LOCAL FORMATTING

Select the appropriate terminal type by typing in the corresponding number. Press ENTER to confirm selection and move to the MAIN MENU. Terminal #1 is the default option. If choosing this terminal, just press ENTER to move to the MAIN MENU.

5-5

Page 42

Series 9000E Scanner/Decoder MAIN MENU

1 = GATE PARAMETERS 2 = FIELD DEFINITIONS 3 = HOST PORT PARAMETERS 4 = LOCAL PORT PARAMETERS 5 = INPUT/OUTPUT PULSE PARAMETERS 6 = SAVE/DEFAULT CONFIGURATION 7 = READ RATE DIAGNOSTICS 8 = COMPARE STRINGS x = EXIT TO READ MODE Enter selection: 1

MAIN MENU Purpose of Screen: This screen allows you to access eight setup and configuration screens. Field Descriptions: Menu items 1 through 8 provide access to detailed menus for setting the configuration parameters to the specific requirement of your system. If the system is being configured for the first time, the menus can be accessed in order. If changes are desired under a particular menu, that menu can be accessed directly. The X = EXIT TO READ MODE selection returns you to a blank screen to display the bar code currently being read. Operator Action:

5-6

Page 43

LOCAL FORMATTING Type the number of the selection you wish to make. The screen for the selection you make will be displayed. GATE CONFIGURATION LASER MODE OF OPERATION: Continuous LASER BEAM TRACKING: NO TRANSMIT AFTER: GATE CLOSES HOST START SCAN CHARACTER: <SI> HOST STOP SCAN CHARACTER: <ETX> ACK/NAK TIMEOUT VALUE: 0 msec (0 - 9999 msec for host port only) DATA CHANGE TIMEOUT VALUE: 0 msec (0 - 30000 msec) GATE INPUT DEBOUNCE TIME: 1 msec (1 - 999 msec for the host port only) GATE TIMEOUT VALUE: 0 msec (0 - 30000 msec) COMPARE MODE: NO (See Main Menu Option 8 for Setup) MASTER/SLAVE MODE: NO (Send Gate Open/Close message to the multiplexor)

KEYS: SPACE=TOGGLE+ BKSPACE=TOGGLE- <CTRL-U>=LAST SCREEN

PAGE 1 - GATE PARAMETERS Purpose of Screen: This screen allows the user to set the mode of operation. The user may select the method of signaling the Series 9000E Bar Code Scanner/Decoder to read a bar code and assure that only a single transmission is made for each bar code read. The user can also control when that transmission is made, and program the laser beam to follow the gate mode and turn on only when a bar code is present. Field Descriptions:

5-7

Page 44

Series 9000E Scanner/Decoder LASER MODE OF OPERATION Choices are: CONTINUOUS - The 9000E Scanner/Decoder reads, decodes and transmits continuously. Data for every read is transmitted. No gating is performed. DATA CHANGE MODE BUFFERED - Once a bar code is read and the data transmitted the 9000E Scanner/Decoder will not recognize that bar code again until a different bar code has been decoded. HARDWARE GATE MODE - Using an external device (as outlined on Page 4-8) the scanner is gated to read when a bar code is present. Only a single data transmission is made for each gate signal. SOFTWARE GATE MODE - The Host computer may be programmed to gate the 9000E Scanner/Decoder via an ASCII start and stop signal. HIGH SPEED MODE - Same as hardware gate mode but only outputs are triggered. The label is not transmitted. DATA CHANGE MODE WITH TIMEOUT - Allows the user to specify a TIMEOUT value before 9000E Scanner/Decoder will read the next bar code. LASER BEAM TRACKING - Selecting NO will leave the laser beam on at all times. Selecting YES will turn the laser beam on when a gate signal is received and turn the beam off when the gate is closed. This will insure that the beam is on only when necessary. TRANSMIT AFTER - The two choices available are CLOSE and READ. Selecting READ will cause the 9000E Scanner/Decoder to transmit the bar code data immediately after a successful read. Selecting CLOSE will cause the data transmission to be delayed until the close of the gate signal. HOST START SCAN CHARACTER - The host computer may be used to gate the 9000E Scanner/Decoder . The start of each gate must be 5-8

Page 45

LOCAL FORMATTING specified as a character or string of characters. Any two character string can be specified but normally ASCII control characters are used. The default character string, SI (Shift In) has been programmed. It may be changed by the user if desired. When these characters are used they must be bracketed by < >. HOST STOP SCAN CHARACTER - The host computer character to end the gate period can also be specified as a character or string of characters. The default string, <ETX> has been programmed. It may be changed to any two character ASCII string if desired. ACK/NAK TIMEOUT VALUE - If ACK/NAK software protocol is selected, each time a label is read it is sent to the Host for confirmation. If the host answers ACK (CTRL-F), the scanner can then immediately read the next label. If the host answers NAK (CTRL-U) or gives no response in the allotted time then the scanner will resend the label. The scanner will resend the label to the host up to 5 times. DATA CHANGE TIMEOUT VALUE - The data change mode can be modified to recognize the same code again after a set period of time. In Data Change timeout, once the label leaves the field, a timer is started. If at the end of the set time the same bar code is presented to the scanner again, it will be considered a new code. Values specified can range from 0 to 30000 msec. GATE INPUT DEBOUNCE TIME - If a switch or relay contact is used for a gate signal, a debounce time may be necessary. Times will have to be determined experimentally. Once a gate signal is detected, it must be present for the set time before it is considered valid. Debounce time may be specified from 1 to 999 msec. GATE TIMEOUT VALUE - The Gate Timeout Value is used to simulate a close gate signal. When the time set has expired the gate will close and either transmit the bar code data or a No Read. The timer will ignore all activity on the Gate line until it has timed out. When a new open gate signal happens the timer starts counting down again. COMPARE MODE - Activates the Compare String mode. (See Page 8 Compare Strings) for information on compare string options. 5-9

Page 46

Series 9000E Scanner/Decoder MASTER/SLAVE MODE - This option will send a multiplexor open message, scanner label data and then a multiplexor close message. The open message consists of STX SI CAN CR. The close message consists of STX SO CAN CR.

Operator Action: 1.

When the cursor is positioned on the LASER MODE OF OPERATION line, press SPACE to display the desired setting. Press ENTER to select the choice and move the cursor to the next line.

2.

Position the cursor on the LASER BEAM TRACKING line. Press SPACE to change the settings from YES to NO. When the desired setting is displayed press ENTER to enter the choice and move the cursor to the next line.

3.

With the cursor positioned on the TRANSMIT AFTER line press SPACE to change the settings from GATE CLOSES to READ and back. When the desired setting is displayed, press ENTER to enter the choice and move the cursor to the next line.

4.

Position the cursor at the end of the HOST START SCAN CHARACTER line. To accept the currently selected character string press ENTER. To change the character press BACK SPACE to delete the present string and enter the new character string. Press ENTER to proceed to the next line. Position the cursor at the end of the HOST STOP SCAN CHARACTER line. To accept the currently selected character string press ENTER. To change the character press BACK SPACE to delete the current string and enter the new character string. Press ENTER to proceed to the next line.

5.

6.

With the cursor on the ACK/NAK TIMEOUT VALUE line enter the desired value in milliseconds. This value can be between 0 and 9999 milliseconds. Press ENTER to enter the value and move to the next line.

7.

With the cursor on the GATE CONTROL DEBOUNCE VALUE line, enter the desired value in milliseconds. 5-10

Page 47

LOCAL FORMATTING

8.

With the cursor on the GATE TIMEOUT VALUE line, enter the desired value in milliseconds.

9.

With the cursor on the COMPARE MODE line use the SPACE BAR to toggle between the choices; YES and NO. Press ENTER to select.

10. With the cursor on the MASTER/SLAVE MODE line use the SPACE BAR to toggle between the choices; YES and NO. Press ENTER to select.

5-11

Page 48

Series 9000E Scanner/Decoder FIELD DEFINITION SUMMARY SCREEN

NO. CODE TYPES 01 C_39

LENGTHS 0

SELECT ACTION:

KEYS: SPACE=TOGGLE+ BKSPACE=TOGGLE- <CTRL-U>=LAST SCREEN

PAGE 2 - FIELD DEFINITION SUMMARY Purpose of Screen: Displays a summary of how each previously entered field is defined. Also lets the user edit, delete or insert fields. Field Descriptions: A total of up to six fields can be listed. Each listing is a summary of how a field is defined. Existing fields can be edited or deleted. New Fields can be inserted. Note:

Think of a field as an encapsulated object that grabs things. the entries made in the Definition for Field Number screen (Page 2, level 2) are simply a set of filters the field uses to filter out the things it does not want to grab. Each field allows up to 10 code 5-12

Page 49

LOCAL FORMATTING type/length combinations (i.e., ten types of code each with their own number of characters). More than one field turned on means that the user is looking for more than one label on a package.

Operator Actions: Choose EDIT, DELETE, or INSERT by typing in E, D, or I at the SELECT ACTION prompt. If DELETE is chosen, the ENTER FIELD NUMBER TO DELETE: prompt will appear. Enter the Field number 1 through 6. The entire Field will be deleted from the display. If a Field is to be edited or inserted, enter the number of the Field at the prompt, and the next screen will be displayed. DEFINITION FOR FIELD NO: 01

CODE TYPE C-39 OFF OFF OFF OFF OFF OFF OFF OFF OFF

MIT S/S

XMIT CHK DGT

NO NO NO NO NO NO NO NO NO NO

NO NO NO NO NO NO NO NO NO NO

LENGTH 0 0 0 0 0 0 0 0 0 0

HOST GOOD READ XMIT FORMAT: LOCAL GOOD READ XMIT FORMAT: HOST NO READ XMIT FORMAT: LOCAL NO READ XMIT FORMAT: HOST READ CONFLICT XMIT FORMAT: LOCAL READ CONFLICT XMIT FORMAT: LABEL MASK:

READ CNTS 1 1 1 1 1 1 1 1 1 1

%D% %D% NR NR %D% %D% NO

KEYS: SPACE=TOGGLE+ BKSPACE=TOGGLE- <CTRL-U>=LAST SCREEN

5-13

Page 50

Series 9000E Scanner/Decoder PAGE 2 - LEVEL 2 FIELD DEFINITION Purpose of Screen: Allows the user to edit existing entries and insert new entries as well as specify data that will be transmitted along with the bar code data. Field Descriptions: CODE TYPE - Any data previously entered will be displayed. If none, all the entries in this column will read OFF. XMIT S/S - (Transmit Start/Stop). This controls transmission of Start and Stop characters with the bar code data. Not all bar codes have start/stop characters. Not all start/stop characters are transmittable. Length (number of characters) in field definition must be set long enough to include start/stop characters if XMIT S/S is enabled. XMIT CHK DGT - (Transmit Check Digit). It may or may not be desirable to transmit the check digits contained in some bar codes. Not all bar codes have check digits. Not all check digits are transmittable. Length (number of characters) in field definition must be set long enough to include check digits, if enabled. LENGTH - Determines the number of characters in the code to be read, including any check characters and start/stop characters if enabled. UPC and EAN codes may not have the first and last digits printed. A 0 allows any length of code to be read. READ COUNT - Number of times the same label must be decoded before it’s transmitted. HOST GOOD READ XMIT FORMAT - This entry determines what data is transmitted from the Host Port when a Good Read occurs. The entry can be a literal string such as GR, a formatted data string or a combination of the two. See Data Formatting Codes page 5-17. LOCAL GOOD READ XMIT FORMAT - This entry determines what data is transmitted from the Local Port when a Good Read occurs. The entry

5-14