Browse
(440) 299-7778
Menu
(440) 299-7778
One-year parts and labor warranty30-day money-back guaranteeEnglishEspañolContact

The text of a manual · part 7 of 7

Webscan Barcode Verifiers TruCheck User Manual, pages 61 to 68

Pages 61 to 68 of 68

The original document is the PDF (5.0 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

Filed in the catalog under the Webscan TruCheck TC-418, Webscan TruRemote TC-403, Cognex TruCheck TC-418 and Cognex TruRemote TC-403

Page 61

The reporting for RSS symbols is different than other symbologies when there is more than one component. Each component has its own grade. The grade of each component will affect the grade for the whole symbol, where the overall symbol grade is the lowest grade scored by any of its components. It is important to note that there is a special flag inside the data of RSS codes that identifies the presence of a secondary component. If this flag is set when there is no composite component found, the grade for the symbol is reported as “F” even if all other parameters receive high grades. This is reported as “link flag set, but composite component not found” in the primary report. The report will contain "Link Flag On" or "Link Flag Off". The report will also indicate whether the symbol is printed in one row, or "stacked".

QR Code, or Quick Response Code, is capable of encoding Japanese Kanji characters and is widely used in Asia. It contains easy to decode finder patterns, alignment patterns, and clock tracks. A small amount of version information is specified by a group of modules surrounding the finder patterns, such as the size of the matrix and the selected level of error correction. QR Code symbols can have varying amounts of error correction, allowing a tradeoff between symbol size and error correction capability to be made when the symbol is created. The general characteristics section will include information specific to QR Code, including the error correction mode and the masking pattern.

The Data Matrix symbology is probably the most common 2D Matrix. It allows for the highest density of data per module, especially for small symbols, so data matrix grading is essential. The finder pattern consists of solid left and bottom sides, which form an “L” pattern and a horizontal and vertical clock track on opposite sides of the “L”. The clock track tells the number of modules in the matrix and is used to decode the grid on which the modules are placed. There is no version information encoded explicitly within the Data Matrix symbology; all the information about the symbol size, encoding and error correction is determined from the size of the matrix. The matrix size shown in the general description section of the report will give the size of the matrix including the “L” patterns and clock tracks, and the size of the inner matrix of data encoding modules in parenthesis.

MaxiCode is a 2D hexagonal grid symbology. “MaxiCode” must be selected in the “type of code” section of the settings menu to decode MaxiCode symbols. Unlike most other barcoding symbologies, all MaxiCode symbols have a necessary specified size. Magnification or shrinkage of the symbol can cause a symbol to become unreadable, even if all elements are properly sized relative to each other. Magnification is not a graded parameter; however, magnification percentage will be reported in the general characteristics tab, along with a warning if the symbol is of a size not allowed in the specification.

Page 62

Data in MaxiCode symbols is encoded across 3 different channels, each with its own independent error correction budget, and each with a different intended interpretation. These three channels are referred to as the high priority message, and the first and second low priority messages. The error detection and correction information for each channel will be listed separately in the general characteristics tab. MaxiCode symbols can be encoded in any of 5 modes: Mode 2 - Mode 2 is often used in the shipping industry and it uses the high priority message to store structured information, including a numeric postal code, a country code, and a service class. The two low priority channels interleave to form an accompanying message. The high priority channel uses “enhanced” error correction, and both low priority channels use “standard”. Mode 3 - Mode 3 is often used in the shipping industry and it uses the high priority message to store structured information including an alphanumeric postal code, a country code, and a service class. The two low priority channels interleave to form an accompanying message. The high priority channel uses “enhanced” error correction, and both low priority channels use “standard”. Mode 4 - Mode 4 uses the full data capacity (including the high priority message channel) to encode a message. The high priority channel uses “enhanced” error correction, and both low priority channels use “standard”. Mode 5 - Mode 5 uses the full data capacity (including the high priority message) to encode a message. In mode 5, all channels use “enhanced” error correction. This means that mode 5 symbols have a greater error detection and correction capacity, but at the cost of a smaller data capacity compared to mode 4 symbols. Mode 6 - Mode 6 symbols encode data used to reprogram reading devices. The high priority channel uses “enhanced” error correction, and both low priority channels use “standard”.

Aztec Code is a 2D symbol with a bull’s-eye pattern at the center to locate the code. The orientation marks at the corners of the code allow the code to be read if rotated in any direction. With an Aztec Code reader, the amount of Reed-Solomon error correction is between 20% and 40% of the data in the code. Like other 2D Matrix symbologies, the size of the Aztec Code depends on the amount of data contained within the code.

Page 63

Webscan strives to provide our customers with the best available service and support in the industry. Our knowledgeable bar code experts are available to assist you with interpretations of verification results and use of the product or software features.

If you need service for your TruCheck™ system, please call Webscan Technical Support at (303) 485-6811. Please be near the TruCheck™ system when you call and have your calibration card available. You may be instructed by Webscan to calibrate the system or to go into the “Advanced” menu in an effort to find and/or fix the problem. If factory service is required for the system, you will be given an RMA # (Return Material Authorization Number) that Webscan will use to track and record service activity. You do not need to ship your power supply unless instructed. Please reference your RMA number on your PO and shipping documents. You can apply for an RMA number on our website, under the resources tab. The address to send equipment to is: Webscan Inc. Attn: RMA#_______ (put in your assigned RMA#) 1341 Sherman Drive Unit D Longmont, CO 80501 Tel (303) 485-6811

If you are having trouble with your Webscan system, please check a few basic things. 1. Turn the system off, then on. See if this corrects the problem. 2. Make sure the system is not in Negative Mode, unless you are in fact trying to verify a Negative. 3. Calibrate the unit. After calibration, verify the calibration barcode. See if the values given by the system match the values for the calibration card. 4. Make sure the Red Cross on your monitor is located within the Data Matrix symbol. 5. If you are not getting the LED’s to illuminate, be sure the power brick is connected well to the power cable. If these actions do not correct the problem, please call Webscan Technical support as described below. Phone: 877.WEBSCAN or 303.485.6811

Email: support@webscaninc.com

NOTE: It is often helpful to save diagnostic information using the “Debug Information” feature in the “Help” menu for email to support@webscaninc.com.

Page 64

Webscan, Inc. warrants that all materials, workmanship, and equipment will be free of defects and will function properly for a period of one (1) year from the date of the Warranty Issue Date under normal and proper installation, start-up, and use in accordance with Webscan’s instructions and user manual. Webscan agrees to repair, or at its option, to replace, modify, or refund the purchase price of the equipment, if proved to be defective in material or workmanship during the warranty period, provided that Buyer gives Seller prompt notice of such defect and returns the defective item to Seller, shipping charges prepaid. Seller shall not be liable for any damage resulting from improper operation, inadequate maintenance, repairs, modification, or changes made by Buyer without Seller’s consent.

NO CONSEQUENTIAL DAMAGES. WEBSCAN, INC. IS NOT RESPONSIBLE FOR ECONOMIC LOSS; PROFIT LOSS; OR SPECIAL, INDIRECT OR CONSEQUENTIAL DAMAGES, INCLUDING WITHOUT LIMITATION, LOSSES OR DAMAGES ARISING FROM RELIANCE ON THE PRODUCT AND ITS REPORTED MEASUREMENTS EVEN IN THE EVENT OF PRODUCT FAILURE FOR ANY REASON. WARRANTY IS NOT TRANSFERABLE. THIS WARRANTY IS NOT ASSIGNABLE AND APPLIES ONLY IN FAVOR OF THE ORIGINAL PURCHASER/USER TO WHOM DELIVERED. ANY SUCH ASSIGNMENT OR TRANSFER SHALL VOID THE WARRANTIES HEREIN MADE AND SHALL VOID ALL WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IMPROPER USAGE. WEBSCAN, INC. ASSUMES NO LIABILITY FOR PARTS OR LABOR COVERAGE FOR COMPONENT FAILURE OR OTHER DAMAGES RESULTING FROM IMPROPER USAGE, INSTALLATION, OR FAILURE TO MAINTAIN PRODUCT AS SET FORTH IN THE USER MANUAL OR OTHER INSTRUCTIONS. IMPROPER ELECTRICAL CONNECTIONS. WEBSCAN, INC. IS NOT RESPONSIBLE FOR THE REPAIR OR REPLACEMENT OF FAILED OR DAMAGED COMPONENTS RESULTING FROM ELECTRICAL POWER FAILURE, THE USE OF EXTENSION CORDS, LOW VOLTAGE, OR VOLTAGE DROPS TO THE UNIT. NO IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE: THERE ARE NO OTHER WARRANTIES, EXPRESSED, IMPLIED OR STATUTORY, EXCEPT THE ONE (1) YEAR PARTS & LABOR AS DESCRIBED ABOVE. THESE WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, INCLUDING IMPLIED WARRANTY AND MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. THERE ARE NO WARRANTIES WHICH EXTEND BEYOND THE DESCRIPTION ON THE FACE HEREOF.

This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

Page 65

Calibration Cards expire after 2 years from the date of issue. Calibration is important for industry standard compliance. If your calibration card is expired, contact Webscan for purchasing information. NOTE: If you have a 2D verifiers (such as a TruCheck Optima or TruCheck Omni), you can choose either a “2D Calibration Card” or a “Linear Calibration Card”. Either of them will allow you to calibrate the reflectance levels which is what calibration does. Importantly, calibration is not specific to the symbology and once the verifier is calibrated using either type of calibration target, the verifier is calibrated for both 1D and 2D barcodes. However, for 1D verifiers, such as TruCheck Laser USB which do not work on 2D barcodes, only a “1D Calibration barcode” can be used. The Webscan calibration card (part number 5136) contains both types of calibration target symbols and can be used to calibrate either 1D or 2D verifiers. The calibration target symbols on the Webscan Calibration Card are “Master” Symbols, similar to the master symbols on the GS1 UPC/EAN Card and GS1 Data Matrix Card. However, this card contains Master Grade Symbols only and does not contain intentionally produced “Conformance Test Symbols” with low grades. Meaning the Webscan calibration card can be used to calibrate and check your calibration values (Rmax and Rmin). Besides Calibration, many of the calibration target cards below can also be used for “Conformance Testing”. Conformance Test Symbols are barcodes produced with intentional defects that give lower than perfect grades. The GS1 Data Matrix calibration card contains Data Matrix symbols with intentional imperfections that you can measure with the verifier to check the reporting capabilities of the verifier and to document conformance to industry standards such as ISO/IEC 15415 and ISO/IEC 15426-2 and GS1 specifications. The GS1 UPC/EAN Card contains UPC/EAN symbols that contain intentional imperfections that you can measure with the verifier to check the reporting capabilities of the verifier and to document conformance to industry standards such as ISO/IEC 15416. Other industry standard calibration cards with linear (1D) barcode conformance tests are available. Call Webscan for more information, 303.485.6811

Page 66

The Webscan Barcode Seminar is a comprehensive learning course in barcode technology aimed at packaging engineers, graphics artists, and other professionals involved in the production of barcoded packaging. Whether you are new to barcoding or have been involved for years, if you want to truly understand the inner workings of barcodes, especially the newer symbologies such as GS1 DataBar (formerly RSS14), Data Matrix, Composite Symbology, and PDF 417, this seminar is for you! Taking you from the most basic concepts of barcode technology all the way through to the most advanced topics such as the actual methods used to scan, decode, and verify all types of barcodes, you will become an expert and a resource to your company. The course is broken into two main parts. The first, Basic Concepts, begins by introducing the basic elements that barcodes are composed of. A vocabulary of barcode terms is developed as the course progresses. The fundamental methodology used to verify barcodes is explained in detail. Also, the structure of common barcode symbologies such as Code 39, Interleaved 2 of 5, and Code 128 is explained in detail, including the mathematical methods used to decode them. The UPC or Universal Product Code is also fully explained along with several variations of the basic UPC symbology such as UPC-E, EAN-13 and EAN-8. The use of special encodation such as GS1/UCC/EAN-128 with Function 1 characters and Application Identifiers (AI’s) is also described. The second part of the course consists of advanced topics such as new 2D symbologies. The new GS1 DataBar and GS1 Composite Symbology family is explained fully, from the goals it seeks to achieve to the various formats allowed. Each of the GS1 DataBar formats, such as GS1 DataBar Limited, GS1 DataBar Stacked, GS1 DataBar with Composite, and many others are fully described, including character structure, check digit calculations, and data content. Data Matrix (ECC200) and PDF 417 are also fully described including structure, decoding methods and quality verification. The quality verification described follows the ISO15415 specification. Attendees will come away from this seminar with an understanding of the relationships between symbologies and the benefits of different symbologies as well as the challenges presented by each symbology to those involved in producing or scanning the codes. Attendees will be able to decipher truth from fiction in the marketing claims of vendors when in the process of evaluating products and be able to troubleshoot problems with barcode printing or scanning systems after implementation. The seminar can be customized to focus more on certain topics of special interest to the intended audience and less on other topics.

This seminar is not about the end use of barcodes such as in retail settings or warehouses. Although it is about the core technology of barcodes used in these settings, this course does not go beyond printing and reading the barcodes. This is also not about how to use any specific hardware or software. For information about arranging for the Webscan Barcode Seminar to be presented at your company, including pricing, please contact Webscan toll free at 877-WEBSCAN.

Page 67

Webscan Validation Overview & Objective Pharmaceutical and medical device manufacturers are very concerned about quality in the manufacturing and packaging of their products as part of their public health responsibilities. FDA regulations and guidance in addition to internal quality and professionalism standards drive part of the concern. Webscan fully supports our customers in their quest for best achievable quality in manufacturing practices. The FDA has issued many documents offering guidance on computer software validation principles generally directed to medical devices. More specifically "General Principles of Software Validation; Final Guidance for Industry and FDA Staff' was issued on January 11, 2002. It is Webscan’s goal to be fully compliant with FDA guidance on validation principles. The Webscan validation protocol is designed to help ensure the highest quality standards are being met and follows the previously mentioned guidance. This validation protocol is expected to be used in combination with the customer's own quality systems to assure themselves and FDA inspectors that their manufacturing practices meet quality and health standards. Please contact Webscan for more information if Validation applies to you.

Page 68 read from a scan by OCR; it may hold reading errors

hd

WEBSCAN

BARCODE VERIFIERS

1341 Sherman Drive Unit D Longmont, CO 80501 303.485.6811 info@webscaninc.com www.webscaninc.com