CiMAX 78xx Scanner Technical Reference Manual: CHAPTER 5 Startup & Operation
Pages 83 to 96 of 201, from CiMAX 78xx Scanner Technical Reference Manual
The original document is the PDF (2.8 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the RVSI CiMatrix 7800, Sick 7850 and Siemens 7800
Page 83
5 Startup & Operation
CHAPTER 5
This chapter describes how to power-up, how to use the Setup port and a terminal to monitor operation, how to enter the Setup and Diagnostic modes and how to change parameters.
Serial Communications Ports The CiMAX 78xx Scanner has a Terminal port that can also be used as the Setup port. Refer to “Communications Inputs & Outputs” on page 3-9 for connector and pin assignment information.
It is necessary, in some master-slave configurations, to take extra steps to tell the scanner that a setup terminal is connected to the Terminal port, in order to enter the Setup/Diagnostics mode. These extra steps are detailed in parameter 002 OPMODE in the Scanner Parameters Reference Guide for more information and in Chapter 8, “Diagnostics”. The Host and Terminal ports can also be used to send commands directly to the scanner's operating system. This direct transmission bypasses the menu-driven user interface associated with the Setup port. Direct transmission is useful for remote setup from a host computer. The commands are described in Chapter 6, “Host Commands & Downloads”. The Starnode port allows the scanner to be connected to an EIA-485 Starnode Local Area Network. This port may be used with the CiMAX 1400 Hand-Held terminal. The CiMAX 78xx Scanner also has an optional Ethernet connector and port. Most functions that can be performed through the serial ports can also be performed through the Ethernet and Starnode ports.
Power Up When you power up or reset the scanner, the Setup LED will flash for a short time. During this time, the scanner will send several status messages to the Setup port.
Status Messages If you have a terminal connected to the port, you will be able to read the messages. The first message displayed is WAITING FOR STABLE MOTOR SPEED... Rev H, February 2002
CiMAX 78xx Technical Reference Manual
5-1
5 Startup & Operation
The port designated for Setup can be used to monitor operation, to change parameters and to perform diagnostic tests. Any RS-232 ASCII terminal, a PC running terminal software or a CiMAX 1400 Hand-Held terminal, can be connected to the Setup port. When connected, the terminal becomes the display and keypad for the scanner.
Page 84
Chapter
5
Startup & Operation
From a stop, the motor may take as much as ten seconds to come to its set speed and stabilize. If the scanner is reset when the motor is already up to speed, motor stabilization may take as little as three seconds. The laser control software prevents the laser from turning on until the motor speed stabilizes above a minimum speed of 100 s/s (scans/second), and will turn the laser off if the motor speed drops below that minimum speed. The motor speed must then re-stabilize above the minimum speed before the laser will turn on again. The laser will not turn on until the motor speed has stabilized. If the motor does not stabilize within one minute, the software will turn on without turning on the laser. During the one minute waiting period, the software will display the WAITING FOR STABLE MOTOR SPEED... message. Table 5–1 shows the minimum acceptable scan (motor) speeds associated with each of the DSP clock rates described in 009 CLOCK RATES in the Scanner Parameters Reference Guide. TABLE 5–1. Minimum Acceptable Scan Speeds
DSP Clock Rate, MHz
Motor Speed, scans/second * 75A, 76, 78
40
610
32
488
25
380
20
305
* Motor speed settings below these thresholds will result in the DSP bar timing being reduced.
You will then see a startup message similar to the following: CIDSP Download Complete No network card detected PREBOOT VX.X FLASH ROM BOOT MONITOR VX.X CIDSP-Non-Stitching Auto-distinguish (5 symbologies) for Scanner: Ver X.X TPU Functions Ver X.X CiMAX 7800 Ver. X.X ***Warning: Checksum 0, NV RAM Defaults Restored*** (C) 1994-2000 CiMATRIX CANTON, MA. ALL RIGHTS RESERVED 256K RAM Main uP:25MHz,DSP:40MHz (bar timing:13MHz) NEW ROM -- RAM CLEARED INITIALIZING FILE SYSTEM 32 Kb File System 32 Kb available to User Program GAP TRACKING DISABLED NEED TERMINAL # The lines in the message have the following meanings: •
5-2
The first line indicates booting is complete.
CiMAX 78xx Technical Reference Manual
Rev H, February 2002
Page 85
Serial Communications Ports
The network card detection line indicates that the hardware has detected a specific network card installed. This will be displayed by the following responses: –
No network card detected
–
Ethernet card detected
–
DeviceNet/SDS card detected
•
The PREBOOT and BOOT lines contain the version number of the FLASH download program. You can determine the software versions installed without rebooting. Refer to Chapter 6, “Host Commands & Downloads” for more information.
•
The CIDSP line provides the DSP barcode decoding description and gives the version of the digital signal processor (DSP) code currently being used.
•
The TPU line gives the version of the internal processor code currently being used.
•
The CiMAX 7800 line gives the version number of the operating system (OS) software.
•
The copyright line will only be displayed during the first power-up/re-boot after a memory clear command or, if the RAM has failed, as described in “Power-up With Customized Values” on page 5-5 and in Chapter 6, “Host Commands & Downloads”.
•
The RAM line indicates how much RAM memory is installed.
•
The Main uP line indicates the processor speeds, determined by the setting of parameter 009 CLOCK RATES. The lines NEW ROM -- RAM CLEARED and INITIALIZING FILE SYSTEM will occur if this is the first reboot since downloading a new FLASH OS, if the structure is corrupt or if you have changed parameter 601 FILESYS. If you have changed the file system size by changing parameter 601 FILESYS, the following warning message will be displayed and a beep, followed by a delay before re-initializing begins: FILESYS ABOUT TO RE-INITIALIZE When the file system is re-initialized, all previous data will be lost. This is equivalent to reformatting a diskette. Refer to the Scanner Parameters Reference Guide for more information about the file system and parameter 601 FILESYS.
•
The sign-on message lines, which follow the INITIALIZE message lines, will appear in the format: 32Kb File System 32Kb available to User Program
Parameter 601 FILESYS sets the file system size in Kb. The default is the minimum size, 32Kb. The largest size is determined by the amount of RAM installed, in either 256Kb or 1Mb blocks. •
Rev H, February 2002
The GAP TRACKING line indicates the setting of parameter 700 TRACKING. The default is DISABLED.
CiMAX 78xx Technical Reference Manual
5-3
5 Startup & Operation
•
Page 86
Chapter
5
Startup & Operation
•
The line NEED TERMINAL # will be displayed if a Starnode network terminal ID number (parameter 401 STARNODE TERMINAL NUMBER) has not been entered. If you are not using a Starnode network (LAN), you may ignore this message. If you are using a Starnode network, you must assign an ID number before the scanner can connect to the network. If an ID number has been assigned, and the network connection is made, the following message will be displayed: WAITING FOR STARNODE... STARNODE ACTIVE If the network connection is not made, the following message will be displayed: WAITING FOR STARNODE... NO STARNODE
Note: There is no corresponding message sent from the Setup port to indicate whether or not an Ethernet server has been connected. Successful connection can only be determined by the host program at the remote computer.
Power-up With Factory Default Values If this is the first power up or reset following a memory clear command, or if the battery-backed RAM is not holding the memory, you will see an additional copyright message: (C) 1994-2000 CiMATRIX CANTON, MA. ALL RIGHTS RESERVED In addition, you will hear three ascending tones, followed by three descending tones, from the beeper (if you have connected a beeper to the I/O CONNECTOR or are using a SICK Auto Ident, Inc. Standard Interface Unit). The three descending tones are an indication that default values have been loaded.
Power-up After Memory Failure If you see the copyright message, it may mean that the battery-backed RAM has failed. To check the battery, restart the scanner by turning the power off, waiting a few seconds and then turning the power on again. If the scanner is still sends this copyright message, contact SICK Auto Ident, Inc. for information regarding service.
Memory Error Detection, Response & Reporting If the checksum for the parameter area of RAM is found to be invalid after a power-down/power-up cycle, the parameters are checked for validity against their maximum and minimum values. The Decode Format Strings are also checked for valid settings. If the checksum is incorrect but no parameter is found to be outside of its allowed range, the parameters will NOT be restored to their default values and the following message will be displayed among the sign-on messages at power-up: ***Warning: Checksum Bad, Params Within Range*** If any parameter is found to be outside of its allowed range, all parameters will be reset to their default values, and the following message will be displayed among the sign-on messages at powerup:
5-4
CiMAX 78xx Technical Reference Manual
Rev H, February 2002
Page 87
Serial Communications Ports
***Warning: Checksum Bad, NV Ram Defaults Restored*** If the memory has never been initialized, or if the default parameter values have been restored, the following message will be displayed among the sign-on messages at power-up: ***Warning: Checksum 0, NV Ram Defaults Restored*** Information about checksum errors can be reviewed from the host computer by invoking the ~DATA~ command. Refer to Chapter 6, “Host Commands & Downloads” for more information on this command.
Power-up With Customized Values If customized parameter values have been entered, or during subsequent power-ups after loading defaults, the copyright message and descending tones will not be sent. Instead, you will hear three ascending tones and the scanner will begin to operate normally.
Power-up After Automatic Reboot The processor frequently checks the operation of the DSPs used for barcode decoding and automatically reboots the scanner if the DSP is found not to be decoding (sleeping). If this happens, the following message is sent to the Setup port:
User Interface - Parameters & Diagnostics The CiMAX 78xx user interface is accessed using a terminal connected to the Setup port. This allows you to change parameters and run diagnostics and calibration procedures. Refer to Chapter 8, “Diagnostics” for more information. You can also change parameters without going through the user interface by using Host Commands. These commands can be sent over any of the communication ports, Host, Terminal or Starnode. Refer to Chapter 6, “Host Commands & Downloads” for more information.
Operating Modes The scanner operates in three modes: •
Normal
•
Setup
•
Diagnostic
In the Normal mode, you can monitor information about the scanner's operation, including the barcode read, input and output status, and barcode quality. In the Setup mode, you can set parameters to program the scanner for differing applications, including the type of terminal you are using, the barcode symbology on your labels, presence configuration, use of output relays, height of boxes and serial communications. In the Diagnostic mode, you can run tests of the hardware and software to check the scanner's performance.
Rev H, February 2002
CiMAX 78xx Technical Reference Manual
5-5
Startup & Operation
5
Reboot: DSP 0 found sleeping
Page 88
Chapter
5
Startup & Operation
Normal Mode In the Normal mode, you can program the scanner to send information such as the barcode read, the input and output status, and barcode quality, to the Host computer using a Normal mode report format. For example: normmd_rpt =1 01234567890123456789 line 1 <--- label data ---> label data line 2 (blank) line 3 ST:01234567890123456 status line 4 IO:87654321-87654321 inputs-outputs Refer to parameter 012 NORMMD RPT in the Scanner Parameters Reference Guide for a description of the Normal Mode Report display.
Setup & Diagnostic Modes Special Host commands, ?S and ?D, are used to place the scanner into the Setup or Diagnostic mode. These commands are only recognized when received at the Setup port. Refer to Chapter 6, “Host Commands & Downloads” for more information. A beginning and ending message character must frame the command. If parameter 403 MSGCHAR has not been changed from its default value of ~, type the commands: ~?S~ (tilde, question mark, S, tilde) to enter the Setup mode. ~?D~ (tilde, question mark, D, tilde) to enter the Diagnostic mode. You can also enter the Setup and Diagnostic modes over the Starnode network. The Starndem program can be used in a rudimentary fashion by manually sending the appropriate messages. To enter Setup or Diagnostics from Starndem, or any program controlling the Starnode network, send a ?S or a ?D message to the terminal. It will respond with the first line in the menu. Use a command of none and begin each message with /M. To exit the menu, send /ME, with the E indicating Exit. You have exited the Setup/Diagnostic mode on the Starnode network when the message EXIT SETUP or EXIT DIAGNOSTICS is sent from the Starnode port when E is received.
Resetting to Default Parameters You can return to SICK Auto Ident, Inc. default parameter values at any time by entering the ?R command through any communication port. If sent through a Host or Terminal port, the command must be sent as ~?R~ (assuming MSGCHAR is a ~ ), or ?R over a Starnode network, to restore default parameters. WARNING! YOUR SCANNER MAY USE NON-DEFAULT PARAMETERS CUSTOMIZED FOR YOUR APPLICATION. THESE PARAMETERS ARE DETAILED IN YOUR KEYSHEET. RESETTING THE SCANNER TO USE DEFAULT PARAMETERS WILL CAUSE YOUR CUSTOMIZED PARAMETERS TO BE LOST AND WILL NEED TO BE RESTORED. Refer to “Restore Keysheet” on page 8-19 for more information on restoring Keysheet parameters. 5-6
CiMAX 78xx Technical Reference Manual
Rev H, February 2002
Page 89
Serial Communications Ports
Terminal Types There are two terminal types supported: •
Non-intelligent ASCII
•
CiMAX 1400 Hand-Held
A PC running terminal emulation software can be used instead of the ASCII terminal. The scanner can be configured to use one or the other by setting parameter 000 TERM DEVICE to the appropriate value. The default value selects the ASCII terminal. A simple keystroke sequence can be used to switch between configurations. This is useful if you accidentally configure the scanner for use with an ASCII terminal and then connect the hand-held terminal. The hand-held terminal will display unusable information. To change the configuration when you cannot see the menus, type the following sequence: <SHIFT> <SETUP> <ENTER> <SPACE> <ENTER> <EXIT>
Non-Intelligent ASCII Terminal
Setup Menu on the ASCII Terminal The Setup menu displays parameter number, parameter name (in abbreviated form) and current value. All parameter abbreviations are listed in the Scanner Parameters Reference Guide. The parameter display format is: III NNNNNN VVVVVV where: III, The three-digit parameter number. NNNNNN - The abbreviated parameter name. VVVVVV - The value currently stored in memory. When you first enter the Setup menu, the following line will be displayed: SETUP - (P)REVIOUS, (G)OTO, (T)OGGLE, (E)XIT This line lists all but one of the scrolling options that allow you to view parameters or leave the Setup menu. The scrolling options function as follows: Enter - Display the NEXT parameter. To view the parameter after the parameter displayed on the last line of the screen, if one is currently shown (or parameter 000 TERM DEVICE if none are shown), press Enter. The next parameter will appear on the next line. If the screen is full, it will scroll up and delete the top line to make room for the new line. Rev H, February 2002
CiMAX 78xx Technical Reference Manual
5-7
5 Startup & Operation
Upon receiving a ~?S~ or ~?D~ command, the scanner will enter the Setup or Diagnostic mode and display a corresponding menu on the ASCII terminal. The Setup menu offers a quick, easy way of viewing and, if necessary, changing parameter values. The Diagnostic menu offers a simple way to run diagnostic tests.
Page 90
Chapter
5
Startup & Operation
For example, when Enter is pressed seven times after the Setup mode is entered, the display will appear as: SETUP - (P)REVIOUS, (G)OTO, (T)OGGLE, (E)XIT 000 TERM DEVICE 2 002 OPMODE 0 003 IOMODE 000 004 MINLEN 04 005 MAXLEN 32 006 THRHLD 01 007 NLLSCN 04 Continuing to press Enter after the screen is filled will cause succeeding parameters to be displayed at the bottom of the list, with preceding parameters deleted from the top. P - Display the PREVIOUS parameter. To view the parameter previous to the parameter displayed on the last line of the screen, if one is currently shown, press P followed by Enter. The previous parameter will appear on the next line. If the screen is full, it will scroll up and delete the top line to make room for the new line. G - GOTO a specific parameter. To GOTO a specific parameter, press G followed by the threedigit parameter number followed by Enter. The parameter and its current value will be displayed on the next line of the display. If the display is full, it will scroll up to make room for the new line. When both a parameter number and a valid value are entered, that parameter's old value will be replaced by the newly entered value. For example, entering G00605 will display parameter 006 THRHLD (VALIDATION THRESHOLD) and enter a value of 05. The change will only become effective when you EXIT the menu. T - TOGGLE between Diagnostic and Setup menus, without changing parameter position. That is, if you TOGGLE to the Diagnostic menu and then TOGGLE back to the Setup menu, you will return to the same parameter that you were viewing before you toggled. To toggle, press T followed by Enter. E - EXIT the Setup/Diagnostic menu and resume operation. All changes made to parameter values prior to the EXIT command will become effective upon exiting. To EXIT, press E followed by Enter.
Data Entry through Setup Menu on the ASCII Terminal New values can be entered as follows. 1.
Move through the Setup menu to display the desired parameter on the last line of the list.
2.
Type in a new value, being careful to enter the required number of digits for the parameter. Press Enter to enter the value. If a valid value (correct number of digits and value within the range for the parameter) was entered, the beeper will sound once and the terminal will show the parameter again, with the new value replacing the old one. If an invalid value was entered, three short beeps will be sounded. The previous value for the parameter will remain displayed and unchanged.
3.
5-8
Exit the Setup menu to accept the changed value or move to the next parameter to be changed.
CiMAX 78xx Technical Reference Manual
Rev H, February 2002
Page 91
Serial Communications Ports
Diagnostics Menu Display on the ASCII Terminal The Diagnostics menu lists all of the diagnostic tests that are available. Operation is similar to Setup menu operation. When you enter the menu from the Normal mode, the following will be displayed: DIAGNOSTICS - (P)REVIOUS, (R)UN, (T)OGGLE, (E)XIT If you toggle from the Setup menu to the Diagnostic menu, you will see the above line plus all of the Diagnostic menu lines you have previously displayed. While in the Diagnostics menu, your choices are: Enter - Display NEXT test on the last line. P - Display the PREVIOUS test on the last line. R - RUN the diagnostic test currently displayed on the last line. Refer to the individual test descriptions described in Chapter 8, “Diagnostics”. T - TOGGLE between the Diagnostics and Setup menus.
Since parameter changes do not take effect until you EXIT the Setup menu, you cannot change a parameter in the Setup menu and then toggle to the Diagnostic menu to observe the effect. Instead, you must exit the Setup menu and then reenter the Diagnostic menu. The Diagnostic menu wraps around. If you back up through the menu from the first test, READ SPEED, you will display the last test SET CLOCK on the last line. If SET CLOCK is already displayed when you advance forward, you will see READ SPEED again on the last line. If you have been working with both the Setup menu and the Diagnostic menu, and have changed parameter values, these changes will become effective upon exiting either the Diagnostic or the Setup menu. The following shows how the Diagnostic menu will look after entering the Diagnostic mode from the Normal mode and pressing Enter repeatedly. DIAGNOSTICS - (P)REVIOUS, (R)UN, (T)OGGLE, (E)XIT READ SPEED READ QUALITY READ EFFICIENCY MIN MAX BAR MOTOR SPEED IOPRTS LABEL POSITION TACH STATUS TACH CALIBRATION AUTOFOCUS DIAGS SET DIGITAL POTS CIDSP DIAGS CAN DIAGS EXHAUSTIVE RAM FLASH UPDATE
Rev H, February 2002
CiMAX 78xx Technical Reference Manual
5-9
5 Startup & Operation
E - EXIT the Diagnostics menu and return to Normal mode.
Page 92
Chapter
5
Startup & Operation
RESTORE KEYSHEET RESTORE PROGRAM SET CLOCK Refer to Chapter 8, “Diagnostics” for complete information about the diagnostic tests and how to run them.
CiMAX 1400 Hand-Held Terminal The CiMAX 1400 Hand-Held terminal is a full ASCII terminal with a 4 line by 20-character liquid crystal display. It has a non-volatile memory for custom function key assignments. It accepts control characters and some (non-ANSI) escape sequences for cursor positioning and other hostcontrolled functions. The same menu that is used for the ASCII terminal is used for the hand-held terminal (HHT). Because the CiMAX 1400 has a 4 by 20 display, some parameters are displayed on two lines. Function keys are programmed to provide entry to the Setup mode (~?S~) and the Diagnostic mode (~?D~). The CiMAX 1400 requires +5 VDC and Ground. The scanner supplies this when the CiMAX 1400 is connected to the Setup port at the scanner. When the CiMAX 1400 is connected to the Setup port, it pulls the TERMinal DETect pin on the port low, to indicate to the scanner that the CiMAX 1400 is connected. In applications where scanners are set up in a master-slave configuration, pulling TERMinal DETect low suspends master-slave operation and allows the scanner to be put into the Setup/Diagnostics mode. Disconnecting the CiMAX 1400 reboots the scanner and restores the system to the master-slave configuration. The keyboard of the CiMAX 1400 is shown in Figure 5–1.
5-10
CiMAX 78xx Technical Reference Manual
Rev H, February 2002
Page 93
Serial Communications Ports
FIGURE 5–1. CiMax 1400 Keyboard
CiMAX 1400 Hand-held Terminal Function Keys Two function keys, F1 and F2, on your keyboard are programmed to provide entry to Setup mode or Diagnostic mode. You must press SHIFT and release it, and then press the Function Key to enter Setup or Diagnostic mode. Setup Mode - Press SHIFT - F1. Sends ~?S~ (tilde, question mark, capital S, tilde). Diagnostic Mode - Press SHIFT - F2. Sends ~?D~ (tilde, question mark, capital D, tilde). User Programmable - Press SHIFT- F3. Refer to the CiMAX 1400 User's Guide for details. User Programmable - Press SHIFT- F4. Refer to the CiMAX 1400 User's Guide for details.
Special-Use Alpha Keys on the CiMAX 1400 Hand-Held Terminal Exits Setup or Diagnostic Mode - Press E . This returns the scanner to normal operation. Displays the Previous Menu - Press P . This returns you to the previous menu, or leaves the Setup/Diagnostic Mode if you are in the topmost menu. Toggles Between Diagnostic Mode and Setup Mode - Pointers are kept to remember the current position in each menu. The toggle function is available at any point in the menu, except during actual parameter entry. Runs a Diagnostic Test - Press R. Pressing Enter will stop the diagnostic test.
Rev H, February 2002
CiMAX 78xx Technical Reference Manual
5-11
Startup & Operation
5
Page 94
Chapter
5
Startup & Operation
Using the Hand-Held Terminal as a Real Time Display Because the CiMAX 1400 terminal supports cursor positioning, it can be used with scanners to display information about barcodes read, I/O status and other data, all in real time. Refer to parameter 012 NORMMD RPT in the Scanner Parameters Reference Guide for more information.
CiMenu32 for Windows Utility (Optional) CiMenu32 is an easy-to-use 32-bit Windows application that makes it convenient to review, set, modify and save parameter values in most SICK Auto Ident, Inc. one-dimensional scanners and terminals. Refer to the CiMenu32 User Manual for complete operating instructions. All parameter values are stored in the CiMenu32 database (CiMenu.mdb) such that you can run the application even when no scanner is connected, to set values now and download later. Parameters are arranged by device and by groups in a device. FIGURE 5–2. CiMenu32 - Device Parameter Setup Application - Main Window
CiMenu32 Features
5-12
•
Parameters and operations are presented graphically.
•
Default values for all parameters are immediately available.
CiMAX 78xx Technical Reference Manual
Rev H, February 2002
Page 95
CiMenu32 for Windows Utility (Optional)
•
Parameter values can be range-checked.
•
Current parameter settings can be quickly reviewed and modified.
•
Parameters can be downloaded by device or by a group in a device.
•
Parameter can be set offline and then downloaded to the scanner later.
•
Serial, Starnode and Ethernet communications are supported, and communications parameters are easily set.
•
Parameters can be saved and loaded to/from a text file.
•
A status bar shows the state of program execution and Help messages and indicates the scanner and communications protocol selected.
•
Supports on-line help. You can press the F1 key any time during a session to view a Help page for the selected parameter.
System Requirements
The PC can be connected to the SICK Auto Ident, Inc. scanner or terminal through a Terminal/Setup serial port or over a Starnode or Ethernet TCP/IP network. To run CiMenu32 over a Starnode network, you must install a SICK Auto Ident, Inc. Starnode OLE server for Windows NT or for Windows 95 in your PC. To run CiMenu32 over an Ethernet network, you must install a SICK Auto Ident, Inc. TCP OLE server in your PC. Refer to OLE Server User Manual for more information.
Rev H, February 2002
CiMAX 78xx Technical Reference Manual
5-13
5 Startup & Operation
CiMenu32 is designed to run on a PC using Windows 95 or later, or Windows NT 4.0 or later. It will not run on a 16-bit Windows operating system.
Page 96
Chapter
5-14
5
Startup & Operation
CiMAX 78xx Technical Reference Manual
Rev H, February 2002