The text of a manual · part 4 of 9
Laser Inspector Model L1000 Operator's Guide, pages 28 to 36
Pages 28 to 36 of 81
The original document is the PDF (324 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the RJS Inspector L1000
Page 28
Model L1000 Laser InspectorTM
Operator’s Guide
Table 2 (Identifier Descriptions for Pass/Fail Analysis Screens) Identifier B-1 B345 LOG HIBC 3 of 9 39+C I25 CC CC8 C128 UPCA UA+2 UA+5 UPCE UE+2 UE+5 EAN8 E8+2 E8+5 EN13 13+2 13+5 C93 CBRV
Symbology Type Code 39; B-1 sub-specification for AIAG Code 39; B3, 4 or 5 sub-specification for AIAG Code 39; LOGMARS sub-specification Code 39; HIBC sub-specification Code 39 Code 39; with Mod 43 Interleaved 2 of 5 Interleaved 2 of 5 Case Code sub-specification 128 Case Code (UCC 128, AI 00) Code 128 UPC version A UPC version A plus 2 digit addendum UPC version A plus 5 digit addendum UPC version E UPC version E plus 2 digit addendum UPC version E plus 5 digit addendum EAN 8 character EAN 8 plus 2 digit addendum EAN 8 plus 5 digit addendum EAN 13 character EAN 13 plus 2 digit addendum EAN 13 plus 5 digit addendum Code 93 Codabar Variant
RJS, Minneapolis, MN
22
002-5568
Page 29
Model L1000 Laser InspectorTM
Operator’s Guide
8.0 Data Analysis Screens After doing a scan, the Data Analysis Screens can be accessed from the Pass/Fail Analysis screen by pressing Enter. (The sequence of displays you see may vary slightly from these examples depending on the symbology and Inspector model.) The following displays are brought up by pressing Select. The first screen to display is the Decoded Symbol Data from the Pass/Fail Screen: If multiple scans were selected in Setup
*1234ABCDEF456* 01 Scans Remain.
(If more than 16 characters are present, those characters will wrap to the next row of the display.) Note: Code 128 is an exception. See Appendix A for details.
Press Select to bring up the next screen (if applicable): Modck: 5 PASS Expect: 5 This display indicates (when applicable) the Mod check character decoded and the Mod check character that should have been decoded (expected) in the symbol. PASS or FAIL is also indicated.
002-5568
23
RJS, Minneapolis, MN
Page 30
Model L1000 Laser InspectorTM
Operator’s Guide
For Code 128, the Mod check character is displayed as the symbology’s character value (a number from 000 to 102). One or two numeric Mod check characters may precede the 3digit Code 128 Mod check described above. These digits represent the Mod 10 check character(s) that might be included, depending on the format. Code 93 will display 2 Mod check characters. See Appendix B for Code 93 character representations. Press Select to bring up the next screen: AIAG B-1 Acceptable Symbology type is indicated on the top row and the analysis overview on the bottom row. Note: A database message screen may be displayed, if database is enabled. The screen displayed will be the first screen shown on the next analysis. This allows you to “program” the first screen to be displayed. If the database is disabled, only the Pass/Fail Analysis screen will display. If the database function is enabled, but the data scanned is not found, the following database message will display: “Data Base Record Does Not Exist” (See Appendix J Downloading the Database)
RJS, Minneapolis, MN
24
002-5568
Page 31
Model L1000 Laser InspectorTM
Operator’s Guide
Press Select to bring up the next screen: Ratio: 2.9 D/bility .81 A
This screen provides the ratio calculation result on the top row (if applicable) and the decodability calculation and grade on the bottom row. If Ratio is not appropriate (UPC, EAN, Code 93 or Code 128), the Ratio field will be omitted. Press Select to bring up the next screen. Bar width deviations are represented graphically on the display. -100 % Tol. +100% --------RRARR+++++
“R” letters represent “range” “A” represents the average of all character bar width deviations
The closer the “A” is to the center of the display, the better. Note: The location of the “A” is also indicated with the 5 LED indicators on the front of the verifier unit. See Table 3.
The following shows the percentage of the tolerance represented by each position on the display: 93
79
Negative numbers to the left indicate bars are narrower Positive numbers to the right indicate bars are wider
002-5568
65
50
36
22
8
25
1
0
+ 8
+ 22
+ 36
+ 50
+ 65
RJS, Minneapolis, MN
+ 79
+ 93
Page 32
Model L1000 Laser InspectorTM
Operator’s Guide
The closer the “A” is to the edge, the more likely a Warning or Reject message will display. A symbol is “Acceptable” until the calculated bar width deviation exceeds 100% of the tolerance for the symbology decoded; for example:
Acceptable -100% Tol. +100% -----RRRARR+++++ All bar widths are within tolerance
Warning -100% Tol. +100% --------------------RRRAR(R) Most deviant element exceeded tolerance
Rejected “A” is off the screen -100% Tol. +100% --------------------RRR(A) Average of all widths exceeded tolerance
RJS, Minneapolis, MN
26
002-5568
Page 33
Model L1000 Laser InspectorTM
Operator’s Guide
Press Select again to display the first technical Data Analysis screen.
Table 3 (LED, Barcode Width Deviations) 5 LED Display Pattern
Average Bar width deviation as a % of tolerance (range) within +/- 21% + 22 to +49% + 50 to +78% + 79 to +99% Out of tolerance (wide) -22 to -49% -50 to -78% -79 to - 99% Out of tolerance, narrow
Note: Solid indicators represent LEDs turned on The 9 possible patterns are displayed
002-5568
27
RJS, Minneapolis, MN
Page 34
Model L1000 Laser InspectorTM
Operator’s Guide
8.1 Continuous and Multi-scan modes In the Continuous Mode, an additional data analysis screen displays: # Decodes 013/019 % Decodes 068%
13 good decodes out of 19 attempted reads Percentage of good reads
There is a limit of 100 good scans in the Continuous Modeor 250 total scans - whichever comes first. When a limit is reached, the unit automatically stops scanning. In continuous mode the unit continues to attempt reads as long as the trigger is depressed. When using this mode the beam may be moved around the symbol to provide extra analysis. The analysis result is based on the average. When multi-scans are selected the number of remaining scans needed to complete the analysis is displayed: *0192837465* 03 Scans Remain After the last scan a double beep or chirp will signal completion of the scan. The resulting analysis is an average of the scans.
RJS, Minneapolis, MN
28
002-5568
Page 35
Model L1000 Laser InspectorTM
Operator’s Guide
9.0 Connect to other Devices Connect to Computer Your Laser Inspector will connect to a computer with the optional laser to PC database downloading cable. Connect to Printer Your Laser Inspector will connect to an RJS printer (models TP-140, TP140A, or TP140B) with a serial interface cable. (Your Inspector will connect to a computer or printer with a serial interface cable; 9600 BPS, 8 bit, no parity, 1 stop bit.)
Contact RJS for details: RJS Technologies 701 Decatur Ave North, Suite 107 Minneapolis, MN 55427 (763) 746-8034 Phone www.rjs1.com Website
002-5568
29
RJS, Minneapolis, MN
Page 36
Appendix A
Operator’s Guide
Appendix A (Code 128) Code 128 Code 128 can encode all the characters currently encodable in the various code formats presently in existence. This includes: all ASCII alphanumeric characters (numbers, letters, special characters, control characters in the 128 character set and the distinction of the 3 subsets, A, B and C). Code subset A includes the standard alphanumeric (upper case only) keyboard characters plus control and special characters. Code subset B includes all the standard alphanumeric keyboard characters and special characters (upper and lower case). Code subset C includes a set of 100 digit pairs from 00 to 99 inclusive, allowing definition of double density numeric digits per symbol, plus special characters. The last 7 characters in Subsets A and B (96 - 102) and the last 3 characters in Subset C (100 - 102) are special characters that are specific to the scanning device.
The Laser Inspector display can accommodate up to 16 characters. However, in order to display symbols greater than 16 (maximum of 32), press Select to display the additional characters. (See Data Analysis Screens). Table A-1 shows the encodable code subsets for 128. The Laser Inspector display (in the shaded boxes) is just below the encodable code 128 character sets.
RJS, Minneapolis, MN
30
002-5568