 – X-PRESS™ Interface: Learn Function
Page 19
RAPID CONFIGURATION
1
STEP 6 – X-PRESS™ CONFIGURATION Once Matrix 400™ has calibrated image density, you can configure it for optimal code reading relative to your application. This configuration can be performed either through the XPRESS™ Interface or the VisiSet™ configuration program.
Locate 1. Enter the Focus function by pressing and holding the X-PRESS™ push button until the Focus LED is on. 2. Release the button to enter the Focus function. The Blue Diamonds™ turn on. 3. Select a code from your application. Position the code at the center of the FOV (equidistant from the Blue Diamonds™). 4. Exit the Focus function by pressing the XPRESS™ push button once. The Blue Diamonds™ turn off.
READY green
LEARN
GOOD green
SETUP
TRIGGER yellow
FOCUS
COM yellow
TEST
STATUS red Figure 12 – X-PRESS™ Interface: Locate Function
Setup 5. Enter the Setup function by pressing and holding the X-PRESS™ push button until the Setup LED is on. 6. Release the button to enter the Setup function. The Setup LED will blink until the procedure is completed. The Setup procedure ends when the Image Acquisition parameters are successfully saved in the reader memory, the Setup LED will remain on continuously and Matrix 400™ emits 3 high pitched beeps. If the calibration cannot be reached after a timeout of about 5 (five) seconds Matrix 400™ will exit without saving the parameters to memory, the Setup LED will not remain on continuously but it will just stop blinking. In this case Matrix 400™ emits a long low pitched beep.
READY green
LEARN
GOOD green
SETUP
TRIGGER yellow
FOCUS
COM yellow
TEST
STATUS red Figure 13 – X-PRESS™ Interface: Setup Function
7. Exit the Setup function by pressing the XPRESS™ push button once.
9
Page 20
MATRIX 400™ REFERENCE MANUAL
1
Learn 8. Enter the Learn function by pressing and holding the X-PRESS™ push button until the Learn LED is on. 9. Release the button to enter the Learn function. The Learn LED will blink until the procedure is completed. The Learn procedure ends when the Image Processing and Decoding parameters are successfully saved in the reader memory, the Learn LED will remain on continuously, the Green Spot is activated and Matrix 400™ emits 3 high pitched beeps 2. If the autolearning cannot be reached after a timeout of about 3 (three) minutes Matrix 400™ will exit without saving the parameters to memory, the Learn LED will not remain on continuously but it will just stop blinking. In this case Matrix 400™ emits a long low pitched beep.
READY green
LEARN
GOOD green
SETUP
TRIGGER yellow
FOCUS
COM yellow
TEST
STATUS red Figure 14 – X-PRESS™ Interface: Learn Function
10. Exit the Learn function by pressing the XPRESS™ push button once.
If you have used this procedure to configure Matrix 400™ go to step 9. NOTE
Reset Reader to Factory Default (Optional) If it ever becomes necessary to reset the reader to the factory default values, you can perform this procedure by holding the X-PRESS™ push button pressed while powering up the reader. You must keep the X-PRESS™ push button pressed until the power up sequence is completed (several seconds) and all LEDs blink simultaneously 3 times. All LEDs remain on for about 1 second, then off for one second, the Configuration and Environmental parameters are reset, and the status LED remains on. If connected through a CBX500 with display module, the message "Default Set" is shown on the display.
2
The Learn procedure will not recognize Pharmacode symbologies.
10