Page 68
Chapter 6
Operation
Operating instructions CLV 43x/44x Bar Code Scanner
Parameterizing oscillating mirror functions Action 1. Choose "Oscillating Mode" (identical angles of deflection with distance configuration switchover/focus positions) – or –
Line scanner with oscillating mirror
Ö OSCILLATING MIRROR tab Ö OSCILLATING MIRROR MODE – Oscillating with fixed amplitude Ö OSCILLATING – Frequency
2. Choose "Oscillating with Variable Amplitude" mode (amplitude can be chosen for each distance configuration with distance configuration switchover/focus control) – or –
Ö "Oscillating Mirror" tab Ö OSCILLATING MIRROR MODE – Oscillating with Variable Amplitude Ö OSCILLATING – Frequency Ö READING Configuration tab Ö DISTANCE CONFIGURATIONS – Oscillating Mirror Amplitude
3. Choose "Fixed Position" mode – or –
Ö OSCILLATING MIRROR tab Ö OSCILLATING MIRROR MODE – Fixed Position Ö FIXED POSITION – Fixed angle CW
4. Choose "One-shot" mode
Ö OSCILLATING MIRROR tab Ö OSCILLATING MIRROR MODE – One-shot Ö ONE-SHOT Ö PHASE 1 Ö
5. With One-shot: choose trigger source
– Mirror Speed – Start Position – Distance Configuration (no.) PHASE 2 – Mirror Speed – Start Position – Distance Configuration (no.)
Ö OSCILLATING MIRROR tab Ö ONE-SHOT Ö TRIGGERMODE
– Reading interval or – Switching input "Sensor 2" or – Serial interface
6. With One-shot and trigger source "Sensor 2" switching input: Choose function for switching input 7. Choose start of activity of oscillating mode "Oscillating with ..."
Ö DEVICE Configuration tab Ö ASSIGNMENT SENSOR 2 – One-shot
Ö OSCILLATING MIRROR tab Ö OSCILLATING MIRROR ACTIVE
– continuous – during reading interval
Table 6-5: Guide: parameterizing the oscillating mirror functions
Explanation of oscillating mirror functions Operating mode "Oscillating with Fixed Amplitude" The oscillating mirror deflects the scan line up to the maximum angle of deflection of ±40 CW (corresponds to ±20°). Fig. 6-2 illustrates the flow when reading from above.
6-10
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 008 568/K949/10-06-2002
Page 69
Operating instructions
Operation
Chapter 6
CLV 43x/44x Bar Code Scanner
Operating mode "Oscillating with Variable Amplitude" The deflection range can be parameterized separately for each of the maximum of 8 distance configurations/focus positions. The minimum/maximum values that can be set for the deflection range for fast applications ensure that the scan line only sweeps areas that contain bar codes during the time available. This also ensures that the scan line density on the object is, for the most part, constant. Fig. 6-3 illustrates this mode when reading from above.
Oscillating frequency constant j0 j0
Scan line density dependent on object height Fig. 6-2:
Oscillating mirror operating mode "Oscillating with Fixed Amplitude"
Oscillating frequency constant Limitation of the deflection range (CW value)
Scan line density is the same, independent of object height Fig. 6-3:
8 008 568/K949/10-06-2002
Oscillating mirror operating mode "Oscillating with Variable Amplitude"
© SICK AG · Division Auto Ident · Germany · All rights reserved
6-11
Page 70
Chapter 6
Operation
Operating instructions CLV 43x/44x Bar Code Scanner
"Fixed Position" operating mode The CLV parks the oscillating mirror at the desired position. 10 CW corresponds to –20°, and 90 CW to +20° (Fig. 4-9, Page 4-7). In the default setting, the fixed position is set to 50 CW (angle of deflection of 105°). The Oscillating mode is switched off. "One-shot" operating mode In the "One-shot" mode, the oscillating mirror only performs one specific oscillating movement per reading interval. This movement consists of a forward phase and a return phase. The start position of the scan line, the mirror speed and the valid distance configuration can be parameterized separately for each phase. The end position of the one phase is also the start position of the other phase. Fig. 6-4 illustrates this mode with object tracking. During the forward phase, the front of the object that moves towards the oscillating mirror is swept once by the scan line from top to bottom in accordance with the conveyor speed. Multiple focus position switchovers are not necessary as the required reading range (DOF) is much smaller than with the Line scanner. Chapter 10.7.3 "Calculation of the start position and mirror speed for the forward and return phase of the one-shot function", Page 10-46 explains how the start position and mirror speed are calculated for a given conveyor speed. Possible trigger sources for One-shot: •
"Sensor 2" switching input
•
Command string (via serial interface)
•
Reading interval start Conventionally required reading range Line scanner
Dt 0 g din Rea ge ran
Reduced reading range via One-shot function Line scanner with oscillating mirror
Dt 1 nge g ra n i d Rea
Fig. 6-4:
6-12
One-shot: object tracking (reading from front)
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 008 568/K949/10-06-2002
Page 71
Operating instructions
Operation
Chapter 6
CLV 43x/44x Bar Code Scanner b) Parameterizing reading pulse source Action 1. Reading pulse: choose source for start
Settings
Ö DEVICE CONFIGURATION tab Ö EDIT READING TRIGGER
– Sensor Input (Active High) or – Sensor Input (Active Low) or – Serial Interface or – Free Running with Timeout or – Reflector polling
2. Pulse source "Sensor Input 1": Select debouncing and response at first trigger
Ö EDIT READING TRIGGER Ö DEBOUNCING SENSOR Ö
3. Pulse source "Serial Interface": choose trigger character
– Standard or – Fast FIRST TRIGGER – Dynamic or – Static
Ö EDIT READING TRIGGER Ö SERIAL INTERFACE
– Standard Trigger or – Single Character
4. For trigger source "Free Running": choose timeout
Ö EDIT READING TRIGGER Ö END OF LABEL / FREE RUNNING – Timeout
5. Reading pulse: choose source for end
Ö EDIT READING TRIGGER Ö END OF READING INTERVAL
– Generated by Trigger Source or – Timer
Table 6-6: Guide: parameterizing the reading pulse source c) Parameterizing laser timeout Action 1. Select the duration for the laser timeout
Settings
Ö DEVICE CONFIGURATION tab Ö EDIT READING TRIGGER Ö LASER TIMEOUT – Duration
2. Deactivate the laser timeout (laser diode is always active when pulsing, regardless of the pulse duration)
Ö LASER TIMEOUT
– Click (deactivate) the ACTIVE checkbox
Table 6-7: Guide: parameterizing the laser timeout
8 008 568/K949/10-06-2002
© SICK AG · Division Auto Ident · Germany · All rights reserved
6-13
Page 72
Chapter 6
Operation
Operating instructions CLV 43x/44x Bar Code Scanner d) Parameterizing evaluation characteristics Selecting decoder type CODE CONFIGURATION tab choose standard decoder!)
•
Activating code types for evaluation Codes"
• • •
Number of bar codes per object 1. Number n = 1
Ö
•
3. Number n > 1: – Code type different – Contents different or identical
Ö CODE CONFIGURATION tab Ö "Codes" and "Edit
Ö DEVICE CONFIGURATION tab Ö "Code Comparison" Mode for teaching in match code 1 via the "Sensor 2" switching input Ö DEVICE CONFIGURATION tab Ö"Teach-In" Setting output time of the reading result Ö DEVICE CONFIGURATION tab Ö EDIT READING TRIGGER Ö output for Good Read Activating code comparison
Line/Raster scanner
Line scanner with oscillating mirror
Stationary/moving conveyor object Stationary conveyor object
Ö CODE CONFIGURATION tab Ö CODE POSITION
Ö CODE CONFIGURATION TAB Ö CODE POSITION
Ö CODE POSITION – Activate "Compare"! Ö NUMBER OF CODES
Ö CODE POSITION – Activate "Compare"! Ö NUMBER OF CODES
– Deactivate "Compare"!
2. Number n > 1: – Same code type – Contents different or identical
Ö "Decoder" (Raster scanner:
– Minimum – Maximum
Ö CODE POSITION – Deactivate "Compare"! Ö NUMBER OF CODES – Minimum – Maximum
– Deactivate "Compare"!
– Minimum – Maximum
Ö CODE POSITION – Deactivate "Compare"! Ö NUMBER OF CODES – Minimum – Maximum
Moving conveyor object
Ö CODE CONFIGURATION TAB Ö CODE POSITION – Activate "Compare"! Ö NUMBER OF CODES
– Minimum – Maximum DEVICE CONFIGURATION tab
Ö Ö EDIT OPC SPATIAL PARAMETERS
– Code distance – Length of the increment vector – Angle of the increment vector
Table 6-8: Guide: settings to be made for the evaluation of identical bar codes e) Parameterizing output characteristics Result status: Switching outputs and beeper •
Defining function of result status output of "Result 1" and "Result 2" switching outputs DEVICE CONFIGURATION tab "Edit Result Outputs"
•
Defining function of result status output and volume of beeper tab "Edit Result Output"
Ö
Ö
Ö
Main data interface (general): • • • •
6-14
Destination of the reading result strings"
Ö DEVICE CONFIGURATION
Ö HOST INTERFACE tab Ö "Destination of result data
Ö DEVICE CONFIGURATION tab Ö "Odette Filter" Sorting of the reading result Ö DATA STRINGS tab Ö "Output Sequence Sort" Masking of the reading result Ö DATA STRINGS tab Ö "Format Mask" Filter reading result
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 008 568/K949/10-06-2002
Page 73
Operating instructions
Operation
Chapter 6
CLV 43x/44x Bar Code Scanner
Host Interface •
Arrangement on the data network arrangement"
•
Physical interface
• •
Ö DEVICE CONFIGURATION tab Ö "Scanner
Ö HOST INTERFACE tab Ö "Data Format" Communication parameters Ö HOST INTERFACE tab Ö "Data Format" Protocol Ö HOST INTERFACE tab Ö "Interface Protocol"
Function of the CAN interface (alternative): See "Application of the CAN Interface" Operating Instructions (no. 8 009 180, English edition) Data output string •
Selecting contents of header, separator and terminator Format
•
Position of separator in data output string
•
Ö DATA STRINGS tab Ö Output
Ö DATA STRINGS tab Ö Separator Position Setting wrong read format Ö DATA STRINGS tab Ö Wrong Read Format, Character Count and Error String • •
Ö DATA STRINGS tab Ö Test String Parameterizing special functions DATA STRINGS tab Ö Specials
If necessary, parameterizing/activating test string f) Parameterizing auxiliary interface (terminal interface) •
AUXILIARY INTERFACE tab
6.4.4
Configuring the CLV with AutoSetup
AutoSetup enables the reading and code configurations of the CLV to be adapted automatically to the application at hand. The following conditions must be fulfilled with regard to this: • one bar code (code type, code length)
• fixed reading distance
• no conveyor movement (static reading)
The CLV determines the code type, code length, the optimum focus position (for CLV 44x) and the most favorable scanning rate for the bar code presented. This takes place in the Free Running mode with the standard decoder in spite of the fact that the SMART decoder may have been preselected beforehand for the Reading mode. Once AutoSetup has been completed successfully, the CLV copies the parameter values determined to its parameter set. All other code types and code lengths are then disabled. If AutoSetup was not completed successfully, the parameter set is not changed. The CLV does not read a Pharmacode or any EAN/UPC types in AutoSetup. It does not react to an external reading pulse. AutoSetup can be activated in three different ways in the CLV:
8 008 568/K949/10-06-2002
•
Via the interface of CLV Setup
•
By presenting profile bar code no. 10 (on the card supplied) to the optical interface (reading window) (see Chapter 10.5.1 "Activating AutoSetup with a profile bar code", Page 10-35)
•
Via a command string (via the serial interface) (see Chapter 10.6 "Configuring the CLV with command strings", Page 10-41)
© SICK AG · Division Auto Ident · Germany · All rights reserved
6-15
Page 74
Operation
Chapter 6
Operating instructions CLV 43x/44x Bar Code Scanner
While AutoSetup is running, the CLV outputs the step-by-step changes made to the parameter values in the search mode and the determined diagnosis data via the terminal interface in the form of a list. The list can be displayed, for example, in the Terminal Emulator of CLV Setup. Once AutoSetup has been completed, appropriate error messages are output for diagnosed malfunctions in the event of a no read. Explanation
AutoSetup procedure For CLV 44x (dynamic focus position) 1.
Starting at F = 20 mm, the CLV moves across the entire focus position range in steps of 20 mm at a scanning frequency of 500 Hz. It evaluates the reading quality of 100 scans at each focus position.
2.
When it reaches the part of the focus position range with the highest reading quality, the CLV determines 5 positions from which it then calculates the maximum reading quality and the focus position.
3.
The CLV approaches this calculated focus position and performs a verification read.
4.
If the reading quality is >75 %, the CLV increases its scanning frequency in steps of 100 Hz, starting at 300 Hz, up to 800 Hz. Once again, it evaluates the reading quality of 100 scans for each step. If the reading quality is lower, the CLV changes further parameter values internally.
5.
The CLV performs verification reads ± 50 Hz about the scanning frequency with the highest reading quality.
6.
It then copies the optimum scanning frequency, the focus position, as well as the code type and code length to its parameter set.
For CLV 43x (fixed focus) 1.
Starting at 300 Hz, the CLV increases its scanning frequency to 800 Hz in steps of 100 Hz. It evaluates the reading quality of 100 scans for each step.
2.
The CLV performs verification reads ± 50 Hz around the scanning frequency with the highest reading quality.
3.
It then copies the optimum scanning frequency, the code type and code length to its parameter set.
Activating AutoSetup via the user interface of CLV Setup In CLV Setup, AutoSetup can be started by choosing the appropriate menu path, via the terminal emulator or as an executable device function (interactively). Due to the better display of the progressing procedure, we recommend starting AutoSetup via the Terminal Emulator. Preparing AutoSetup: 1.
Switch on the power supply of the AMV/S 40 Connection Module (of the CLV). After the CLV has started, it emits an acoustic signal to indicate that the self-test was successful. After a short while, it outputs two consecutive acoustic signals to indicate that it has switched to Reading mode. The "Device Ready" LED lights up. Line scanner with oscillating mirror: In AutoSetup mode, the CLV behaves as follows: – in "Oscillating" mode (default setting: oscillating with fixed Amplitude) switches to Oscillating mode and positions the scan line at the angle CW = 50 (corresponds to an angle of emergence under 105°). This position cannot be changed. – in "One-shot" mode, it also positions the scan line under the angle CW = 50
6-16
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 008 568/K949/10-06-2002
Page 75
Operating instructions
Operation
Chapter 6
CLV 43x/44x Bar Code Scanner
– in "Fixed Position" mode, on the other hand, the selected scan line position remains unchanged. 2.
Upload the current parameter set from the CLV with CLV Setup (the parameter set initially contains the factory default setting). CLV Setup displays the current values on the tabs.
3.
AutoSetup overwrites important values in the current parameter set of the CLV. Save the existing parameter set as a configuration file "*.scl" in CLV Setup before you change the parameter settings. (See Chapter 6.4.1 "Configuring CLV with the user interface of CLV Setup", Page 6-5.)
Performing AutoSetup:
8 008 568/K949/10-06-2002
1.
Click in the toolbar. The Terminal Emulator is launched. The CLV is in Reading mode.
2.
Present the application-specific bar code to the CLV at the maximum reading distance for the application. Note the limit values of the CLV here (see reading field graphs in Chapter 10.2 "Specification diagrams", Page 10-1). To prevent surface reflections, rotate the bar code towards the CLV approximately 15° away from the vertical axis (see Fig. 4-8, Page 4-6).
3.
Click AUTO SETUP under DEVICE MODE. The CLV cancels Reading mode. The "Device Ready" LED is extinguished. The CLV then starts AutoSetup and temporarily activates the standard decoder. The scan line appears when the "Laser On" LED lights up.
4.
The CLV then runs AutoSetup and continuously displays the changes to the parameter values in the terminal emulator. Fig. 6-5, Page 6-18 contains an example of this.
5.
The CLV confirms that AutoSetup has been completed successfully by outputting two consecutive acoustic signals via the beeper and activating the "Result" LED for 100 ms (default setting). To begin with, it only overwrites the appropriate parameter values in the RAM. The CLV returns to the Reading mode with the decoder parameterized for this purpose. The "Device Ready" LED lights up. The READING MODE radio button in the terminal emulator is not, however, displayed as the active device mode again. The CLV only evaluates the presented code type and length the next time a bar code is read.
6.
Perform an upload from the CLV. This is done by clicking in the toolbar. CLV loads the current parameter set from the RAM of the CLV and displays the new values in the tabs.
7.
Perform a download to the CLV. This is done by clicking The DOWNLOAD PARAMETER dialog box is displayed.
8.
Confirm the dialog box by selecting the PERMANENT save option. The CLV saves the changes made to the parameter set permanently in the EEPROM.
9.
Close the Terminal Emulator.
© SICK AG · Division Auto Ident · Germany · All rights reserved in the toolbar.
6-17
Page 76
Chapter 6
Operation
Operating instructions CLV 43x/44x Bar Code Scanner with: FocPos = focus position Step = motor step for dynamic focus control Amp = amplifier setting ScanFreq = scanning frequency ReadRate = reading quality CodeType = code type CodeLen = code length CodeStr = data contents of the bar code Fig. 6-5:
CLV Setup: AutoSetup displayed in the Terminal Emulator for CLV 44x
Cancelling AutoSetup by the CLV in the event of a no read: If the CLV was not able to execute AutoSetup successfully, it cancels the search mode after the last increase made to the parameter values by outputting a long acoustic signal via the beeper. Depending on the reason for AutoSetup being terminated, it outputs one of the following error messages via the Terminal Interface: "No valid code found" (no valid bar code found or reading quality less than 75 %) "Cancel AutoSetup" (AutoSetup terminates) – or – "More than one code found" "Cancel AutoSetup" (AutoSetup terminates) The current parameter set in the CLV does not change. The CLV returns to the Reading mode with the decoder parameterized for this purpose. The "Device Ready" LED lights up. Running AutoSetup again:
6-18
1.
Click READING MODE under DEVICE MODE. The CLV switches to Reading mode.
2.
Click AUTO SETUP under DEVICE MODE again. The CLV then starts AutoSetup.