The text of a manual · part 7 of 9

SV Series Scanner/Verifier Operator's Guide: SV COMMAND LANGUAGE

Pages 25 to 36 of 54, from SV Series Scanner/Verifier Operator's Guide

The original document is the PDF (666 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the RJS SV 100

Page 25

Symbology Type (scanner decoder function) % Decode (multiple scanning parameter) Global Threshold (ANSI method parameter) Modulo Check Digits (mandatory symbology and optional application parameters) Encoded Data (scanner decoder function)

SV COMMAND LANGUAGE All commands to the verifier begin with a tilde, ‘~’, followed by one or more alphanumeric characters. The ‘~’ identifies the subsequent character string as a verifier command. The string is parsed from left to right. The first character after the ‘~’ identifies the command category. Each command requires at least one additional character that identifies the particular command from that category. Some commands require additional data, e.g., a data value, or an option ID character, or a text string. These values are included at the end of the command string. The command string format is shown below. As each character is received by the SV unit, it echoes back the same character. Echo-back is used to determine when the verifier has completed the command action. The verifier is ready to receive the next command when the last character of the command is transmitted (echoed-back) to the host. ~CPx… ~ - indicates the start of a command string. C - a character that identifies the command Category P - a character that specifies the Particular command in the category x… - a number or text string (as necessary)

Data Match Commands – B A ‘~’ followed by a ‘B’ specifies a command from the Data Match category. All Data Match commands include a second alphabetic character that identifies the particular diagnostic command. The Data Match Commands are listed below. An important consideration when using data match type commands is the data format. These commands incorporate a “scanner” philosophy rather than a “verifier” philosophy. A scanner philosophy is to report the data encoded in a symbol while a verifier philosophy is to report all symbol characters whether they include data or not. Symbol characters typically include stop character, start character, symbology mod check digit(s), internal control characters, etc. depending on the symbology type. The data fields described in the B commands below include data characters only, therefore the ~B commands have a “scanner” philosophy. The one possible exception is in UPC or EAN symbols the symbology mod 10 check digit is also considered a data character. ~BCa##bc{…bbb…}{…mmm…}{…bbb…}… {etc. up to 32 characters max.} Check for specific data This command sets (enables) or clears (disables) one of 10 available data match arrays. All ten arrays may be active simultaneously. A data match error occurs if at least one array is active and the data scanned does not match the data programmed in the array. Each array can be up to 32 characters in length. Multiple portions of each array can be masked and the mask character is defined in the command. The data in the array must consist of printable ASCII characters with ASCII values less than 128. ~BC - Data Match Command a - 1 digit. A number between 0 and 9. It identifies the data match array being programmed or cleared. ## - Number of data characters in the Data Match array. The value can be 01 through 32. A value of 00 clears (disables) this array. b - This character defines the fill character for masking purposes. This character cannot be a character to be matched in the match field mmm. c - This character defines the size of the data match field to be fixed (“f”) or variable (“v”). If variable, the data in the field is left justified in a 32 character possible field.

22

Page 26

…bbb…

- Fill characters prior to the Data Match text. These characters must match the character defined by b in the command. …mmm… - Data Match Character Field. This can be from 1 to 32 characters. These define the exact characters that must be matched. …bbb… - Fill characters after the Data Match text (if needed). Note: Other Data Match Fields and Fill characters can follow for up to 32 characters total length. Examples: ~BC005xfABCDE -

Data match array #0 with 5 characters of “ABCDE” of fixed length with ‘x’ as the ignore or fill character. In this case, the ‘x’ must be in the command, but is not used. ~BC210xfxxxABCDExZ - Data match array #2 with 10 characters of fixed length. ‘x’ as the ignore or fill character. The 4th through 8th characters must be ABCDE respectively and the 10th character must be a Z in a 10 character field for a data match. ~BC004xvx111 - Data match array #0 is defined as variable length, therefore can be up to 32 characters long. ‘x’ is the fill character. The second through 4th characters all must be 1 for a match ~BC000bc - Disable Data match array #0. Any proper character as described above can be used for the b and c characters. The remainder of the array data is not required. Note: In the above examples, the character ‘x’ (defined as the fill character) cannot be included in the portion of the bar code data that is programmed to be matched. ~ Brnbb{ .. + 0-9.. ! .... ) Define increment or decrement data check One array is available for checking a numeric (base 10) or alphanumeric (base 36) incrementing or decrementing field. The data field being checked must be fixed length. Portions of a field can be checked through masking. An initial (starting) value for the field may also be programmed in the array. Total field length can be up to 32 characters maximum. Characters analyzed can be numeric or alphanumeric (0-9, AZ). This command can be used in conjunction with the ~BC command. ~B - Command initiation for checking an incrementing data field r - Indicates incrementing (I) or decrementing (D) function n - Required digit following ~Br. If set to a ‘0’ (zero), the check is performed for digits 0-9 only. If set to 1, the check is performed for alphanumeric characters 0-9, A-Z. bb - Number of characters in array. A value of 00 clears (disables) this array. …!!! - Mask characters for defining positions in the data field not being checked. …+++ - Characters in the data field being checked. If + characters are used, this marks the field and defines its size. Maximum field lengths are 8 characters for numeric and 6 characters for alphanumeric. If a proper numeric or alphanumeric value is placed in the field, this additionally become the initial value expected on the first data scanned after a power-up or reset. Incrementing Examples: ~BI010!!!!!!!+++ - Increment is to be done in the last three characters of the 10 character array. ~BI010!!!!!!!234 -

Increment is to be done in the last three characters of the 10 character array. Initial value scanned must be 234.

-

Increment is to be done in location 6 through 8 characters of the 10 character array.

~BI010!!!!!+++!!

23

Page 27

Decrementing Examples: ~BD010!!!!!!!+++ - Decrement is to be done in the last three characters of the 10 character array. ~BD110!!!!!!!C34 ~BD010!!!!!+++!! -

Decrement is to be done in the last three characters of the 10 character array. The initial value scanned must be C34 (base 36 alphanumeric field). Decrement is to be done in location 6 through 8 characters of the 10 character array.

~BI000 or ~BD000 - Disable the Increment/Decrement array. The remainder of the array data is not required.

Relationship Between ~BC and ~Br Commands When programming both data match and increment/decrement arrays, the data match array(s) have higher priority. This is useful when analyzing a label containing more than one bar code, and one of the codes contains an incrementing or decrementing field. In this case, a data match array(s) can be used to match (or in this case – logically “mask”) the bar code(s) containing static data and the increment/decrement array can analyze the incrementing or decrementing bar code data. ~ BT Lists all Data Match strings and Increment/Decrement fields.

Diagnostic Commands - D A ‘~’ followed by a ‘D’ specifies a command from the Diagnostic category. All Diagnostic commands include a second alphabetic character that identifies the particular diagnostic command. The Diagnostic Commands are listed below. ~DF Get the Scan Rate Possible returned frequency values are 270 through 800. 1) ~D 2) hexadecimal value ‘4’ - identifies the beginning of the datapacket response 3) a data sequence that repeats the following format: a space, then three digits (a ‘counts’ value), then a ‘\r’ character (carriage return) and a ‘\n’ character (a line feed). 4) hexadecimal value ‘5’ - identifies the end of the datapacket response 5) F - the last character of the sent command string Example return using Scanvision: Scan Rate: 400 ~DN# Inquire port activation data. This command instructs the verifier to report all parameters of the event that caused this port to go active. # = 1 through 8 (port ID) ~DV Inquire software version. Commands the verifier to transmit the SV Series firmware version to the host in the following datapacket format: 1) ~D

24

Page 28

2) hexadecimal value ‘4’ - identifies the beginning of the datapacket response 3) The following character string between the quotes: ‘Version: ‘ 4) Five alphanumeric characters that specify the version number. They can include trailing spaces. (See example below.)

5) hexadecimal value ‘5’ - identifies the end of the datapacket response 6) V Example return using Scanvision: Version: X244

Hardware Configuration Commands - H A ‘~’, immediately followed by an ‘H’, specifies a command from the ‘Hardware’ category. All Hardware Commands require one additional character that identifies the particular Hardware Command. The Hardware Commands are listed below. ~HB# Set the Baud Rate. Possible values for # are:

1 (specifies 9600) 2 (specifies 19200) 3 (specifies 38400) 4 (specifies 57600) 5 (specifies 115200… the default) The default value is set at the factory (before shipping). Returns the following sequence: 1) ~H 2) hexadecimal value ‘4’ - identifies the beginning of the datapacket response. 3) Six hexadecimal digits (upper case) - the memory address 4) \r - carriage return character 5) \n - line feed character 6) hexadecimal value ‘5’ - identifies the end of the datapacket response 7) B Example return using Scanvision: (inputted ~HB5) BAUD SET TO: 115200 ~HL# Select “Ladder” or “Picket Fence” bar code travel direction mode. # = 1 (Ladder) 0 (Picket Fence) ~HO# Operational Mode # = 0 (inhibit output) Use this state only when downloading. 1 (moving bar codes mode) Use for normal operation. 2 (moving bar codes mode, fast code exit) Use only when scanning one code across in each scan. 3 (reserved) 4 (stationary bar codes mode) ~HQ# Transmission Mode # = 0 = Standard mode 1 – 9 = reserved

25

Page 29

~HT Display the verifier parameters The format of the reply is: 1) ~H 2) hexadecimal value ‘4’ - identifies the beginning of the datapacket response. 3) A list of text strings like the one shown in the example. Each string is terminated by a two character sequence: ‘\r’ (the carriage return character), ‘\n’ (the line feed character). 4) hexadecimal value ‘5’ - identifies the end of the datapacket response 5) T Example: (The numbers of parameters reported varies with the commands included in the firmware revision) ~H {hex 4} [~HS###]SOS= 060 {This is the SOS delay value} [c]HIGHcal= 088 {This is a CAL value} [c]LOWcal= 004 {This is a CAL value} [c]SCfact= 046 {This is a CAL value} [c]min cal= 004 {This is a CAL value} [~HB#]baud= 054 {This is the baud rate value} [~LL]smt_decode= 535 {This enables decoders} [~LM##]chars= 003 {Minimum number of characters} [~LN##]num_bc= 003 {Number of barcodes per sync} [~HQ#]output_mode= 000 {Serial Output Format} [~HO#]oper_mode= 001 {Scanning Mode} [~LA##]ansi= 005 {Minimum Overall grade} [~LS##]min_scn= 002 {Number of scans per barcode} [~LT#]sync_in= 001 {diff. Or ttl input} [~LP#]sync_pol= 000 {Sync polarity set by this} [~LX#]sync_typ= 001 {Env. Or Edge sync} [~LV##]sync_md= 011 {Custom sync modes; 09 STD} [~PO#]port_sel= 005 { } port_clear_md= 000 {Clears ports per read if 2} port_time= 099 { } port_pol= 000 { } 10_mil_cnts= 120 { } [~LD##] %dec= 070 {Percent decode} [~PR####]PRST= 000210 {Port reset state} [~HL#]ladder_code= 000 {Ladder mode flag} [~HJ####]ladd_start= 0400 {Start ladder reads here} [~HK####]ladd_end= 3500 {End ladder reads here} [~HC####]ana_len= 2800 {Profile Sampling size} [~HG####]ana_lenh= 1600 {Laser shutoff point} [~HN####]pana_len= 1600 {Processing shutoff point} [~HU####]freq= 400 {Scan Rate} [~HV###]GTLA_len= 199 {Processing limit} [~HY###]snoise= 070 {Signal to Noise threshold} [~HX###]dist= 078 {Presets the distance} [~H=#]dyn_out= 000 {Dynamic Zone output} [~H!###]zone_tol= 350 {Tolerance of starting locations in a zone} [~H@##]gt_%= 043 {Finder Global Threshold} [~SN###]#scans= 100 {Number of scans in ~HO4 mode} [~Lp#]part= 001 {Enables partials at the decoder level} [~Lt#]itfp= 001 {Enables or DisablesITF partials at the decoder level} [DACs]= 125 126 255 255 {D/A converter levels, Rev.F} {hex 5} T ~Hdx01nnn

26

Page 30

Sets scanner gain or offset. x = value of 1 or 2. If =1, scanner gain is set. If = 2, scanner offset is set. nnn = gain or offset setting. For gain, the higher the value, the higher the gain. 225 is the max allowed value. Example: ~Hd101195 sets the scanner gain setting to a value of 195. For offset, the higher the value, the lower the offset. Example: ~Hd201145 sets the scanner offset value to 145. These commands are best used in conjunction with RJS ScanVision software. Please refer to the Setup Procedures Section in this guide for descriptions on the use of these commands. ~Hx Saves the parameters to FLASH (non-volatile memory) from RAM. ~H#nn Sets the number of signal transitions to search for finding a bar code symbol nn = values of 10 through 99. This indicates the number of high to low transitions of the optical signal to search before ending a search for a bar code. High values of nn are useful when complex graphic fields or small fonts are closely adjacent to a bar code and therefore in the scan path. The higher the number, however, the greater chance the analysis rate will be lowered. For normal situations with minimum graphics in the same path as the bar code, a value of 10 is recommended (~H#10.)

Label Setup Commands - L A ‘~’ followed by an ‘L’ specifies a command from the Label Setup category. The next character identifies the particular Label Setup Command. In general, the commands in this category terminate with one or more numeric digits. The Label Setup Commands are listed below. ~LA## Overall passing grade. ## is the numeric grade value – 00 through 40. Example ~LA32 causes an ANSI Overall Symbol Grade of 3.1 or lower to set a failure condition. ~LC## Minimum number of symbol characters to decode in an Interleaved 2 of 5 symbol. This command sets the minimum number of symbol characters to be expected (## in the command) by the Interleaved 2 of 5 decoder. This adds extra filtering in applications where I 2 of 5 symbols are used and there is a large amount of other graphics such as fonts in the scan path. Example: ~LC03 sets the unit to decode I 2 of 5 symbols (assuming the I 2 of 5 decoder is enabled) containing three or more symbol characters. Note: each I 2 of 5 symbol character contains two data digits, therefore the symbol in the example must contain a minimum of six data digits to be decoded.

~LD## Passing percent decode. This command sets the % decode threshold for a passing condition. If analyzing poorly printed symbols, it is recommended that partial decodes be enabled at the decoder level when this command is enabled.

27

Page 31

xx = failure threshold. Example ~LD75: if 74 % or less of scans on a code were not fully decoded, this sets a failure for this parameter. ~LE# Exclude the Specified Symbology for a Label: (‘#’ identifies the excluded symbology: #= 1(UPC/EAN) 2(Code 39) 3(Code 128) 4(Interleaved 2 of 5) 5(Code 93) 6(Codabar.) ~LFxy Analyze special subsymbology parameters This command enables special subsymbology parameters within a symbology. This does not affect the automatic calculations and analyses for mandatory symbology check digits such as those required in Code 128, UPC/EAN and Code 93. x = symbology type 1(UPC/EAN) 2(Code 39) 3(Code 128) 4(Interleaved 2 of 5) 5(Code 93) 6(Codabar.) y = subsymbology choice for a particular symbology. Current choices include: 0 = disabled (default condition) 1 = Mod 10 check digit for Interleaved 2 of 5 Note: additional subsymbology choices will be added as applications require. Contact RJS for details. Example: ~LF41 = check mod 10 check digit in last location for any I 2 of 5 symbol analyzed ~LF40 = disable any subsymbology analysis for I 2 of 5 symbols ~LI# Include the Specified Symbology for a Label: ‘#’ identifies the included symbology: #= 1(UPC/EAN) 2(Code 39) 3(Code 128) 4(Interleaved 2 of 5) 5(Code 93) 6(Codabar.) ~LL Report the Excluded Symbologies and the Included Symbologies Specified for all the Defined Bar Codes. 1) Hexadecimal value ‘4’ - identifies the beginning of the datapacket response 2) The mode for each symbology is reported in a character string, which terminated with either ‘IN’ (for included) or ‘EX’ (for excluded) followed by a ‘\r’ character (carriage return) and a ‘\n’ character (a line feed). See the example below. The next two columns indicate statuses of the ~LM and ~LN commands respectively. (The ~LM command is not used in normal operation.) 3) Hexadecimal value ‘5’ - identifies the end of the datapacket response 4) L - the last character of the sent command string Example return using ScanVision: 1- UPC/EAN- IN 2- C39- IN 3- C128- IN 4- CITF- IN 5- C93- IN

28

Page 32

6- CBAR- IN min chars: 032 number bc: 099 L ~LN## Set Minimum Number of Bar Codes Per Label (i.e., per sync period) This command instructs the verifier on how many bar codes it is expected to read and analyze during a sync period. If the programmed number of codes are not read, a (robust) NO READ condition is internally flagged for communications or output port activation. Except for mode ~LV03, this command sets the minimum bar codes expected. Mode ~LV03 uses this value for the EXACT number of codes read during the sync interval. Note: See ~LZ## command for other options. ## = 01 through 99 Default = 01. ~LP# Set Sync Polarity (See Sync Inputs Section for examples of sync input polarities.) # = 0 (falling edge of signal is active) 1 (rising edge of signal is active) ~LQ# Enable or Disable Port Activations by Partial Decodes 0- Disable Port Activation by Partial Decodes 1- Enable Port Activation by Partial Decodes ~LR# Enable or Disable No Read Transmission A No Read transmission message can be sent via the serial port if the SV unit is receiving sync inputs. See the Serial Port Transmission Format Section for details. Fields y, M and N (data positions 52 through 55) set to 0 indicate a No Read condition. # = 1, enables the No Read message to be transmitted # = 0, disables the No Read transmission. The transmission is available in any output interface mode that includes a no read condition. The transmission data format enabled (~OS#) does not affect the format of the No Read transmission. ~LS## Set minimum number of scans for valid read ## = 02 – 99 This command sets the minimum number of times a full decode must be achieved on a symbol in order to be a valid read. The system acts slightly differently, depending on what operation mode is set – ~HO1 or ~HO2. With ~HO1 enabled, if the minimum number of full decodes is not accomplished, a No Read condition is set, even though the bar code is read, analyzed and reported. The report includes a No read status indication. With ~HO2 enabled, if the minimum number of full decodes is not accomplished, the code is not recognized nor reported. This essentially creates a No Read condition when a sync input is being received. When used in conjunction with the ~LDxx command, the ~LS## command can create a powerful area check of a symbol. ~LT# Sync Input Source Isolated Input (Pins 1(+), 2(-)) = TTL Input (Pin 4) =

0 1

29

Page 33

~LV## Output Interface Mode ## = 00 - 99: This command sets the output interface mode of the unit. Modes have fixed port activation logic defined by specific applications. Some activation parameters are programmable. Custom modes are available – contact RJS technical Support. Modes included in standard SV Series are described in the Output Interface Modes Descriptions Section. Example: ~LV01 sets mode 01. ~LX# Sync Type 0- Envelope sync 1- Edge sync ~LZ## Set Number of Codes per Label (i.e., per sync period) This command sets the exact number of codes to be read during a sync period. ## = the number of codes. For example: ~LZ02 causes a No Read condition to be set if exactly 2 bar codes are not fully decoded during a sync interval. This command overrides the ~LN## command if the number of codes set does not = 00. If ~LZ## is set to ~LZ00, the ~LN## command takes precedence. ~Lp# Turn partials on and off at the decoder level for all symbologies. 0- Disable 1- Enable ~Lt# Turn partials on and off at the decoder level for Interleaved 2 of 5 codes only. 0- Disable 1- Enable

Output Mode Selection Commands - O A ‘~’ followed by an ‘O’ specifies a command from the Output Mode Selection category. These commands include one additional character that identifies the particular Output Mode Command. The Output Mode Selection Commands are listed below. ~OL## Set an LED on/off. (Does not over-ride normal operation; recommended for use as diagnostic in Mode 0 (~LV00) only) First ‘#’ - Possible values are ‘1’,’2’. Identifies the LED. Second # - Possible values are ‘1’ (on), ‘0’ (off). ~OP## Set a PORT on or off. (Does not over-ride normal operation; recommended for use as diagnostic in Mode 0 (~LV00) only) First # - Possible values are ‘1’,’2’,’3’,’4’,’5’. Identifies the port. Second # - Possible values are ‘1’ (on), ‘0’ (off).

~OS# Set Transmission Data Format This command is affective only if the standard transmission mode is enabled via the ~HQ0 command. 0 = Standard transmission format as described in the Serial Port Transmission Format Section

30

Page 34

1 =Standard transmission format modified; analysis parameters a through N included, but encoded data is formatted as data characters only rather than all symbol characters. The data consists only of printable ASCII characters and does not include any mandatory symbol characters such as stop, start and symbol mod check characters. One exception is in UPC/EAN codes, the mod 10 check digit is included.

Output Port Setup Commands - P A ’~’ immediately followed by a ‘P’ specifies a command from the Output Port Setup category. NOTE: The following descriptions of the ~PB and ~PP commands imply all ports are individually programmable. All SV port logic is currently programmed via the Output Interface Mode command ~LV##. See the Output Interface Mode Descriptions section for details of port logic. The ~PB and ~PP commands are used to program activation parameters available in the mode settings and are limited to the case where the value of the variable p in the command is always a value of 8. “Port 8” is a general port used for internal purposes. This structure reserves individual port programming to be available in the future. Many of the activation parameters are ANSI method calculations. Please see the ANSI Parameter Grade Thresholds Section for details on the values of these parameters. ~PBpiiaaabbb Port barcode decision block parameters. piiaaabbb = Nine decimal digits representing: Port- p ID- ii (See Table of Analysis Parameters for ID descriptions) Low passing threshold- A (aaa) High passing threshold- B (bbb) Values for aaa and bbb can be found in the following table: Values “A” and “B” determine the passing thresholds. If the measured parameter is less than A or greater than B the relevant error flag would be set. See Output Interface Modes Descriptions Section for ~PB command examples

Table of Analysis Parameters IDs

Analysis Parameters

Low

High

Units

Low Value

High Value

01 02 03 04 05 06 07 08 09 10 11 12 13 14 15 16 17 18 19 20 21

Overall Grade Decodability Modulation Symbol Contrast EC min Defects Rmin/Rmax Rw Rb PCS Z within TOL of VAL AVG BWD MIN BWD MAX BWD Ratio QZ (Percent good QZ) ICG Application Mod CHK Min Scans Min Chars Percent Decode

0.0 0 0 0 0 0 0 0 0 0 1 -100 -100 -100 1.8 0 0.1 NA 2 1 0

4.0 100 100 100 100 100 100 100 100 100 255 100 100 100 8.0 100 8.0 NA 200 64 100

Grade % % % % % % % % % Mils % of X % of X % of X Ratio % X-dim Flag Scans # %

000 000 000 000 000 000 000 000 000 000 001 000* 000* 000* 018 000 001 000 002 001 000

040 100 100 100 100 100 100 100 100 100 255 200* 200* 200* 080 100 080 000 200 064 100

31

Page 35

22 23 24 25 26

Ref. Decode NA NA Flag 000 001 Character Format NA NA Flag 000 001 Partial Decode NA NA Flag 000 001 Data Match NA NA Flag 000 001 Illegal Position NA NA Flag 000 001 *The data format for negative numbers ranging from –100 to –1 is 0 to 99. For example: −100=0; −95=5; −10=90; 0=100; 25=125; 95=195; 100=200

~PPpiiaaa Port action parameters. piiaaa = Six decimal digits representing: Port- p ID- ii (See Table of Action Parameters for description of ID) Value of action- aaa

Table of Action Parameters ID s 01 02 03

Description

Low Value

Pulse width (0.1s to 10.0s; > 10.0= latch) Active State high or low Set port on Good/Bad Evaluation

000 000 (active low) 000(Bad)

High Value 255 001 001(Good)

~PLpii List all Port setup parameters. (Parameters are set by the ~PB command.) pii= Three decimal digits representing: Port- p ID- ii ~PT# Show <Port 1 to 8> status and all the Port Parameters (Parameters are set by the ~PB command.) Transmission is in the following format for command ~PT1: 1) ~PT 2) hexadecimal value ‘4’ - identifies the beginning of the datapacket response. 3) A list of text strings like the one shown in the example. Each string is terminated by a two character sequence: ‘\r’ (the carriage return character), ‘\n’ (the line feed character). 4) hexadecimal value ‘5’ - identifies the end of the datapacket response 5) 1 Example return using Scanvision: Port Status: Global HW: 001 Global SW MIRR: 001 PORT STATE: 000 PASS/FAIL PARAMETERS: 001 - 255 255 002 - 255 255 003 - 255 255 004 - 255 255 005 - 255 255 006 - 255 255 007 - 255 255 008 - 255 255

32

Page 36

009 010 011 012 013 014 015 016 017 018 019 020 021 022 023 024 025 026

-

255 255 255 255 255 255 255 255 255 255 255 255 255 255 255 255 255 255

255 255 255 255 255 255 255 255 255 255 255 255 255 255 255 255 255 255

~PR#### Port/LED reset state #### = four hexadecimal digits Each port and led are represented by a binary digit in the four hexadecimal digits as follows: Indicator Digit value Power indicator 0x0001 Read LED 0x0002 LED1 0x0004 LED2 0x0008 Port 1 0x0010 Port 2 0x0020 Port 3 0x0040 Port 4 0x0080 Port 5 0x0100 * Sync Indicator 0x0200 * Inverted logic • Except for the sync indicator, if a bit is clear, the inactive state of the port is OFF. Example: ~PR0210 = sync indicator goes “ON” when a sync is detected and Port 1 goes “OFF” when activated. All other ports go “ON” when activated. ~Px Save the new port configuration by writing it onto the flash. ~PY Cancel the current port setup (restore the old Port configuration by getting the old values from the flash). ~PZpii List the parameter action. (Parameter actions are set by the ~PP command.) Port- p ID- ii

System Control Commands - S A ‘~’ followed by an ‘S’ specifies a command from the System Control category. These commands include one additional character that identifies the particular System Control Command. The System Control Commands are listed below.