Page 61
Communications
LRC Status (Longitudinal Redundancy Check) Usage: Definition:
Serial Cmd: Default: Options:
Used when extra data integrity is required. An error-checking routine that verifies the accuracy of transmissions. It is the exclusive OR of all characters following the STX (start of text) up to and including the ETX (end of text). What this means is that the binary representation of all the characters in a transmission are cumulatively added in a column and each resulting odd integer is assigned a 1 and each even integer a 0 (two 1s = 0, two 0s = 0, a 1 and a 0 = 1). The extra LRC character is then appended to the transmission, and the receiver (usually the host) performs the same addition and compares the results. <K145,status> Disabled 0 = Disabled 1 = Enabled
FIS-6300 ESD Imager User’s Manual
3-21
Page 62
Protocol Configuration Examples
Protocol Configuration Examples Point-to-Point (Main Port) <K100,8,0,1,1> <K140,0> <K102,0>
Baud Rate: 115.2K; Parity: None; Stop Bits: 2; Data Bits: 8 Point-to-Point RS-232 enabled; <K102,1> = RS-422 enabled
Polling Mode (Main Port) <K100,4,0,1,1> <K140,5,23> <K102,0> <K143,30>
Baud Rate: 9600; Parity: None; Stop Bits: 2; Data Bits: 8 Polling Mode; Address: 23 RS-232 Point-to-Point polling; <K102,1> = RS-422/RS-485 true multidrop (switched transmitter) 30 mS Response Timeout
“User-Defined” Polling Mode (Main Port) <K100,4,0,1,1> Baud Rate: 9600; Parity: None; Stop Bits: 2; Data Bits: 8 <K140,5,12> Polling Mode; Address: 12 <K148,,08,09,18,0B,0C,0D> Default RES (0x04), REQ=0x08; EOT=0x09; STX=0x18; ETX=0x0B; ACK=0x0C; NAK=0x0D <K102,0> RS-232 Point-to-Point polling; <K102,1> = RS-485 (true multidrop with switched transmitter) <K143,40> 40 mS Response Timeout
ACK / NAK (Main Port) <K100,9,0,1,1> <K140,4> <K147,,,01,1B,2E,1F> <K102,1> <K143,50>
Baud Rate: 230K; Parity: None; Stop Bits: 2; Data Bits: 8 ACK/NAK Default RES and REQ (00, disabled); STX=0x01; ETX=0x1B; ACK=0x2E; NAK=0x1F RS-422 enabled; <K102,0> = RS-232 enabled 50 mS Response Timeout
Auxiliary Port Listed below are several examples of Auxiliary Port commands. <K101,2,4,0,1,1> <K101,1,5,0,1,1> <K101,3,7,0,1,1> <K101,5,6,0,1,1>
3-22
Half Duplex; Baud Rate: 9600; Parity: None; Stop Bits: 2; Data Bits: 8 Transparent; Baud Rate: 19.2K; Parity: None; Stop Bits: 2; Data Bits: 8 Full Duplex; Baud Rate: 57.6K; Parity: None; Stop Bits: 2; Data Bits: 8 Command Processing; Baud Rate: 230K; Parity: None; Stop Bits: 2; Data Bits: 8 FIS-6300 ESD Imager User’s Manual
Page 63
Communications
ASCII Character Entry Modifier Commands that require ASCII text fields, such as Preamble and Postamble commands, can be sent to the Imager as hex pairs (see the ASCII Conversion Values table in Appendix E for conversions). Serial Cmd Format: <Knnnh,00-FF> To enter ASCII fields as hex values (00 to FF), add a lower-case h directly after the command’s K number, and then enter the hex value that corresponds with the desired ASCII character. Example: Consider the Postamble command: Serial Cmd:
<K142,status,postamble character(s)>
Imagine that your application requires the ASCII character > to be the postamble in your symbol decode output. The ASCII characters <, >, and , can only be entered as hex pairs. So, to make > the postamble in your symbol decode output, enter the Postamble command as follows: <K142h,,3E> Note that the “status” field contains only a , . This is because the only field that is being changed is the “postamble character(s)” field. (See Serial Configuration Command Conventions on page A-10 for a more detailed explanation of this command shortcut.)
FIS-6300 ESD Imager User’s Manual
3-23
Page 64
Preamble
Preamble Preamble Status Usage: Definition: Serial Cmd: Default: Options:
Useful for identifying and controlling incoming data. For example, defining the preamble as a carriage return and a line feed causes each decoded message to be displayed on its own line. Defines a one to four character data string that can be added to the front of the decoded data. <K141,status,preamble character(s)> Disabled 0 = Disabled 1 = Enabled (within any protocol)
Preamble Characters Serial Cmd: Default: Options:
<K141,status,preamble character(s)> ^M corresponds to: carriage return. To enter control characters within a serial command, hold down the Ctrl key while typing the desired character. Example: <K141,1,CNTL-m> to enter the control character ^M.
Important: The ASCII characters <, >, and , can only be entered as hex pairs: For <: <K141h,,3C> For >: <K141h,,3E> For , : <K141h,,2C> See ASCII Character Entry Modifier on page 3-23 for a detailed explanation of how to enter ASCII characters as hex values.
3-24
FIS-6300 ESD Imager User’s Manual
Page 65
Communications
Postamble Postamble Status Usage: Definition: Serial Cmd: Default: Options:
Useful for identifying and controlling incoming data. For example, defining the postamble as a carriage return and a line feed causes each decoded message to be displayed on its own line. Allows the user to enable or disable up to four postamble characters that can be added to the end of the decoded data. <K142,status,postamble character(s)> Enabled 0 = Disabled 1 = Enabled (within any protocol)
Postamble Characters Serial Cmd: Default: Options:
<K142,status,postamble character(s)> ^M^J corresponds to: carriage return/line feed. To enter control characters within a serial command, hold down the control key while typing the desired character. Example: <K142,1,CNTL-m CNTL-j> to enter ^M^J.
Important: The ASCII characters <, >, and , can only be entered as hex pairs: For <: <K142h,,3C> For >: <K142h,,3E> For , : <K142h,,2C> See ASCII Character Entry Modifier on page 3-23 for a detailed explanation of how to enter ASCII characters as hex values.
FIS-6300 ESD Imager User’s Manual
3-25
Page 66
Postamble
3-26
FIS-6300 ESD Imager User’s Manual
Page 67
4 Calibration Contents Calibration Serial Commands ....................................................................................................... 4-2 Calibration Overview..................................................................................................................... 4-2 Calibration Options ....................................................................................................................... 4-3 Calibration by ESP..................................................................................................................... 4-11 Initiating Calibration ................................................................................................................... 4-13 Additional Notes about Calibration............................................................................................. 4-19
This section shows the FIS-6300 ESD Imager’s calibration options, and explains the different ways that those options can be configured.
FIS-6300 ESD Imager User’s Manual
4-1
Page 68
Calibration Serial Commands
Calibration Serial Commands Calibration Options Autocalibrate
<K529,gain,shutter speed,focus position,symbol type,WOI framing,WOI margin,line scan height,processing> <@CAL>
Calibration Overview Autocalibration is one of the most powerful features of the FIS-6300 ESD. The calibration process can be initiated by serial command, EZ button, or ESP user interface. When the Imager enters calibration, it runs through an optimization cycle that sets the ideal parameters for reading symbols at the highest possible level of performance. Autocalibration can be specially configured to optimize specific parameters, such as focus, gain, shutter speed, and symbol type. The calibration process consists of five separate phases: quick focus, search pass, focus pass, medium pass, and fine-tune pass. During this process the Imager quickly locates configuration setup, defines a calibration range for the configuration parameters, and then tunes those parameters for optimal performance.
4-2
FIS-6300 ESD Imager User’s Manual
Page 69
Calibration
Calibration Options This command specifies the operation of the Imager’s calibration feature. The default configuration is set up to perform calibration on gain, focus, and symbol type. The default shutter speed setting is disabled, because not all applications require shutter speed, and when it is required, the application-specific settings tend to vary widely. Note: Background Color (page 6-35) is part of the calibration process regardless of the calibration setup. After successful completion of the calibration routine, the background color will be set to the appropriate value, depending on the symbol in the field of view. The Mirrored Image parameter (page 10-28) is not part of the calibration process, and it must be configured appropriately prior to calibration.
Gain Definition: Serial Cmd: Default: Options:
When enabled, gain will be calibrated to provide the best available image quality and performance. When disabled, gain is fixed and is not part of the calibration process. <K529,gain,shutter speed,focus position,symbol type,WOI framing,WOI margin,line scan height,processing> Enabled 0 = Disabled 1 = Enabled
FIS-6300 ESD Imager User’s Manual
4-3
Page 70
Calibration Options
Shutter Speed Definition:
Unless an application is a static setup, the shutter speed setting should be configured by the user based on the application. For dynamic applications, the user should configure the shutter speed setting so the calibration process can optimize the gain for that setting. Serial Cmd: <K529,gain,shutter speed,focus position,symbol type,WOI framing,WOI margin,line scan height,processing> Default: Disabled Options: 0 = Disabled 1 = Enabled 2 = Fast Shutter Note: The following table shows general guidelines for shutter speed configurations based on various line speeds. The configurations shown depend on the Imager’s optical configuration, and on symbol element size. Shutter Speed 0 - 250 250 - 750 750 - 1500 1500 - 2000 2000 - 2500
Line Speed static 5” / sec 10” / sec 15” / sec 20” / sec
Disabled When this feature is disabled, the shutter speed is fixed and is not part of the calibration process.
Enabled When this feature is enabled, shutter speed will be calibrated to provide the best possible image quality and performance.
Fast Shutter The calibration process will concentrate on achieving the fastest possible shutter setting that will still provide good performance. The image quality or contrast may not be as good as what would be achieved with the Enabled setting. The calibration process is not designed to choose the fastest shutter speed that can decode a symbol, but rather to optimize for the fastest shutter speed that still provides good image quality.
4-4
FIS-6300 ESD Imager User’s Manual