Terminal Configuration: Building and Sending the Hex Image
Icon for 16 seconds. Both methods have the same effect, but selecting “Run System” is faster. Once the selection is made, IPL proceeds to the Area Selection Menu. IPL Ver 1.0 Prev. Menu Single Image Multiple Images BIOS Disk 1 Disk 2 Disk 3 Disk 4 EXEC SETUP IPL Partition Map
3. Use the up and down cursor keys or the up and down arrows on the touch screen to select the area to be received. You have the following options on this screen: Prev. Menu
Returns you to the Baud Rate menu to modify the baud rate.
Single Image
Indicates that a single image is to be received, without specifying the area to be received.
Multiple Images (default selection)
Indicates that more than one area is to be received contiguously. Note: TCM must be set up to respond to RTS/CTS handshaking. See Sending the Hex Image on page 7-12 for more information on this setting in TCM.
BIOS Disks 1 - 4
Indicates that a new BIOS is to be received. Indicates the specific disk volume(s) to be received.
EXEC
For factory diagnostic use only.
SETUP
Indicates that a new graphical setup program is to be received.
IPL Partition Map
For service use only. Indicates that the partition map is to be received.
7-17
Page 103
PDT 7200 Series Product Reference Guide
4. After selecting the area, press Enter. IPL checks for the presence of external power. If the system is running off battery power, IPL displays the Connect to Power Screen: Insert in Cradle Or Connect Power Supply
5. Place the terminal in the cradle or connect the external power supply and serial connection. IPL waits at this screen until external power is supplied. If the terminal is already running from external power, this screen does not display. While this screen is displayed, pressing Enter or the Cursor Up/Down keys on the keyboard returns IPL to the Area Selection screen. 6. After the power and serial connections are made, the following screen displays: Waiting for data Multiple Images
Note: If a cradle is the input source, serial parameters are fixed at 38400 baud, no parity, 8 data bits, and 1 stop bit. These parameters cannot be changed. The second and fourth lines of this screen reflect the selections made on the Area Selection and Baud Rate menus, respectively. This screen continues to display until the first character of the image to be downloaded is received from the host. While this screen is displayed, pressing the Enter or Cursor Up/Down keys on the keyboard returns IPL to the Area Selection screen. As soon as the first character of data is received, IPL displays the receiving screen:
Multiple Images Pages to Erase Pages to Receive Page to Write
7-18
Page 104
Terminal Configuration: Building and Sending the Hex Image
The top line of this screen identifies the area selected in the Area Selection screen. Once a character has been received, IPL stays in this screen until an entire image has been received, or until an error is detected. The only other way to exit from this screen is to reset the terminal by holding the power switch down for 16 seconds. As more data is received, the Receiving Screen is updated to reflect the current status. The first record in the image identifies the area to be processed. When IPL receives the record, it verifies that the area is valid, and updates the Receiving screen as follows:
BIOS Pages to Erase16 Pages to Receive16 Page to Write16
The top line of the display identifies the actual area being processed. The remainder of the display identifies the number of 8Kbyte pages to erase, receive and write. To process the data, IPL must receive, erase, and write the entire image. With the exception of the disk images, IPL receives the entire image before starting the erase, and completes the erase before starting to write. With disk images, IPL multiplexes all three operations as data is received. Eventually, (assuming no errors), all three counts reduce to 0. When the entire image is received, and the flash is updated, IPL displays the Verify screen: BIOS Verifying Image
This screen continues to display while the image is being verified. Once the image is verified, IPL displays the Success screen:
BIOS Download Complete No Errors Detected
7-19
Page 105
PDT 7200 Series Product Reference Guide
If the Multiple Images option was selected on the Area Selection screen, IPL then immediately returns to the Waiting for Data screen to wait for the next image. If any other selection was made on the Area Selection screen, then IPL stays at the success screen until the operator acknowledges the message by pressing Enter (or the trigger, or taps Enter on the touchscreen). Once the screen has been acknowledged, IPL returns to the Area Selection screen to wait for a new selection.
IPL Error Detection While receiving data, IPL performs many checks on the data to ensure that the data is received correctly. If any error is detected, IPL immediately aborts the download, and reports the error on an error screen:
Download Failed! BIOS Error Message
This screen is displayed until the operator acknowledges the error by pressing Enter (or trigger, or Enter on the touch screen). Once the screen has been acknowledged, IPL returns to the Area Selection screen to wait for a new selection. The second line of the displays the area that was being processed when the error was detected. The third line of the screen identifies the cause of the error. The errors that can be reported, and the probable cause of the error, are as follows: Table 7-2. IPL Errors
7-20
Error
Explanation
Erase Error
Either the flash is faulty, and the terminal needs to be returned for service, or an attempt was made to erase IPL. (IPL is write-protected, and cannot be erased or programmed without special hardware.)
Page 106
Terminal Configuration: Building and Sending the Hex Image
Table 7-2. IPL Errors Error
Explanation
Line Stat Err
A framing error was detected in the incoming data. The most probable cause of this is a parameter mismatch between the terminal and the device used to transmit the data. The baud rate on both devices must be set to the same value, and the transmitting device must be set to 8 data bits, no parity, and at least one stop bit.
Baud too High
This error could be caused by transmitting large compressed images at a high baud rate, or by faulty flash.
Image Corrupt
This error indicates that the image received by the terminal is corrupt. The most probable cause of this error is inserting an IPL-enabled terminal into a cradle while the host is transmitting an image.
Wrong Area
This error indicates that the image being received has an incorrect area number. This could be caused by specifically selecting an area, and loading an image intended for another area, or by receiving an image for an area that is not defined in the partition map.
Not Hex File
This error is caused by transmitting the wrong file format to IPL. IPL can only receive Hex files supplied by Symbol, or generated by TCM.
Write Error
This error is caused by a failure of the flash device used to hold the image. If this error is seen, then the terminal requires maintenance.
Checksum Err
This error is occurs when one of the critical partitions (BIOS, Partition Map, or Setup) has been received with no apparent errors, but the received image does not pass the checksum check. The most probable cause of this error is a failure of the tool used to generate the image. If this error is displayed, then the original contents of the partition will not have been changed as the error is detected before any flash is updated.
7-21
Page 107
PDT 7200 Series Product Reference Guide
Table 7-2. IPL Errors Error
Explanation
Verify Failed:
This error is caused when the image has been written to flash, but cannot be read back correctly. It only applies to critical sections of the flash, and will prevent an exit from IPL as a critical section of the flash is damaged. If this error occurs, the terminal needs maintenance. Contact Symbol Support for help.
Lost Connection
This error displays when the terminal is removed from the cradle during an IPL session. The terminal should be replaced in the cradle when the host is no longer transmitting an image, and the image should be resent to the terminal that lost the connection. Disconnecting the serial cable does not generate this error, and attempting to reconnect the cable will result in an Image Corrupt error.
Cradle Error
This error indicates that corrupted data is being continually received from the cradle. The most probable cause of the error is an obstruction on the IRDA window on the terminal or the cradle. If cleaning the windows fails to correct the problem, then the terminal requires service. Contact Symbol Support for help.
To exit and restart IPL, hold the reset button for 12 seconds (or whatever value was set in Setup).
Restarting after Download Fails If the download fails, exit IPL by rebooting the terminal. If possible, determine what caused the download to fail. If a communications failure occurred, check all the cables between the terminal and the host to ensure they are secure. If the hex file is too large to fit in the terminal’s Flash disk, reset the hex file’s size (refer to Defining Script Properties on page 7-6) and rebuild the image.
Exiting IPL To exit IPL, cold boot the terminal (hold the reset switch for 15 seconds, or whatever value was set in Setup). The terminal should boot to an application or to a login screen.