Page 131
Parameters
Polling Mode Options These parameters only take effect for Polling Mode <K140,5> on the main RS-232 or RS-422 ports (not on the Auxiliary Port), and are completely independent of the ACK/NAK Options <K147>. The values of protocol characters can be changed, but the protocol events cannot be disabled. The polling mode address is configured in the <K140> command. To enable true multidrop protocol, the RS422/485 port must be enabled, <K102,1>, in order to turn the transmitter on and off. If RS-232 is enabled instead of RS422/485, <K102,0>, then Polling Mode will operate as a Point-to-Point polling protocol. This is because the RS-232 transmitter is always left on when enabled. Note: See ACK/NAK Data Flow Examples in Appendix F for sample Polling Mode communication scenarios.
RES-NAK Defaults RES: (Reset) REQ: (Request) STX: (Start of Text) ETX: (End of Text) ACK: (Acknowledge) NAK: (Negative Acknowledge)
FIS-6801 Industrial Scanner User’s Manual
EOT (0x04) ENQ (0x05) STX (0x02) ETX (0x03) ACK (0x06) NAK (0x15)
5-29
Page 132
Communication
Response Timeout Only used when a response is required from the host. While in Multidrop, if the Scanner does not receive an ACK or NAK from the host after sending polled data, it will act on a fault. The Scanner can be set to wait indefinitely by setting Response Timeout to zero. The time that the Scanner will wait before timing out if ACK, NAK, and ETX are enabled, and a host response is expected.
5-30
FIS-6801 Industrial Scanner User’s Manual
Page 133
Parameters
LRC Status Used when extra data integrity is required. An error-checking routine that verifies the accuracy of transmissions. It is the exclusive OR of all characters following the STX (start of text) up to and including the ETX (end of text). What this means is that the binary representation of all the characters in a transmission are cumulatively added in a column and each resulting odd integer is assigned a 1 and each even integer a 0 (two 1s = 0, two 0s = 0, a 1 and a 0 = 1). The extra LRC character is then appended to the transmission, and the receiver (usually the host) performs the same addition and compares the results.
FIS-6801 Industrial Scanner User’s Manual
5-31
Page 134
Communication
External Data Routing External Data Routing settings configure the global operation of all external data port settings.
External Data Routing Options Transparent Mode When Transparent Mode is enabled, the following conditions apply: Symbol Data to Source = Fixed to Enabled Ambles to Source = Fixed to Disabled Echo to Source = Fixed to Enabled Output at End of Read Cycle = Fixed to Enabled Output at ETX = Fixed to Enabled with user-defined characters. Default character will be used if this setting is set to Disabled. Output at Timeout = Fixed to Enabled with user-defined timeout. Default time will be used if this setting is set to Disabled.
5-32
FIS-6801 Industrial Scanner User’s Manual
Page 135
Parameters Mode of Operation • Source Port data will echo back to itself. • Source Port data will always pass through even when the Destination Port is in Polling Mode with the host. • Whenever ETX is received on the Source Port or symbol data is generated, the data will be send to the Destination Port with its data appearing between the Preamble and symbol data. • Source Port data will always be sent to the Destination Port with a Preamble and Postamble. • Symbol data will be sent to the Source Port on a good read without Preamble or Postamble in Point-to-Point protocol even if the Destination Port is in Polling Mode with the host. • Destination Port data always echoes to the Source Port even if the Destination Port is in Polling Mode.
FIS-6801 Industrial Scanner User’s Manual
5-33
Page 136
Communication
Half Duplex Mode When Half Duplex Mode is enabled, the following conditions apply: Symbol Data to Source = Fixed to Enabled Ambles to Source = Fixed to Enabled Echo to Source = Fixed to Disabled Output at End of Read Cycle = Fixed to Disabled Output at ETX = Fixed to Disabled with user-defined characters. Default character will be used if this setting is set to Disabled. Output at Timeout = Fixed to Disabled with user-defined timeout. Default time will be used if this setting is set to Disabled. Mode of Operation • Source Port data is not echoed back to itself. • Source Port data is ignored when the Destination Port is in Polling Mode. • Source Port data or symbol data is sent to the Destination Port whenever it is received. • Source Port data is sent to the Destination Port without a Preamble or Postamble. • Symbol data is sent to the Source Port and the Destination Port at the same time, and conforms to the communication parameters. • Destination Port data is echoed to the Source Port in an un-polled mode.
5-34
FIS-6801 Industrial Scanner User’s Manual
Page 137
Parameters
Full Duplex Mode When Full Duplex Mode is enabled, the following conditions apply: Symbol Data to Source = Fixed to Disabled Ambles to Source = Fixed to Disabled Echo to Source = Fixed to Disabled Output at End of Read Cycle = Fixed to Disabled Output at ETX = Fixed to Disabled with user-defined characters. Default character will be used if this setting is set to Disabled. Output at Timeout = Fixed to Disabled with user-defined timeout. Default time will be used if this setting is set to Disabled. Mode of Operation • Source Port data is not echoed back to itself. • Source Port data is ignored when the Destination Port is in Polling Mode. • Source Port data or symbol data is sent to the Destination Port whenever it is received. • Source Port data is always sent to the Destination Port without a Preamble or Postamble. • Symbol data is not sent to the Source Port. • Destination port data is echoed to the source port in an un-polled mode.
Custom Mode Custom allows for user-defined communication (Ambles to Source, Echo to Source, Output at End of Read Cycle, Output at ETX, Output at Timeout).
FIS-6801 Industrial Scanner User’s Manual
5-35
Page 138
Communication
Destination Port Determines the port to which data will be sent.
Ambles to Source Enables or Disables the ability to send Preambles and Postambles to the Source port.
Echo to Source Enables or Disables the ability to send an Echo to the Source port.
Output at End of Read Cycle Enables or Disables the ability to output data at the end of read cycle.
5-36
FIS-6801 Industrial Scanner User’s Manual
Page 139
Parameters
Output at ETX Determines the output at ETX.
Output at Timeout Determines the Timeout value for output.
FIS-6801 Industrial Scanner User’s Manual
5-37
Page 140
Communication
Array Communication Modes
Mode Daisy Chain When set to Daisy Chain, follows wenglor Daisy Chain protocol. Note: Daisy Chain can also be autoconfigured by sending the Daisy Chain Autoconfigure serial command.
Source Defines the communication port.
Daisy Chain ID Status When enabled, the Scanner will append a two-character prefix to each Scanner in the array. This allows the user to identify which Scanner sent the data. Note: Daisy Chain ID will automatically disable Symbol Data Output, Extra Symbol Information, and Diagnostics Output for the Source Port. Symbol Data Output will be automatically enabled, Extra Symbol Information will be automatically enabled, and Diagnostics Output will be automatically disabled for the Destination Port.
5-38
FIS-6801 Industrial Scanner User’s Manual