For the Videx LaserLite Mx and Videx LaserLite Pro

Page 71

To avoid problems when using multiple Base Stations, you need to be aware of the following cautions: •

Each LaserLite Pro or LaserLite Mx used in a multiple Base Station arrangement must have a unique ID.

•

When the data transfer process begins, do not remove the data collector from the Base Stations, quit the program, or shut down the computer until the transfer process is complete. Caution: When sending an operating system and/or application to a LaserLite Pro or LaserLite Mx, you must have only one LaserLite Pro or LaserLite Mx at a time in the multiple Base Station set up. If there is more than one, the data collector will not be damaged, but it will not be properly programmed. When transferring data from a LaserLite Pro or LaserLite Mx, do not interrupt the process by removing the unit from its Base Station or aborting the program. The system is designed to avoid data loss, but observing this caution provides added protection.

If you remove a LaserLite Pro or LaserLite Mx while it is in the process of transferring data, the software displays a message that it was unable to communicate with the unit and then continues to try to transfer the data from all of the other units. The data in the interrupted unit will not clear and can be transferred after the others have completed their data transfer.

Appendix E Using Multiple Data Collectors and Base Stations

65

Page 72

Changing a Data Collector’s ID with a Windows Computer When using multiple Base Stations, each data collector must be given a unique ID; this allows the computer to access each data collector and verify that it has transferred the data from each one. To assign an ID, you must first create a commands text file, then click and drag the file onto the Vxcom icon. Vxcom then uses the commands file to unlock the unit and change the ID. (For complete information on Vxcom and using a commands text file, see Chapter 1 of the Developer’s Reference Manual.) Note: The original Vxcom filename was Vxcomm.exe; newer versions of Vxcom have names in the form Vxcomxxx.exe, where xxx corresponds to the version number (for example, Vxcom138.exe). The Vxcom file is accompanied by a DLL with a name in the form Vxcomxxx.dll. The original DLL file (Vxcomm.dll) is also included because the Application Builder program (Appbuild.exe) requires Vxcomm.dll to communicate with the unit. During normal operation, a unit is given an ID of ten zeros. To change the ID, create a text file consisting of the following three command lines: 1) an I command, followed by a space, followed by ten zeros to unlock the unit; 2) a C command, followed by a space, followed by the new ID for the unit; 3) and an L command to relock the unit.

66

Appendix E Using Multiple Data Collectors and Base Stations

Page 73

I 0000000000

Sends an unlock command to the data collector. (Note: Ten zeros will unlock any data collector.)

C new ID#

Changes the unit’s ID to the given ID. The ID can be any alphanumeric combination of up to ten characters in length. For example, if you want an ID of 1, this line should be C 1; if you want an ID of BuildingD, this line should be C BuildingD; if you want an ID of 123#_AB, this line should be C 123#_AB, and so on.

L

Sends a LOCK command and puts the data collector to sleep. When the unit is awakened with a keypress, it immediately runs the current application.

For example, to change a unit’s ID to 12345: 1.

Use a word processing program to create a text file containing the following three lines: I 0000000000 C 12345 L

2.

Save the file as a text file with a .txt, .vdx, or no extension and quit the program.

3.

Place a data collector into the Base Station slot. IMPORTANT: When sending a file to a data collector, it must be the ONLY data collector attached to the computer.

4.

Click and drag the text file you created onto the Vxcom icon.

5.

Vxcom executes the commands listed in the text file: it unlocks the unit, changes the ID to what is listed after the C command (in this case, 12345), and then relocks the unit.

6.

Remove the data collector from the Base Station slot; it is ready to use. Appendix E Using Multiple Data Collectors and Base Stations

67

Page 74

To continue changing other data collector’s IDs: 1.

Reopen the text file.

2.

Change the ID listed after the C command by selecting 12345 and typing in a new ID. (Note: There must be a space character between the C and the ID.)

3.

Save and close the file.

4.

Insert a data collector into the Base Station slot.

5.

Click and drag the edited text file to the Vxcom icon.

6.

Vxcom executes the commands listed in the text file: it unlocks the unit, changes its ID to what is listed after the C command, and then relocks the unit.

7.

Continue these same steps until a unique ID is assigned to each unit.

Note: Keep a list of the IDs; you will need this list in the section “Transferring Data from Multiple Data Collectors.”

68

Appendix E Using Multiple Data Collectors and Base Stations

Page 75

Changing a Data Collector’s ID with a DOS Computer To assign a unique ID when using a DOS computer, create a commands file, and then enter the name of the commands file on the Download.exe command line. Download.exe then uses the commands file to unlock the unit and change the ID. (For complete information on Download.exe and using a commands file, see Chapter 1 of the Developer’s Reference Manual.) During normal operation, a unit is given an ID of ten zeros. To change the ID, you must create a commands file consisting of the following three command lines: 1) an I command, followed by a space, followed by ten zeros to unlock the unit; 2) a C command, followed by a space, followed by the new ID for the unit; 3) and an L command to relock the unit.

I 0000000000

Sends an unlock command to the data collector. (Note: Ten zeros will unlock any data collector.)

C new ID#

Changes the unit’s ID to the given ID. The ID can be any alphanumeric combination of up to ten characters in length. For example, if you want an ID of 1, this line should be C 1; if you want an ID of BuildingD, this line should be C BuildingD; if you want an ID of 123#_AB, this line should be C 123#_AB, and so on.

L

Sends a LOCK command and puts the data collector to sleep. When the unit is awakened with a keypress, it immediately runs the current application.

Appendix E Using Multiple Data Collectors and Base Stations

69

Page 76

For example, to change a unit’s ID to 12345: 1.

Use a word processing program to create a commands file containing the following three lines: I 0000000000 C 12345 L

2.

Save the file as a text file with a .txt or no extension and quit the program.

3.

Place a data collector into the Base Station slot.

IMPORTANT: When you are sending a file to a data collector, it must be the ONLY data collector attached to the computer. 4.

Enter the name of the commands file on the Download.exe command line.

5.

Download.exe executes the commands listed in the file: it unlocks the unit, changes the ID as instructed by the C command (in this case, 12345), and then relocks the unit.

6.

Remove the data collector from the Base Station slot and it is ready to use.

70

Appendix E Using Multiple Data Collectors and Base Stations

Page 77

To continue changing other data collector’s IDs: 1.

Reopen the commands file.

2.

Change the ID listed after the C command by selecting 12345 and typing in a new ID. (Note: There must be a space character between the C and the ID.)

3.

Save and close the file.

4.

Insert a data collector into the Base Station slot.

5.

Reactivate the Download.exe command line containing the name of the commands file.

6.

Download.exe executes the commands listed in the file: it unlocks the unit, changes its ID as instructed by the C command, and then relocks the unit.

7.

Continue these same steps until you have assigned a unique ID to each of your units.

Note: Keep a list of the IDs you are using; you will use this list in the section “Transferring Data from Multiple Data Collectors.”

Appendix E Using Multiple Data Collectors and Base Stations

71

Page 78

Changing a LaserLite Pro’s ID with a Macintosh Computer To assign a unique ID to a LaserLite Pro when using a Macintosh computer, you must first create a commands file, then click and drag the file onto the Videx Download icon. Videx Download then uses the commands file to unlock the unit and change the ID. (For complete information on Videx Download and using a commands file, see Chapter 1 of the Developer’s Reference Manual.) (Note: LaserLite Mx is not Macintosh compatible.) During normal operation, a LaserLite Pro is given an ID of ten zeros. To change the ID, create a commands file consisting of the following three command lines: 1) an I command, followed by a space, followed by ten zeros to unlock the LaserLite Pro; 2) a C command, followed by a space, followed by the new ID for the unit; 3) and an L command to relock the LaserLite Pro For example, to change a LaserLite Pro’s ID to 12345: 1.

Use a word processing program to create a file containing the following three lines: I 0000000000 C 12345 L

2.

Save the file as a text file with a .txt, .vdx, or no extension and quit the program.

3.

Place a LaserLite Pro into the Base Station slot.

IMPORTANT: When you are sending a file to a data collector, it must be the ONLY data collector attached to the computer.

72

Appendix E Using Multiple Data Collectors and Base Stations

Page 79

4.

Click and drag the commands file onto the Videx Download icon.

5.

Videx Download executes the commands listed in the text file: it unlocks the unit, changes the ID to what is listed after the C command (in this case, 12345), and then relocks the unit.

6.

Remove the LaserLite Pro from the Base Station slot; it is now ready to use.

To continue changing other LaserLite Pro’s IDs: 1.

Reopen the commands file.

2.

Change the ID listed after the C command by selecting 12345 and typing in a new ID. (Note: There must be a space character between the C and the ID.)

3.

Save and close the file.

4.

Insert another LaserLite Pro into the Base Station slot.

5.

Click and drag the edited commands file to the Videx Download icon.

6.

Videx Download executes the commands in the commands file: it unlocks the unit, changes its ID as instructed by the C command, and then relocks the unit.

7.

Continue these same steps until you have assigned a unique ID to each LaserLite Pro.

Note: Keep a list of the IDs; you will use this list in the next section “Transferring Data from Multiple Data Collectors.”

Appendix E Using Multiple Data Collectors and Base Stations

73

Page 80

Transferring Data from Multiple Data Collectors To transfer data from a group of data collectors in one transfer process, you must: 1. Create a commands text file that uses the FOREVER, FOREACH, or FORALL looping commands to transfer the data from the data collectors. 2. Insert all of the data collectors into their Base Stations. 3. If using Windows, click and drag the commands text file onto the Vxcom icon; if using DOS, enter the name of the commands text file on the Download.exe command line; or if using Macintosh (LaserLite Pro only), click and drag the commands text file onto the Videx Download icon. The three looping commands provide transfer methods for three different situations. The FOREVER loop is useful if you have a computer dedicated to transferring the data from the data collectors. The FOREACH loop is useful if you are transferring data from only a portion of the data collectors. The FORALL loop is useful if all of the data collectors are required to be transferred during the process. The following table describes the looping commands: FOREVER

For use with a computer that is dedicated to transferring data. This loop will transfer the data from any data collector it locates. This loop attempts to execute commands within a loop indefinitely.

FOREACH

This loop will search for each data collector in the list of IDs. If it locates the data collector with the ID, it executes the commands within the loop once; if it does not locate the data collector with the ID, it goes on to the next ID. This loop attempts to execute commands within a loop once for each ID listed.

FORALL

This loop will look for each listed data collector’s ID until it has located and transferred the data from each ID in the list. This loop attempts to execute commands within a loop indefinitely, until all listed IDs have been located.

Table E-1 Looping Commands

74

Appendix E Using Multiple Data Collectors and Base Stations