CLV 43x / 44x Bar Code Scanner Operating Instructions: 10 Appendix, pages 137 to 146

Pages 137 to 146 of 194, from CLV 43x / 44x Bar Code Scanner Operating Instructions

The original document is the PDF (5.7 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Sick CLV430, Sick CLV431, Sick CLV432, Sick CLV442 and Sick CLV440

Page 137

Appendix

Operating instructions

Chapter 10

CLV 43x/44x Bar Code Scanner

Deflection range [mm]

CLV 440-6010: focus position F = 215 mm

400 300

40

200

20 mm m 5 . 0 5m 0.3 5 mm 0.2

100 0 -100 -200 -200

-100

0

100

200

300

400

500

Reading distance [mm] Fig. 10-19: CLV 431-/440-6010: deflection range as a function of the reading distance, deflection angle and resolution

Deflection range [mm]

CLV 440-6010: focus position F = 140 mm

400 300

40

100

20 mm 5 . 0 mm 5 0.3 2 mm 0.

200

0 -100 -200 -200

-100

0

100

200

300

400

500

Reading distance [mm] Fig. 10-20: CLV 432-/440-6010: deflection range as a function of the reading distance, deflection angle and resolution

8 008 568/K949/10-06-2002

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-21

Page 138

10-22

© SICK AG · Division Auto Ident · Germany · All rights reserved

0

50

100

150

200

250

40

80

120

200

25°

0.15 mm

240 Focus position [mm] min. max.reading readingdistance distanc Aperture Aperture angle angle 25 25°° max. reading reading distance distanc Aperture angle 25°° Aperture angle 25

Aperture angle:

Resolution:

Scanning frequency/reading distance characteristics: see Fig. 10-24, Page 10-25

160

Reading conditions: see Table 10-1, Page 10-1

Appendix

Reading range (DOF) = max. reading distance minus min. reading distance

0

Reading distance [mm]

CLV 442-0010 (front reading window)

Chapter 10 Operating instructions

CLV 43x/44x Bar Code Scanner

10.2.15 Reading field diagrams for CLV 442 Line scanner (front reading window)

Fig. 10-21: CLV 442-0010: min. and max. reading distance for Line scanner as a function of the focus position at resolution 0.15 mm

8 008 568/K949/10-06-2002

Page 139

8 008 568/K949/10-06-2002

© SICK AG · Division Auto Ident · Germany · All rights reserved

0

50

100

150

250

300

25°

0.20 mm

Focus position [mm]

350 min. reading max. readingdistance distanc Aperture Aperture angle angle25 25°° max. reading reading distance max. distanc Aperture angle Aperture angle25 25°°

Aperture angle:

Resolution:

Scanning frequency/reading distance characteristics: see Fig. 10-24, Page 10-25

200

Reading conditions: see Table 10-1, Page 10-1

Appendix

Reading range (DOF) = max. reading distance minus min. reading distance

0

50

100

150

200

250

300

350

400

Reading distance [mm]

CLV 442-0010 (front reading window)

Operating instructions Chapter 10

CLV 43x/44x Bar Code Scanner

Fig. 10-22: CLV 442-0010: min. and max. reading distance for Line scanner as a function of the focus position at resolution 0.20 mm

10-23

Page 140

10-24

© SICK AG · Division Auto Ident · Germany · All rights reserved

80

120

160

Reading range (DOF) = max. reading distance minus min. reading distance

40

240

280

25°

0.25 mm

320

360 Focus position [mm] min. reading max. readingdistance distanc Aperture angle Aperture angle25 25°° max. reading max. reading distance distanc Aperture angle angle25 25°°

Aperture angle:

Resolution:

Scanning frequency/reading distance characteristics: see Fig. 10-24, Page 10-25

200

Reading conditions: see Table 10-1, Page 10-1

Appendix

0 0

50

100

150

200

250

300

350

400

Reading distance [mm]

CLV 442-0010 (front reading window)

Chapter 10 Operating instructions

CLV 43x/44x Bar Code Scanner

Fig. 10-23: CLV 442-0010: min. and max. reading distance for Line scanner as a function of the focus position at resolution 0.25 mm

8 008 568/K949/10-06-2002

Page 141

Appendix

Operating instructions

Chapter 10

CLV 43x/44x Bar Code Scanner

10.2.16 CLV 442 characteristics field of scanning frequency

Scanning frequency [Hz] Resolution

800

0.15 mm

700

0.20 mm

600

0.25 mm

500 400 300 200 100 0

0

100

200

300

400

500

600

700

800

Reading distance [mm] Reading conditions: see Table 10-1, Page 10-1

Fig. 10-24: CLV 442-0010: scanning frequency as a function of the reading distance and resolution

8 008 568/K949/10-06-2002

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-25

Page 142

Chapter 10

Appendix

Operating instructions CLV 43x/44x Bar Code Scanner

10.3

System messages

The CLV outputs all system messages in plain text via the terminal interface. The messages are in English and can be displayed in extended mode of the Terminal Emulator in the CLV Setup user interface (see Chapter 6.6 "CLV messages", Page 6-32). Message

Meaning

"CLV 440 Soft.Vers. V2.20 Rev. no. J934 Host: RS232"

The software version and revision is displayed when the power supply is switched on.

"no code"

In Reading/Parameter Evaluation mode, the CLV indicates that no codes matching the parameterized evaluation conditions were detected during the reading interval. (error status ST = 2 output on the host interface).

"laser safety timeout"

The CLV has deactivated the laser diode 10 minutes (default setting) after a reading pulse was initiated. The reading pulse is still active, even though the CLV is no longer reading. The reading pulse is to be terminated by resetting the pulse signal. The laser diode is activated again with the next reading pulse.

"cancel AutoSetup"

The CLV has terminated the AutoSetup, no valid bar code was found.

Table 10-5: CLV system messages

10.4

Installation and operating instructions for the PC-based "CLV Setup" program

The scope of delivery of the CLV includes a CD-ROM containing the "CLV Setup" program. The following is required to run the program: a PC with at least an 80486 running at 66 MHz, 16 MB RAM, CD-ROM drive, serial interface, mouse (recommended) and Windows 95TM/ 98TM, Windows NTTM or Windows XPTM. Approximately 26 MB of hard disk space is required to install "CLV Setup", "CLV Assistant", "CLV Setup Help" and the offline browser "I-ViewProTM". 10.4.1

10-26

Preparing for installation

1.

Make sure you have the CD-ROM at hand.

2.

Connect the CLV to the AMV/S 40 Connection Module (see Chapter 5.5.3 "Connecting the power supply", Page 5-4).

3.

Connect the PC to the terminal interface (port) of the CLV using an RS 232 data connection cable (AMV/S 40: connect the PC to the internal, 9-pin "Service" plug). See Chapter 5.5.6 "Connecting the PC", Page 5-6.

4.

Switch on the power supply to the AMV/S 40. Once the CLV has started, it outputs a tone to indicate that the self-test was successfully completed. Shortly afterwards, it outputs two tones to indicate that it has assumed Reading mode. The "Device Ready" LED lights up.

5.

Switch on the PC and start Windows.

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 008 568/K949/10-06-2002

Page 143

Operating instructions

Appendix

Chapter 10

CLV 43x/44x Bar Code Scanner

10.4.2

Performing installation

The CLV Setup installation program creates a directory called "CLV", a series of subdirectories and the necessary links. Using the included uninstaller, CLV Setup can be completely removed from the PC. As of version V2.6, CLV Setup will only be available and further developed as a 32-bit application. Initial installation 1.

Close all Windows applications.

2.

Insert the CD-ROM in the CD drive.

3.

Call up the RUN dialog box in Windows. In the OPEN drop-down list, enter: "X\setup.exe" (X= character of CD drive). Confirm the dialog box with "OK".

– or – Open Windows Explorer, open the CD drive and double-click the "setup.exe" file. The installation program is started and guides you step by step through the installation procedure. The program asks you for your user name and company name. This information appears as a header in printouts. "CLV Setup", "CLV Assistant" and the online help "CLV Setup Help" are installed based on the selected configuration. 4.

Install the "I-ViewProTM" HTML browser if necessary (the system asks you whether you want to do so).

If you choose not to install "I-ViewProTM", the system searches your hard disk for the "Netscape NavigatorTM" and uses it to display "CLV Setup Help". 5.

Please read the Readme file, as it contains the latest information and problems regarding CLV Setup.

6.

Then confirm the final installation message by clicking "OK".

7.

Restart the PC. DLL files may be updated in Windows.

The "CLV Setup" program and "CLV Setup Help" are installed and ready. The program logs all of the files that are installed to the "install.log" file. This list is used by the uninstaller to remove the program and must not be deleted if you want to use the uninstaller at some time in the future. Updating the program There are two options for installing a new version of CLV Setup: •

You can install the new version and still keep the old version (parallel installation)

– or – •

8 008 568/K949/10-06-2002

You can install the new version over the old version (overwrite)

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-27

Page 144

Chapter 10

Appendix

Operating instructions CLV 43x/44x Bar Code Scanner

Installing new version and retaining old version If you want to install the new version and still keep the old version of CLV Setup, follow the procedure described under Initial installation. When the program asks you for the target directory, you must specify a new directory. Both versions of the software are then available but may not be started simultaneously. The configuration files "*.scl" of the old version can be used in the new version and contain the CLV parameter sets. In order to do so, copy the configuration files from the "data" directory of the old version to the "data" directory of the new version. Installing new version over old version When installing the new version over of the old version, the files of the old version must first be deleted (exception: the "*.scl" configuration files). They contain the parameter sets of the CLV. The uninstaller of the old version removes the program files. You can use the uninstaller to remove all of the files (with the exception of the configuration files) or only selected files. The default setting is a complete uninstall (the configuration files are not deleted). With the custom uninstall, the files are listed and only selected files are uninstalled. If you choose this uninstall method, ensure that the main directory containing the configuration files "*.scl" is not deleted. We recommend that you move these files to a different directory and copy them back to the "data" directory later. 1.

From the Windows Start menu, choose "Programs" and the uninstaller for CLV Setup. The uninstaller starts and guides you step by step through the uninstall procedure.

2.

Select the uninstall method (complete or custom).

3.

Install the new version of CLV Setup as described under Initial installation. Select the same directory.

The new version of CLV Setup is installed. The configuration files of the old version can be used again.

10-28

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 008 568/K949/10-06-2002

Page 145

Operating instructions

Appendix

Chapter 10

CLV 43x/44x Bar Code Scanner

10.4.3

Starting "CLV Setup"

The "CLV Setup" program is launched with the following default setting: Communication

COM 1, 9 600 bits/s, 8 data bits, 1 stop bit, no parity

Transmit control character

Start: STX, stop: ETX

Receive control character

Start: STX, stop: ETX

Unit of measure

Metric (depending on the selected installation)

Browser

I-ViewProTM (if installed), alternatively Netscape NavigatorTM

Language

As selected for the installation

Company name

As entered for the installation

User name

As entered for the installation

Establish connection to CLV At startup: yes automatically On scanner selection: yes When the Terminal Emulator is started/closed: yes Initial screen

Yes

Storage confirmation prompt

Yes

Last type selected

CLV 41x

File storage location

"data" (configuration files for CLV)

Table 10-6: Default settings of CLV Setup (extract)

1.

Switch on the PC and start Windows.

2.

Choose CLV Setup from the Start menu. The introductory dialog box is displayed after the identifier for the SICK software.

3.

Confirm the initial screen with "OK". CLV Setup then checks whether a CLV is connected to the COM 1 port on the PC and whether the communication parameters on the PC match those on the CLV. If this is the case, the "Connected" status is displayed with the CLV specification on the right of the screen, below the status bar. The detected CLV type is displayed in the toolbar in the DEVICE drop-down list (in this case: CLV 43x/CLV 44x).

The software then loads the internal device description of the CLV and the defaults settings of the parameter values from its database and displays these on the tabs. The software then copies the current parameter set from the memory (RAM) of the CLV. This is displayed on the tabs instead of the default setting. You can edit the current parameter set on the tabs. The default setting is displayed when the device is used for the first time. Troubleshooting If CLV Setup cannot establish a connection with the CLV, it displays "No connection" in the status field on the bottom, right-hand side. There are two possible causes for this. Either the CLV is not connected or the communication parameters on the CLV do not match those on the PC. In this case, CLV Setup enters the CLV type of the device it last communicated with in the DEVICE drop-down list on the toolbar. The default setting is CLV 41x the first time CLV Setup is started. The software then loads the internal device description for this type and the default setting of the parameter values from its database and displays these on the tabs.

8 008 568/K949/10-06-2002

© SICK AG · Division Auto Ident · Germany · All rights reserved

10-29

Page 146

Chapter 10

Appendix

Operating instructions CLV 43x/44x Bar Code Scanner

1.

Connect the PC to the terminal interface (port) of the CLV using an RS 232 data connection cable (AMV/S 40: connect the PC to the internal, 9-pin "Service" plug). See Chapter 5.5.6 "Connecting the PC", Page 5-6.

2.

Click (AutoBaud detect) in the toolbar or choose Options, AUTOBAUD DETECT. CLV Setup scans the serial interface by varying the communication parameters and sends a telegram to the CLV repeatedly. As soon as a reply from the CLV is detected, CLV Setup conveys the values found for the communication parameters. Fig. 10-25 shows an example of the AutoBaud detect result display. CLV Setup displays "Connected" in the status bar on the bottom, right-hand side.

3.

Confirm the AUTO DETECT dialog box with "OK". CLV Setup displays the detected CLV type in a separate dialog box and asks you whether you want to upload the current parameter set from the CLV.

4.

Confirm the dialog box with YES. CLV Setup then copies the current parameter set from the RAM of the CLV to its database and displays the values on the tabs.

You can edit the current parameter set on the tabs.

Fig. 10-25: CLV Setup: results of the AutoBaud detect function

– or – 1.

Select the menu item SERIAL INTERFACE under EXTRAS in the menu bar.

CLV Setup displays the current communication parameter settings of the PC in the COM PARAMETERS dialog box. 2.

Ensure that the communication parameters of the PC and CLV are identical (connected COM port, 9 600 bits/s, 8 data bits, 1 stop bit, no parity).

3.

Confirm the dialog box with "OK". CLV Setup attempts to communicate with the CLV again. If it is successful, CLV Setup displays "Connected " in the status bar on the bottom, righthand side.

4.

Click in the toolbar. CLV Setup then copies the current parameter set from the RAM of the CLV to its database and displays the values on the tabs.

You can edit the current parameter set on the tabs.

10-30

© SICK AG · Division Auto Ident · Germany · All rights reserved

8 008 568/K949/10-06-2002