Page 197

Troubleshooting

Table 10-1. PPT 4300 Troubleshooting Symptom Battery life too short (continued from page 10-10).

Possible Cause Battery Smart chip out of synchronization.

Action Run the battery through a conditioning cycle to resynchronize the capacity value.

Power Management warning Check the Power Management settings on limits have been set at too high the Power tab of the Control Panel. By a level. default, the Low warning is set at 8%, and the Critical warning is set at 4%. Power Management has been turned off.

Check Power Management settings on the Power tab on the Control Panel. Some transitions may have been turned off, or set very long, resulting in lower power savings than expected.

Pen and Direct Digitizer driver is not loaded. Access Panel icons not working.

Boot to a command line prompt and verify that the driver (mutohpen.com or touch.com) is being loaded when the terminal starts. The autoexec.bat file must include one of these lines: For a PPT 4300-00E: loadhigh c:\bin\mutohpen.com /int=5 /io_base=220 /load_init /rate_div=3 For a PPT 4300-00R: loadhigh c:\bin\touch.com /int=e /io_base=220 /load_init

Digitizer is not initialized.

Run c:\bin\sfinit.exe.

Pen battery is dead.

Change pen battery.

Cursor not aligned with pen tip.

Pen Digitizer not properly calibrated.

Re-calibrate the Pen, by pressing <Fn-F8> or by using the Calibration utility on the Control Panel (Display tab). For this utility to work, sftsr and the pen driver must be loaded.

Pen Calibration screen doesn’t appear.

Driver not loaded.

Boot to a DOS prompt and verify that the sftsr and pen drivers (sftsr.com and mutohpen.com or touch.com) are loaded when the terminal starts. Make sure autoexec.bat includes lines that load the drivers.

10-11

Page 198

PPT 4300 Series Product Reference Guide

Table 10-1. PPT 4300 Troubleshooting Symptom Control Panel doesn’t load.

Possible Cause

Action

Not enough low memory or the Limit the number of drivers loaded. You Control Panel .exe executable may run the Control Panel during the boot file is missing. sequence by pressing the Settings icon or <Fn-F7> before the drivers load. You may also reboot and choose option 5 from the ROM-DOS Start-up Menu. In either case, the Control Panel should start because there are fewer drivers loaded. File ctrlpnl.exe not loaded. Check that file is present in c:\bin\ctrlpnl.exe.

Keyboard is locked.

Application crash or program failure.

Reset the terminal. Press <Ctrl-Alt-Del> to reboot, or you can press <Ctrl-Fn-F12> to power off immediately. You can also toggle the Run/Ship switch in the battery well. If these don’t work, use the Master Reset switch, as described on page 10-6.

In the VT220 terminal Press <Ctrl-Q> to resume operation. emulator environment, if a user presses <Ctrl-S> in an application, it will appear as if the terminal’s keyboard has locked up. External Keyboard doesn’t work.

10-12

The keyboard you are using is You must use a Symbol-approved low incompatible with the terminal. power/low voltage keyboard: ! 5V for PPT 43xx-xxE ! 3.3V for PPT 43xx-xxR

Page 199

Troubleshooting

Beep Codes Some problems with the terminal will cause a series of audible beeps, in a sequence. Table 102 describes these beep codes.

Table 10-2. Beep Codes Beep Code

Beep Sequencea

Meaning

0

S-S-S-P-S-S-L-P

The DMA page registers are faulty.

1

S-S-S-P-S-L-S-P

The refresh circuitry is faulty.

2

S-S-S-P-S-L-L-P

The ROM checksum is incorrect.

3

S-S-S-P-L-S-S-P

The CMOS RAM test failed.

4

S-S-S-P-L-S-L-P

The DMA controller is faulty

5

S-S-S-P-L-L-S-P

The interrupt controller failed.

6

S-S-S-P-L-L-L-P

The 8042 keyboard controller failed.

7

S-S-L-P-S-S-S-P

No video adapter was found.

8

S-S-L-P-S-S-L-P

No RAM is installed. No message is displayed.

a. S= Short, L=Long, P=Pause

10-13

Page 200

PPT 4300 Series Product Reference Guide

Diagnostic Utilities The PPT 4300 Utilities PC Card includes a number of diagnostic utilities. If a terminal is not functioning properly, these utilities can help identify the problem. The Utilities PC Card is available from Symbol. Call your Symbol representative at 1-800-653-5350. The utilities include diagnostics for: !

System Board

!

Memory

!

Serial/COM Port

!

IrDA Port

!

Video Controller

!

Keyboard

!

Battery

!

Radio

For information on using the diagnostic utilities, see Diagnostic Test Utilities on page 11-6.

Bypassing and Debugging config.sys and autoexec.bat To bypass config.sys and autoexec.bat, press <F5> when you see Starting ROM-DOS as the terminal is booting. For information on editing these files, see Editing autoexec.bat and config.sys on page 2-8. Note that if you bypass these files, no drivers will load (e.g., the digitizer driver), and your path will not be set. You can step through the files selectively by pressing <F8> at the “Starting ROM-DOS” message shown as the terminal is booting. The process gives you the option of executing or skipping each line in config.sys and autoexec.bat.

Network Troubleshooting Table 10-3 describes some of the most likely reasons that a terminal might stop communicating. For information on troubleshooting the Access Point, see the Spectrum24 documentation. For information on troubleshooting the Network or Host computer, see the documentation specific to the equipment you are using.

10-14

Page 201

Troubleshooting

Table 10-3. Network Troubleshooting Problem Source PPT 4340 terminal

Possible Cause

Action

Incorrect terminal configuration.

Check the terminal configuration on the Network tab of the Control Panel and inspect \net\net.cfg or \lanman.dos\protocol.ini.

Drivers not loaded.

The drivers normally load when the terminal boots. As the terminal powers up, check the boot messages for errors as the drivers load.

Interference source between Identify and remove the source of the terminal and AP that prevents interference. See Interference on page 10-19, communication. for more details. Faulty radio card.

For Spectrum24 cards: • Use the Diagnostics Utilities described on page 11-6. • Use the Spectrum24 Access Point to ping your radio and test the MAC layer. See the Spectrum24 documentation. For other card types, see the documentation that accompanies the card.

Out of range of any Access Point.

You must be within Effective Range of a Spectrum24 Access Point with a NetID or ESS ID that matches the NetID or ESS ID of the terminal. See Effective Range on page 10-19. You can use the built-in RF/AP Diagnostic Utility to check the RF link (described on page 10-17).

Radio drivers not compatible with radio.

Use s_ver to determine the radio protocol type your terminal is using. The 1 Mbps Spring radios (s_ver=3.x) use the slaodi.exe driver. The 2 Mbps 802.11 radios (s_ver=4.x) use the sl8odipc.com driver.

10-15

Page 202

PPT 4300 Series Product Reference Guide

Table 10-3. Network Troubleshooting Problem Source

Possible Cause

Action

Access Point

Incorrect Access Point configuration.

• Check Access Point status lights and configuration. Ensure that the NetId or ESS ID matches between all Access Points and terminals. • Ping or telnet to an Access Point that has been configured with an IP address. It can be accessed from the wired network, or on the RF side using a PPT 4300 VT220 telnet emulator or the Utility card radio ping diagnostic (see page 11-10). See the Spectrum24 User Manuals for more information. • Use the built-in RF/AP Diagnostic Utility to check the RF link (described on page 10-17).

Network

Incorrect network/LAN configuration.

The network may have problems that prevent communication between Access Points and the host computer. Check the network equipment and their logging mechanisms; actions to be taken depend on the network at your facility. • The host can ping the Access Point if the Access Point has been configured with a valid IP address. • The host can ping a terminal if the terminal has loaded the appropriate radio and stack drivers.

Host Computer

Incorrect host computer configuration, or other host computer malfunction.

• Verify that a wired terminal connects to the host successfully. • If the host is configured incorrectly or is malfunctioning, it cannot communicate with the network. Check the host; actions to be taken depend on the host computer you are using at your facility.

10-16

Page 203

Troubleshooting

RF/AP Diagnostic Utility This diagnostic utility provides wireless network information from the Spectrum24 radio driver in the upper left corner of the screen. It is designed to be a diagnostic or troubleshooting tool, and is not meant to be used on an everyday basis. It can temporarily affect colors during screen updates. To use this tool, you must first enable it in ppt4300.ini: 1. Boot the terminal to a DOS prompt and open the ppt4300.ini file in the NED editor: ned ppt4300.ini

2. In the [System] section set the following parameters as shown here: PermitS24Info=1 S24InfoWantAP=1 PermitS24Info determines whether the on-screen RF/AP Diagnostic Tool is available to the user. If set to “1”, the on-screen information is allowed. If set to “0”, the feature is disabled. The default is “0”. S24InfoWantAP controls whether the S24 information will be extended to contain

Access Point identification. The default is “0” (disabled). 3. Save the changes and reboot the terminal. 4. After the terminal boots, press the Settings icon to display a pop-up window. A checkbox appears in the Settings pop-up. When the checkbox is selected, a status line appears in the upper left corner of the window, as illustrated in Figure 10-3. 5. To turn the RF/AP Diagnostic Tool off, de-select the checkbox in the Settings pop-up window. You may also disable the settings by directly editing ppt4300.ini with the NED editor and running sfinit from a DOS prompt. The RF/AP utility requires the Spectrum24 drivers. On a standard terminal, Spectrum24 drivers are only loaded under Citrix ICA and Terminal Emulators. To run this utility from DOS: 1. Boot to DOS. 2. Load the Link Support Layer: \net\lsl.com 3. Load the Spectrum24 driver manually: \net\slaodi.exe (Spring protocol) or \net\sl8odipc.exe (802.11 protocol). 4. Run sftsr and then sfinit. 5. Turn the RF/AP utility on from the Settings pop-up window.

10-17

Page 204

PPT 4300 Series Product Reference Guide

Using the RF/AP Diagnostic Utility The Transmit and Receive lights appear in the top left corner of the display window. !

Both lights are red when no data is being transmitted or received.

!

The Transmit light flashes green when data is being transmitted by the terminal.

!

The Receive light flashes green when the terminal is receiving data.

!

RSSI value shows the strength of the RF signal. Normal operating range is 25 - 65. Below 25, the terminal is likely to roam for a better connection. “RR” displays while the terminal is roaming.

!

AP ID shows the ID of the Access Point with which the terminal is currently communicating. This corresponds to the AP_ID setting on the Access Point. If you are using an 802.11 radio, the AP_ID will be the last two bytes of the Access Point’s MAC address in hexadecimal (e.g., 044E).

Transmit

Receive 51-18

AP ID

RSSI value Upper left corner of display screen

Figure 10-3. The RF/AP Diagnostic Tool

10-18

Page 205

Troubleshooting

Troubleshooting RF Interference If there is a problem with the wireless network connection: !

you may have encountered an area of interference

!

you may be standing in a null spot

!

you may be too far from the Access Point (out of range)

!

the Access Point may have stopped working

Move to a location where you know the terminal works. If the problem is range, a null spot or interference, you may solve the problem simply by moving. If moving does not solve the problem, then check the Access Point.

Interference Certain structures, objects, machines or transmissions may interfere with radio signals passing between PPT 4340 terminals and Spectrum24 Access Points. Once a network in a facility has been installed with Access Points and PPT 4340 terminals, radio interference usually becomes the most common problem experienced. Symptoms of interference include: !

slow response time

!

no response

!

sudden loss of connection with the host computer

Problem areas and objects differ from place to place. Symbol can perform a site survey to pinpoint problem areas before any Access Points or terminals are installed. This allows your facility to arrange its equipment so that interference can be minimized. You can also use the RF/AP Diagnostic Utility, as described on page 10-17.

Null spots A Null spot is an area in which a terminal behaves as though it has lost connection with the host computer. It is caused by reflection and cancellation of radio waves. Null spots tend to be small; usually if you move 2 - 3 feet in any direction you will move out of the null area.

Effective Range For a terminal to work reliably, it must be within radio range of an Access Point. Maximum effective range indoors is about 300 feet, but doors, walls, corners and construction materials may combine to decrease this range. Outdoor effective range increases to about 800 feet. Response time may deteriorate as you reach the end of your range. You can greatly extend a terminal’s functional range by setting up a network of Spectrum24 Access Points. 10-19

Page 206

PPT 4300 Series Product Reference Guide

Terminal Emulators Keyboard Lock In the VT220 terminal emulator, if a user presses <Ctrl-S> in an application, it will appear as if the terminal’s keyboard has locked up. Press <Ctrl-Q> to resume operation. Other standard ASCII Control Functions remain active. See ASCII Control Functions on page 7-12 for a list. If the terminal stops responding, check the status line. If XTWAIT (in 3270) or X (in 5250) appears on the status line, then the terminal is waiting for data or acknowledgment from the Host. To see if the pen is responding, tap the Battery icon on the Direct Access Panel. If the Battery window pops up, the pen is still working. To see if the keyboard has stop responding, press the Caps Lock key. If the Caps Locks light responds, the keyboard is still working. If both pen and keyboard are still working, then the Host has started to ignore the client, or some part of the network connection between the two has problems. You can try probing the host by pressing the Attention key (<Esc> in 5250, <Alt-A> in 3270). If the keyboard and pen stop responding, press <Ctrl-Alt-Del> to attempt to reboot the terminal. Also see Switches on the Terminal on page 10-6.

Status Line Warning Messages Table 10-4. Status Line Warning Messages Message

Cause

Action

Session may be lost in 90 seconds. Network or Host problems exist:

Terminal is having communication problems. Stop entering data immediately, to avoid overfilling keyboard buffers.

Wait to see if the connection with the host can be re-established. When the time count reaches zero, a pop-up error message appears on the screen in an enclosed box, indicating the nature of the problem. Note the error message and the action it suggests.

Re-establishing network link

The terminal is reestablishing a connection (may take 10-15 seconds).

Continue your session when you see: Wireless link re-established.

10-20