FIS-0870 Industrial Sweep Raster Scanner User's Manual: Utilities, pages 298 to 307
Pages 298 to 307 of 381, from FIS-0870 Industrial Sweep Raster Scanner User's Manual
The original document is the PDF (4.9 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Wenglor FIS-0870
Page 298
Appendices Master Symbol Database Size New Master Pin Sequence Step Master Symbol Match Replace Mismatch Replace Diagnostics High Temperature Threshold Low Temperature Threshold Counts (Read-only) Hours Since Reset (Read-only) Service Message Laser Current Warning Message User-Defined Name Output Format Format Extract Format Insert Format Assign Output Format Status Output Filter Configuration Output Filter Enable
<K224,number of master symbols> <K225,status> <K228,sequence step> <K231,index,master symbol data> <K735,status,replacement string> <K736,status,replacement string> <K402,status,message> <K403,status,message> <K406> (returns: power-on,resets, power-on saves,custom default saves) <K407> (returns: hours,minutes) <K409,status,service message, threshold,resolution> <K411,laser high status,laser high message,laser low status, laser low message> <K412,user-defined name> <K740,output index,start location,length> <K741,output index,length,hex string> <K742,symbol number,status> <K743,output format status> <K744,filter number,symbology,length,wildcard,placeholder, data,decode direction,database index> <K745,number of filters>
FIS-0870 Industrial Raster Scanner User’s Manual
A-11
Page 299
Serial Command Format wenglor readers are controlled by two types of serial commands: configuration commands and utility commands. Rules that apply to both configuration and utility commands • Less than ‘<‘ and greater than ‘>’ angle bracket characters enclose the commands. • Commands and data are case sensitive. Characters must be entered as upper or lower case, as specified. Serial Utility Commands Serial Utility Commands are sent during operations and are not followed by <A> or <Z>. Serial Configuration Commands (K Commands) wenglor’s serial configuration commands begin with a single “K” character followed by a 3digit numeric character, comma-separated command fields, and an initializing command, as follows: <Knumeric character,data,data,...etc.><initializing command> An initializing command <Z> or <A> may follow the command. • <Z> initializes the scanner’s memory and saves for power-on. • <A> initializes the scanner’s memory but does not save for power-on. For example, to enable UPC and save the change for power-on, send <K473,1><Z>. To change Baud Rate and reset without saving changes for power-on, send <K100,3><A>.
Serial Configuration Command Conventions • All command fields (except the last) must be followed by a comma (without a space). • NULL cannot be used. The characters <, >, and , can be used, but only if entered as hex pairs (see ASCII Character Entry Modifier). • All fields preceding a modified field must be included. • If there is no change in preceding fields, then commas alone can be entered in these fields. For example, if only the last field in the following command is changing, <K100,4,1,0,0> can be entered as <K100,,,,0>. • All fields following a modified field can be omitted. For example, to change Baud Rate only, send <K100,3>.
A-12
FIS-0870 Industrial Raster Scanner User’s Manual
Page 300
Appendices
Concatenating Configuration Commands Commands can be concatenated (added together) in a single string. For example, <K145,1><K220,1><K450,1><A> enables LRC, sets End of Read Cycle mode to New Trigger, enables Narrow Quiet Zone, and resets the data buffers without saving the changes for power-on.
Serial Command Status Request To ensure that any command was received and accepted, send the Show Reader Status command: <?>. The status of a specific serial command can be requested by entering the command followed by a question mark. For example, send <K142?> to request the status of Postamble.
Entering Control Characters in Serial Commands To enter control characters within a serial command, hold down the Ctrl key while typing the desired character. Example: To enter a carriage return and line feed (^M^J), enter <K141,1,CNTL-m CNTL-j>
FIS-0870 Industrial Raster Scanner User’s Manual
A-13
Page 301
Serial Configuration Commands The following serial commands can be entered through ESP’s Terminal to control FIS0870 functions. Detailed descriptions of command parameters are available in Scanner Parameters.
Communication RS-232 A RS-232 B RS-422 Ethernet Ethernet TCP Ports EtherNet/IP RS-232 A Data Type RS-232 B Data Type RS-422 Data Type Ethernet TCP Port 1 Data Type Ethernet TCP Port 2 Data Type EtherNet/IP Data Type Preamble Postamble Response Timeout LRC Status ACK/NAK Options Polling Mode Options Autoconfiguration Daisy Chain Protocol Selection External Data Routing Array Communication Modes
<K100,baud rate,parity,stop bits,data bits> <K101,status,baud rate,parity,stop bits,data bits> <K102,status,baud rate,parity,stop bits,data bits> <K126,status,IP address,subnet,gateway,IP address mode> <K127,TCP Port 1,TCP Port 2> <K129,status> <K130,symbol data output,extra symbol information,diagnostic output,external source processing mode> <K131,symbol data output,extra symbol information,diagnostic output,external source processing mode> <K132,symbol data output,extra symbol information,diagnostic output,external source processing mode> <K133,symbol data output,extra symbol information,diagnostic output,external source processing mode> <K134,symbol data output,extra symbol information,diagnostic output,external source processing mode> <K136,symbol data output,extra symbol information,diagnostic output,external source processing mode> <K141,status,preamble> <K142,status,postamble> <K143,response timeout> <K145,status> <K147,RES,REQ,STX,ETX,ACK,NAK> <K148,RES,REQ,STX,ETX,ACK,NAK> <K150DAISY> <K160,protocol,address,protocol port> <K161,mode,destination port,ambles to source,echo to source,output at end of read cycle,output at ETX,output at timeout> <K162,mode,source,daisy chain i.d. status,daisy chain i.d.>
RS-232 A Baud Rate, RS-232 A Serial Cmd: Default: Options:
A-14
<K100,baud rate,parity,stop bits,data bits> 8 = 115.2K 0 = 600 1 = 1200 3 = 4800 4 = 9600 6 = 38.4K 7 = 57.6K 9 = 230K
2 = 2400 5 = 19.2K 8 = 115.2K
FIS-0870 Industrial Raster Scanner User’s Manual
Page 302
Appendices Parity, RS-232 A Serial Cmd: Default: Options:
<K100,baud rate,parity,stop bits,data bits> 0 = None 0 = None 1 = Even
2 = Odd
Stop Bits, RS-232 A Serial Cmd: Default: Options:
<K100,baud rate,parity,stop bits,data bits> 0 = One 0 = One 1 = Two
Data Bits, RS-232 A Serial Cmd: Default: Options:
<K100,baud rate,parity,stop bits,data bits> 1 = Eight 0 = Seven 1 = Eight
RS-232 B Status, RS-232 B Serial Cmd: Default: Options:
<K101,status,baud rate,parity,stop bits,data bits> 1 = Enabled 0 = Disabled 1 = Enabled
Baud Rate, RS-232 B Serial Cmd: Default: Options:
<K101,status,baud rate,parity,stop bits,data bits> 8 = 115.2K 0 = 600 1 = 1200 2 = 2400 3 = 4800 4 = 9600 5 = 19.2K 6 = 38.4K 7 = 57.6K 8 = 115.2K 9 = 230K
Parity, RS-232 B Serial Cmd: Default: Options:
<K101,status,baud rate,parity,stop bits,data bits> 0 = None 0 = None 1 = Even 2 = Odd
Stop Bits, RS-232 B Serial Cmd: Default: Options:
<K101,status,baud rate,parity,stop bits,data bits> 0 = One 0 = One 1 = Two
FIS-0870 Industrial Raster Scanner User’s Manual
A-15
Page 303
Data Bits, RS-232 B Serial Cmd: Default: Options:
<K101,status,baud rate,parity,stop bits,data bits> 1 = Eight 0 = Seven 1 = Eight
RS-422 Status, RS-422 Serial Cmd: Default: Options:
<K102,status,baud rate,parity,stop bits,data bits> 0 = Disabled 0 = Disabled 1 = Enabled
Baud Rate, RS-422 Serial Cmd: Default: Options:
<K102,status,baud rate,parity,stop bits,data bits> 8 = 115.2K 0 = 600 1 = 1200 2 = 2400 3 = 4800 4 = 9600 5 = 19.2K 6 = 38.4K 7 = 57.6K 8 = 115.2K 9 = 230K
Parity, RS-422 Serial Cmd: Default: Options:
<K102,status,baud rate,parity,stop bits,data bits> 0 = None 0 = None 1 = Even 2 = Odd
Stop Bits, RS-422 Serial Cmd: Default: Options:
<K102,status,baud rate,parity,stop bits,data bits> 0 = One 0 = One 1 = Two
Data Bits, RS-422 Serial Cmd: Default: Options:
<K102,status,baud rate,parity,stop bits,data bits> 1 = Eight 0 = Seven 1 = Eight
Ethernet Serial Cmd: Default: Options:
A-16
<K126,status,IP address,subnet,gateway,IP address mode> 1 = Enabled 0 = Disabled 1 = Enabled
FIS-0870 Industrial Raster Scanner User’s Manual
Page 304
Appendices IP Address Serial Cmd: Default: Options:
<K126,status,IP address,subnet,gateway,IP address mode> 192.168.0.100 0.0.0.0 – 255.255.255.255
Subnet Serial Cmd: Default: Options:
<K126,status,IP address,subnet,gateway,IP address mode> 255.255.0.0 0.0.0.0 – 255.255.255.255
Gateway Serial Cmd: Default: Options:
<K126,status,IP address,subnet,gateway,IP address mode> 0.0.0.0 0.0.0.0 – 255.255.255.255
IP Address Mode Serial Cmd: Default: Options:
<K126,status,IP address,subnet,gateway,IP address mode> 1 = DHCP 0 = Static 1 = DHCP
Ethernet TCP Ports TCP Port 1 Serial Cmd: Default: Options:
<K127,TCP Port 1,TCP Port 2> 2001 1024 – 65535
TCP Port 1 Serial Cmd: Default: Options:
<K127,TCP Port 1,TCP Port 2> 2003 1024 – 65535
EtherNet/IP Serial Cmd: Default: Options:
<K129,status> 1 = Enabled 0 = Disabled
1 = Enabled
FIS-0870 Industrial Raster Scanner User’s Manual
A-17
Page 305
RS-232 A Data Type Symbol Data Output, RS-232 A Serial Cmd: Default: Options:
<K130,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
Extra Symbol Information, RS-232 A Serial Cmd: Default: Options:
<K130,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
Diagnostic Output, RS-232 A Serial Cmd: Default: Options:
<K130,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
External Source Processing Mode, RS-232 A Serial Cmd: Default: Options:
<K130,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Command 0 = Disabled 1 = Command 2 = Data
RS-232 B Data Type Symbol Data Output, RS-232 B Serial Cmd: Default: Options:
<K131,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
Extra Symbol Information, RS-232 B Serial Cmd: Default: Options:
A-18
<K131,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
FIS-0870 Industrial Raster Scanner User’s Manual
Page 306
Appendices Diagnostic Output, RS-232 B Serial Cmd: Default: Options:
<K131,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
External Source Processing Mode, RS-232 B Serial Cmd: Default: Options:
<K131,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Command 0 = Disabled 1 = Command 2 = Data
RS-422 Data Type Symbol Data Output, RS-422 Serial Cmd: Default: Options:
<K132,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
Extra Symbol Information, RS-422 Serial Cmd: Default: Options:
<K132,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
Diagnostic Output, RS-422 Serial Cmd: Default: Options:
<K132,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
External Source Processing Mode, RS-422 Serial Cmd: Default: Options:
<K132,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Command 0 = Disabled 1 = Command 2 = Data
FIS-0870 Industrial Raster Scanner User’s Manual
A-19
Page 307
Ethernet TCP Port 1 Data Type Symbol Data Output, Ethernet TCP Port 1 Serial Cmd: Default: Options:
<K133,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
Extra Symbol Information, Ethernet TCP Port 1 Serial Cmd: Default: Options:
<K133,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
Diagnostic Output, Ethernet TCP Port 1 Serial Cmd: Default: Options:
<K133,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
External Source Processing Mode, Ethernet TCP Port 1 Serial Cmd: Default: Options:
<K133,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Command 0 = Disabled 1 = Command 2 = Data
Ethernet TCP Port 2 Data Type Symbol Data Output, Ethernet TCP Port 2 Serial Cmd: Default: Options:
<K134,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
Extra Symbol Information, Ethernet TCP Port 2 Serial Cmd: Default: Options:
A-20
<K134,symbol data output,extra symbol information,diagnostic output, external source processing mode> 1 = Enabled 0 = Disabled 1 = Enabled
FIS-0870 Industrial Raster Scanner User’s Manual