Page 101

ADVANCED DATA FORMAT

ADD NEW STRING a)

Define New String String Character b)

Read the Hex value from the Hex/Numeric table identifying the starting characters. Valid values are in the range 00-7F.

End String and Define Procedure End String Plus Procedure

OR

Read the code to end the string selection and continue defining a new procedure belonging to Method 1.

End String and Format

End String & Format

Read the code to end the string and the format definition.

85

Page 102

ADVANCED DATA FORMAT

2.13.2

Match Conditions

By setting one or more of the following conditions it is possible to select the codes to be formatted. Follow the given steps to define the desired condition/s:

MATCH WITH PREDEFINED SUBSTRING Define Matching Substring Match with Substring Read the above code and: 1. read a number in the range 1-5 corresponding to the desired format number; 2. set the number of characters defining the matching string in the range 00-10; 3. read the corresponding character as Hex value from the Hex/Numeric table identifying the substring character/s. Valid values are in the range 00-7F.

(optional)

Define Substring Position

Matching Substring Position

Read the above code and: 1. read a number in the range 1-5 corresponding to the desired format number; 2. read the number corresponding to the substring position in the range 0-255;

AND/OR

MATCH CODE LENGTH Define Code Length Match Code Length i&-& i&-& i&-& i&-&

86 k k k k

Read the above code and: 1. read a number in the range 1-5 corresponding to the desired format number; 2. read the number in the range 0-255;

Page 103

ADVANCED DATA FORMAT

AND/OR

MATCH SYMBOLOGY Define Code Symbology Match Symbology i&-'#k i&-'#k i&-'#k i&-'#k

Read the above code and: 1. read a number in the range 1-5 corresponding to the desired format number; 2. set the number of the matching code symbologies in the range 0-4; 3. select the Datalogic Standard Code Identifier from the Code Identifier Table in Appendix B.

87

Page 104

ADVANCED DATA FORMAT

2.13.3

Format Enable/Disable

Format 1

Format 2 i&-"Xk i&-"Xk i&-"Xk i&-"Xk

0 = disabled 1 = enabled

0 = disabled 1 = enabled

Format 3

Format 4 i&-" `k i&-" `k i&-" `k i&-" `k i&-"-dk i&-"-dk i&-"-dk i&-"-dk

0 = disabled 1 = enabled

0 = disabled 1 = enabled

Format 5

Disable All Formats i&-" Tk i&-" Tk i&-" Tk i&-" Tk

0 = disabled 1 = enabled

88

Page 105

ADVANCED DATA FORMAT

2.13.4

Mismatch Result

The result of each format may be set in case the match conditions previously selected are not satisfied. Once the desired formats have been enabled and a code has been read, the results corresponding to each format will be concatenated together and transmitted in the output message. For this reason, it is strongly advised to set the mismatch result for each format. Example Decoded Code: <DATALOGICproduct> Format definition: Format Format 1

Enable/Disable Enabled

Format 2 Format 3

Disabled Enabled

Format 4

Enabled

Format 5

Enabled

Match Condition Function Code having a length Select field from of 16 characters position1 to position3 / / Code having a length Substitute string of 25 characters "ab" with string "12" string Code having the Insert substring "AT" in "789" in position 7 position 2 Code belonging to the Delete string PDF417 symbology "DA" and "pr"

Mismatch Result No string / Unformatted read code Unformatted read code Unformatted read code

Output message: <DATDATALOGICproductDATALO789GICproductDATALOGICproduct> Format 1

Format 3

Format 4

Format 5

Define Mismatch Result Mismatch Result i&-(&k i&-(&k i&-(&k i&-(&k

Read the above code and: 1. read a number in the range 1-5 corresponding to the desired format number; 2. 0 = empty string as output 1 = unformatted read code as output.

89

Page 106

LYNX™

3

3

TEST BARCODE SYMBOLS

Use these test 1D and 2D symbols to check that the reader is imaging and decoding properly, according to your configuration. UPC-A

1

52637 48509

2

UPC-E

0

123456 5

EAN-8

1234 5670

EAN-13

1 234567 000992

90

Page 107

TEST BARCODE SYMBOLS

3

Code 39 (Normal)

1

7

1

6

2

Code 39 (Full ASCII)

a b

{

Codabar

A8 0 0 5 4 7 2 5 0 7 B

Code 128 t e s t

91

Page 108

LYNX™

3 Interleaved 2 of 5

01234567891234

PDF417

QR

92

Page 109 read from a scan by OCR; it may hold reading errors

TEST BARCODE SYMBOLS r 3)

gc

¥: ° Xe t9 SR, BAA coarse 6 ext aA soe are te A TaN coy

DATA MATRIX NORMAL

INVERSE DATA MATRIX

93

Page 110

LYNX™

4

4

TROUBLESHOOTING & MAINTENANCE

4.1

CABLE CONNECTIONS

Ensure your reader-to-host interface cable and the reader power supply cable are correctly and firmly attached (see "System Connections" in par. 1.3). If the reader is simply not functioning, contact your nearest Datalogic distributor.

4.2

CONFIGURATION TOOLS

If initial communication with the LYNX™ reader and the computer failed when using the LYNX™ Configuration Tools software, then your serial port may be configured differently. 1.

If your serial port is configured differently, choose Options > RS232 Host Setting to make the necessary settings for your COM port.

2.

If you still can not establish host-reader communication, try one or more of the following solution:

• restore the Reader's settings by reading the Restore System Fields Default symbol to set all communication settings to their default. 3.

94

If you choose Actions > Dump Reader Status, then text for the current reader configuration should appear in the text window. If nothing happens, copy and complete the following Worksheet form and contact your service representative.