Page 81

Calibration

Calibration Progress Output to Terminal When initiated by serial command or EZ button, the autocalibration operation will output progress and process information to a terminal if the system is configured to do so.

Terminal Depending upon how the calibration options are configured, the autocalibration process can have up to 5 steps (Quick Focus, Search Pass, Focus Pass, Medium Pass, Fine-Tune Pass). Calibration data is shown in a simple table format under 8 category headings, each of which are defined below. Heading Prog Rating Decode Quality Focus Shutter Gain Locate

Definition Progress indicator with a range of 0 - 100. Indicates percent complete. Rating indicator; higher ratings are more favorable. Number of successful decodes for the pass. Quality indicator; higher ratings are more favorable. Focus position under operation. Shutter setting under operation. Gain setting under operation. Time required to locate symbol, expressed in milliseconds.

Calibration Progress Output Examples (Terminal) Quick Focus This mode is only run if Quick Focus calibration mode is enabled (see Focus Position on page 4-5). Quick Focus calibration mode is designed to quickly locate the focus setting for an object at the center of the Imager’s field of view. There is no terminal output for this operation.

Search Pass The purpose of Search Pass is to decode a symbol in as few configurations as possible during calibration. This process is designed to quickly identify symbology type, and any image processing parameters that are already calibrated, as well as determining focus position. Running Search Pass Prog

Rating

Decode

Quality

Focus

Shutter

Gain

Locate

12 14 16 18

0 0 0 2174

0 0 0 1

0 0 0 72

313 313 313 313

250 250 250 250

0 9 18 27

0 0 0 24

FIS-6300 ESD Imager User’s Manual

4-15

Page 82

Initiating Calibration

Focus Pass The purpose of Focus Pass is to determine the optimal focal position of the optical system. This is accomplished by determining the inside and outside focus positions and then recording the center position as the desired focal position. Running Focus Pass Prog

Rating

Decode

Quality

Focus

Shutter

Gain

Locate

38 38 38 38 38 38 38 38 38 38 38 38 38 38 38 38 38 38 38 38 42 42 42 42 42 42 42 42 42 42 46

0 0 4256 0 4236 0 4236 0 4232 0 2128 0 0 2144 0 0 0 0 0 0 4280 4280 4280 4280 4280 4280 4280 4280 4280 4280 0

0 0 2 0 2 0 2 0 2 0 1 0 0 1 0 0 0 0 0 0 2 2 2 2 2 2 2 2 2 2 0

0 0 48 0 28 0 28 0 24 0 24 0 0 40 0 0 0 0 0 0 72 72 72 72 72 72 72 72 72 72 0

353 353 353 409 409 447 447 492 492 536 536 572 572 572 601 601 601 601 601 601 297 283 270 258 247 236 227 217 209 200 601

250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250 250

0 9 18 0 9 0 9 0 9 0 9 0 9 18 27 36 45 45 45 45 45 45 45 45 45 45 45 45 45 45 45

0 0 5 0 5 0 4 0 4 0 5 0 0 4 0 0 0 0 0 0 6 5 5 5 5 6 5 5 6 6 0

Inside Focus = 200; Outside Focus = 572; Center Focus = 386

4-16

FIS-6300 ESD Imager User’s Manual

Page 83

Calibration

Medium Pass The theory behind Medium Pass is to develop a readable range from which to dial in a decoded symbol. Running Medium Pass Prog

Rating

Decode

Quality

Focus

Shutter

Gain

Locate

51 53 55 57 59 61 63 65 67 69 71 73 75

4216 8436 6336 8448 8456 8464 8472 8488 8504 8512 8528 8584 8644

2 4 3 4 4 4 4 4 4 4 4 4 4

8 20 24 32 40 48 56 72 88 96 112 168 228

386 386 386 386 386 386 386 386 386 386 386 386 386

250 250 250 250 250 250 250 250 250 250 250 250 250

0 4 8 12 16 20 24 28 32 36 40 44 48

5 4 4 4 5 4 5 5 5 5 6 5 5

Rating List (Best): Rating = 8606; Focus = 386; Shutter = 250; Gain = 48 Rating = 8581; Focus = 386; Shutter = 250; Gain = 44 Rating = 8553; Focus = 386; Shutter = 250; Gain = 40 Rating = 8527; Focus = 386; Shutter = 250; Gain = 36 Rating = 8505; Focus = 386; Shutter = 250; Gain = 32 Rating = 8488; Focus = 386; Shutter = 250; Gain = 28 Rating = 8476; Focus = 386; Shutter = 250; Gain = 24

FIS-6300 ESD Imager User’s Manual

4-17

Page 84

Initiating Calibration

Fine-Tune Pass The theory behind Fine-Tune Pass is to dial in a decoded symbol based on the range supplied by the Medium Pass. Running Fine Tune Prog

Rating

Decode

Quality

Focus

Shutter

Gain

Locate

77 79 81 84 86 88 90 93 95 97 100

6416 6424 6424 6432 6448 6464 6476 6488 6506 6524 6538

3 3 3 3 3 3 3 3 3 3 3

104 112 112 120 136 152 164 176 194 212 226

386 386 386 386 386 386 386 386 386 386 386

250 250 250 250 250 250 250 250 250 250 250

38 39 40 41 42 43 44 45 46 47 48

7 6 6 6 5 5 5 5 6 5 6

Rating List (Best): Rating = 6527; Focus = 386; Shutter = 250; Gain = 48 Rating = 6517; Focus = 386; Shutter = 250; Gain = 47 Rating = 6505; Focus = 386; Shutter = 250; Gain = 46 Rating = 6491; Focus = 386; Shutter = 250; Gain = 45 Rating = 6476; Focus = 386; Shutter = 250; Gain = 44 Rating = 6462; Focus = 386; Shutter = 250; Gain = 43 Rating = 6449; Focus = 386; Shutter = 250; Gain = 42 Calibration PASSED.

4-18

FIS-6300 ESD Imager User’s Manual

Page 85

Calibration

Additional Notes about Calibration The following conditions apply to the Imager’s calibration process. Some of these items are noted at various points throughout this section, or in other sections of the Imager’s documentation. 1. The Mirrored Image parameter <K514> is not part of the calibration process. 2. Background Color <K451> is calibrated. 3. If Window of Interest (WOI) Framing is enabled, the WOI will be set to full frame when calibration begins. If WOI Framing is disabled, the current WOI configuration will be used for the Search Pass. 4. IP Mode <K527> will not be altered during calibration. 5. Threshold Mode <K512> will be set to Adaptive during calibration. At the end of a successful calibration, Threshold Mode will remain set to Adaptive regardless of its previous setting. Otherwise, the value will revert to its original state. 6. If Symbol Type is enabled for calibration (Autodiscriminate), then Interleaved 2 of 5 Range Mode Status (Interleaved 2 of 5) <K472> will be enabled. This allows variablelength Interleaved 2 of 5 symbols to be decoded. If an Interleaved 2 of 5 symbol is decoded during calibration, then code length # 1 will be set to the decoded symbol length at the end of calibration. Otherwise, the symbol lengths will be restored to their original configurations. 7. Pharmacode is not calibrated. 8. The symbol size, dimension, and orientation parameters of the 2D Symbology Qualification command <K718> will be updated if calibration is completed successfully. 9. All symbol types that were enabled before calibration will still be enabled after calibration. For example, if Data Matrix ECC 200 was enabled before calibration, and the calibration routine was performed on a Code 128 symbol, then after calibration is completed both Data Matrix ECC 200 and Code 128 will be enabled. 10. The calibration process does not modify the global Composite status <K453>. The global Composite status must be configured properly before calibration. 11. If the user requires calibration of an RSS Expanded or RSS-14 symbology <K482> or <K484>, that symbology must be enabled and configured appropriately before calibration. 12. The Search process will use the configurable Window of Interest for image captures. However, when the search process is completed, the WOI will be reduced to include only the symbol of interest and some additional boundary area.

FIS-6300 ESD Imager User’s Manual

4-19

Page 86

Additional Notes about Calibration

4-20

FIS-6300 ESD Imager User’s Manual

Page 87

5 Read Cycle Contents Read Cycle by ESP ...................................................................................................................... 5-2 Read Cycle Serial Commands...................................................................................................... 5-3 Read Cycle Setup......................................................................................................................... 5-4 Multisymbol................................................................................................................................... 5-5 Trigger Mode and Duration........................................................................................................... 5-6 External Trigger Polarity ............................................................................................................. 5-11 Serial Trigger .............................................................................................................................. 5-12 Start Trigger Character ............................................................................................................... 5-13 Stop Trigger Character ............................................................................................................... 5-14 End of Read Cycle...................................................................................................................... 5-15 Capture Mode ............................................................................................................................. 5-17 Capture Timing ........................................................................................................................... 5-22 Image Processing Timeout ......................................................................................................... 5-24 Image Storage ............................................................................................................................ 5-25 Minimum Good Reads ................................................................................................................ 5-27

After you’ve established communications and completed basic read rate testing, you will need to address the spatial and timing parameters associated with your application. This section explains those parameters.

FIS-6300 ESD Imager User’s Manual

5-1

Page 88

Read Cycle by ESP

Read Cycle by ESP

Click this button to bring up the Read Cycle menu. To change a setting, double-click the setting and use your cursor to scroll through the options.

To open nested options, single-click the +.

5-2

FIS-6300 ESD Imager User’s Manual

Page 89

Read Cycle

Read Cycle Serial Commands Trigger Mode/Duration Serial Trigger Character External Trigger State End of Read Cycle Minimum Good Reads Multisymbol Start Trigger Character Stop Trigger Character Capture Mode Capture Timing Image Storage IP Timeout

<K200,trigger mode,trigger filter duration>

<K201,serial trigger character> <K202,external trigger state> <K220,end of read cycle,read cycle timeout> <K221,minimum good reads> <K222,number of symbols,multisymbol separator> <K229,start character> <K230,stop character> <K241,capture mode,number of captures,rapid capture mode, pre-processing> <K242,time before first capture,time between first and second captures,,,,,,,,time between seventh and eighth captures> <K244,image storage type,image store mode> <K245,image processing timeout>

FIS-6300 ESD Imager User’s Manual

5-3

Page 90

Read Cycle Setup

Read Cycle Setup Setting up read cycle and triggering parameters will involve a series of decisions based on your particular application, as follows: 1. Select the number of symbols to be read in a single cycle. The FIS-6300 ESD can read multiple symbols in a single image frame. 2. Decide on the trigger type to be used: if serial, choose a serial character; if external, choose either Level or Edge. 3. Designate how the read cycle should end (Timeout, New Trigger, Last Frame). 4. Select Capture Mode, Continuous Mode, or Rapid Mode. 5. Select Number of Captures (if in Rapid Capture Mode). 6. Set the Time Before First Capture and Time Between Captures, if any. Note: The Imager’s camera captures images at a rate of about 16/second (based on a full-size 1280 x 1024 image). The capture rate increases as frame sizes decrease.

5-4

FIS-6300 ESD Imager User’s Manual