FIS-6801 Industrial Scanner User's Manual, pages 371 to 380

Pages 371 to 380 of 454

The original document is the PDF (7.5 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Wenglor FIS-6801

Page 371

Appendices Options:

Any 7-bit ASCII character

Start Trigger Start Character (Non-Delimited) Serial Cmd: Default: Options:

<K229,start character> NULL (0x00) Two hex digits representing any ASCII character except <, >, XON, and XOFF.

Stop Trigger Stop Character (Non-Delimited) Serial Cmd: Default: Options:

<K230,stop character> NULL (0x00) Two hex digits representing any ASCII character except <, >, XON, and XOFF.

Processing Timeout Serial Cmd: Default: Options:

<K245,processing timeout> 200 (0.2 seconds) 1 to 65535

FIS-6801 Industrial Scanner User’s Manual

A-43

Page 372

Serial Commands

Configuration Database Number of Active Indexes <K252,number of active indexes,sort database> <K255,index,shutter speed,gain,focal distance,sub-sampling,row pointer,column Configuration Database pointer,row depth,column width,symbol type,morphology operation,morphology (Standard FIS-6801) size> Configuration Database <K255,index,shutter speed,gain,unused,sub-sampling,row pointer,column (FIS-6801 C-Mount) pointer,unused,column width,unused,unused,unused> Load Current Setting from <K255–,index> Configuration Database Request Database Index <K255?,index> Setting Request All Database <K255?> Settings <K256,switch mode,frame count/time,image process looping, image dimensions> Database Mode

Number of Active Indexes Number of Active Indexes Serial Cmd: Default: Options:

<K252,number of active indexes,number of database cycles> 0 (Disabled) 0 to 10

Number of Database Cycles Serial Cmd: Default: Options:

A-44

<K252,number of active indexes,number of database cycles> 0 (Disabled) 0 to 255

FIS-6801 Industrial Scanner User’s Manual

Page 373

Appendices

Symbologies Composite Aztec Micro QR Code Postal Symbologies Code 39 Codabar Interleaved 2 of 5 UPC/EAN Code 128 / EAN 128 Code 93 PDF417 Pharmacode Data Matrix QR Code BC412 DataBar Omnidirectional DataBar Limited DataBar Expanded MicroPDF417

<K453,symbology status,separator status,separator> <K458,status> <K459,status> <K460,postal type,POSTNET status, PLANET status,USPS4CB status,POSTNET allow B and B’ fields,Australia Post allow 0 FCC> <K470,status,check character status,check character output status, large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set> <K471,status,start/stop match status,start/stop output status, large intercharacter gap,fixed symbol length status,fixed symbol length,check character type,check character output status> <K472,status,check character status,check character output status, symbol length 1,symbol length 2,guard bar,range mode status> <K473,UPC status,EAN status,supplementals status,separator status,separator character,supplementals type,UPC-E as UPC-A> <K474,status,fixed symbol length status,fixed symbol length, EAN status,output format,application record separator status, application record separator character,application record brackets, application record padding> <K475,status,fixed symbol length status,fixed symbol length> <K476,status,unused,fixed symbol length status,fixed symbol length,unused,codeword collection> <K477,status,fixed bar count status,fixed bar count,minimum bar count,bar width mode,direction,fixed threshold value,background> <K479,ECC 200 status,ECC 000 status,ECC 050 status,ECC 080 status,ECC 100 status,ECC 140 status,ECC 120 status> <K480,status> <K481,status,check character output,fixed symbol length status, fixed symbol length> <K482,status> <K483,status> <K484,status,fixed symbol length status,fixed symbol length> <K485,status,unused,fixed length status,fixed symbol length>

Composite Symbology Status (Composite) Serial Cmd: Default: Options:

<K453,symbology status,separator status,separator> 0 = Disabled 0 = Disabled 1 = Enabled 2 = Required

Separator Status (Composite) Serial Cmd: Default: Options:

<K453,symbology status,separator status,separator> 0 = Disabled 0 = Disabled 1 = Enabled

FIS-6801 Industrial Scanner User’s Manual

A-45

Page 374

Serial Commands Separator (Composite) Serial Cmd: Default: Options:

<K453,symbology status,separator status,separator> , (comma) Any 7-bit ASCII character

Code 39 Serial Cmd: Default: Options:

<K470,status,check character status,check character output status,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set> 1 = Enabled 0 = Disabled 1 = Enabled

Check Character Status (Code 39) Serial Cmd: <K470,status,check character status,check character output status,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set> Default: 0 = Disabled Options: 0 = Disabled 1 = Enabled Check Character Output Status (Code 39) Serial Cmd: Default: Options:

<K470,status,check character status,check character output status,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set> 0 = Disabled 0 = Disabled 1 = Enabled

Large Intercharacter Gap (Code 39) Serial Cmd: Default: Options:

<K470,status,check character status,check character output status,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set> 0 = Disabled 0 = Disabled 1 = Enabled

Fixed Symbol Length Status (Code 39) Serial Cmd: Default: Options:

<K470,status,check character status,check character output status,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set> 0 = Disabled 0 = Disabled 1 = Enabled

Fixed Symbol Length (Code 39) Serial Cmd:

A-46

<K470,status,check character status,check character output status,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set> FIS-6801 Industrial Scanner User’s Manual

Page 375

Appendices Default: Options:

10 1 to 128

Full ASCII Set (Code 39) Serial Cmd: Default: Options:

<K470,status,check character status,check character output status,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set> 0 = Disabled 0 = Disabled 1 = Enabled

Codabar Serial Cmd: Default: Options:

<K471,status,start/stop match,start/stop output,large intercharacter gap,fixed symbol length status,symbol length,check character type,check character output> 0 = Disabled 0 = Disabled 1 = Enabled

Start/Stop Match (Codabar) Serial Cmd: Default: Options:

<K471,status,start/stop match,start/stop output,large intercharacter gap,fixed symbol length status,symbol length,check character type,check character output> 1 = Enabled 0 = Disabled 1 = Enabled

Start/Stop Output (Codabar) Serial Cmd: Default: Options:

<K471,status,start/stop match,start/stop output,large intercharacter gap,fixed symbol length status,symbol length,check character type,check character output> 1 = Enabled 0 = Disabled 1 = Enabled

Large Intercharacter Gap (Codabar) Serial Cmd: Default: Options:

<K471,status,start/stop match,start/stop output,large intercharacter gap,fixed symbol length status,symbol length,check character type,check character output> 0 = Disabled 0 = Disabled 1 = Enabled

Fixed Symbol Length Status (Codabar) Serial Cmd: Default: Options:

<K471,status,start/stop match,start/stop output,large intercharacter gap,fixed symbol length status,symbol length,check character type,check character output> 0 = Disabled 0 = Disabled 1 = Enabled

FIS-6801 Industrial Scanner User’s Manual

A-47

Page 376

Serial Commands Symbol Length (Codabar) Serial Cmd: Default: Options:

<K471,status,start/stop match,start/stop output,large intercharacter gap,fixed symbol length status,symbol length,check character type,check character output> 10 1 to 128

Check Character Type (Codabar) Serial Cmd: Default: Options:

<K471,status,start/stop match,start/stop output,large intercharacter gap,fixed symbol length status,symbol length,check character type,check character output> 0 = Disabled 0 = Disabled 1 = Mod 16 2 = NW7 (Mod 11) 3 = Both Mod 16 and NW7 (Mod 11)

Check Character Output (Codabar) Serial Cmd: Default: Options:

<K471,status,start/stop match,start/stop output,large intercharacter gap,fixed symbol length status,symbol length,check character type,check character output> 0 = Disabled 0 = Disabled 1 = Enabled

Interleaved 2 of 5 Serial Cmd: Default: Options:

<K472,status,check character status,check character output status,symbol length #1,symbol length #2,guard bar status,range mode status> 1 = Enabled 0 = Disabled 1 = Enabled

Check Character Status (Interleaved 2 of 5) Serial Cmd: Default: Options:

<K472,status,check character status,check character output status,symbol length #1,symbol length #2,guard bar status,range mode status> 0 = Disabled 0 = Disabled 1 = Enabled

Check Character Output Status (Interleaved 2 of 5) Serial Cmd: Default: Options:

A-48

<K472,status,check character status,check character output status,symbol length #1,symbol length #2,guard bar status,range mode status> 0 = Disabled 0 = Disabled 1 = Enabled

FIS-6801 Industrial Scanner User’s Manual

Page 377

Appendices Symbol Length #1 (Interleaved 2 of 5) Serial Cmd: Default: Options:

<K472,status,check character status,check character output,symbol length #1,symbol length #2,guard bar status,range mode status> 16 0 to 128

Symbol Length #2 (Interleaved 2 of 5) Serial Cmd: Default: Options:

<K472,status,check character status,check character output,symbol length #1,symbol length #2,guard bar status,range mode status> 6 0 to 128

Guard Bar Status (Interleaved 2 of 5) Serial Cmd: Default: Options:

<K472,status,check character status,check character output,symbol length #1,symbol length #2,guard bar status,range mode status> 0 = Disabled 0 = Disabled 1 = Enabled

Range Mode Status (Interleaved 2 of 5) Serial Cmd: Default: Options:

<K472,status,check character status,check character output,symbol length #1,symbol length #2,guard bar status,range mode status> 1 = Enabled 0 = Disabled 1 = Enabled

UPC/EAN Serial Cmd: Default: Options:

<K473,UPC status,EAN status,supplementals status,separator status, separator character,supplemental type,format UPC-E as UPC-A> 3 = Both Standard and Edge-to-Edge 0 = Disabled 1 = Standard 2 = Edge-to-Edge 3 = Both Standard and Edge-to-Edge

EAN Status Serial Cmd: Default: Options:

<K473,UPC status,EAN status,supplementals status,separator status, separator character,supplemental type,format UPC-E as UPC-A> 1 = Enabled 0 = Disabled 1 = Enabled

Supplementals Status (UPC/EAN) Serial Cmd: Default:

<K473,UPC status,EAN status,supplementals status,separator status, separator character,supplemental type,format UPC-E as UPC-A> 0 = Disabled

FIS-6801 Industrial Scanner User’s Manual

A-49

Page 378

Serial Commands Options:

0 = Disabled

1 = Enabled

2 = Required

Separator Status (UPC/EAN) Serial Cmd: Default: Options:

<K473,UPC status,EAN status,supplementals status,separator status, separator character,supplemental type,format UPC-E as UPC-A> 0 = Disabled 0 = Disabled 1 = Enabled

Separator Character (UPC/EAN) Serial Cmd: Default: Options:

<K473,UPC status,EAN status,supplementals status,separator status, separator character,supplemental type,format UPC-E as UPC-A> , (comma) Any 7-bit ASCII character

Supplemental Type (UPC/EAN) Serial Cmd: Default: Options:

<K473,UPC status,EAN status,supplementals status,separator status, separator character,supplemental type,format UPC-E as UPC-A> 0 = Both 0 = Both 1 = 2 characters only 2 = 5 characters only

Format UPC-E as UPC-A (UPC/EAN) Serial Cmd: Default: Options:

<K473,UPC status,EAN status,supplementals status,separator status, separator character,supplemental type,format UPC-E as UPC-A> 0 = Disabled 0 = Disabled 1 = Enabled

Code 128/EAN 128 Serial Cmd:

Default: Options:

<K474,status,fixed symbol length status,fixed symbol length,EAN 128 status, output format,application record separator status,application record separator character,application record brackets,application record padding,separation factor> 3 = Both Standard and Edge-to-Edge 0 = Disabled 1 = Standard 2 = Edge-to-Edge 3 = Both Standard and Edge-to-Edge

Fixed Symbol Length Status (Code 128/EAN 128) Serial Cmd:

Default: A-50

<K474,status,fixed symbol length status,fixed symbol length,EAN 128 status, output format,application record separator status,application record separator character,application record brackets,application record padding,separation factor> 0 = Disabled FIS-6801 Industrial Scanner User’s Manual

Page 379

Appendices Options:

0 = Disabled

1 = Enabled

Fixed Symbol Length (Code 128/EAN 128) Serial Cmd:

Default: Options:

<K474,status,fixed symbol length status,fixed symbol length,EAN 128 status,output format,application record separator status,application record separator character,application record brackets,application record padding, separation factor> 10 1 to 128

EAN 128 Status (Code 128/EAN 128) Serial Cmd:

Default: Options:

<K474,status,fixed symbol length status,fixed symbol length,EAN 128 status, output format,application record separator status,application record separator character,application record brackets,application record padding,separation factor> 1 = Enabled 0 = Disabled 1 = Enabled 2 = Required

Output Format (Code 128/EAN 128) Serial Cmd:

Default: Options:

<K474,status,fixed symbol length status,fixed symbol length,EAN 128 status, output format,application record separator status,application record separator character,application record brackets,application record padding,separation factor> 0 = Standard 0 = Standard 1 = Application

Application Record Separator Status (Code 128/EAN 128) Serial Cmd:

Default: Options:

<K474,status,fixed symbol length status,fixed symbol length,EAN 128 status, output format,application record separator status,application record separator character,application record brackets,application record padding,separation factor> 0 = Disabled 0 = Disabled 1 = Enabled

Application Record Separator Character (Code 128/EAN 128) Serial Cmd:

Default: Options:

<K474,status,fixed symbol length status,fixed symbol length,EAN 128 status, output format,application record separator status,application record separator character,application record brackets,application record padding,separation factor> , (comma) Any 7-bit ASCII character

FIS-6801 Industrial Scanner User’s Manual

A-51

Page 380

Serial Commands Application Record Brackets (Code 128/EAN 128) Serial Cmd:

Default: Options:

<K474,status,fixed symbol length status,fixed symbol length,EAN 128 status, output format,application record separator status,application record separator character,application record brackets,application record padding,separation factor> 0 = Disabled 0 = Disabled 1 = Enabled

Application Record Padding (Code 128/EAN 128) Serial Cmd:

Default: Options:

<K474,status,fixed symbol length status,fixed symbol length,EAN 128 status, output format,application record separator status,application record separator character,application record brackets,application record padding,separation factor> 0 = Disabled 0 = Disabled 1 = Enabled

Separation Factor Serial Cmd:

Default: Options:

<K474,status,fixed symbol length status,fixed symbol length,EAN 128 status, output format,application record separator status,application record separator character,application record brackets,application record padding,separation factor> 0 = Normal 0 = Normal 1 = High 2 = Highest

Code 93 Serial Cmd: Default: Options:

<K475,status,fixed symbol length status,symbol length> 3 = Both Standard and Edge-to-Edge 0 = Disabled 1 = Standard 2 = Edge-to-Edge 3 = Both Standard and Edge-to-Edge

Fixed Symbol Length Status (Code 93) Serial Cmd: Default: Options:

<K475,status,fixed symbol length status,symbol length> 0 = Disabled 0 = Disabled 1 = Enabled

Symbol Length (Code 93) Serial Cmd: Default: A-52

<K475,status,fixed symbol length status,symbol length> 10 FIS-6801 Industrial Scanner User’s Manual