Page 151

HOST NOT RESPONDING HIT ANY KEY_ IDENTIFIER ERROR HIT ANY KEY_ IN DIAG MODE HIT ANY KEY_ INIT FAILED CYCLE POWER_ INCOMPLETE PGM PRESS ANY KEY_ INVALID DATA LEN HIT ANY KEY_ INVALID DEVICE HIT ANY KEY_ INVALID FILE # HIT ANY KEY_ INVALID HEADER HIT ANY KEY_ INVALID TABLE # HIT ANY KEY_ INVALID PASSWORD HIT ANY KEY_ INVALID PGM # HIT ANY KEY_

INVALID STMT. # HIT ANY KEY_ ITEM NOT FOUND n HIT ANY KEY_ LITHIUM BATTERY LOW Press any key_ Memory has been lost Press any key_ NMI ERROR Cycle Power_ NONPORTABLE MODE _

NOT ALLOW TO RCV HIT ANY KEY_

OUT OF MEMORY HIT ANY KEY_ PLEASE ENTER DIAG COMMAND_ Possible Data Loss Press Any Key_ PROGRAMMING MODE? KEY [YES/NO]?_

The TriCoder requested a table but the host is not responding. You have entered data that doesn’t match the DATA ID specified. TriCoder was shipped without being re-initialized. See Troubleshooting section. Initialization failed. Unit was shut off or went to sleep during programming. You have tried to enter 00 for Max Data Length. Entry must be at least 01. You have specified a character other than K W or ENTER for the INPUT DEVICE. You have entered a character other than 0,1,2,or 3 at the Data File No. Prompt. The Header Record in a table has an invalid format. You have entered a character other than 1,2, or 3 for the Table Number prompt You have not entered the correct password. You have entered something other than 0,1,2, or 3 in Programming Mode or 1,2, or 3 for Data Collection Mode Your program has an undefined statement # in it. An item entered in the TriCoder has not been found in table n. The backup lithium coin battery is low and needs replacing BIG memory problems. This is not a low lithium battery problem. This is not a problem unless it repeats itself. You are operating in Non Portable mode - the TriCoder transmits data immediately to the host computer without saving in memory. You are trying to receive a program on the TriCoder but there is data still present in the TriCoder. You have filled the memory and must upload before you can collect any more data. TriCoder is in diagnostic mode, waiting for a command. Slight memory problem. Not a problem with the lithium battery. Watch TriCoder closely. Do you want to write, edit , review or delete a program? Also allows access to voice operations.

14-3

Check host computer. Re-enter your data or change the DATA ID. Re-initialize TriCoder

Power OFF/ON. Check program for errors. HIT ANY KEY. Re-enter data length. HIT ANY KEY. Enter a valid character. HIT ANY KEY. Enter a valid number. Check header record, fix and retry. HIT ANY KEY. Enter a valid number. HIT ANY KEY. Enter the correct password - WDTRI. HIT ANY KEY. Enter the correct number.

HIT ANY KEY to go to error. HIT ANY KEY. Enter another item number or check table data. Press any key. Change lithium battery. Call for repair. Cycle Power. If error repeats itself, call for support. Continue to enter data. Press F1 to get to Mode Menu. HIT ANY KEY. You must upload and delete any data in TriCoder before you can receive a program. HIT ANY KEY. Upload data and continue with data collection. Enter the correct diagnostic code or press F1 to exit. Call for support if problem continues. Press YES to enter PROGRAMMING MODE. Press NO to return to Mode Menu.

Page 152

PROGRAMMING MODE PASSWORD?_

PROGRAM FUNCTION KEY Create Programs---->1-3 Voice Operations---->4 Change Setup-------->5 PROGRAM NOT SET HIT ANY KEY_

PROGRAM NO.? KEY [0-3]?_ PROGRAM RECEIVE KEY [Y/N]?_ PROGRAM TRANSMIT KEY [Y/N]?_

RECEIVE PROGRAM KEY

Clone TriCoder?->0 RECEIVING PROG. PLEASE WAIT> RECEIVING INTERRUPTED Key Pressed In Error Restart Transmission Hit any Key_ RTC ERROR Call 831-458-9938 SEARCH PATTERN _ SENDING PROGRAM PLEASE WAIT_ SRAM WRITE ERROR CYCLE POWER_ Start Host Program Then Hit Any Key Start Host USB Firmware Download Program First and Then Hit Any Key. TABLE DOWNLOAD? KEY [YES/NO]?_

TABLE FORMAT ERR PRESS ANY KEY_ TABLE NOT AVAILABLE HIT ANY KEY_

You are trying to enter Programming Mode and need to enter the password. The password prevents programming accidents. You have correctly entered the password and are in Programming Mode. Select the function you want by pressing the correct key. You have requested to collect data in a program that has not been defined, or you are trying to transmit a program that has not been defined. You are in Data Collection mode and need to select a program to use. Do you want to receive a program from another TriCoder or host computer? Do you want to transmit a program to another TriCoder or a host computer? You have answered YES to program receive mode. Choose the program to receive into or choose to clone. A program is being received by the TriCoder. A key was pressed on the TriCoder after program transfer had started, aborting the transmission. TriCoder contains invalid program information. This error should not occur and indicates a firmware problem. Try to record what led to this error. The SEARCH key was pressed. You are transmitting a program from the TriCoder. Wait for completion. A serious error was detected. If persistent, call for RMA. TriCoder is ready to receive from the host via USB, (i.e. Table, Program, or Firmware Download) Firmware download has been requested. USB enumeration has been completed. Now TriCoder is requesting the user to start PC TriCoder Firmware Downloader. A table has been defined in the memory partitions - you can now download a table into the TriCoder. The table header is in an invalid format or EOF Ctrl Z (ASCII 26) is missing. The host program has told the TriCoder it has no table to download.

14-4

Key in the password WDTRI

Key the number that corresponds to what you want to do. HIT ANY KEY. Enter the correct program number or go to Programming Mode to create a new program. Key 0, 1,2 or 3.

Press YES to go to Receiving Menu. Press NO to go to the Mode Menu. Press YES to go to Transmit Menu. Press NO to go to Mode Menu. Press the Program # or 0 to Clone.

Wait for next message. Hit any Key. Restart transmission. Do not press any keys on TriCoder once transmission has begun. Cycle Power to clear, Call for support. Enter data to search for or press ENTER key to exit. Wait for next message.

Turn TriCoder OFF/ON. Call for support if error continues to occur. Start the host computer program, i.e. TriCoder Utilities, then Hit Any key. Start the Windows program, TriCoder Firmware Downloader. Available on TriCoder Utilities CD ROM, or on www.barcodehq.com. Press YES to continue with table download. Press NO to return to Mode Menu. Check table file for format errors. Try again. Try again later.

Page 153

TABLE NOT DEFINED HIT ANY KEY_ TABLE NOT FOUND HIT ANY KEY_ TABLE NUMBER KEY [1-3]?_ TABLE DOWNLOAD? KEY [YES/NO]?_

TABLE TOO BIG PRESS ANY KEY_ TRANSMIT PROGRAM KEY Programs [1-3]------->1-3 Clone TriCoder?----->0 TRICODER x9nnn HIT ANY KEY_ UPLOADING DATA PLEASE WAIT_ USB Enumeration Please wait…..

USB not Connected Connect Cable, Then Press a Key Or F1 to Exit

You are trying to download a table to a table number that is not defined in the Partitions. A table referenced in a data collection program does not exist. You have requested to download a table. Now choose which table number you want to download to. A table is defined and ready for download. Do you want to download? Not enough memory was allocated to accept the table you are downloading. You have answered YES to Transmit Program? prompt. Choose program to transmit or clone entire TriCoder. TriCoder opening screen. Displays firmware version. Collected data is being uploaded to host. At power on or when a cable is plugged in, TriCoder must enumerate as a USB device. Also happen when changing modes. Computer Interface is set to USB. A mode has been selected to transfer data. The USB cable is not plugged in.

USB not Connected Connect, Hit any Key or F1 – Exit or F2 – Display Only

You have answered YES to Non Portable Mode when the TriCoder is configured for USB interface and NOT connected to the USB port on the computer.

USB not Connected Connect, Hit any Key or F1 – Exit or F2 – Delete File

You have answered YES to Data Upload Mode when the TriCoder is configured for USB Interface and is NOT connected to the USB port on the computer.

VOICE OPERATIONS KEY Record/Playback------>1 Assign Errors----------->2

Voice Operations Menu accessed through Programming Mode.

Waiting on Host Program Or F1 to Exit WAITING ON PMPT_

USB cable is plugged in. Upload has been started. TriCoder is waiting on Tri Utilities to be started. TriCoder is in Non-Portable Mode, operating with HOST RESPONSE ACCEPTED and is waiting from a prompt from the host.

14-5

Go to Setup and define the table in the Partitions parameter. Download table or correct program to reference correct table. Choose a defined table number to download to. Press YES to proceed with download process. Press NO to return to Mode Menu. Re-allocate memory or change table size. Press 1,2, or 3 to transmit that program or Press 0 to clone entire TriCoder. HIT ANY KEY. Wait for upload to finish. Just wait a few seconds. There is nothing to do. Advisory only. Connect USB cable and press a key or F1 to Exit to Mode Menu.

If you want to see the data on the computer, connect to the USB port, then press any key. Press F1 to exit Non Portable Mode or press F2 to simply display the data on the TriCoder screen without transmitting to the computer. If you want to upload via the USB port, connect the TriCoder to the USB port and press any key. Press F1 to exit Data Upload Mode or press F2 to delete the data file from the TriCoder without transmitting it to the computer Press 1 to record or playback voice messages. Press 2 to assign error messages. Start the TriCoder Utilities.

Wait for prompt from Host or press F1 to bypass.

Page 154

WARNING 2K LEFT HIT STATUS KEY_

You are running out of memory. Finish what you are doing and upload your data.

WRONG INPUT DEVICE HIT ANY KEY_

The input device you are using is not what is specified in your program. A key was pressed on a transmitting TriCoder during the program transmission. The transmission has been aborted and the receiving TriCoder had invalid program information.

XMISSION INCOMPLETE Key Pressed in Error Restart Transmission Hit any Key_

Press STATUS key to see exact amount of memory left. Finish what you are doing and go upload your data Use the correct input device or change the specification in your program. Hit any Key. Restart program transmission.

General Troubleshooting If you want to what the settings in the TriCoder are, you can scan Start Setup, scan the Parameter(s) in which you are interested, and then read the display settings referring to the TriCoder Setup Menu for explanation of the letter codes. You can return the TriCoder to the default shipped settings by scanning:

• • •

Start Setup Reset End Setup then reconfigure the TriCoder for your system using the bar coded TriCoder Setup Menu or by using the keypad in PROGRAMMING MODE. Use the following information to troubleshoot any problems you might be having:

USB won’t upload •

Be sure you are not trying to use USB with Windows 95 or NT. There is not support for USB devices in these operating systems.

•

When you first plug your TriCoder into a USB port, Windows will sense new hardware and attempt to install driver(s) for the TriCoder. If you fail to complete the driver(s) installation, the TriCoder will not upload or transfer programs, etc. See Appendix G, USB Driver Installation Recovery for how to recover from a failed driver installation.

Serial or Keyboard Wedge won’t upload •

If everything else is working well, it has been uploading just fine, and now it has suddenly quit uploading, open the TriCoder up and push the Keyboard/Serial jumper block down fully. After dropping, the jumper block can sometimes become slightly unseated, destroying the upload. See Appendix F: Jumpers, Switches, and Adjustments.

14-6

Page 155

Strange Behavior •

Low batteries that are not detected can exhibit some bizarre TriCoder symptoms, including laser beams that shut off, continuous beeping, shutting off when you scan, etc. Try the power supply. Try new batteries. Be sure if you are using NiCad or NiMH rechargeable batteries, you have the Speaker/Batteries set to 3 and not 2.

The TriCoder won’t read your bar codes… •

Recheck all hardware connections using Chapter 5; TriCoder Connection as a guide. Try reading a bar code you know is valid and readable – try the Test Label in Chapter 3; Non Portable Mode.

•

If you are reading Code 39 bar codes with leading spaces (see the Barpad in Appendix H; Code 39, Accumulate Mode) and you have enabled Accumulate Mode in the setup, the reader will beep twice upon reading a bar code and will not transmit data to the computer until it reads a bar code without a leading space.

•

Recheck the TriCoder setup and make sure you have enabled the bar code type you are trying to read.

•

If you are having problems reading Interleaved 2 of 5 codes, make sure the data length of the bar code is the same as the 2 of 5 Length set on the TriCoder.

•

Make sure you have not enabled check digits for Code 39 or Interleaved 2 of 5 if your bar codes do not have check digits.

•

Make sure the scanner is plugged into the SCANNER port and not the COMPUTER port.

The TriCoder transmits extra characters at the start or end of bar code data… •

Clear the preamble and postamble.

•

Check to make sure you have not setup the TriCoder to transmit start/stop characters, check digits, NSC characters or terminator characters that you don’t want transmitted.

•

For UPC-E, select Compressed transmission if you don’t want your data padded out with extra zeroes.

14-7

Page 156

Poor read rates… •

Try reading the bar code below as an example of a known “good” bar code. This bar code is a Code 39 bar code and is readable by the TriCoder set to its default settings.

•

Make sure your bar codes have clearly defined dark bars and clean white spaces. If the bars are so light that they are gray instead of black or are so dark that they “bleed” into the white spaces, the printing of the bar codes is the problem. Whoever is printing the bar codes needs to make the necessary corrections.

•

The bar code should also have a “quiet zone” of at least ¼” to the left of the first bar and to the right of the last bar. Make sure to start your scanning to the left of the bar code in the “quiet zone”, moving the scanner quickly and smoothly as if drawing a line through the bar code. If using a laser scanner or CCD scanner, make sure the beam covers the bar code completely.

•

Make sure you are using the correct type of scanner for the type of bar code you are reading. If you are using an infrared scanner, your bar code must be printed using infrared-quality ink.

•

If using a wand scanner, clean the tip. You would be amazed at how many wands we see where the only problem is that it needs to be cleaned. You would also be amazed at what we have found inside the tips of some of these wands. We cannot stress this enough – CLEAN YOUR WAND. In fact, make it a habit to clean your wand on a regular basis. Use only warm soapy water and a soft cloth; alcohol and other solvents can fog the lens of the tip – do not use them. If using a laser or CCD, clean the window of the scanner with a soft cloth.

Sometimes I get long beeps and sometimes short beeps… •

When the TriCoder is in NON PORTABLE MODE or using the power adapter, is uses a longer beep that can vary in tone – see Chapter 6; TriCoder Setup for details on changing the beep tone. If the TriCoder is in a “portable” mode (i.e. Data Collection), it uses a shorter beep to conserve power.

14-8

Page 157

USB troubleshooting (for Windows Users) The bulk of USB problems are due to driver(s) installations not being completed and the data transfer(s) won't work. Whether you have selected USB HIGH Speed Data Upload or USB Keyboard Data Upload, you will be required to install a driver for Human Interface Device (HID) keyboard emulation as well as the Worth Data drivers for High Speed data transfers. If the driver (s) are not completely installed, when you plug the TriCoder into a USB port, two things can happen: 1.

Windows will recognize the device but flag it in the Windows Device Manager with a "!" icon (if you cannot find a HID entry for the TriCoder - double click on Human Interface Devices - or an entry for the TriCoder after double clicking on the Universal Serial Bus Controllers, the device has been disabled). Assuming you find the device with the “!”, right click on the device, look at properties, and then click on the Update Driver button – only this time being sure Windows finds the driver files it needs. For the HID driver, you will probably have to insert the original Windows CD and point the installer to the CD ROM drive; for the High Speed driver, you will have to insert the TriCoder Utilities CD and point the installer to the CD ROM drive.

2.

The device will not show up in the Windows Device manager and the hardware will not be recognized when it is plugged in, because Windows has marked the device as "Disabled". To fix this, you will have to go into the Control Panel, go to "Add Hardware" and search for new devices, (with the TriCoder plugged into a USB port). When the TriCoder device is found, you need to uncheck the "Disabled" box and then proceed to click on "Update Driver", this time furnishing the original Windows CD (for HID) or the Worth Data Utilities CD (for the High speed driver).

See Appendix G, USB Driver Installation Recovery for additional details. Once the all the drivers are installed correctly, there should not be any further problems.

Keyboard Wedge mode troubleshooting The following problems are specific to a TriCoder used in keyboard wedge model T64/LT64.

You get a “keyboard error” or a “301 keyboard error” when you power up your system… •

Check the cables to make sure all connections are secure and the TriCoder is ON.

•

Make sure the keyboard splitter cable is plugged into the keyboard port and not the mouse port. 14-9

Page 158

•

Remove any keyboard extension cables or adapters. Plug directly into the port using the TriCoders keyboard splitter cable to eliminate the possibility that the extensions or adapters may be causing the problem.

The TriCoder transmits incorrect or no data to the PC’s screen… •

If the TriCoder is transmitting punctuation characters (!@#$%^&*) when reading numeric bar codes, or transmitting letters in the wrong case (upper vs. lower), you could have a Num Lock, Caps Lock, Shift or timing delay problem. Check your keyboard to see if your Num Lock or Caps Lock is on. Try typing numbers (the upper row of keys on your keyboard) and see if you get the correct output on your screen. If you don’t, make sure to reboot your PC before changing any delays on the TriCoder. See Chapter 6; TriCoder Setup for details on the Intercharacter Delay setting.

•

If you scan a bar code and correct data shows up on your PC screen but some is missing, you need to change the Intercharacter Delay. See Chapter 6; TriCoder Setup for details on how to change this setting. Test your keyboard after every change made to the TriCoder setup. If the keyboard does not work correctly, you will have to reboot before you make another change to the reader.

•

Make sure you have configured the TriCoder to use the correct Computer Interface.

The TriCoder reads one bar code and stops reading after that… •

Double-check the Computer Interface setting. Selecting Macintosh as the interface when you are not physically connected to a Mac will hang up the TriCoder.

Serial mode troubleshooting The following problems pertain mostly to TriCoders setup in serial mode – T54.

Using the 5 volt power adapter, the TriCoder does not beep when turned on… •

Plug something else into the wall outlet to make sure the outlet is working or plug the TriCoder’s power adapter into an outlet that is working with something else. If the outlet is good, either the power adapter or the TriCoder is not working correctly. If you have more than one TriCoder, try the other TriCoder with the suspect power adapter. If you find a problem with either the TriCoder or the power adapter, call for repair or replacement. 14-10

Page 159

The TriCoder beeps, but nothing appears on your computer screen… • Recheck the installation instructions – Chapter 5; TriCoder Connection – to make sure all cables are properly connected. If you are connecting directly to a serial port on a PC, use the following checklist: 1.

What kind of computer and port are you plugging into? • A 25 pin female port is a parallel port. • PC – Male 9 pin port. A 9 pin female port is a video port. • Laptop COM ports are almost always 9 pin male.

2.

What cable are you using? • F34 – 25 pin null-modem cable • F36 – 9 pin female cable • If you have a 9-pin COM port and have the F34 cable, use a 9 to 25 null modem adapter or get the right cable (F36).

3.

What Computer Interface is the TriCoder set for? • “3” is the correct setting for RS-232

4.

What Baud Rate, Parity and Start and Stop bits are you using? • Make sure the settings on the TriCoder match what the host program is expecting. When all else fails, try the default settings.

The TriCoder reads one bar code and stops reading after that… •

Double-check the Computer Interface setting. Selecting Mac ADB as the interface when you are not physically connected to a Mac will hang up the TriCoder.

•

If you selected polling protocol, the TriCoder needs to be polled by your computer before it will read another bar code. If you selected Polling with ACK/NAK, you also need to send an ACK or NAK back to the TriCoder after each read (unless you specify a finite time-out) as well as the poll, before it will read another code.

The TriCoder transmits data to your screen or serial port but some characters are garbled or missing… •

Check your serial parameters. Do the baud rate, parity, start and stop bits match on both the TriCoder and the serial port?

•

Make sure you have selected the correct Data Format. RS-232 ASCII is the most commonly used.

•

If you are using your own software to read the serial port, consider 14-11

Page 160

that not all software can keep up with the serial port at the highest baud rate. Try setting the TriCoder and serial port to a lower Baud Rate. If you have cables over 100 feet long, try lowering the baud rate.

I can’t communicate with my RS-232 port on the host… •

This is one of the most common problems we hear. Check your cables first. Get rid of all cables and connectors except the Worth Data cable. If that works, your cables are suspect.

The TriCoder transmits a few messages then hangs… •

If you are using a terminal/ host configuration and are using XON/XOFF protocol on the terminal, disable XON/XOFF on the TriCoder – it may be missing the XON after intercepting a XOFF.

Program troubleshooting The following examples have to do with programming the TriCoder.

When editing a program, all I can see are the upload prefixes and suffixes… •

If there is a data file currently in the TriCoder, you cannot edit any part of the programs except for the Upload Prefix or Upload Suffix. This is to protect the integrity of the data that is currently still in the TriCoder. You will know if you still have a data file in the TriCoder if you see the DATA FILE EXISTS PRESS ANY KEY_ message when you enter PROGRAMMING MODE and enter the password. To edit or create a program, upload and delete the data from the TriCoder.

I can’t delete a programming statement while editing a program in the TriCoder… •

You cannot actually “remove” a statement from a program when editing a program in the TriCoder. You can clear all of the program prompts for the statement but the statement number will remain “occupied” by a blank programming statement. The best thing to do is simply not “go to” that statement in your program. If that bothers you, the alternative is to delete the entire program and re-enter it without the offending statement. See Chapter 8; Creating custom programs for details.

14-12