Page 113

Operating Instructions

Appendix

Chapter 10

CLX490

10.3.2

Installation and electrical connection

The pin assignments for the connection cables of the parameter memory no. 2020307/no. 2021689 are shown in Table 5-5 and Table 5-6, Page 5-7. ¾

Mount the parameter memory on the CLX as shown in Chapter 5.5.3 Connecting the supply voltage, Page 5-13 and connect the cables to the AMV/S60 Connection Module.

The color assignment of the wiring for the open ends of the connection cables for the parameter memory no. 2020981 is shown in Table 5-12 and Table 5-13, Page 5-11. ¾

Mount the parameter memory on the CLX as shown in Chapter 5.5.3 Connecting the supply voltage, Page 5-13 and connect the cables to the external power supply.

10.3.3

Operation

The parameter memory connected to the CLX is accessed • using the selected CLX start option

• after the parameters have been downloaded to the CLX with the permanent storage function and the START WITH EXTERNAL PARAMETERS option

The start option is set at the bottom right of the DEVICE CONFIGURATION tab as shown in Fig. 10-8 (red frame). The START WITH EXTERNAL PARAMETERS option is selected by default.

Fig. 10-8:

CLV-Setup: "Device configuration" tab with the CLX start options

Start with internal parameters When it starts, the CLX loads the internal parameter set to its RAM. The "Device Ready" LED lights up constantly. Since the data is stored permanently, it is not lost when the device is switched off. Start with external parameters (default setting) When it starts, the CLX loads the external parameter set to its RAM. The "Device Ready" LED blinks for approx. 10 s and then lights up constantly. The data in the CLX is lost when the device is switched off.

8009237/RB47/2007-08-03

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-9

Page 114

Chapter 10

Appendix

Operating Instructions CLX490 Compact OMNI Scanner

Start with copy internal -> external When it starts, the CLX loads the internal parameter set to its RAM. It then copies the parameter set to the external memory and resets the start option to START WITH EXTERNAL PARAMETERS. The two "Device Ready" and "Read Result" LEDs blink alternately for approx. 10 s. The "Device Ready" LED then lights up constantly. The data is not lost when the device is switched off since it is stored permanently in the CLX and is identical to the internal parameter set. 10.3.4

Switching on the device for the first time

Since the external parameter memory in the connector cover is empty (e. g. on delivery) after the CLV has started, the CLV copies its current parameter set to the external parameter memory. Then the "Device Ready“ LED lights up. The CLX is ready for operation. If an external parameter memory containing a corresponding parameter set is connected to the CLX, the "Device Ready" LED blinks for 10 s when the device starts up correctly and then lights up constantly. 10.3.5

Adjusting the parameter set in the external parameter memory after it has been downloaded to the CLX

If the CLX is started with he START WITH EXTERNAL PARAMETERS option with an external parameter memory and if the parameter set is downloaded from CLV-Setup to the CLX with the permanent storage option, CLV-Setup asks you whether the new parameter set is also to be copied to the external parameter memory (Fig. 10-9). This function prevents the CLX from reloading the old parameter set the next time it starts.

Fig. 10-9:

¾

10-10

CLV-Setup: dialog box for adjusting the external parameter memory

Adjust the parameter set in the external parameter memory as described in Chapter 6.4.1 Configuring the CLX with CLV-Setup, Page 6-4.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03

Page 115

Operating Instructions

Appendix

Chapter 10

CLX490

10.3.6

Meaning of the LEDs

The function of the two "Device Ready" and "Read Result" LEDs, as defined by the selected start option, is shown in Table 3-3, Page 3-9. Their function in the event of an error accessing the external memory as well as troubleshooting measures are described in Table 8-2, Page 8-3. 10.3.7

Error messages

If errors occur accessing the external parameter memory, the CLX outputs error messages on the auxiliary interface. When the CLX is started, these messages can be displayed using the "CLV-Setup" program. 1.

Connect the PC to the auxiliary interface on the CLX using a 3-core RS 232 data cable (null modem cable)(AMV/S60: connect the PC to the internal 9-pin "Service" plug). See Chapter 5.5.6 Connecting the PC, Page 5-15.

2.

Switch on the PC and start Windows.

3.

Start the CLV-Setup program and open the Terminal Emulator.

4.

Choose the extended display mode.

5.

Switch on the power supply for the AMV/S60. The CLX starts and loads the parameter set from the external memory.

6.

Monitor the messages in the main window of the Terminal Emulator.

Table 8-3, Page 8-5 lists the error messages with the appropriate troubleshooting measures. 10.3.8

Replacing a CLX

The device is replaced by starting the new CLX with the START WITH EXTERNAL PARAMETERS option (default). This is the case when the device is put into service for the first time, e. g. when it is taken from the warehouse. The device must be replaced with a device of the same type (e. g. CLX490-0010 with CLX490-0010). You do not need to connect a PC to replace the device. ¾

Replace the device as described in Chapter 10.11.2 Importing the parameter set from the external memory, Page 10-34.

The CLX imports the external parameter set into the RAM. This is sufficient to ensure that the device operates correctly. If the parameter set is to be stored permanently in the CLX (EEPROM), a PC with the CLV-Setup program must be connected to upload the parameter set from the CLX to CLV-Setup and then download it again to the CLX with the PERMANENT storage option. When asked to do so, adjust the external memory again.

8009237/RB47/2007-08-03

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-11

Page 116

Chapter 10

Appendix

Operating Instructions CLX490 Compact OMNI Scanner

10.4

Optional heating

10.4.1

Features

•

Integrated heater (permanently installed)

•

Wider range of applications for CLX: up to max. –30 °C without air draft

•

Supply voltage 24 V DC +20 %/–10 %

•

CLX enabled via internal temperature switch (power-up delay approx. 35 to 40 minutes at 24 V DC and min. ambient temperature of –30 °C)

•

Required cable cross-section (power supply): min. 0.75 mm2

10.4.2

Design

The heating system comprises two parts: • the front window heater

• the housing heater

The optional heating system is installed and tested at the factory. The user cannot install it on site. 10.4.3

Function

When the 24 V DC supply voltage is applied, a temperature switch first supplies electric current to the heater only. This switch enables the supply voltage for the CLX when the internal temperature rises above 7 °C during the warm-up phase (approx. 35 min). The device then performs a self-test and switches to Reading mode. The "Device Ready" LED lights up to indicate that the device is ready. When the internal temperature has reached approx. 25°C, a further temperature switch deactivates the housing heater and reactivates it if necessary. Reading mode is not interrupted as a result. The front window heater remains active constantly. The temperature curve inside the housing is shown in Fig. 10-10. Temperature inside housing (°C)

The temperature enables the operating voltage for the CLX internally

Time (min)

Fig. 10-10: CLX with heater: Temperature curve inside the housing

10-12

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03

Page 117

Operating Instructions

Appendix

Chapter 10

CLX490

10.4.4

Electrical installation

Cable cross-section The connection cables for the power supply must have a cross-section of at least 0.75 mm2. See also Chapter 5.4.2 Supply voltage, Page 5-8. Power consumption The power consumption of the CLX with heater is typically 75 W and max. 90 W. These values apply to operation without the switching outputs connected. AMV/S60 Connection Module The AMV/S60 Connection Module is not suitable to power the CLX with heater. An external power supply pack with a higher power output should be used for this purpose. See Chapter 3.1.3 System requirements, Page 3-1. 10.4.5

Outdoor applications

If the CLX with integrated heater is used outdoors, it should be installed in a protective housing to prevent the front window from being damaged by rain, snow, or dust. The housing also acts as a wind protector.

8009237/RB47/2007-08-03

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-13

Page 118

Chapter 10

Appendix

Operating Instructions CLX490 Compact OMNI Scanner

10.5

System messages

The CLX outputs all system messages in plain text on the auxiliary interface. The messages are in English and can be displayed in the Terminal Emulator mode in the "CLV-Setup" program. See also Chapter 6.6.1 Displaying messages, Page 6-25. 10.5.1

CLX without external parameter memory

Message

Meaning

"CLX490 Soft.Vers. Vx.xx Aend. Nr. 0000 CLV 490/OTS"

The software version and revision is displayed when the power supply is switched on.

"internal parameters loaded"

When the CLX is started, it imports the parameter set that was last stored permanently in the internal EEPROM to the RAM.

"laser safety timeout"

The CLX has shut off the laser diode 10 minutes (default setting) after a reading pulse has begun. The reading interval is still active, even though the CLX is no longer reading. The reading interval has to be terminated by a corresponding pulse signal. The laser diode is activated again with the next reading pulse.

"no code"

In Reading/Percentage Evaluation mode, the CLX indicates that no codes matching the parameterized evaluation conditions were detected during the reading interval (error status ST = 2 output on the host interface).

"no ext. parameter memory connected"

No external parameter memory was connected to the CLX when it was started.

Table 10-3: CLX system messages

10.5.2

CLX with external parameter memory connected

In addition to the messages above, the CLX also indicates whether it was able to access the external parameter memory Message

Meaning

"diff. SW version of parameters" When the CLX was started, the software version of the parameter set in the external memory was different from that in the CLX. The parameter set is still transferred. "external parameters loaded"

When the CLX was started, the external parameter set was imported successfully to the RAM.

"internal parameters copied to external memory"

When the CLX was started, the internal parameter set was successfully copied to the external parameter memory.

Table 10-4: Additional CLX system messages for the connected parameter memory

10-14

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03

Page 119

Appendix

Operating Instructions

Chapter 10

CLX490

10.6

Installing and operating the "CLV-Setup" program

The scope of delivery of the CLX includes a CD-ROM ("Manuals & Software") containing the "CLV-Setup" program. The software can be installed on a PC with the following minimum configuration: Pentium II processor, 350 MHz, and 64 MB RAM, CD drive, serial interface and mouse (recommended) as well as an operating system Windows 95TM /98TM, Windows NTTM, Windows 2000TM or Windows XPTM. Approximately 25 MB of hard disk space is required to install CLV-Setup (V 4.10) and CLV Assistant, approximately 20 MB to install CLV-Setup Help (V 4.10). Note

For the lastest version of the software, see www.sick.com. 10.6.1

Preparations

1.

Make sure you have the CD-ROM ("Manuals & Software") at hand.

2.

Connect the CLX to the AMV/S60 Connection Module (see Chapter 5.5.3 Connecting the supply voltage, Page 5-13).

3.

Connect the PC to the auxiliary interface on the CLX using a 3-core RS 232 data cable (null modem cable) (AMV/S60: connect the PC to the internal 9-pin "Service" plug). See Chapter 5.5.6 Connecting the PC, Page 5-15.

4.

Switch on the power supply to the AMV/S60. Once the CLX has started, the "Device Ready" LED lights up.

5.

Switch on the PC and start Windows.

10.6.2

Installing the software

The "CLV-Setup" installation program creates a directory "CLX", a series of subdirectories, and the necessary links. A deinstaller is also installed automatically so that you can remove the program from the PC at any time. As of version V 2.6, CLV-Setup will only be available and further developed as a 32-bit application. Note

The following instructions describe how to install the software on Windows XPTM. These instructions may differ depending on the operating system you are using (for example, administrator rights are required to install the software on Windows XP). Initial installation

8009237/RB47/2007-08-03

1.

Shut down all the applications running in Windows.

2.

Insert the CD-ROM into the CD-ROM drive.

3.

If the auto run function is enabled, the start page (table of contents) of the CD-ROM is automatically displayed in the your browser.

4.

Select the SOFTWARE folder on top left.

5.

Select CLV, ICR, OPS SETUP SOFTWARE in the listing. A table with CLV-Setup information is displayed.

6.

Select DOWNLOAD under SOFTWARE FILE.

7.

In the FILE DOWNLOAD dialog box select the option to install directly from the CD-ROM. Confirm with OK. The software is automatically saved in the “Programs\CLV“ directory on your hard disk. – or –

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-15

Page 120

Chapter 10

Appendix

Operating Instructions CLX490 Compact OMNI Scanner

Select the option to save the software locally on your local hard disk and confirm with OK. In the FILE DOWNLOAD dialog box select the desired target directory on your hard disk. The “CLVSetupxx.exe“ file is then saved at this location (xx = version number). Select OPEN in the dialog box to start the file. 8.

The installation program starts and guides you through the installation with screen messages. The program asks you for your user name and company name. This information appears as a header in printouts. “CLV-Setup“ and CLV Assistant are installed. The software will be entried in the Windows start menu under “Programs\CLVSetup“.

9.

Please read the Readme file, as it contains the latest information and problems regarding CLV-Setup.

10. Then confirm the final installation message by clicking “OK. CLV-Setup Help (Online): 1.

Return to the start page of the CD-ROM with the opened folder “Software“.

2.

Select CLV, CLV HELP SOFTWARE in the listing. A table with CLV Help information is displayed.

3.

Select DOWNLOAD under SOFTWARE FILE.

4.

In the FILE DOWNLOAD dialog box select the option to install directly from the CD-ROM. Confirm with OK. The software is automatically saved in the “Programs\CLV“ directory on your hard disk. – or – Select the option to save the software locally on your local hard disk and confirm with OK. In the FILE DOWNLOAD dialog box select the same target directory as selected for CLV-Setup on your hard disk. The “CLVHelp.exe“ file is then saved at this location.

5.

After downloading the software, select OPEN in the dialog box to start the software. The software is installed in the sub directory “Help“ of the main directory “CLV“. CLV Help will also be entried in the Windows start menu under “Programs\CLV-Setup“.

6.

Restart the PC. The Windows DLL files are possibly updated.

The “CLV-Setup“ program, the CLV Assistant and the online “CLV-Setup Help“ are installed and ready. The program logs all of the files that are installed to the "install.log file". This list is used by the deinstaller to remove the program and must not be deleted if you want to use the deinstaller at some time in the future. Updating the program There are two ways of installing a new version of CLV-Setup: •

You can install the new version and still keep the old version (parallel installation) – or –

•

You can install the new version over the old version (overwrite)

Installing the new version while still keeping the old one If you want to install the new version and still keep the old version of CLV-Setup, follow the procedure described under Initial installation. When the program asks you for the target

10-16

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03

Page 121

Operating Instructions

Appendix

Chapter 10

CLX490 directory, you must specify a new directory. Both versions of the software are then available but must not be started simultaneously. The configuration files "*.scl " (scanner configuration), "*.npj" (network project), "*.opj" (OTS project) or "*.rpj" (redundant OTS project) of the old version can be used in the new version and contain the CLX parameter sets as well as project data if serveal scanners had been organized logically by projects. In order to do so, copy the configuration files from the "data" directory of the old version to the "data" directory of the new version. Installing the new version over the old one Before you can install the new version of the software, you must remove the files of the old version – with the exception of the configuration files "*.scl " (scanner configuration), "*.npj" (network project), "*.opj" (OTS project) or "*.rpj" (redundant OTS project). They contain the CLX parameter sets as well as project data if serveal scanners had been organized logically by projects. The deinstaller of the old version removes the program files. You can use the deinstaller to remove all of the files (with the exception of the configuration files) or only selected files. The default setting is a complete deinstallation (the configuration files are not deleted). With the custom deinstallation, the files are listed and only selected files are deinstalled. If you choose this deinstallation method, make sure that the main directory containing the configuration files "*.scl" is not deleted. We recommend that you move these files to a different directory and copy them back to the "data" directory later. 1.

From the Windows Start menu, choose PROGRAMS and the deinstaller for CLV-Setup. The deinstaller starts and guides you step by step through the deinstallation procedure.

2.

Choose the deinstallation method (complete or custom).

3.

Install the new version of CLV-Setup as described under Initial installation, making sure to choose the same directory.

The new version of CLV-Setup is installed. The configuration files of the old version can be used again.

8009237/RB47/2007-08-03

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-17

Page 122

Chapter 10

Appendix

Operating Instructions CLX490 Compact OMNI Scanner

10.6.3

Starting CLV-Setup

The "CLV-Setup" program is launched with the following default settings: Communication

COM 1, 9,600 bd, 8 data bits, 1 stop bit, no parity

Transmit control character

Start: STX, Stop: ETX

Receive control character

Start: STX, Stop: ETX

Unit of measure

Metric (depending on the selected installation)

Browser

User dependent

Language

As selected for the installation

Company name

As entered for the installation

User name

As entered for the installation

Automatic connection to CLX

On startup: yes On scanner selection: yes When the Terminal Emulator is started/closed: yes

Initial screen

Yes

Storage confirmation prompt

Yes

Last type selected

CLV41x

File storage location

"data" (configuration files for CLX)

Table 10-5: Default settings in CLV-Setup

1.

Switch on the PC and start Windows.

2.

Choose CLV-SETUP from the Start menu. The SICK logo is then followed by the initial screen.

3.

Confirm the initial screen. CLV-Setup then checks whether a CLX is connected to the COM 1 port on the PC (default setting in CLV-Setup) and whether the communication parameters on the PC match those on the CLX. If this is the case, the CONNECTED status is displayed with the CLX specification in the status bar right in the bottom of the screen. The CLX type (here: CLX490) is displayed in the DEVICE drop-down list in the top right of the toolbar.

The program then loads the internal description of the CLX as well as the default parameter values and displays them on the tabs. Finally, CLV-Setup uploads the current parameter set from the RAM of the CLX and displays the values on the tabs instead of the default values. You can edit the current parameter set on the tabs. The default values are displayed when the device is used for the first time. Troubleshooting If the program cannot establish communication, it outputs a timeout warning and displays NO CONNECTION in the status field. This may be because the CLX is not connected or because the communication parameters on the CLX do not match those on the PC. In this case, CLVSetup enters the CLX type of the device it last communicated with in the DEVICE field on the toolbar. The default setting the first time the device is used is the CLV41x. The software then loads the internal device description for this CLV type and displays the default values on the tabs.

10-18

© SICK AG · Division Auto Ident · Germany · All rights reserved

8009237/RB47/2007-08-03