Page 21
INSTALLATION
2.3.1
2
Power Supply
Power can be supplied to the scanner through the pins provided on the 25- or 9-pin connector used for communication with the host (Figure 6): DS2100 9-pin
DS2100 25-pin
7
13
USER INTERFACE VS
VS
V+ (10 - 30 Vdc) 5 or
25
GND
1
CHASSIS
GND 1 CHASSIS
VGND CHASSIS
Earth Ground
Figure 6 - Power supply connections
The power must be between 10 and 30 Vdc only. It is recommended to connect pin 1 (CHASSIS) to a common earth ground.
2.3.2
Main Serial Interface
The signals relative to the following serial interface types are available on the input/output connector of DS2100 depending on the DS2100 model (see par. 1.3). A passive 20-mA Current Loop interface is available if the optional INT-22 accessory is installed. This accessory interface replaces the main serial interface of the DS2100 model. The INT-22 accessory board is not available for DS2100-X3X0 models. If the interface type is not compatible with the current communication handshaking, then the system forces the handshake to none. The parameters relative to the interface selected (baud rate, data bits, etc.) can be defined using the WinHost utility program or "Host Mode" programming procedure through ESC sequences installed from the diskette. Details regarding the connections and use of the interfaces are given in the next paragraphs.
9
Page 22
DS2100
2
RS232 Interface The serial interface is used in this case for point-to-point connections; it handles communication with the host computer and allows both transmission of code data and the programming of the scanner. This is the default setting. The following pins are used for RS232 interface connection: Connector 9-Pin
25-Pin
2 3 6 4 5
2 3 4 5 7
Name
Function
TX232 RX232 RTS232 CTS232 GND/SGND transmitted data received data request to send clear to send signal ground
It is always advisable to use shielded cables. The overall maximum cable length must be less than 15 m (49.2 ft).
DS2100 25-pin
DS2100 9-pin TX232 RX232 RTS232 CTS232 SGND
USER INTERFACE
2
2
TX232
3
3
RX232
TXD
4
RTS232
DCD
5
CTS232
DTR
7
SGND
6 4 5 or
RXD
SGND Earth Ground
Figure 7 – RS232 main interface connections using hardware handshaking
The RTS232 and CTS232 signals control data transmission and synchronize the connected devices.
10
Page 23
INSTALLATION
2 START OF TRANSMISSION
END OF TRANSMISSION
+V RTS -V
DATA TRANSMISSION
DATA TRANSMISSION
+V TX DATA -V
C1 C2
C3
C4
C5
TRANSMISSION STOPPED ENABLED
+V CTS -V
ENABLED
DISABLED
IDLE
IDLE
Figure 8 - RS232 control signals
If the RTS/CTS handshaking protocol is enabled, the DS2100 activates the RTS232 output to indicate a message is to be transmitted. The receiving unit activates the CTS232 input to enable the transmission.
RS485 Full-Duplex Interface The RS485 full-duplex (5 wires + shield) interface is used for non-polled communication protocols in point-to-point connections over longer distances (max 1200 m / 3940 ft) than those acceptable for RS232 communications or in electrically noisy environments. The connector pinout follows: Connector 25-Pin 2 4 3 5 7
Name
Function
TX485+ TX485RX485+ RX485SGND
RS485 transmitted data + RS485 transmitted data RS485 received data + RS485 received data signal ground
11
Page 24
DS2100
2 DS2100 25-pin
USER INTERFACE
2
TX485+
4
TX485-
3
RX485+
5
RX485-
7
SGND
+ RX485 + TX485 +
SGND
Earth Ground
Figure 9 - RS485 full-duplex connections
RS485 Half-Duplex Interface The RS485 half-duplex (3 wires + shield) interface is used for polled communication protocols. It can be used in a master/slave layout or for Multidrop connections with a Datalogic Multiplexer, (see par. 2.5.4 and 2.5.5) exploiting a proprietary protocol based on polled mode called MUX32 protocol, where a master device polls slave devices to collect data. The connector pinout follows: Connector
Name
Function
RTX485+ RTX485SGND
RS485 transmitted/received data + RS485 transmitted/received data signal ground
25-Pin 2 4 7
DS2100 25-pin
MULTIPLEXER 2
RTX485 +
4
RTX485 -
7
SGND
RTX485 + RTX485 RS485REF
Figure 10 - RS485 half-duplex connections
This interface is forced by software when the protocol selected is MUX32 protocol.
12
Page 25
INSTALLATION
2
In a Multiplexer layout or for slaves, the Multidrop address must also be set via serial channel by the WinHost utility or by ESC sequences. Figure 11 shows a multidrop configuration with DS2100 scanners connected to a Multiplexer. 120 Ohm max 2 m DS2100 #x (up to 31)
max 1200 m
DS2100 #1 RTX485 +
DS2100 #0 CHASSIS
RTX485 SGND
Three wires + shield RTX485 + RTX485 MULTIPLEXER RS485 REF SHIELD
120 Ohm
Figure 11 - DS2100 Multidrop connection to a Multiplexer
13
Page 26
DS2100
2
20 mA Current Loop Interface (INT-22 Accessory Only) When the INT-22 accessory board is installed, the DS2100 is equipped with a 20 mA passive current loop. The following pins of the 25-pin connector are used: Connector 25-Pin 5 3 4 2
Name
Function
CLINCLIN+ CLOUTCLOUT+ current loop input current loop input + current loop output current loop output +
DS2100 25-pin
USER INTERFACE
5
C.L. IN-
3
C.L. IN+
4
C.L. OUT-
2
C.L. OUT+
I = 20 mA
I = 20 mA
MAX. 300 METERS
Figure 12 - 20 mA C.L. connections
14
Page 27
INSTALLATION
2.3.3
2
Auxiliary RS232 Interface
The auxiliary serial interface is used exclusively for RS232 point-to-point connections. The parameters relative to the aux interface (baud rate, data bits, etc.) as well as particular communication modes such as LOCAL ECHO can be defined using the WinHost utility program or "Host Mode Programming", installed from the diskette. The following pins of the 25-pin connector are used to connect the RS232 auxiliary interface:
Connector
Name
Function
RXAUX TXAUX CTSAUX RTSAUX SGND received data transmitted data clear to send request to send signal ground
25-Pin 20 21 23 24 7
DS2100 25-pin
USER INTERFACE
20 RXAUX
TXD
21 TXAUX RXD 23 CTSAUX 24 RTSAUX 7
SGND
DTR DCD SGND
Figure 13 - RS232 auxiliary interface connections using hardware handshaking
When the auxiliary interface is permanently connected as part of the system cabling, it is recommended to connect the cable shield to earth ground. The RTSAUX and CTSAUX signals control data transmission and synchronize the connected devices. If the RTS/CTS handshaking protocol is enabled, the DS2100 activates the RTSAUX output to indicate a message is to be transmitted. The receiving unit activates the CTSAUX input to enable the transmission.
15
Page 28
DS2100
2
Code Verifier If the DS2100 is used as a Code Verifier, it is possible to indicate to the scanner what code to store as the verifier code by means of an external hardware input. The Code Verifier parameter must be enabled and the configuration parameters to allow correct Code Type reading must be saved to the scanner in order to read the verifier code. To activate the input, connect together pins 23 and 24 (CTSAUX and RTSAUX) of the 25-pin connector (for example with a push-button), before the active edge of the External Trigger input (or before the code passes under the laser beam for the Automatic operating mode). The next read code will be stored as the verifier code in the scanner's RAM and NON-VOLATILE (EEPROM) memory by default. Then, the two pins must be disconnected. Since it uses part of the RS232 auxiliary serial interface, this interface is limited when using this option and the Handshake selection must not be set to RTS/CTS.
2.3.4
Inputs
The inputs available on the connector supplied with the scanner are the pins relative to the External Trigger, as indicated below:
Connector 9-Pin
25-Pin
8 9
18 19
Name
Function
EXT TRIG+ EXT TRIG- external trigger + external trigger -
The External Trigger input is used in the On-Line operating Mode and tells the scanner to scan for a code. The active state of this input is selected in software. Refer to the WinHost Help On Line. The yellow LED (Figure A, 4) is on when the External Trigger forces a current flow through the EXT TRIG+ and EXT TRIG- pins. This input is optocoupled and can be driven by both an NPN or PNP type command. The connections are indicated in the following diagrams:
16
Page 29
INSTALLATION
2
DS2100 9-pin
DS2100 25-pin 8
EXT TRIG +
18
EXT TRIG+
19
EXT TRIG-
V or
9
EXT TRIG -
Vext 30 Vdc max. EXTERNAL TRIGGER
+ 5V
Signal
Figure 14 - Input NPN command using external power
DS2100 9-pin
DS2100 25-pin
7
VS
EXTERNAL TRIGGER 9
+ 5V
VS
V
18 EXT TRIG+
8 EXT TRIG+ or
9
EXT TRIG-
19 EXT TRIG25
5
Signal GND
Ground
GND
Figure 15 - Input NPN command using DS2100 power
DS2100 9-pin
Vext 30 Vdc max.
DS2100 25-pin
EXTERNAL TRIGGER V
+ 5V
8
EXT TRIG+ or
9
EXT TRIG-
Signal
18 EXT TRIG+ 19
EXT TRIGGround
Figure 16 - Input PNP command using external power
DS2100 9-pin
EXTERNAL TRIGGER
DS2100 25-pin 9
7 VS
+ 5V 8
VS
18
EXT TRIG+
19
EXT TRIG-
25
GND
V Signal
EXT TRIG+ 9 or
EXT TRIG5 GND
Ground
Figure 17 - Input PNP command using DS2100 power
17
Page 30
DS2100
2 Electrical features:
Maximum voltage: Maximum current:
30 Vdc 25 mA
An anti-disturbance hardware filter is implemented on the External Trigger input (< 5 milliseconds delay). An additional 15 ms (typical) delay can be implemented through a dedicated software parameter (refer to WinHost Help On Line).
2.3.5
Outputs
Two general purpose outputs are available which can be connected only in open collector configurations. The outputs are available only on 25-pin connector models. The following pins are present on the 25-pin connector of the scanner:
Connector
Name
Function
OUT1+ OUT2+ OUT REF output 1 + output 2 + output reference
25-Pin 8 11 12, 22
The meaning of the two outputs OUT1 and OUT2 can be defined by the user (No Read, Right or Wrong). Refer to the WinHost Help On Line. By default, OUT1 is associated with the No Read event, which activates when the code signaled by the external trigger is not decoded, and OUT2 is associated with the Right event, which activates when the code is correctly decoded.
DS2100 25-pin
USER INTERFACE
8/11
+
12/22
OUT REF
Vext 40 Vdc max
Figure 18 - Output open collector connections
These outputs are both level or pulse configurable.
18