Page 21

o

Activate Feature – This is a method for loading a license file that activates a special or optional feature of the software.

o

Calibration Log – A complete log of date and time the verifier has been calibrated.

o

Revision Log – A list of software feature additions and bug fixes, organized by software version. These can be selected and printed.

o

Debug Information –

Save Now - Used to save a file of problematic codes for evaluation by Webscan. Select this option after the scan and save the file to a known location. This file can be emailed to Webscan for evaluation.

Save Before Verify – Used to save a file of problematic codes for evaluation by Webscan in cases where the verification process causes the software to crash or otherwise prevents the saving debug information after the scan has taken place.

o

About – Lists the current software revision.

o

Update – If the PC that the verifier is connected to has internet access, this option will initiate a check of more current software and download the software to the machine for installation. Details for upgrading software are in section of this manual. To find out what the most current version of software is, please visit www.webscaninc.com and review the FAQ section.

Page 22

Illumination Icons

These icons only pertain to the TruCheck™ 2D USB-DPM imagers and the TruCheck™ DPM Tower. For more information, please refer to the Using a TruCheck 2D USB DPM and TruCheck DPM Tower Verifier section. Diagnostic Icons: The Magnifier Icon – The Magnifier enhances or enlarges the view of a decoded symbol. This is beneficial for examining problem areas of a symbol. After you verify a symbol, the default view is magnified. Clicking this icon will zoom out to the actual image size. If you click the image again, it will zoom back in. NOTE: You can also draw a section in the zoomed in image for additional magnification. Right clicking with your mouse will return the view to the actual image size. The Grid Icon – This either removes or overlays the barcode with a grid. In the case of 2D symbols, the grid displays the intersections and apertures, yellow or red indicators, representing warning or failing sections of the barcode. There are options on the grid to view the real grid, the ideal grid, have mod circles filled, or mod circles with no fill. The circles represent the test aperture size and if shown in green, they indicate an “A” or “B” graded area. Yellow circles indicate a “C” level graded area and red circles indicate a failing region. The Original Icon – This restores the original view of the barcode. The Black and White Icon – This will display the image as an imager might see the barcode. The black and white image is the result of thresholding the blurred image with the global threshold. The black and white image is what is actually used for decoding. The Blurred Image Icon – The blurred image represents the result of processing the original image with the verification aperture. Small defects such as spots and voids may be “filled in” by this processing and edges will appear to be “softened”. This image is then used to calculate the global threshold, which is applied to the blurred image to get the black and white image. The Excel Icon – Included with the TruCheck software is a feature where verification results can be captured into a Microsoft Excel™ Spread Sheet. Excel™ files can be saved on the local hard drive or any network path where the user has permission to write data to. The Excel option can be initiated by clicking on the Excel Icon displayed on the interface or by selecting it from the “Options” menu -> “Save results to Excel”. The spreadsheet will open and can then be minimized to run in the background. All subsequent verifications will be posted to the sheet. The sheet should be saved as a different name (job, date, etc.) so that the template remains a blank sheet. The Print Icon – This icon will print the results of the last barcode verified to the Default Windows Printer.

Page 23

DPM (Direct Part Mark) verifiers have multiple lighting options to grade according to various grading methodologies and standards (such as AIM DPM). Depending on the model that you have, the Lighting buttons on the toolbar will be available to quickly select the lighting option of your choice. The available lighting options may be affected by your application settings.

NOTE: Some lighting options are not allowed in some application standards so you should only select options that provide you with a result that is valid or your application. It is common to use AIM DPM as the basis for grading when working with DPM symbols and lighting other than 45 degree light from four sides. Illumination Icons 45Q Icon – This icon selects 45-degree illumination. 45-degree four-sided illumination is used primarily for labels and is the only option with the standard USB and Wide Angle imagers. This is the illumination needed for linear 1D and 2D barcode verification. This illumination is either not reported or is reported as 45Q. (If illumination is not reported, it is assumed to be 45Q) 30Q Icon – This icon selects 30-degree illumination from four sides. It may reduce glare from some substrates, which cause poor Symbol Contrast or Modulation grading. This illumination is reported as 30Q. 30T Icons – These icons select 30-degree two-sided illumination. These are again for substrates causing issues with the 45Q illumination or on curved surfaces. This illumination can be from either the North/South or East/West. This illumination is reported as 30T. 90 Icon – This icon selects 90-degree diffuse illumination and works well on very shiny substrates and Dot Peen Applications. It is reported as 90. The Illumination Icons are quick shortcuts to select the most common lighting choices. The Lighting menu, under Options, shown here, provides a way to select even more options (if available on your verifier model) and to choose a sequence of multiple lighting options to be used automatically when verifying. Select each of the lighting options that you would like to use during verification, and then select your cycling option. When verifying, the system will proceed to verify, repeatedly using each selected light (Cycle All) or until one of the lighting options results in a passing grade.

Page 24

The DPM Tower allows the height of the imager to be adjusted to facilitate verification of barcode symbols that can be on parts of various size and thickness. When a part is placed under the imager, the height of the imager can be adjusted by turning the handle at the top. Turning the handle in one direction raises the imager and turning it in the opposite direction lowers it.

When you select “Go Live”, if the imager is not at the correct distance over the target, the software will automatically go into focusing mode. In the focusing mode, you will see a target circle and a spot of light.

The spot of light is the image of a light source that is projected down from the imager. By moving the imager up and down you place the spot of light into the target circle. When the spot of light is in the target circle, the circle will turn green, indicating the correct focal distance.

Now when you press “Go Live”, the focusing light will be turned off and a live image will be shown. Whenever you initiate a verification, the focus will be checked again, and if not in focus, the software will automatically go into the focusing mode.

Page 25

Your TruCheck™ verifier is a precision optical electronic instrument. Periodic calibration is necessary to account for changes in the environment. Ambient light and temperature are typical environmental conditions that frequently change. Electrical components also drift (change characteristics) over time and this can affect the accuracy of measurements made by the TruCheck™ verifier. Webscan recommends that you calibrate on a monthly basis; this is set as the default calibration reminder. Please locate your calibration card for the next steps in this process. Calibration cards do not come standard with verifiers. In most cases, your sales representative should have explained the different card options and one should be included in your packaging. NOTE: Calibration cards have an expiration date listed on the card. It is very important to only calibrate your unit with a valid card. When working with a verifier that can handle 2D codes, you can calibrate with either a 1D or a 2D calibration card. If you would like to order a calibration card for your unit, please contact support@webscaninc.com or call 303.485.6811. Part Number

Description

Verifier models

4899

3MIL Calibration and Conformance Symbol (UPC/EAN)

3mil lasers

1556

6MIL Calibration and Conformance Symbol (UPC/EAN)

6mil Lasers

2633

10MIL (GS1/UCC/EAN-128) Calibration and Conformance Symbol

10mil Lasers

2222

20MIL Calibration and Conformance Symbol (ITF-14)

20 mil Lasers

4744

GS1 Data Matrix Calibration and Conformance Symbol

All Camera Based verifiers

5136

Webscan Calibration Symbol

All Cameras and 6mil lasers

Page 26

1. When calibrating with GS1 standards, locate your copy of the “Calibrated Conformance Standard” sheet that was included with your calibration target. This sheet contains the Rmin and Rmax values you will log into the software prior to calibrating. If using the Webscan Calibration card, the Rmin/Rmax values are logged directly on the card. 2. Click “Go Live” and position the calibration symbol under the guide plate as shown in the image below. Set the laser sweep so that the laser is within the code beginning to end. 3. Select “Options → Calibrate Remote” from the main menu. 4. Enter the Rmax and Rmin values in the respective textboxes. These values can be found by using the chart on page 1 of your “Calibration Conformance Standard-ANSI Print Quality Analysis” sheet. On the Webscan calibration card, these are logged below the UPC symbol. 5. You may log in a calibration interval reminder at this time if desired. 6. Click on “Calibrate”. The system is now being calibrated; this may take a few seconds. NOTE: The report included with the calibration barcode gives all the values in tenths. i.e. 69 would be considered a 6.9. The TruCheck accepts whole numbers only so you must round this value to the nearest integer. For the previous example of 6.9, you would enter 7. Likewise, the value 824 represents 82.4 so you would enter 82. Once the Rmin and Rmax values are entered during the initial calibration, they will be stored in the system for future use. 7. At the end of the calibration process, the TruCheck™ USB calibration screen will show, “Successfully set Rmax and Rmin”. 8. Check that the calibration settings are within range by verifying the calibrated symbol and confirming the results. Results are displayed in the “Quality Parameters” tab on the User Interface. Beside the SC (Symbol Contrast) grade, RL/ RD (Reflectance Light element/ Reflectance Dark element) values are displayed and should be equivalent or within tolerance to the Rmax /Rmin values logged in at calibration. Tolerance is referenced in the Annex –Engineering Specs. DPM NOTE: When checking conformance with a calibration card that is calibrated to ISO 15415 using 45 degree light from 4 sides, it is necessary to select the 45Q lighting option, and set the aperture at Auto 80% on your DPM verifier in order to get conforming results.

Page 27

1. Note the Rmin/Rmax values logged on your calibration card. If using a linear one-dimensional calibration symbol on a GS1 calibration and conformance symbol, locate the “Calibrated Conformance Standard” sheet that was included with your calibration target. 2. Click “Go Live” and position the calibration Master Symbol as close to the red target in the live image as possible. 3. Select “Options → Calibrate Remote” from the main menu. 4. Select the bullet appropriate to the type symbol your “Calibration Symbol” represents: Data Matrix or Linear. 5. Enter the Rmax and Rmin values in the respective textboxes. These values can be found by using the chart on page 1 of your “Calibration Conformance Standard-ANSI Print Quality Analysis” sheet GS1 Linear Calibrated Conformance symbols. On the GS1 Data Matrix Calibrated Conformance standard and on the Webscan calibration card, these are logged directly on the card. 6. You may log in a calibration interval reminder at this time if desired. 7.

Click on “Calibrate”. The system is now being calibrated; this may take a few seconds.

NOTE: The report included with the calibration barcode gives all the values in tenths. i.e. 69 represents the number 6.9. The TruCheck™ accepts whole numbers only so you must round this value to the nearest integer. For the previous example of 6.9, you would enter 7. Likewise, the value 824 represents 82.4, so you would enter 82. Once the Rmin and Rmax values are entered during the initial calibration, they will be stored in the system for future use. 8. At the end of the calibration process, the TruCheck™ USB calibration screen will show, “Successfully set Rmax and Rmin and Pixels/Mil”. 9. Check that the calibration settings are within range by verifying the calibrated symbol and confirming the results. Results are displayed in the “Quality Details” tab on the User Interface. Beside the SC (Symbol Contrast) grade, RL/ RD (Reflectance Light element/ Reflectance Dark element) values are displayed and should be equivalent or within tolerance to the Rmax /Rmin values logged in at calibration. Tolerance is referenced in the Annex –Engineering Specs.

Page 28

This is an example of the 1D settings screen. You can access this screen from the options menu in the upper left hand corner of the interface.

Alternately, you can simply click the 1D button (depicted below) also located in the upper left hand corner of the interface.

Select the appropriate box for the type of symbol you are verifying. For best results, deselect all boxes for code types you are not verifying. Common 1D linear symbols are automatically selected.

SETTINGS SPECIFIC TO GS1 DATABAR:

GS1 DataBar can report as the legacy name ‘RSS’ or the current designation ‘GS1 DataBar’.

The symbology specification changed in 2011, so your verifier will grade according to the new rule or legacy rule. Application Specification Acceptance Criteria

Page 29

GS1- Turns on the acceptance criteria defined by GS1, including minimum grade, detailed parsing of the GS1 data strings and minimum X Dimension/Magnification. When GS1 is activated by clicking its radio button, the selections are: Always – to always check, and Auto – to check only when a GS1 Symbology is decoded (but not fail a Code 128 symbol which does not contain a leading Function 1 character). When a non-GS1 symbology is decoded, the pass grade rule will be applied instead. Note: you can select GS1 Format Check without selecting GS1 Application standard (select Pass Grade instead) if you want to check data format but not other rules, such as minimum Magnification or minimum pass grade. (UPC/EAN Symbols will be checked for Magnification and a warning will be issued when Magnification is less than 80% but more than 75%. A failure of GS1 Acceptance criteria will be issued when Magnification is less than 75%. If you do not want to be warned or receive a failure when violating these magnification minimums you should set your Application to Pass Grade, not GS1). Pass Grade- Establishes the minimum acceptable grade for the user’s application. When activated, this displays a new field on the main screen that not only displays the grade, but turns red if the grade received is lower than the minimum set, and green if the grade is higher. This is useful for providing an at-a-glance acceptance or failure. UDI/HIBC – This application standard checks symbols that meet UDI requirements using either GS1 or HIBCC guidelines. Format Check ISO/IEC 15434 – This application standard checks data for many industry standards which encode information using ISO/IEC 15434 data structures. Typical applications using this data structure include parcel shipping and other material handling applications using PDF417 or other high capacity barcode formats. Laser Emulation – This is a special feature to make imager based verifiers perform in a similar way to Laser based verifiers.

# Scans- Default value is 10 scans, but it is adjustable from 1 to 100. This is the number of scans that will be analyzed and graded. Aperture- This is the synthetic aperture size the system will use to process the raw image into the reference (blurred) image. The aperture can be selected with the drop down box or manually entered by typing an aperture size in ten thousandths of an inch. For example, a 6-mil aperture is set either by selecting it from the drop down by typing 60.

Page 30

NOTE: 6mil aperture is required for verifying many linear barcodes, such as GS1 –DataBar (RSS), so when Linear/RSS is selected, the aperture will default to 6mil. The correct aperture size to choose is always governed by an application standard. Wavelength- This setting cannot be adjusted. This is the wavelength of the light used to verify the symbol. I25/ITF14- When evaluating per ANSI/ISO I25 standard or GS1 ITF14, you must select “Standard I25” to evaluate per the GS1 criteria. UPC Sup- Used to evaluate UPC Supplemental symbol extensions. Options include none (default), Auto, 2digit, and 5-digit. Choose “Auto” to let the software detect when a supplement exists. Mode- This option is used to verify barcodes that are either inverted (white on black) or pre-press (such as film). Options are Normal, Pre-Press Negative, Pre-Press Positive, or Inverted Image. Linear Auto Detect- This option allows the user to create test areas larger than the code or no test area selection at all. The process will find the decodable area and divide the scans evenly throughout the code. Symbol height is reported in the event only a small area of a poorly printed code is decodable. This option can be deselected if the user wants precise diagnostics of poorly printed symbols. Ladder Orientation- Option for verification of linear codes aligned in the field of view in a ladder orientation, where the code’s bars are oriented left to right. MSI Plessy Decode Options Mode- Mode can be set to 1 or 2 and controls the check digit calculation so that it can match either of two ways of calculating the check digit in various MSI/Plessey applications. The mode selects which characters (even or odd) are weighted by 1 or 2 in the check digit calculation. Truncate Check Digit- When selected, the last digit (which is a check digit) will not be shown as part of the decoded data. This is done to make the printed data appear to match the human readable information in applications in which the check digit is not shown in the human readable data. Detail that will be included in Archived or Printed reports Quality Parameters- The nine ANSI grades can be listed for each scan. Per Scan Results- Includes a table of values for each of the individual scans. Element Width- Prints the Bar Space width. Measured widths and deviations of the bar space combinations that make up each character encoded. SRP- Prints a scan reflectance profile on the report. Traditional Parameters- Turns on the reporting of traditional parameter values on the PDF report. Traditional parameter will appear in the other reports and in the user interface regardless of this option when they apply.