Page 28
DS4600
2
RS485 Full-Duplex Interface The RS485 full-duplex (5 wires + shield) interface is used for non-polled communication protocols in point-to-point connections over longer distances (max 1200 m / 3940 ft) than those acceptable for RS232 communications or in electrically noisy environments. The following pins of the 25-pin connector are used for RS485 full-duplex communications: Pin 2 4 3 5 7
Name TX485+ TX485RX485+ RX485SGND Main
Function RS485 transmitted data + RS485 transmitted data RS485 received data + RS485 received data signal ground main interface
DS4600
USER INTERFACE 2
TX485+
4
TX485-
3
RX485+
5
RX485-
7
SGND Main
RX485
TX485
RS485REF Earth Ground Figure 15 - RS485 full-duplex connections
16
Page 29
INSTALLATION
2
RS485 Half-Duplex Interface The RS485 half-duplex (3 wires + shield) interface is used for polled communication protocols. It can be used for Multidrop connections in a master/slave layout or with a Datalogic Multiplexer, (see par. 2.6.4 and 2.6.5). The following pins of the 25-pin connector are used for RS485 half-duplex communications: Pin 2 4 7
Name RTX485+ RTX485SGND Main
Function RS485 transmitted/received data + RS485 transmitted/received data signal ground main interface
DS4600
MULTIPLEXER 2
RTX485+
4
RTX485-
RTX485+ RTX485-
7 SGND Main
RS485REF
Figure 16 - RS485 half-duplex connections
17
Page 30
DS4600
2
The figure below shows a multidrop configuration with DS4600 scanners connected to a Multiplexer. max. 2 m.
1
DS4600
2
#x
4
(up to 31)
7
DS4600
2
#1
4
1
CHASSIS
120 Ohm
7 DS4600
2
#0
4
1 max. 1200 m.
RTX485+ RTX485SGND Main
7 three wires + shield RTX485+
MULTIPLEXER
RTX485RS485REF SHIELD
120 Ohm Figure 17 - DS4600 multidrop connection to a Multiplexer
18
Page 31
INSTALLATION
2
20 mA Current Loop (INT-24 Accessory Only) When the INT-24 accessory board is installed, the DS4600 is equipped with a 20 mA active current loop interface. The following pins of the 25-pin connector are used for the connections: Pin 4 5 7
Name CLOUTCLINSGND Main
Function current loop output current loop input signal ground main interface
DS4300/DS4600
USER INTERFACE
+ I = 20 mA
4
C.L. OUT -
7
SGND Main
5
C.L. IN -
+ I = 20 mA
GND
MAX. 300 METERS
Figure 18 - 20 mA active C.L. connections
For 20 mA passive current loop interface connections, contact your local Datalogic representative.
19
Page 32
DS4600
2
2.4.3
Auxiliary RS232 Interface
The auxiliary serial interface is used exclusively for RS232 point-to-point connections. The parameters relative to the auxiliary interface (baud rate, data bits, etc.) as well as particular communication modes such as Local Echo can be defined using the WinHost utility program or Host Mode programming. For more details refer to paragraph 2.6 and to the section "Auxiliary Interface Menu" in the WinHost Help On Line. The following pins of the 25-pin connector are used to connect the RS232 auxiliary interface: Pin 20 21 23
Name RXAUX TXAUX SGND Aux
Function auxiliary RS232 received data auxiliary RS232 transmitted data signal ground auxiliary interface
DS4600
USER INTERFACE 20
RXAUX TXD
21 TXAUX RXD 23
SGND Aux SGND
Figure 19 - RS232 auxiliary interface connections
When the auxiliary interface is permanently connected as part of the system cabling, it is recommended to connect the cable shield to earth ground.
2.4.4
Inputs
There is an input available on the DS4600 scanner relative to the External Trigger. There are also 2 general purpose inputs: IN1 can be used to store the code verifier (see "Store verifier HW" in the WinHost Help On Line). IN2 can be used as the Reading Field Control signal which determines either the Code Reading Condition or the Code Resolution parameter for each reading phase. Refer to the WinHost Help On Line.
20
Page 33
INSTALLATION
2
The pinouts are indicated below: Pin 18 19 6 10 14 15
Name EXT TRIG+ EXT TRIGIN1+ IN1IN2+ IN2-
Function external trigger + external trigger input 1 + input 1 input 2 + input 2 -
The EXT TRIG inputs are used in the On-Line Operating mode and tells the scanner to scan for a code. The active state of this input is selected in software. Refer to the WinHost Help On Line. The yellow led, (Figure A, 7), is on when the External Trigger signals the active reading phase. These inputs are optocoupled and can be driven by both an NPN or PNP type command. The connections are indicated in the following diagrams: Vext 30 Vdc max.
DS4600
USER INTERFACE
+ 5V 18/6/14
+ V
19/10/15
Signal
Figure 20 - Input NPN command using external power
DS4600
USER INTERFACE 9
VS
18/6/14
+
19/10/15
-
25
GND
V
+ 5V
Signal Ground
Figure 21 - Input NPN command using DS4600 power
21
Page 34
DS4600
2 Vext 30 Vdc max.
USER INTERFACE
DS4600
V + 5V 18/6/14
+
19/10/15
-
Signal
Ground
Figure 22 - Input PNP command using external power
DS4600
USER INTERFACE 9
VS
18/6/14
+
19/10/15
-
25
GND
+ 5V
V Signal
Figure 23 - Input PNP command using DS4600 power
The electrical features of the inputs are: Maximum voltage = 30 Vdc Maximum current = 25 mA. An anti-disturbance hardware filter is implemented on the External Trigger input (< 5 milliseconds delay). An additional 15 ms (typical) delay can be implemented through a dedicated software parameter (refer to WinHost Help On Line).
22
Page 35
INSTALLATION
2.4.5
2
Outputs
The following pins are present on the 25-pin connector of the scanner: Pin 8 22 11 12
Name OUT1+ OUT1OUT2+ OUT2-
Function output 1 + output 1 output 2 + output 2 -
The meaning of the two outputs OUT1 and OUT2 can be defined by the user (No Read, Right or Wrong). Refer to WinHost Help On Line. By default, OUT1 is associated with the No Read event, which activates when the code signaled by the External Trigger is not decoded. OUT2 is associated with the Right event, which activates when the code is decoded correctly.
DS4600
USER INTERFACE Vext 40 Vdc max
8/11
+
22/12
-
Figure 24 - Open collector output connection
These outputs are both level or pulse configurable.
23
Page 36
DS4600
2
2.5
POSITIONING
The DS4600 scanner is able to decode moving barcode labels at a variety of angles, however significant angular distortion may degrade reading performance. When mounting the DS4600 take into consideration these three ideal label position angles: Pitch 0°, Skew 10° to 30° and Tilt 0°. Follow the suggestions below for the best orientation: The Pitch angle is represented by the value P in Figure 25. Position the reader in order to minimize the Pitch angle. P
Figure 25 - Pitch angle
The Skew angle is represented by the value S in Figure 26. Position the reader to assure at least 10° for the Skew angle. This avoids the direct reflection of the laser light emitted by the DS4600. For raster models, this angle refers to the most inclined or external raster line, so that all other raster lines assure more than 10° Skew.
S
Figure 26 - Skew Angle
24