Page 118
Chapter 10
Appendix
Operating Instructions CLV 41x bar code scanner
6.
Start Windows and "CLV Setup" on your PC. See Chapter 10.3.3 Starting "CLV Setup", Page 10-8.
7.
Upload the current parameter set from the CLV. CLV Setup displays the current values on the tabs.
8.
If you have not already done so, save the parameter set as a configuration file "*.scl" in CLV Setup.
a) Dynamic teach-in procedure Configuring the dynamic teach-in procedure: •
Accompanying measure: change the reading pulse for teach-in
1.
Select the DEVICE CONFIGURATION tab.
2.
Click the DYNAMIC option in the TEACH-IN group.
3.
If necessary proceed as follows to activate the output filter in the CODE COMPARISON group: – Click the CODE #1 ACTIVE checkbox. – Click the FILTER FOR OUTPUT checkbox. – Click the CODE #1 ACTIVE checkbox again. The code comparison is now no longer permanently activated via the user interface. The output filter, however, remains active.
4.
Click the READING TRIGGER PARAMETERS button.
5.
Click the SERIAL INTERFACE radio button in the READING TRIGGER SOURCE group.
6.
Comfirm the dialog box with "OK".
7.
Click the RESULT OUTPUT PARMETERS button.
8.
In the dialog box under RESULT FUNCTIONS, choose the MATCH 1 option for e.g. Result 2 in the drop-down list.
9.
Select the CODE CONFIGURATION tab.
10. Activate the code type of the match code in the SYMBOLOGIES group. 11. Click the EDIT button. 12. Choose the code type tab and click the FREE radio button in the CODE LENGTH group. 13. Close the dialog box. 14. Perform a download to the CLV. This is done by clicking The DOWNLOAD PARAMETER dialog box is displayed.
in the toolbar.
15. Confirm the dialog box by selecting the PERMANENT save option. Carrying out the dynamic teach-in procedure: 1.
Click in the toolbar. The terminal emulator is launched. The CLV is in Reading mode.
2.
Close the switch to connect the power supply to the "Sensor 2" switching input.
3.
Click PARAMETERIZE under DEVICE MODE. The CLV switches to PARAMETER MODE.
4.
Click READING MODE under DEVICE MODE. The CLV returns to Reading mode and polls the condition of the "Sensor 2" switching input. It also activates code comparison at this time.
– or –
10-28
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 009 974/O079/01-09-2004
Page 119
Operating Instructions
Appendix
Chapter 10
CLV 41x bar code scanner
As an alternative to steps 3 and 4, switch the power supply to the connection module (CLV) off and on again. The CLV outputs two consecutive tones to indicate that it has switched to Reading mode. 5.
Open the switch to interrupt the power supply to the "Sensor 2" switching input. The CLV activates teach-in mode for match code 1.
6.
Click the SW-TRIGGER ON button or press [F7] key. The CLV starts the reading interval and waits for the match code.
7.
Present the match code.
8.
Click the SW-TRIGGER OFF button or press [F8] key. The CLV confirms that the read was successful by emitting a tone and then ends the reading interval. Match code 1 is now stored in the RAM.
9.
Close the switch to reconnect the power supply to the "Sensor 2" switching input. The CLV saves match code 1 (code type, code length, and data content) permanently in the parameter set stored in the EEPROM. It compares every bar code it reads with match code 1 until the power supply is switched off. In doing so, it outputs the result of the comparison via the "Result 2" output with the predefined pulse length. The "Result“ LED only lights up (until the end of the next reading pulse) to indicate a successful code comparison.
10. Keep the switch closed. The CLV will activate teach-in mode again if you open the switch! Replacing the match code: To replace the current match code 1 with a new match code using the teach-in method, simply repeat steps 5 to 9. Checking the programmed match code in the CLV: 1.
Switch the power supply of the connection module (of the CLV) off and on again.
2.
Perform an upload from the CLV. This is done by clicking
3.
The code type and data content of match code 1 are shown in the CODE COMPARISON group on the DEVICE CONFIGURATION tab.
in the toolbar.
Resetting the reading pulse: 1.
Select the DEVICE CONFIGURATION tab.
2.
Click the READING TRIGGER PARAMETERS button.
3.
Click the current option in the READING TRIGGER SOURCE group in the dialog box.
4.
Perform a download to the CLV. This is done by clicking
5.
Confirm the dialog box by selecting the PERMANENT save option. The CLV then uses the reading pulse type that was set before the teach-in procedure.
in the toolbar.
b) Static teach-in procedure Configuring the static teach-in procedure:
8 009 974/O079/01-09-2004
1.
Select the DEVICE CONFIGURATION tab.
2.
Click the STATIC WITHOUT PHARMACODE, or STATIC WITH PHARMACODE to read a Pharmacode, in the TEACH-IN group.
3.
If necessary, click the WITH COUNTER RESET option in the COUNTER RESET group.
4.
If necessary proceed as follows to activate the output filter in the CODE COMPARISON group:
© SICK AG · Division Auto Ident · Germany · All rights reserved
10-29
Page 120
Appendix
Chapter 10
Operating Instructions CLV 41x bar code scanner
– Click the CODE #1 ACTIVE checkbox. – Click the FILTER FOR OUTPUT checkbox. – Click the CODE #1 ACTIVE checkbox again. The code comparison is now no longer permanently activated via the user interface. The output filter, however, remains active.
Note
5.
Click the RESULT OUTPUT PARAMETERS button.
6.
In the dialog box under RESULT FUNCTIONS, choose MATCH 1 in the drop-down list for e.g. Result 2.
7.
Perform a download to the CLV. This is done by clicking
8.
Confirm the dialog box by selecting the PERMANENT save option.
in the toolbar.
Before you can use the static teach-in method, the parameter sets in the RAM and EEPROM of the CLV must be identical. This is done with the PERMANENT saving option. If you select the TEMPORARY option instead, the CLV loads the default settings after the teach-in function has been carried out and the power supply switched off. Carrying out the static teach-in procedure: 1.
Click in the toolbar. The terminal emulator is launched. The CLV is in Reading mode.
2.
Close the switch to connect the power supply to the "Sensor 2" switching input.
3.
Click PARAMETERIZE under DEVICE MODE. The CLV switches to Parameter mode.
4.
Click READING MODE under DEVICE MODE. The CLV returns to Reading mode and polls the condition of the "Sensor 2" switching input. It also activates code comparison at this time.
– or – As an alternative to steps 3 and 4, switch the power supply to the connection module (of the CLV) off and on again. The CLV outputs two consecutive tones to indicate that it has switched to Reading mode.
10-30
5.
Open the switch to interrupt the power supply to the "Sensor 2" switching input. The CLV activates teach-in mode for match code 1. It starts the reading interval automatically and waits for the match code.
6.
Present the match code. The CLV confirms that the read was successful by outputting a tone and then ends the reading interval. Match code 1 is now stored in the RAM.
7.
Close the switch to reconnect the power supply to the "Sensor 2" switching input. The CLV saves match code 1 (code type, code length and data content) permanently in the parameter set stored in the EEPROM. It compares every bar code it reads with match code 1 until the power supply is switched off. In doing so, it outputs the result of the comparison via the "Result 2" output with the predefined pulse length. The "Result“ LED only lights up (until the end of the next reading pulse) to indicate a successful code comparison. If the CLV was not able to read the match code, the reading interval is terminated when the switch is closed. The previously set code configuration remains intact.
8.
Keep the switch closed. The CLV will activate teach-in mode again if you open the switch!
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 009 974/O079/01-09-2004
Page 121
Operating Instructions
Appendix
Chapter 10
CLV 41x bar code scanner
Replacing the match code: To replace the current match code 1 with a new match code using the teach-in method, simply repeat steps 5 to 7. Checking the programmed match code in the CLV: Proceed as described in steps 1 to 3 in "a) Dynamic teach-in procedure", Page 10-28. Activating/deactivating the code comparison via the "Sensor 2" switching input Activating the code comparison function: 1.
Close the switch before switching on the power supply to the CLV.
2.
Switch on the power supply. The CLV activates the code comparison function.
3.
Do not open the switch while the CLV is in operation.
Deactivating the code comparison function: 1.
Switch off the power supply to the CLV.
2.
Open the switch.
3.
Switch on the power supply. The CLV starts Reading mode without the code comparison.
Assigning output functions to the "Result 1" to "Result 3" switching outputs and beeper: CLV can output the result of the code comparison in the following functions: –
No match (code does not match either match code)
–
Match 1 (matches match code 1)
–
Mismatch 1 (does not match match code 1)
–
Match 2 (matches match code 2)*)
–
Mismatch 2 (does not match match code 2)*)
–
Match 1 or 2 (matches match code 1 or 2)
–
Match 1 and 2 (matches match code 1 and 2)*)
*)
only relevant if match code 2 is active
When configuring the teach-in function for match code 1, define the following settings:
8 009 974/O079/01-09-2004
1.
Select the DEVICE CONFIGURATION tab.
2.
Click the RESULT OUTPUT PARAMETERS button.
3.
Choose the required option for Result 1 to Result 3 and beeper from the RESULT FUNCTIONS drop-down list.
4.
Perform a download to the CLV. This is done by clicking in the toolbar. Confirm the dialog box by choosing the PERMANENT storage option.
© SICK AG · Division Auto Ident · Germany · All rights reserved
10-31
Page 122
Chapter 10
Appendix
Operating Instructions CLV 41x bar code scanner
10.8.2
Structure of a daisy chain configuration (data forwarding or master/slave arrangement)
See Technical Information "Master/Slave with Bar Code Scanners from the CLV Series" (order no. 8 007 675, English edition) 10.8.3
Structure of a SICK network (RS 485)
See "CLX 200 Network Controller" Technical Description (order no. 8 007 157, German/English edition) 10.8.4
Connection to Profibus DP
See operating instructions "BMV/BMH 10 for Profibus DP" (order no. 8 008 825, English version) 10.8.5
Connection to DeviceNet
See operating instructions "BMV/BMH 10 for DeviceNet" (order no. 8 008 972, English version) 10.8.6
Connection to Interbus S
See technical information "BMS 10/20 for Interbus S" (order no. 8 007 546, English version) 10.8.7
Connection to Ethernet
See operating instructions "BMV/BMH 10 for Ethernet" (order no. 8 009 398, English version)
10-32
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 009 974/O079/01-09-2004
Page 123
Operating Instructions
Appendix
Chapter 10
CLV 41x bar code scanner
10.9
Replacing a CLV (copying the parameter set)
There are two methods of replacing a CLV and transferring the current parameter set to the new device: • by using the profile bar codes
• by downloading the parameters from CLV Setup
The parameter set can only be transferred between identical device types (e.g. from one CLV 410 to another CLV 410, etc.). 10.9.1
Transferring the parameter set using the profile bar codes
To transfer the parameter set using the profile bar codes, you do not need to connect the device to a PC. The parameter set of the CLV that is to be replaced must be available in the form of archived profile bar codes or at least as a configuration file in CLV Setup. Otherwise, the replacement device must be configured from scratch. The procedure for printing out the profile bar codes is described in Chapter 10.4.2 Profile programming, Page 10-18. 1.
Switch off the power supply of the CDB 410 or CDM 410 connection module (of the CLV).
2.
Disconnect the CLV connector from the CDB 410 or CDM 410 connection module.
3.
Remove the CLV from the bracket. Before doing so, mark the position and alignment on the bracket/station.
4.
Install, align, and connect the replacement device.
5.
Switch on the power supply to the CDB 410 or CDM 410. The "Device Ready" LED of the CLV lights up. The CLV outputs a tone by the beeper to indicate that it is ready to read the profile bar codes.
6.
Present the first profile bar code within 5 seconds. The CLV confirms that the bar code has been read successfully by outputting a tone via the beeper and waits 10 seconds for the next profile bar code.
7.
Present all of the other profile bar codes consecutively in the printed sequence.
8.
Then allow the final wait time (10 seconds) to elapse. The CLV outputs two consecutive tones to indicate that it has switched to Reading mode.
The new CLV is now ready to use the transferred parameter set. Troubleshooting If the software version on the replacement (new) device is older than the version on the old device (e.g. new device is from warehouse stock), the profile bar codes may contain coded parameters or parameter values that cannot be interpreted by the older software of the replacement device. The CLV, however, reads all of the profile bar codes. But this does not indicate that the CLV can use all the parameters or parameter values. ¾
8 009 974/O079/01-09-2004
Use the CLV Setup program to check whether the new CLV is functioning correctly. To do so, connect the PC to the host interface 2 (RS 232) on the CLV. Temporarily disconnect any existing electrical data connection to the host! Check whether all of the parameter settings used are necessary to configure the CLV for the current reading application. If necessary, replace the software version in the new device.
© SICK AG · Division Auto Ident · Germany · All rights reserved
10-33
Page 124
Chapter 10
Appendix
Operating Instructions CLV 41x bar code scanner
If the software version on the new device is more recent than that on the old one, the default settings in the imported parameter set will include parameters that were not contained in the old version. ¾
Upload the parameter set with the CLV Setup program and check whether the new parameters are relevant for configuring the CLV for the current reading application. If necessary, change the parameter values and download the parameter set to the CLV again. Save the new parameter set as a configuration file in CLV Setup and print it out as profile bar codes for archiving purposes.
See also Chapter 10.3.10 Dealing with unknown parameters, Page 10-14. 10.9.2
Downloading the parameter set
To download the parameter set, you must connect a PC running the CLV Setup software to the replacement device. The parameter set of the CLV that is to be replaced must be available as a configuration file in CLV Setup, otherwise the device must be configured from scratch. 1.
Switch off the power supply of the CDB 410 or CDM 410 connection module (of the CLV).
2.
Disconnect the CLV connector from the CDB 410 or CDM 410.
3.
Remove the CLV from the bracket. Before doing so, mark the position and alignment on the bracket/station.
4.
Install, align, and connect the replacement device.
5.
Connect the PC to the host interface 2 (RS 232) of the CLV using a 3-core RS 232 data cable (null modem cable). (CDB 410 or CDM 410: connect the PC to the internal, 9-pin "Aux" plug). (See Chapter 5.5.5 Connecting the PC to the host interface 2 (RS 232), Page 5-7.) Temporarily disconnect any existing electrical data connection to the host!
6.
Switch on the power supply to the CDB 410 or CDM 410. After the CLV has started, it emits a tone to indicate that the self-test was successful. After a short while, it outputs two consecutive tones to indicate that it has switched to Reading mode. The "Device Ready" LED lights up.
7.
Switch on your PC and start Windows.
8.
Start "CLV Setup". If the communication parameters on the PC and CLV are identical, CLV Setup uploads the parameter set from the CLV and displays the values on the tabs. If CLV Setup can not establish communication with the CLV, carry out the procedure described in Chapter 10.3.3 Starting "CLV Setup", Page 10-8, Troubleshooting.
9.
Open the archived parameter set as a configuration file "*.scl" in CLV Setup and download it to the CLV.
10. Confirm the dialog box by choosing the TEMPORARY storage option. 11. Check whether the CLV reads the bar codes correctly. Correct the parameter settings if necessary. 12. Finally, download the tested parameter set to the CLV and save it with the PERMANENT storage option. The parameter set is transferred permanently to the replacement device. 13. Save the modified parameter set as a new configuration file in CLV Setup.
10-34
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 009 974/O079/01-09-2004
Page 125
Appendix
Operating Instructions
Chapter 10
CLV 41x bar code scanner
Troubleshooting If the software version on the replacement (new) device is older than the version on the old device (e.g. new device is from warehouse stock), the parameter set transferred may contain coded parameters or parameter values that cannot be interpreted by the older software. The new device, however, outputs an error message in CLV Setup for each of these parameters/parameter values when the parameter set is downloaded. ¾
Check whether all of the parameter settings used are necessary to configure the CLV for the current reading application. If necessary, replace the software version in the new device.
If the software version on the new device is more recent than that on the old one, the default settings in the imported parameter set will include parameters that were not contained in the old version. ¾
Upload the parameter set from CLV to CLV Setup and check whether the new parameters are relevant for configuring the CLV for the current reading application. If necessary, change the parameter values and download the parameter set to the CLV again. Save the new parameter set as a configuration file in CLV Setup.
See also Chapter 10.3.10 Dealing with unknown parameters, Page 10-14. Tip
8 009 974/O079/01-09-2004
If a parameter set is duplicated on other devices with the same application conditions, you may have to change the device number of the individual CLVs subsequently using the DEVICE CONFIGURATION tab for this parameter set so that it can be evaluated in the host and then downloaded the parameter set again.
© SICK AG · Division Auto Ident · Germany · All rights reserved
10-35
Page 126
Appendix
Chapter 10
Operating Instructions CLV 41x bar code scanner
10.10 Accessories 10.10.1 Installation accessories Order no.
Description
Fig.
2 020 077
Small mounting bracket with installation accessories
Fig. 10-13, Page 10-38
2 020 078
Large mounting bracket with installation accessories
Fig. 10-14, Page 10-38
Table 10-5: Accessories: installation accessories
10.10.2 Connection modules Part No.
Type
Description
1 023 813
CDB 410-001
Connection module for one CLV 41x, with 15-pin D-Sub HD device socket, terminal strips (signal distributors) for wiring the data and functional interfaces, internal 9-pin D Sub “Aux“ plug, cable grips (dia. 4.5 to 10 mm (0.18 to 0.39 in)), polycarbonate housing, enclosure rating max. IP 65, operating temperature 0 to +40 °C (32 to 104 °F), operating voltage 6 to 30 V DC.
1 025 361
CDM 410-0001
Connection module for one CLV 41x, with 15-pin D-Sub HD device socket, terminal strips (signal distributors) for wiring the data and functional interfaces, internal 9-pin D Sub “Aux“ plug, cable grips (dia. 4.5 to 10 mm (0.18 to 0.39 in)), polycarbonate housing, enclosure rating max. IP 65, operating temperature –351) to +40 °C (–311) to 104 °F), operating voltage 6 to 30 V DC, power consumption 0.5 W without CLV 41x.
2 029 468
CMP 400
Power Supply Module (Connection Module Power) as plug-in module for the CDM 410 Connection Module. Power supply for CLV 41x from an AC power line. Input 100 to 250 V AC/50 to 60 Hz, output 24 V DC, max. 10.8 W (short-circuit proof). Connection to the CDM 410 via ribbon cable. Operating temperature 0 to +40 °C (32 to 104 °F).
1) Without any mounting or electrical installation work, otherwise –20 °C (–4 °F)
Table 10-6: Accessories: Connection modules
10.10.3 Bus connection modules Order no.
Type
Description
6 020 896
BMV 10-0111
Bus connection module for connection of max. two CLVs to Profibus DP; for the CLV 41x each with a 15-pin D Sub HD device socket. With a 9-pin D Sub bus socket, terminal strips (signal distributors) for wiring the RS 232 data interface and the functional interfaces, two internal 9-pin D Sub "Service" plugs, cable grips, die-cast aluminum housing, enclosure rating max. IP 54 (enclosure rating max. IP 65 with a plug hood), operating voltage 20 to 30 V DC
6 020 893
BMH 10-0111
Bus module for connection of a CLV to Profibus DP; with terminal screws, a 9-pin D-sub bus socket, no housing, for top-hat-rail installation, enclosure rating max. IP 20, operating voltage 10 to 30 V DC
6 021 190
BMV 10-0311
Bus connection module for connection of max. two CLVs to DeviceNet, terminal strips for bus connection, otherwise, as BMV 10-0111
6 021 188
BMH 10-0311
Bus module for connection of one CLV to DeviceNet, otherwise as BMH 10-0111
1 012 672
BMS 10-0113
Bus connection module for the connection of one CLV to Interbus-S; for the CLV 41x with a 15-pin D Sub HD device socket, two 9-pin D Sub bus connections (in/out), terminal strips (signal distributors) for wiring the data and functional interfaces, internal 9-pin D Sub "Service" plug, integrated power supply 230 V AC 50 Hz/ 24 V DC, cable grips, polycarbonate housing, enclosure rating max. IP 54
1 012 680
BMS 10-0112 as BMS 10-0113, but with integrated 115 V AC 50 to 60 Hz/24 V DC power supply
6 025 757
BMV 10-0411
Bus connection module for the connection of max. two CLV 41x to Ethernet TCP/IP. Data transmission rate 10 MBit/s. For each CLV 41x with a 15-pin D Sub HD device socket. With a 8-pin RJ-45 bus socket, terminal strips (signal distributors) for wiring the RS 232 data interface and the functional interfaces, two internal 9-pin D Sub "Service" plugs, cable grips, die-cast aluminum housing, enclosure rating max. IP 54 (enclosure rating max. IP 65 with a plug hood), operating voltage 20 to 30 V DC
Table 10-7: Accessories: bus connection modules
10-36
© SICK AG · Division Auto Ident · Germany · All rights reserved
8 009 974/O079/01-09-2004