The text of a manual · part 10 of 17
Operating Instructions: CLV 42x Bar Code Scanner, Standard line, pages 77 to 88
Pages 77 to 88 of 146: 7 Maintenance; 8 Troubleshooting
The original document is the PDF (3.7 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Sick CLV420 and Sick CLV422
Page 77
Maintenance
Operating Instructions
Chapter 7
CLV 42x bar code scanner
7
Maintenance 7.1
Cleaning during operation
To ensure optimum reading performance, the reading window of the CLV must be clean at all times. For this reason, we recommend that you check the reading window at regular intervals, particularly if the device is operated under extreme conditions (dust, abrasion, humidity, finger prints). Laser radiation can seriously damage your eyesight The CLV uses a class 2 red-light laser. Looking directly at the laser beam can seriously damage the retina in your eyes. As with sunlight, never look directly into the laser beam. Switch off the device while it is being cleaned. Never look directly into the reading window while the device is in operation.
Damage to the reading window The reading window is made of glass. Scratches and streaks on the window will reduce the reading performance. Use a mild, non-scouring detergent. Never use scratching or scouring motions to clean the reading window.
Cleaning reading window Clean the reading window at regular intervals with a mild, non-scouring detergent, e.g. an antistatic glass cleaning fluid. Fig. 7-1 shows the areas that require cleaning. Use a soft, lint-free cloth. If necessary, clean the LEDs on the rear of the device. Line/Raster scanner
Line/Raster scanner with 105° angle attachment clean here: Fig. 7-1:
Note
8 009 981/O078/16-08-2004
Cleaning the reading window
Don’t open the device. The producer warranty will be forfeited if the device is opened.
© SICK AG · Division Auto Ident · Germany · All rights reserved
7-1
Page 78
Chapter 7
Maintenance
Operating Instructions CLV 42x bar code scanner
Cleaning other optical surfaces: If an external reading pulse generator with a sensor (e. g. photoelectric reflex switch) is used, make sure to clean the relevant optical surfaces (Fig. 7-2). Contamination on these surfaces can cause switching errors. If the reading pulse is generated by reflector polling, make sure to clean the surface of the reflector.
Fig. 7-2:
7.2
Cleaning the external optical sensor (reading pulse generator)
Maintenance
The CLV does not require any maintenance except the described cleaning measures. Its self-monitoring function ensures fault-free operation over long periods of time. Any faults or malfunctions are displayed as error messages via the terminal interface. If they are plain text, they can be displayed on the PC screen in the extended mode of the Terminal Emulator in the CLV Setup program. Additional errors are displayed as number combinations in the fourth field from the left on the bottom, in the status line (see Chapter 8.3 Error messages, Page 8-2) .
7.3
Disposal
Irreparable devices or devices that are no longer required are to be disposed of in an environmentally-friendly manner: 1.
Always observe the applicable national waste disposal regulations.
2.
Remove the CLV housing.
3.
Remove the electronic modules and the connection cable.
4.
Remove the glass panel from the reading window and send it to be recycled.
5.
Send the chassis and cover (zinc die-cast) to be recycled.
6.
Send the electronic modules and connection cable for disposal as special waste.
At present, SICK AG does not accept any unusable or irreparable devices.
7-2
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 009 981/O078/16-08-2004
Page 79
Operating Instructions
Troubleshooting
Chapter 8
CLV 42x bar code scanner
8
Troubleshooting 8.1
Overview of errors and malfunctions which could occur
8.1.1
Installation errors
•
CLV aligned incorrectly with the object carrying the bar code (e. g. masking)
•
Reading pulse sensor positioned incorrectly
8.1.2 •
Electrical connection errors
Interfaces on the CLV connected incorrectly (wiring error in CDB 420 or CDM 420)
8.1.3
Parameterization errors
•
Functions not adjusted to local conditions, e. g. communication parameters on the host interface are set incorrectly
•
Technical limits of the device exceeded, e. g. relationship between the scanning frequency and the reading distance
8.1.4
Malfunctions during operation
•
Laser timeout for laser diode elapsed
•
Device error (hardware/software)
8.2
Monitoring errors and malfunctions
The CLV is self-monitoring:
8 009 981/O078/16-08-2004
•
After the power supply has been switched on, the CLV automatically carries out a selftest, in which it checks important hardware components, before it is initialized (loading of parameter set and initializing of device functions). The self-test can be repeated at any time by canceling reading mode. For calling up, see Chapter 6.5.9 Self-test, Page 6-23.
•
If the CLV detects an error during the self-test or at any other time, it outputs the error status ST=3 via the host interface. Prerequisite: the error status is enabled for transmission. By default, transfer is blocked. The CLV displays the diagnosed error as a numeric combination in the fourth field from the left on the bottom, in the status line of CLV Setup (Table 8-1).
•
While it is in operation, the CLV constantly monitors the laser diode and speed of the polygon mirror wheel. In addition, a watchdog circuit responds to error statuses.
•
A timer automatically deactivates the laser diode 10 minutes (default setting) after a continuous reading pulse is initiated in reading mode with switching input pulse modes "Sensor Input" and "Serial Interface". In this case, the CLV outputs the message: "Laser safety timeout". The reading pulse must be terminated by resetting the pulse signal. The laser diode is activated again by the next reading pulse.
© SICK AG · Division Auto Ident · Germany · All rights reserved
8-1
Page 80
Chapter 8
Troubleshooting
Operating Instructions CLV 42x bar code scanner
8.3
Error messages
The CLV outputs error messages in plain text (English) on the terminal interface only. Additional errors are displayed as number combinations in the status line of CLV Setup. Table 8-1 contains a list of the messages in alphabetical order. For warnings, see Chapter 6.6.3 Warnings, Page 6-25 Displaying messages: To display the messages on the PC in the CLV Setup user interface, proceed as follows: 1.
Click in the toolbar. The Terminal Emulator window is then displayed.
2.
In the TERMINAL MODE group, activate the EXTENDED radio button.
Message
Meaning
Possible cause
"CLV SYS-Failure: xxx" 011 RAM error 012 RAM error 013 RAM error 051 Mirror wheel speed out of tolerance 061 Mirror interval out of tolerance 071 Error: laser shutter
The CLV has diagnosed a Device fault system error. The "Device Ready" LED does not light up or it extinguishes. The CLV restarts the system.
If this system error occurs again after the CLV is restarted: Contact the SICK Service department.
"EEPROM Check fault, default parameters loaded"
The CLV has diagnosed an irregularity in the EEPROM and downloads the entire default settings to the RAM instead of the permanently stored application-specific parameters.
Device fault
Contact the SICK Service department.
"EEPROM - reading fault"
The CLV cannot download the Device fault application-specific parameters, permanently stored in the EEPROM, to the RAM.
Contact the SICK Service department.
"EEPROM - writing fault"
The CLV cannot write the current parameters in the RAM to the EEPROM.
Contact the SICK Service department.
"Host – receive fault"
The host has requested the • Host receive set incorrectly data output string last sent by • Data transmission interrupted the CLV three times; NAK has e.g. by cable break been issued with each attempt.
"Host – transfer too slow"
In the CLV, data is available for output faster than it is transmitted to the host.
Device fault
• Baud rate too low • Incorrect protocol type selected • Protocol timeout too long for the reading pulse frequency
Remedy
• Check host. • Check connecting cables and monitor data transmission at the host interface. To do so, upload from the CLV using the CLV Setup program, choose the AUXILIARY INTERFACE tab and select MONITOR HOST INTERFACE from the drop-down list. Download to the CLV! Start the Terminal Emulator. Check the configuration of the host interface.
Table 8-1: Error message output on the terminal interface
8-2
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 009 981/O078/16-08-2004
Page 81
Operating Instructions
Troubleshooting
Chapter 8
CLV 42x bar code scanner
Message
Meaning
Possible cause
Remedy
"Host – transmit fault"
The CLV cannot correctly receive the data sent by the host.
• Host interface of the CLV configured incorrectly • Data transmission interrupted e. g. by cable break
• Check the CLV and host.
"More than one code found"
AutoSetup was not successful.
More than one bar code was presented to the CLV
Repeat AutoSetup. Make sure that only one bar code is located within the visible range of the CLV!
"No valid code found"
AutoSetup was not successful.
No bar code was presented to Repeat AutoSetup. the CLV or the reading quality did Make sure that the bar code is not reach 75% located within the reading range of the CLV.
• Check connecting cables and monitor data transmission at the host interface. To do so, upload from the CLV using the CLV Setup program, choose the AUXILIARY INTERFACE tab and select MONITOR HOST INTERFACE from the drop-down list. Download to the CLV! Start the Terminal Emulator.
Table 8-1: Error message output on the terminal interface (contd.)
8 009 981/O078/16-08-2004
© SICK AG · Division Auto Ident · Germany · All rights reserved
8-3
Page 82
Troubleshooting
Chapter 8
Operating Instructions CLV 42x bar code scanner
8.4
Error status ST in read result of a bar code
Value
Meaning
Possible cause
Remedy
0
"Good read"
– not applicable
1
Incorrect check digit
The check digit calculated by the CLV during the reading process does not match that printed in the bar code.
Check whether the check digits in the bar code which were generated by the print process are correct.
2
No bar code of the enabled code type • No bar code in the CLV's reading range • Synchronize the pulse of the CLV found according to the evaluation during the reading pulse with the entry of a bar code into the criteria reading area. • Code type/length in the CLV not • Correct the code configuration enabled for decoding parameters. • Reading window obstructed/dirty • Check the reading window.
3
Device fault
Device fault
Call up the self-test! See Chapter 6.5.9 Self-test, Page 6-23. If result ≠ 000: Contact the SICK Service department.
5
Required number of successful multiple reads for the bar code not reached
• Conveyor speed of the object is too high • Scanning frequency is too low • Ladder-type arrangement for bar code relative to the conveyor direction: code height (bar length) too poor • Print quality too poor
• Check the conveyor speed. • Adjust the scanning frequency. • Check the code height.
• Check the print quality.
6
Master: • The parameterized number of slaves • Check the entered number of Number of slave reading results does does not match the actual number of slaves. slaves available. not match the parameterized number of slaves • Timeout for the master is selected too • Check the master timeout setting. short. • Data connection between slaves and/ • Check the cables between the or the master has been interrupted. slaves and the master.
7
The source of the reading result is the The bar code was not detected by the not applicable auxiliary input via the terminal reading function of the CLV, but was interface entered subsequently with the auxiliary function of the terminal interface and sent to the host in a separate data string.
9
The OUTPUT FILTER function is also The scanned object does not have a bar activated for the code comparison. code that matches the specified match The CLV has detected valid bar codes. code. However, these do not match the active match code(s)
not applicable
A
The CHECK MAX. NUMBER OF CODES One object in a set of objects with a function is active. constant number of bar codes, for The number of valid bar codes example, contains more bar codes than detected by the CLV in reading mode defined in the application. exceeds that specified under NUMBER OF CODES: MAXIMUM. It repeatedly outputs the defined error string until the number specified under NUMBER OF CODES: MINIMUM is reached.
This message is used to indicate errors on the objects (e.g. check whether objects are homogeneous: mix with incorrect objects).
D
The "Code 32" evaluation option is activated for Code 39. The CLV is attempting to interpret 6-digit C39 bar codes as C32 bar codes (output as 9-digit decimal values).
The read 6-digit bar code is not a C32 bar not applicable code. The CLV outputs the defined error string instead.
Table 8-2: Meaning of the ST error status in the reading result
8-4
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 009 981/O078/16-08-2004
Page 83
Operating Instructions
Troubleshooting
Chapter 8
CLV 42x bar code scanner
Value
Meaning
Possible cause
Remedy
F
The CHECK MIN. NUMBER OF CODES function is active. The number of valid codes detected by the CLV in Reading mode is less than that specified under NUMBER OF CODES: MINIMUM. Instead of the code contents, it outputs the defined error string in the quantity defined under NUMBER OF CODES: MINIMUM
One object in a set of objects with a constant number of codes, for example, contains fewer codes than defined in the application.
This message is used to indicate errors on the objects (e. g. check whether objects are homogeneous: mix with incorrect objects).
Table 8-2: Meaning of the ST error status in the reading result (contd.)
8.5
Troubleshooting
The following are required for troubleshooting purposes using the tables below: •
These Operating Instructions
•
Tools
•
A tape measure (up to 1 m (39.4 in))
•
A digital measuring device (ammeter/voltmeter)
•
A PC running "CLV Setup"
•
A 3-core RS 232 data cable (null modem cable), pins 2 and 3 crossed, e. g. no. 2 014 054
8.5.1
General malfunction: CLV not ready
Malfunction
Possible cause
The "Device Ready" LED is not lit. The "Device Ready" switching output (default setting: "Good read") is not outputting a pulse.
After switching on the power supply (10 to 30 V DC): • CLV not supplied with operating voltage
Remedy
• Check wiring (power connector fitted securely to the CDB 420 or CDM 420 connection module). Measure voltage at pin 1 and pin 5. • The CLV has diagnosed a device error in • Switch the CLV off and on again. Does the self-test the LED light up? If not, contact the SICK Service department. During operation: • The CLV is not in "reading" mode
• Return to reading mode using the CLV Setup program See Chapter 6.5.1 Reading mode (standard operating mode), Page 6-14. • The CLV has deactivated the laser diode • Terminate reading pulse. Check the reading pulse function. 10 min (default setting) after the start of Generate new pulse or parameterize the current reading pulse (pulse mode: new laser timeout duration using the CLV sensor input/serial interface) Setup user interface. See Chapter 6.4.3 Parameterization guide, Page 6-7. Table 8-3: Troubleshooting: restoring operation (reading mode)
8 009 981/O078/16-08-2004
© SICK AG · Division Auto Ident · Germany · All rights reserved
8-5
Page 84
Troubleshooting
Chapter 8
Operating Instructions CLV 42x bar code scanner
8.5.2 Prerequisite
Malfunctions in reading mode: reading pulse errors
The "Device Ready" LED lights up. If not, see Table 8-3, Page 8-5.
Malfunction
Possible cause
1. Pulse mode: • Incorrect reading pulse source sensor input parameterized on the CLV (external sensor). The CLV cannot be pulsed: – the "Laser On" LED does not light up • Polarity of switching input inverted – the scan line does not appear • Sensor not connected or connected to incorrect switching input
Remedy • In the CLV Setup program: Select the DEVICE CONFIGURATION tab and click the READING TRIGGER PARAMETERS button. READING TRIGGER SOURCE group: sensor input selected? • Active high: reading pulse starts when power applied • Connect sensor to "Sensor 1" switching input. Check sensor wiring (Fig. 5-4, Page 5-8). Measure output signal on sensor. • Insert jumper between pin 5 and pin 15
• If sensor supplied by CLV: sensor not connected to ground potential • Photoelectric switch not aligned with the • Align sensor. Check functioning. reflector (photoelectric switch/inductive sensor not dampened) • Incorrect reading distance (polling reflector) • Incorrect reflector type used
2. Pulse mode: Reflector polling The CLV cannot be pulsed: – the "Laser On" LED flickers but does not light up constantly – the scan line is very faint
• Incorrect pulsing sequence (reflector is not blocked to start the reading interval on the CLV)
• Check: reading distance of reflector in specified reading range? • Check reflector type, Table 4-2, Page 4-11. • Check sequence: Read: 1. Block reflector, 2. Read bar code, 3. Unblock reflector.
3. Pulse mode: • Incorrect reading pulse source Serial Interface parameterized on the CLV The CLV cannot be pulsed: – the "Laser On" LED does not light up – the scan line does not appear
• In the CLV Setup program: Select the DEVICE CONFIGURATION tab and click the READING TRIGGER PARAMETERS button. In the READING TRIGGER SOURCE group: is serial interface selected? Group SERIAL INTERFACE: is trigger type correct? • The CLV is not receiving any command • Check data connection to host. In the CLV Setup program: Select the string to start the reading interval on the HOST INTERFACE tab. data interface In DATA FORMAT group: are the correct interface type (hardware) and data format selected? Group: INTERFACE PROTOCOL: are the correct start and stop characters selected? • Use the CLV Setup program to check the host command strings: Choose AUXILIARY INTERFACE tab. Choose MONITOR HOST INTERFACE function. See Chapter 6.5.7 Monitor host interface, Page 6-21. Download temporarily to the CLV.
4.
CLV does not respond to an external signal that ends the reading pulse (sensor input, reflector polling, serial interface)
• Incorrect end of reading pulse parameterized on the CLV
• No sensor signal
• In the CLV Setup program: Check by choosing the DEVICE CONFIGURATION tab, click the READING TRIGGER PARAMETERS button. END OF READING INTERVAL group: is "Generated by Trigger Source" selected? • See Malfunction 1, Remedy
Table 8-4: Troubleshooting: reading pulse errors in reading mode
8-6
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 009 981/O078/16-08-2004
Page 85
Operating Instructions
Troubleshooting
Chapter 8
CLV 42x bar code scanner
8.5.3
Malfunctions in reading mode: result output errors
Check whether the "Device Ready" LED illuminates. If the LED does not illuminate, follow Table 8-3, Page 8-5. For the following, the number of bar codes to be read is 1. Malfunction 1.
Possible cause
The "Result" LED • (default setting: "Good read") does not light up at the end of the reading pulse. – The "Result 2" switching output (default setting: "Good read") is • not outputting a pulse – The CLV outputs the status ST = 2 in the reading result on the host interface (if ST is enabled for the separator; this is not selected in the default setting) •
• • •
•
Remedy
Read not successful, since there was no • Synchronize reading with an object in the bar code in the reading field during the reading field. reading pulse See Chapter 8.5.2 Malfunctions in reading mode: reading pulse errors, Page 8-6. Scan line positioned incorrectly • Align CLV. Is bar code at center of scan line? In the CLV Setup program: check reading quality. Choose PERCENTAGE EVALUATION. See Chapter 6.5.2 Percentage evaluation, Page 6-17. Evaluation range of the scan line is • In the CLV Setup program: Check by limited improperly (CP values) choosing the READING CONFIGURATION tab, CODE POSITION group: are the min. and max. code position values correct? Check with SHOW CP-LIMITS. See Chapter 6.5.4 Show CP-limits, Page 6-19. Bar code presented at incorrect reading • Check: reading distance of bar code in distance specified reading range? The bar code is tilted excessively with • Align bar code accordingly. respect to the scan line The reading angles at which the bar code • In the CLV Setup program: Start PERCENTAGE EVALUATION, select DISTANCE appears to the CLV are too large CONFIGURATION (if used). Monitor reading quality (reading quality >70 %!). Realign CLV if necessary. Bar code is in total reflection zone • Optimize CLV alignment (scan line should strike the bar code at an skew of approx. 15°). In the CLV Setup program: check reading quality. Start PERCENTAGE EVALUATION. See Chapter 6.5.2 Percentage evaluation, Page 6-17.
Table 8-5: Troubleshooting: result output errors in reading mode
8 009 981/O078/16-08-2004
© SICK AG · Division Auto Ident · Germany · All rights reserved
8-7
Page 86
Chapter 8
Troubleshooting
Operating Instructions CLV 42x bar code scanner
Malfunction
Possible cause
1.
• Evaluation criteria for bar code incorrect • In the CLV Setup program: Choose the CODE CONFIGURATION tab, SYMBOLOGIES group. Enable all code types (except Pharmacode), set code lengths to FREE. Download temporarily to CLV! Start PERCENTAGE EVALUATION. Code type and length are displayed in the reading result. Enable recorded code type in the SYMBOLOGIES group and disable all other types. Enter the code length under FIXED on the code type tab. Download to CLV! • Check: Are sufficient blank zones • Bar code quality not adequate present? Is the print contrast adequate? Are the specified print tolerances exceeded? In the CLV Setup program: Check by choosing READING CONFIGURATION tab, QUIET ZONE group. Is "Start/Stop" selected? Perform test read with a reference code in good condition.
The "Result" LED does not light up at the end of the reading pulse. (default setting: "Good read") – continued –
Remedy
2.
The CLV is not transferring any reading • Reading pulse is not functioning correctly • See Chapter 8.5.2 Malfunctions in results to the host. reading mode: reading pulse errors, Page 8-6. • Wiring fault in data connection • Check wiring as shown in Fig. 5-2, Page 5-5. • Voltage level incorrect • Measure voltages. • Host interface in CLV parameterized • In the CLV Setup program: incorrectly Check by choosing HOST INTERFACE tab, DATA FORMAT group. Are the correct interface type (hardware) and data format selected? Check settings in INTERFACE PROTOCOL group.
3.
The CLV transfers the status ST = 3 to The CLV has diagnosed a device error in the host in the reading result the self-test. (if ST is enabled for the separator; this is not selected by default).
4.
The data content of the bar code output by the CLV is incorrect or incomplete.
Switch the CLV off and on again. Does the "Device Ready" LED light up? If not, contact the SICK Service department.
• The data format on the host interface is • In the CLV Setup program: Select the parameterized incorrectly in the CLV HOST INTERFACE tab, DATA FORMAT group. Are the values correct? • The CLV suppresses the last character in • In the CLV Setup program: Select the the bar code. CODE CONFIGURATION tab. Click the SYMBOLOGY PARAMETERS button. Choose the tab for the relevant code. Is Transmit Check Digit activated? Change if necessary. Download to CLV!
Table 8-5: Troubleshooting: result output errors in reading mode (contd.)
8-8
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 009 981/O078/16-08-2004
Page 87
Operating Instructions
Troubleshooting
Chapter 8
CLV 42x bar code scanner
8.5.4 Prerequisite
Malfunctions in reading mode: errors when outputting the result status
The "Device Ready" LED lights up. If not, see Table 8-3, Page 8-5.
Malfunction
Possible cause
Remedy
1.
The "Result 1" (default setting: "Device Ready") and "Result 2" (default setting: "Good read") switching outputs are not outputting any pulses.
• The event of the assigned function for • not applicable outputting the result status does not occur during the reading process • Incorrect switching mode parameterized • In the CLV Setup program: Check by in CLV choosing the DEVICE CONFIGURATION tab. Click the RESULT OUTPUT PARAMETERS button. Are the outputs set to "Invert Output"? Change if necessary. Download to CLV!
2.
The beeper is not confirming the output of the reading result (default setting: "Good read").
Beeper deactivated
In the CLV Setup program: Check: Check by choosing DEVICE CONFIGURATION tab, group BEEPER VOLUME: is "Low" or "High" selected?
Table 8-6: Troubleshooting: errors in the result status output in reading mode
8.5.5 Prerequisite
The "Device Ready" LED lights up. If not, see Table 8-3, Page 8-5.
Malfunction 1.
Malfunctions: configuration errors (parameterization)
Possible cause
AutoSetup: • CLV not in "AutoSetup" mode The CLV does not read the presented, application-specific bar code (no confirmation from beeper)
•
• •
• 2.
Remedy
Profile programming: CLV does not read the presented profile bar code (no confirmation from beeper)
• Check: Does the "Device Ready" LED light up? If so, the CLV is in "reading" mode again. In CLV Setup program: Restart AUTOSETUP from the VIEW pull-down menu. If AutoSetup is started with profile bar • Switch the CLV off and on again. code no. 10: wait time of 5 s after powerPresent profile bar code no. 10 within up has been exceeded 5 s. Then present the application-specific bar code. Bar code presented at incorrect reading • Check: Is the reading distance of the bar distance code within the specified reading area? Bar code is in total reflection zone • Optimize bar code alignment (scan line should strike the bar code at an angle of approx. 15°). Bar code quality not adequate • See Malfunction 1, Table 8-5, Page 8-7
• Wait time of 5 s after power-up has been • exceeded • Wait time of 10 s after the first profile bar • code was presented has been exceeded • In reading mode: CLV not pulsed
Switch the CLV off and on again. Present profile bar code within 5 s. Once the first profile bar code has been read successfully, present the next bar code within 10 s, etc. • Clock the CLV accordingly and present the first profile bar code. Present the next bar code within 10 s, etc.
Table 8-7: Troubleshooting: configuration errors (parameterization)
8 009 981/O078/16-08-2004
© SICK AG · Division Auto Ident · Germany · All rights reserved
8-9
Page 88
Chapter 8
Troubleshooting
Operating Instructions CLV 42x bar code scanner
8.6
SICK Support
If a system error occurs repeatedly ("CLV SYS-Failure: xxx", xxx = 000 to 071) or if you cannot correct an error using the above measures, the CLV may be defective. The CLV contains no user-serviceable parts. Please contact our local SICK office or subsidary: •
The telephone numbers and email addresses are listed on the back page of this manual. For addresses see also www.sick.com. Do not send the device to the SICK service without first contacting us.
8-10
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 009 981/O078/16-08-2004