Page 37

~SA Software sync for communication output mode #4 – Commanded Read Mode ( ~HO4). ~SC Execute the calibration procedure. The SV unit will respond to this command as follows: CP505050 will be transmitted if the calibration is successful. CF000000 will be transmitted if the calibration was unsuccessful ~SD Disable Verification This command turns the laser beam off. ~SE Enable Verification This command turns the laser beam on. ~SN### Set the number of scans for Software Commanded Read Mode. ### - The number of scans up to 200. ~SQ Raster Check. This command shows beams for all eight polygon facets for 4 seconds, then shows a beam for one facet for 4 seconds. Ideally, the beam widths will look identical in non-raster type scanners. Acceptable difference in most cases is 1/16th inch. ~SR Reset the System. This re-initialization procedure recalls all of the current setup parameters from the flash. ~SY Software Sync for communication output modes #1 and #2 (~HO1, ~HO2); trigger a sync response from the verifier, and enable the verifier ~Sh## Set the high target reflectance value for calibration. ## - The high target reflectance value (70 to 99). This sets a new high reflectance target value for a calibration symbol. To use the target values encoded in the calibration symbol set this value to 00. ~Sl## Set the low target reflectance value for calibration. ## - The low target reflectance value (1 to 20). This sets a new low reflectance target value for a calibration symbol. To use the target values encoded in the calibration symbol set this value to 00.

OUTPUT INTERFACE MODES DESCRIPTIONS Output interface modes are used to set the SV Series for particular ways to activate output ports. The mode is set via the ~LV## command. The following list describes the standard output interface modes available in SV firmware versions x270 and higher. Each description includes the commands for programming the

34

Page 38

particular parameters that can activate the proper ports for that mode. Custom modes are available – contact RJS Technical Support. After programming the SV unit for the desired port activation parameters, it is recommended to review them. When using ScanVison, this can be done via the Transmit SV Commands Window by sending the ~HT command to review all scanner settings, and sending the ~PT8 command to review all quality parameter failure threshold settings.

Mode 00 (~LV00) This mode does not activate output ports. Only the serial communications port is active.

Mode 01 (~LV01) A. Mode 01 I/O operation 1. System will operate in Edge Sync or Envelope sync mode (~LX#) 2. Sync polarity programmable (~LP#) 3. Use ~PR0210 for proper port initialization. 4. Port 1 will go active OFF (no current) on an error condition. 5. Port 2 will go active ON (sink current) on an error condition. 6. Pushing the reset button or power re-cycle will place Ports 1 and 2 in their inactive states and reset the # codes per sync counter.

7. Error conditions available are: a. Partial Read (Programmable, ~LQ#, ~Lp#) b. % decode (Programmable, ~LD##) c. Bad Quiet Zone (~PB816xxx100) d. ANSI Defects grade (Programmable, ~PB806xxx000) e. ANSI Decodability (Programmable, ~PB802xxx100) f. Symbol Contrast grade (Programmable, ~PB804xxx100) a. Overall ANSI Grade (Programmable, ~LAxx) g. No Read (if sync is received) h. Number of codes per sync programmable (~LN##, ~LZ##) B. Mode 01 LED Operation 1. LED1 will turn on if Outputs 1 and 2 go active due to ANSI, contrast, or quiet zone failure 2. LED2 will turn on if Outputs 1 and 2 go active due to a partial or no read condition Note; Both LEDs can be on in cases where multiple bar codes are analyzed in a sync period.

C. SV Commands Important to Mode 01 This system is operating in a “mode” rather than fully programmable logic. The port activation parameters are programmable via the ~PB8 rather than ~PB1, ~PB2, etc. for each individual port. Other commands, such as mode commands are also available. The commands most useful for this system application are described below. ~LV01 This command sets this mode of operation. ~Lp0 This command turns off all partial decode logic at the decoder level ~Lp1 This command turns on partial decode logic at the decoder level ~LQ1

35

Page 39

This command allows partial decodes to activate the output ports ~LQ0 This command disables partial decodes to activate output ports. NOTE: In SV firmware versions x238 and lower, ~LQ commands are not implemented and the command ~HQ1 must be used to disable partial decodes from activating ports. In this case the standard data transmission format is modified and ScanVision will not display data characters correctly. ~LDxx This command sets the % decode threshold for a passing condition. If analyzing poorly printed symbols, it is recommended that partial decodes be enabled at the decoder level when this command is enabled. xx = failure threshold. Example ~LD75: if 74 % or less of scans on a code were not fully decoded, this sets a failure for this parameter. ~PB816xxx100 This command sets the minimum percent of scans on a code which calculate good quiet zones to determine an acceptable quiet zone analysis. The field xxx is the minimum passing threshold. Example: if xxx = 030, a minimum of 30% of all fully decoded scans on a code must calculate a good quiet zone, or a failure condition is set for this parameter. ~PB806xxx000 This command sets the threshold for the ANSI Defects calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 025, a Defects analysis of 26% or higher will cause a failure condition for this parameter. ~PB802xxx100 This command sets the threshold for the ANSI Decodability calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 037, a Decodability analysis of 36% or lower will cause a failure condition for this parameter. ~PB804xxx100 This command sets the threshold for the ANSI Symbol Contrast calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 020, a Symbol Contrast analysis of 19% or lower will cause a failure condition for this parameter. ~LN## This command sets the minimum number of codes to be read during a sync period. ## = the number of codes. For example: ~LN02 causes a No Read condition to be set if less than 2 bar codes are fully decoded during a sync interval. ~LZ## This command sets the exact number of codes to be read during a sync period. ## = the number of codes. For example: ~LZ02 causes a No Read condition to be set if exactly 2 bar codes are not fully decoded during a sync interval. This command overrides the ~LN## command if the number of codes set does not = 00. If ~LZ## is set to ~LZ00, the ~LN## command takes precedence. ~LAxx This command sets the passing threshold for the Overall ANSI method Symbol Grade. For example: ~LA28 causes an Overall Symbol Grade of 2.7 or lower calculated for the code analyzed to set a failure condition for this parameter. ~Hx This command stores all parameters into non-volatile memory. This command should be the last command sent after parameters are programmed via the above commands.

36

Page 40

Mode 02 (~LV02) A. Mode 02 I/O operation 1. Envelope type sync input on pin 4 2. Port 3: per sync type. The state of this output must be stable within 12.5 milliseconds after the sync input goes low (trailing edge.) Typical time for stability will be less than 7.5 milliseconds if the sync is received when a bar code or complex graphic is not in the beam. The port shall go active (sink current) for the following conditions; a. ANSI Defects grade programmable (~PB806xxx000) b. ANSI Decodability grade programmable (~PB802xxx100) c. Overall ANSI Grade (Programmable, ~LAxx) d. Bad quiet zone (~PB816xxx000) e. No Read (if sync is received) f. Partial decodes and % decode logic programmable (~Lp#, ~Lq#) g. Number of codes per sync programmable (~LN##, ~LZ##) 3. Sync polarity - Leading edge high going; time between trailing (low going) edge to leading edge - 12.5 milliseconds minimum 4. Port 1; acts as strobe (signal ensuring Port 3 condition is stable) in response to trailing edge of sync input. This signal will go “on” (sink current) between 20 and 100 microseconds after Port 3 condition is stable. 5. Ports 1 and 3 will go high (no current) within 12.5 milliseconds after the sync input goes high (leading edge). 6. Use ~PR0200 for proper port initialization. B. Mode 02 LED Operation 1. LED1 will turn on if Port 3 goes active due to ANSI or quiet zone failures 2. LED2 will turn on if Port 3 goes active due to a no read or partial condition 3. The leading edge of a sync input will turn LEDS 1 and 2 OFF.

C. SV Commands Important to Mode 02 This system is operating in a “mode” rather than fully programmable logic. The port activation parameters are programmable via the ~PB8 rather than ~PB1, ~PB2, etc. for each individual port. Other commands, such as mode commands are also available. The commands most useful for this system application are described below. ~LV02 This command sets this mode of operation. ~Lp0 This command turns off all partial decode logic at the decoder level ~Lp1 This command turns on partial decode logic at the decoder level ~LQ1 This command allows partial decodes to activate the output ports ~LQ0 This command disables partial decodes to activate output ports NOTE: In SV firmware versions x238 and lower, ~LQ commands are not implemented and the command ~HQ1 must be used to disable partial decodes from activating ports. In this case the standard data transmission format is modified and ScanVision will not display data characters correctly.

37

Page 41

~LDxx This command sets the % decode threshold for a passing condition. If analyzing poorly printed symbols, it is recommended that partial decodes be enabled at the decoder level when this command is enabled. xx = failure threshold. Example ~LD75: if 74 % or less of scans on a code were not fully decoded, this sets a failure for this parameter. ~PB816xxx100 This command sets the minimum percent of scans on a code, which calculate good quiet zones to determine an acceptable quiet zone analysis. The field xxx is the minimum passing threshold. Example: if xxx = 030, a minimum of 30% of all fully decoded scans on a code must calculate a good quiet zone, or a failure condition is set for this parameter. ~PB806xxx000 This command sets the threshold for the ANSI Defects calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 025, a Defects analysis of 26% or higher will cause a failure condition for this parameter. ~PB802xxx100 This command sets the threshold for the ANSI Decodability calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 037, a Decodability analysis of 36% or lower will cause a failure condition for this parameter. ~LAxx This command sets the passing threshold for the Overall ANSI method Symbol Grade. For example: ~LA28 causes an Overall Symbol Grade of 2.7 or lower calculated for the code analyzed to set a failure condition for this parameter. ~LN## This command sets the minimum number of codes to be read during a sync period. ## = the number of codes. For example: ~LN02 causes a No Read condition to be set if less than 2 bar codes are fully decoded during a sync interval. ~LZ## This command sets the exact number of codes to be read during a sync period. ## = the number of codes. For example: ~LZ02 causes a No Read condition to be set if exactly 2 bar codes are not fully decoded during a sync interval. This command overrides the ~LN## command if the number of codes set does not = 00. If ~LZ## is set to ~LZ00, the ~LN## command takes precedence. ~Hx This command stores all parameters into non-volatile memory. This command should be the last command sent after parameters are programmed via the above commands.

Mode 03 (~LV03) This mode is identical to Mode 01 with the exception of the meaning of the ~LN command: ~LN## This command sets the exact number of codes to be read during a sync period. ## = the number of codes. For example: ~LN02 causes a No Read condition to be set if less or more than 2 bar codes are fully decoded during a sync interval. (This system is using edge sync mode, therefore a sync interval is the time between sync inputs.) ~LN00 disables a no read condition.

Mode 12 (~LV12) A. Mode 12 I/O operation 1. System will operate Envelope or Edge sync mode (~LX#) 38

Page 42

2. 3. 4. 5. 6.

Sync polarity – programmable (~LP#) Use ~PR0210 for proper port initialization. Port 1 will go active OFF (no current) on an error condition. Port 2 will go active ON (sink current) on an error condition. Port 3 will operate in per analysis pulse mode. Active polarity ON (sink current), approximately 500 millisecond duration. Port 3 will activate (pulse) on an error condition. 7. Port 4 will operate in per analysis pulse mode. Active polarity ON (sink current), approximately 50 millisecond duration. Port 4 will activate (pulse) on an error condition. 8. Pushing the reset button or power re-cycle will place Ports 1- 4 in their inactive states. 9. Error conditions available are: a. Partial Read (~Lp#, ~LQ#) b. % decode programmable, ( ~LD##) c. Bad Quiet Zone (~PB816xxx100) d. ANSI Defects grade programmable (~PB806xxx000) e. ANSI Decodability grade programmable (~PB802xxx100) f. Symbol Contrast grade programmable (~PB804xxx100) b. Overall ANSI Grade (Programmable, ~LAxx) g. No Read (if edge sync is received) h. Number of codes per sync programmable (~LN##, ~LZ##)

B. Mode 12 LED Operation 1. LED1 will turn on if Ports 1-4 go active due to ANSI, contrast, or quiet zone failure 2. LED2 will turn on if Ports 1-4 go active due to a partial or no read condition Note; Both LEDs can be on in cases where multiple bar codes are analyzed in a sync period.

C. SV Commands Important to Mode 12 This system is operating in a “mode” rather than fully programmable logic. The port activation parameters are programmable via the ~PB8 rather than ~PB1, ~PB2, etc. for each individual port. Other commands, such as mode commands are also available. The commands most useful for this system application are described below. ~LV12 This command sets this mode of operation. ~Lp0 This command turns off all partial decode logic at the decoder level ~Lp1 This command turns on partial decode logic at the decoder level ~LQ1 This command allows partial decodes to activate the output ports ~LQ0 This command disables partial decodes to activate output ports. NOTE: In SV firmware versions x238 and lower, ~LQ commands are not implemented and the command ~HQ1 must be used to disable partial decodes from activating ports. In this case the standard data transmission format is modified and ScanVision will not display data characters correctly. ~LDxx

39

Page 43

This command sets the % decode threshold for a passing condition. If analyzing poorly printed symbols, it is recommended that partial decodes be enabled at the decoder level when this command is enabled. xx = failure threshold. Example ~LD75: if 74 % or less of scans on a code were not fully decoded, this sets a failure for this parameter. ~PB816xxx100 This command sets the minimum percent of scans on a code which calculate good quiet zones to determine an acceptable quiet zone analysis. The field xxx is the minimum passing threshold. Example: if xxx = 030, a minimum of 30% of all fully decoded scans on a code must calculate a good quiet zone, or a failure condition is set for this parameter. ~PB806xxx000 This command sets the threshold for the ANSI Defects calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 025, a Defects analysis of 26% or higher will cause a failure condition for this parameter. ~PB802xxx100 This command sets the threshold for the ANSI Decodability calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 037, a Decodability analysis of 36% or lower will cause a failure condition for this parameter. ~PB804xxx100 This command sets the threshold for the ANSI Symbol Contrast calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 020, a Symbol Contrast analysis of 19% or lower will cause a failure condition for this parameter. ~LN## This command sets the minimum number of codes to be read during a sync period. ## = the number of codes. For example: ~LN02 causes a No Read condition to be set if less than 2 bar codes are fully decoded during a sync interval. ~LZ## This command sets the exact number of codes to be read during a sync period. ## = the number of codes. For example: ~LZ02 causes a No Read condition to be set if exactly 2 bar codes are not fully decoded during a sync interval. This command overrides the ~LN## command if the number of codes set does not = 00. If ~LZ## is set to ~LZ00, the ~LN## command takes precedence. ~LAxx This command sets the passing threshold for the Overall ANSI method Symbol Grade. For example: ~LA28 causes an Overall Symbol Grade of 2.7 or lower calculated for the code analyzed to set a failure condition for this parameter. ~Hx This command stores all parameters into non-volatile memory. This command should be the last command sent after parameters are programmed via the above commands.

Mode 16 (~LV16) A. Mode 16 I/O operation 1. 2. 3. 4. 5.

System will operate in Edge Sync or Envelope sync mode (~LX#). Sync polarity programmable (~LP#) Use ~PR0200 for proper port initialization. Port 1 will pulse active ON for a bar code quality error condition. Port 2 will pulse active ON for a NO READ, partial or low % decode error condition.

40

Page 44

6. 7. 8. 9.

Port 3 will pulse active ON for a non-error condition. Port pulse times programmable. (~PP801xxx) Pulses act like re-triggerable 1-shot logic Pushing the reset button or power re-cycle will place Ports 1, 2 and 3 in their inactive states and reset the # codes per sync counter. 10. Error conditions available are: a. Partial Read (Programmable, ( ~Lp#, LQ#) b. % decode (Programmable, ~LD##) c. Bad Quiet Zone: < 10X (~PB816xxx100) d. ANSI Defects grade less than B (Programmable, ~PB806xxx000) e. ANSI Decodability grade less than C (Programmable, ~PB802xxx100) f. Symbol Contrast grade less than D (Programmable, ~PB804xxx100) g. Overall ANSI Grade (Programmable, ~LAxx) h. No Read (if sync is received) i. Number of codes per sync programmable (~LN##, ~LZ##) j. X dimension (Programmable ~PB811xxxyyy)

B. Mode 16 LED Operation 1.

LED1 will turn on during the time a Port goes active due to any bar code quality or dimensional error. 2. LED2 will turn on during the time a Port goes active due to a partial or no read condition Note; Both LEDs can be on in cases where multiple bar codes are analyzed in a sync period.

C. SV Commands Important to Mode 16 This system is operating in a “mode” rather than fully programmable logic. The port activation parameters are programmable via the ~PB8 rather than ~PB1, ~PB2, etc. for each individual port. Other commands, such as mode commands are also available. The commands most useful for this system application are described below. Please see the SV Series Operator’s Guide for additional command description details. ~LV16 This command sets the proper mode for this operation. ~Lp0 This command turns off all partial decode logic at the decoder level ~Lp1 This command turns on partial decode logic at the decoder level ~LQ1 This command allows partial decodes to activate the output ports ~LQ0 This command disables partial decodes to activate output ports.

~PP801xxx This command sets the “active” time for the port pulses. The value “xxx” sets the time in .1 second intervals. Any value greater than 100 (10 seconds) creates a latching output. ~LDxx This command sets the % decode threshold for a passing condition. If analyzing poorly printed symbols, it is recommended that partial decodes be enabled at the decoder level when this command is enabled.

41

Page 45

xx = failure threshold. Example ~LD75: if 74 % or less of scans on a code were not fully decoded, this sets a failure for this parameter. ~PB816xxx100 This command sets the minimum percent of scans on a code which calculate good quiet zones to determine an acceptable quiet zone analysis. The field xxx is the minimum passing threshold. Example: if xxx = 030, a minimum of 30% of all fully decoded scans on a code must calculate a good quiet zone, or a failure condition is set for this parameter. ~PB806xxx000 This command sets the threshold for the ANSI Defects calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 025, a Defects analysis of 26% or higher will cause a failure condition for this parameter. ~PB802xxx100 This command sets the threshold for the ANSI Decodability calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 037, a Decodability analysis of 36% or lower will cause a failure condition for this parameter. ~PB804xxx100 This command sets the threshold for the ANSI Symbol Contrast calculation on a code to set a failure condition. The field xxx is the passing threshold for the calculation. For example: if xxx = 020, a Symbol Contrast analysis of 19% or lower will cause a failure condition for this parameter. ~PB811xxxyyy This command sets the threshold for acceptable X dimension ranges. The value for xxx is the narrowest acceptable X dimension. The value yyy is the widest acceptable X dimension. Units are in .001 inch (mils). For example: ~PB811010020 will cause a failure for n X dimension less than 10 mils or greater than 20 mils. ~LAxx This command sets the passing threshold for the Overall ANSI method Symbol Grade. For example: ~LA28 causes an Overall Symbol Grade of 2.7 or lower calculated for the code analyzed to set a failure condition for this parameter. ~LN## This command sets the minimum number of codes to be read during a sync period. ## = the number of codes. For example: ~LN02 causes a No Read condition to be set if less than 2 bar codes are fully decoded during a sync interval. (This system is using edge sync mode, therefore a sync interval is the time between sync inputs.) ~LZ## This command sets the exact number of codes to be read during a sync period. ## = the number of codes. For example: ~LZ02 causes a No Read condition to be set if exactly 2 bar codes are not fully decoded during a sync interval. This command overrides the ~LN## command if the number of codes set does not = 00. If ~LZ## is set to ~LZ00, the ~LN## command takes precedence. ~Hx This command stores all parameters into non-volatile memory. This command should be the last command sent after parameters are programmed via the above commands.

Mode 17 (~LV17) A. I/O operation

42