For the Vernier LabPro
Page 19
1. Set up LabPro for remote data collection. a. Put fresh batteries in the LabPro. b. Set up the sensors, LabPro, and Logger Pro just as you normally would to collect data in Time Based mode. c. Instead of clicking the button, choose Remote Remote Setup from the Experiment menu.10 A summary of your setup will be displayed. d. Click to prepare the LabPro. e. Disconnect the LabPro from the computer. f. If it has not already been saved, save the experiment file so it can be used to later retrieve the data from LabPro. 2. Collect data. a. When you are ready to collect data, press the START/STOP button. b. When data collection is complete, the yellow LED will flash briefly. You can also stop data collection early by pressing the START/STOP button before data collection is finished. 3. Retrieve the data. a. Start Logger Pro if it is not already running. Choose Continue Without Interface in the Connect LabPro dialog and click . b. Open the experiment file used to set up LabPro. c. Attach LabPro to the computer. d. If a Remote Data Available window appears, click the YES button. Click , which accepts the default to retrieve remote data into the current file. If a window does not appear when the interface is reconnected, choose Remote Retrieve Remote Data from the Experiment menu and follow the on-screen instructions. e. The data will be retrieved. Set Up for Selected Events Method Using the Set Up for Selected Events Method, you will use a computer to set up LabPro, detach it to collect individual data points, and then reattach it to retrieve the data. With this method, you can use any sensor (not just auto-ID), and collect individual data points whenever you press the START/STOP button. 1. Set up LabPro for remote data collection a. Put fresh batteries in the LabPro. b. Set up the sensors, LabPro, and Logger Pro just as you normally would to collect data using the Selected Event mode. c. Instead of clicking the button, choose Remote Remote Setup from the Experiment menu. A summary of your setup will be displayed. d. Click to prepare the LabPro. e. Disconnect the LabPro from the computer. f. If it has not already been saved, save the experiment file so it can be used to later retrieve the data from LabPro. 2. Collect data. Every time you press the START/STOP button, LabPro will collect one data point. The green LED will remain on and the yellow LED will flash with each reading. Repeat as often as you want, for up to 99 readings.
10
Sensors that require a warm-up time may have additional on-screen instructions.
Vernier LabPro User’s Manual
15
Page 20
3. Retrieve the data. a. Start Logger Pro if it is not already running. Choose Continue Without Interface in the Connect LabPro dialog and click . b. Open the experiment file used to set up LabPro. c. Attach LabPro to the computer. d. If a Remote Data Available window appears, click the YES button. Click ,which accepts the default to retrieve remote data into the current file. If a window does not appear when the interface is reconnected, choose Remote Retrieve Remote Data from the Experiment menu and follow the on-screen instructions. e. The data will be retrieved.
Remote Data Collection with DataMate Quick Setup Method The Quick Setup feature is very simple: 1. Set up LabPro for remote data collection. a. Connect your auto-ID sensor(s) to LabPro. (Sensors that are not auto-ID cannot be used with this method. The next two methods allow you to use auto-ID and/or non auto-ID sensors.) b. Press the QUICK SETUP button. The yellow LED will blink twice and two beeps will be heard verifying setup. 2. Collect data. a. When you are ready to collect data, press the START/STOP button. The green LED will flash during data collection. b. When data collection is complete, the green LED will stop blinking.11 You do not need to press the START/STOP button to end data collection. (You can, however, stop data collection early by pressing the START/STOP button before data collection is finished.) 3. Retrieve the data. a. You can retrieve the data by connecting the interface to a TI graphing calculator and starting DataMate. A message will appear stating “DATA COLLECTION IS DONE. . .” b. Press ENTER to proceed to the main screen. c. Select TOOLS from the main screen. d. Select RETREIVE DATA from the TOOLS menu. The data will be retrieved and one or more graphs made available. Set Up for Time-Based Data Collection This method gives you more control of the data collection than the Quick Setup method, because you have control of which sensors and what data collection sample interval will be used. This is how it works: 1. Attach one or more sensors to LabPro. These sensors can be auto-ID or non auto-ID sensors. 2. Start DataMate. The program will detect auto-ID sensors. Non auto-ID sensors can be set up manually. (Refer to the DataMate Guidebook for help.) a. Select SETUP from the main screen. 11
The duration of data collection will vary from sensor to sensor.
16
Vernier LabPro User’s Manual
Page 21
b. Press to select MODE and press ENTER . c. Select TIME GRAPH from the SELECT MODE menu. d. Select CHANGE TIME SETTINGS from the TIME GRAPH SETTINGS menu to set the data collection rate. e. Enter the time between samples in seconds (remember to finish your entry with ENTER ). f. Enter the number of samples. 3. Set up the interface for manual triggering. a. Select ADVANCED from TIME GRAPH SETTINGS. b. Select CHANGE TRIGGERING from ADV. TIME GRAPH SETTINGS. c. Select MANUAL from SELECT TRIGGERING. d. Select OK three times to return to the main screen. 4. Select START to prepare LabPro for data collection. The yellow light will come on, two beeps will be made, and a message will appear stating that remote data collection has been set up. 5. Press ENTER to return to the main screen and choose Quit. 6. Disconnect LabPro from the calculator. To start data collection, press the START/STOP button on LabPro. The yellow light will go off and the green light will blink during data collection. 7. When the green LED stops blinking and data collection is done, attach LabPro to the same calculator and start the DataMate program. The program will detect that data has been collected by LabPro and will display a message about retrieving the data. Press ENTER on the calculator. On the main screen, select TOOLS, and then select RETRIEVE DATA on the TOOLS menu. The data will be retrieved and one or more graphs made available.
Remote Data Collection with EasyData There are two methods of triggering (remote data collection) available for use with EasyData and LabPro: software triggering and manual triggering. Software Triggering In software triggering, you choose a measurement reading that will trigger data collection, e.g., when the temperature is increasing and reaches a value of 28°C. Follow these steps to set up a software trigger. 1. From the EasyData main screen, select
, and then select Time Graph...
2. The current Time Graph settings are displayed. Select
.
3. The Select Trigger Type screen is displayed. Choose the appropriate channel, e.g., CH1, and select . 4. The Measurement screen is displayed. Choose whether triggering should occur when the readings . increase or decrease (Increasing or Decreasing), and select to clear the old threshold and type in the 5. The Trigger Threshold screen is displayed. Press . Note: trigger threshold is the sensor reading, e.g., 28 for 28°C, that will new value, then select trigger data collection. to clear the old prestore and type in the new 6. The Trigger Prestore screen is displayed. Press . Note: trigger prestore is the percentage of all data collected before the value, then select trigger level is reached, e.g., 10 for 10%.
Vernier LabPro User’s Manual
17
Page 22
7. The updated Time Graph Settings are displayed. Select 8. EasyData is now set for triggering. When set trigger value is met.
to return to the main screen.
is selected, data collection will not begin until the
Manual Triggering In manual triggering, data collection starts when the Start/Stop or Trigger button on the data-collection device (LabPro, CBL 2, CBR, or CBR 2) is pressed. Follow these steps to set up for manual triggering. 1. From the EasyData main screen, select
, and then select Time Graph...
2. The current Time Graph settings are displayed. Select
.
3. The Select Trigger Type screen is displayed. Choose Manual and select 4. The updated Time Graph Settings are displayed. Select 5. EasyData is now set for triggering. Select ready to start data collection.
18
.
to return to the main screen.
and follow the on-screen instructions when you are
Vernier LabPro User’s Manual
Page 23
Sensor Warm-Up Times Some sensors require a certain warm-up period before an accurate measurement can be taken. LabPro handles this automatically in remote modes by warming up each sensor for the appropriate amount of time. The Turbidity Sensor, for example, requires a 2-second warm-up period. The table below lists all sensors with warm-up times of 2 seconds or longer. If a sensor is not on this list, its warm-up time is negligible. Sensor Ammonium Ion-Selective Electrode Calcium Ion-Selective Electrode Chloride Ion-Selective Electrode CO2 Sensor Colorimeter Conductivity Probe Dissolved Oxygen Probe Dual-Range Force Sensor Electrode Amplifier Flow Rate Sensor Force Plate Nitrate Ion-Selective Electrode Oxygen Gas Sensor pH Sensor Turbidity Sensor UVA Sensor UVB Sensor
Warm-Up Time (s) 2 2 2 90 300 2 600 2 30 2 2 2 2 30 2 2 2
The following sensors cannot be set up to collect remote data: Digital Control Unit Drop Counter Heat Pulser Photogate Radiation Monitor Student Radiation Monitor Rotary Motion Sensor Digital Radiation Monitor
Vernier LabPro User’s Manual
19
Page 24
Page 25
Appendix A: Troubleshooting Tips Refer to http://www.vernier.com/til/1613/ for additional troubleshooting tips. Problem
Cause/Solution
When LabPro is used with the computer the Logger Pro 3 Collect button is faded (or the Logger Pro 2 Collect button does not appear).
The Collect button does not appear because the computer could not communicate with the interface. Check each of the following: Make sure the cables are firmly attached. Make sure LabPro is connected to the correct port on the computer. Confirm that the port on the computer is working properly. Make sure the interface has power - either AC or battery.
When LabPro is used with a TI graphing calculator, I get a link error message
This message appears when the calculator cannot communicate with the interface. Check the following: Make sure the cables are firmly attached to the calculator and LabPro. Make sure the interface has power - either AC or battery.
The sensor is not working.
This may be happening because the sensor is not plugged into the correct port. Since LabPro has 4 analog and 2 digital ports, make sure that the sensor is connected to the proper port.
The readings are noisy.
LabPro may be picking up interference from other electrical devices, such as a computer. Move the LabPro interface at least 30 cm from another computer.
When LabPro is plugged into AC power, the lights and beeps are not generated.
When LabPro powers up, lights and beeps signal a successful self-test. If batteries are already installed in LabPro, the selftest will not be performed.
When connected to USB, a light can be seen blinking inside the interface.
This is normal when LabPro is used through a USB port.
Vernier LabPro User’s Manual
A-1
Page 26
Page 27
Appendix B: Beep Sequences LabPro makes four kinds of sounds or “beeps.” A “good” beep is a medium tone followed by a high tone. A “caution” beep is a medium tone followed by another medium tone. A “bad” beep is two high tones. A “tick” sound is made when a key is pressed. The following bullets explain when beep sequences occur and what the beep sequences mean: When LabPro completes initialization, you will hear the startup sequence: bad beep, caution beep, good beep (6 total beeps, plus LEDs light up in this order: red LED, yellow LED, and green LED) When you press the QUICK SETUP button, the caution beep sounds if a sensor is attached to LabPro. the bad beep sounds if no sensors are attached to LabPro. When LabPro is connected to a calculator during sampling commands, the caution beep sounds when initializing data collection. the caution beep sounds when starting data collection (transition from pre-store to store). the caution beep sounds when completing data collection. Note 1: If the sampling timing causes the beeps to run together, the LabPro software may not sound all the beeps. Note 2: You will not get all the beeps when Fast Sampling is enabled. Note 3: You will not get all the beeps when using triggering. When you set LabPro for manual trigger and press the START button, a caution beep sounds. When you press the TRANSFER button, the good beep sounds when the transfer succeeds. the bad beep sounds if the transfer fails for any reason.
When an overcurrent condition is detected, five bad beeps sound. (This causes an error, which causes even more beeps to sound.) When LabPro begins a full self-test, three good beeps sound. When self-test completes, the good beep sounds if self-test passes. the bad beep sounds if self-test fails.
When LabPro’s base code detects an error in the commands sent from the host, a bad beep sounds twice. When LabPro powers up, two bad beeps sound if the base code is not loaded. three bad beeps sound if the power-up self-test fails.
During base code download, three bad beeps sound when any errors occur. (The unit resets and then the two bad beeps mentioned in the previous bullet sound.)
Vernier LabPro User’s Manual
B-1