Page 35
X p l o r e r
G L X
U s e r s ’
G u i d e
29
If the chosen data source does not contain data, the data source name will appear at the top of the column, but the cells of the column will remain empty until data collection starts. Also from the data source menu, you can select Properties to edit the name and other properties of the currently displayed data set. See “Data Properties” on page 69 for more information. For more information on selecting data from a data source menu when you are working with more than one sensor, or with a multiple-measurement sensor, see “Data Source Menus” on page 89.
Units Select the units field21 to choose different units (if available) for the chosen data source. Run Number Select the run number field21 to choose a different run, or select No Data to display nothing. Units menu
Run number menu
By default, the Table is set to change the displayed run number in all columns whenever you change it in one column. Refer to “Lock Run Selection” on page 31 to change that setting.
The last two options in the run number menu are Delete Run and Rename Run. See “Run Number” on page 15 for more information.
Table Scrolling At any time, up to eight cells of data are visible in each column. Press the up or down arrow key to scroll the Table and make other rows of data visible. Press and hold the up or down arrow key to make the Table scroll continuously. Press the left arrow key to jump to the beginning of the table; press the right arrow key to jump to the end.
Scroll up Jump to the top
Jump to the bottom Scroll down
The rules for scrolling are different when an editable column is visible. See page 35
Table Function Keys F1
F2
F3
In the Table, the function keys are used to control the display of statistics, to edit data, and to set the number of columns.
F1 Statistics From the Statistics menu, you can choose to display any of the following statistics for the displayed data sets: •
Minimum (Min), the lowest value in the data set,
•
Maximum (Max), the highest value in the data set,
•
Average (Avg), the average of all values in the data set,
•
Standard Deviation (σ), a measure of how much the data varies from the average, and
•
Count (#), the number of values in the data set.
Statistics menu
F4
Page 36
30
T a b l e
The chosen statistics will be displayed at the bottom of every column. To choose a statistic, select it from the menu.22 To turn off a displayed statistic, select it again. Select Show All or Hide All to turn on or off all of the statistics.
22To select a statistic from the Statis- tics menu:
Keypad 1. Press F1 to open the Statistics menu. 2. Use the arrow keys to move the highlight to the desired statistic and press ; or press the number on the keypad corresponding to the desired statistic. Mouse 1. Click “Statistics” at the bottom of the screen to open the Statistics menu. 2. Click the desired statistic.
Table with statistics displayed
F2 Edit Cell When you are navigating in an editable column, press F2 to edit the selected cell. See “Manually Entered Data” on page 32 for instructions on editing data.
F3 Edit The options in the Edit menu are used when editing or entering data. See “Manually Entered Data” on page 32 for more information. New Data Column Select New Data Column from the Edit menu to create an editable data set. See “Manually Entered Data” on page 32 for more information. Insert Cell Insert Cell works only when the dashed Navigating Box is visible23 (see page 34). When you select Insert Cell, a new cell is added directly above the Navigating Box. If the column contains numeric data, the new cell will contain the number 0. If the column contains text data, the new cell will be blank. Every cell below the new cell shifts down by one position. Delete Cell Delete Cell works only when the dashed Navigating Box is visible23 (see page 34). When you select Delete Cell, the cell that the Navigating Box is on is deleted. Every cell below the deleted cell shifts up by one position. Num Lock Num Lock is only available in the Edit menu when an editable column is displayed. When Num Lock is turned on (which it is by default), a check mark ( ) appears next to it in the menu, and the keypad operates in numbersonly mode. When Num Lock is turned off, the keypad can be used in multipress text input mode (see page 90) to enter numbers and letters. To turn Num Lock on or off, select it from the Edit menu. Caps Lock Caps Lock is only available in the Edit menu when an editable column is displayed. When Caps Lock is turned on, a check mark ( ) appears next to it in the menu. When used in multipress text input mode, the keypad inputs capital letters.
Edit menu
23To make the Navigating Box appear in an editable column, press
Esc
.
Page 37
X p l o r e r
G L X
U s e r s ’
G u i d e
31
F4 Tables
F4
Tables Menu
From the Tables menu, you can select to display one, two, three, or four columns.24 If you increase the number of columns, new columns are added to the right side (the GLX automatically selects data sources, if available, for new columns). If you decrease the number of columns, they will be removed from the right side (the data displayed in those columns disappears, but is not be permanently deleted).
24To select an option from the Tables menu
Keypad 1. Press F4 to open the Tables menu. 2. Use the arrow keys to move the highlight to the desired menu option and press ; or press the number on the keypad corresponding to the desired menu option. Mouse 1. Click “Tables” at the bottom of the screen to open the Tables menu. 2. Click the desired menu option.
Tables with one, two, three, and four columns
Other options in the Tables menu are described below. 25
Show Time Select Show Time to add a column of time data in the left-most position.25 If there are four columns displayed when you select Show Time, the fourth column will be removed and the other three columns will shift to the right to make room for the time column on the left.
To show time in a column other than the first one, select it from the data source menu at the top of the desired column. See “Data Source” on page 28 for more information.
Date/Time To turn on this option, select it from the Tables menu. With the option turned on, times displayed in the Table appear as the date and time of day at which each data point was recorded. With the Date/Time option turned off, time data appears as measured from the start of data collection. The Date/Time format is set in the Settings screen. See page 85.
Lock Run Selection When Lock Run Selection is on (which it is by default), all data columns will display the same run number, and changing the run number
Table with Date/Time option
Page 38
32
T a b l e of one column will cause all of them to change.26 To turn Lock Run Selection off or on, select it from the Tables menu. Export All Data If you have a USB storage device27 (such as a USB flash drive), you can use this option to export data in tab-delimited text format. The GLX creates a text file on the storage device that you can open on a computer in a spreadsheet, mapping software, DataStudio, and other programs. See page 83 for other ways to save and open data on a USB storage device.
Connect the storage device to the GLX’s USB port and select Export All Data from the Tables menu. A dialog box appears showing which data sets will be exported. Highlight any of these data sets and press to exclude it. To select other data sources for exporting press F3 to add a single data source or press F4 to add all data sources that have been set to “visible” in the Sensors screen. Press F1 (OK) to create the export file on the USB storage device.28 Before exporting data for use with mapping software, display time in the first column and turn on the Date/Time option (see page 31).
To view the exported data, disconnect the storage device from the GLX and connect it to a computer. You will find the text file at the root level of the storage device. Its filename encodes the date and time at which the data was exported; for instance, if you exported data on October 28, 2005 at 11:22:53 AM, the filename will be “Export10282005112253.txt”. You can open this file in a text editor, spreadsheet, or other program designed to import tab-delimited data.29
26
See “Run Number” on page 29 for instructions on changing the run number displayed in a column. 27
The USB storage device must be formatted using the FAT16 or FAT32 file system. Most flash drives are formatted this way. Other mass-storage devices such as portable hard drives and audio players may also be compatible. You can use a USB card reader, however the card must be in the reader before you connect the reader to the GLX.
28
If you export data while data collection is in progress, the text file will contain the data collected up to that point, and the GLX will continue to collect data.
29To import the file into DataStudio, open the File menu and select Import Data.
Print If there is a printer connected to the GLX, select Print to print one page of data. The GLX will print the data that are visible on the screen, plus as much additional data as will fit on one page.30
30For more information on printing, see
“Printing” on page 91.
Print All Select Print All to print the entire contents of the displayed data sets, including data that are not currently visible on the screen (data that you would have to scroll up or down to see). If necessary, the GLX will print multiple pages.
Manually Entered Data In addition to collecting data from sensors, the GLX allows you to enter data in the Table display via the keypad or an attached keyboard. Any data that you have entered can later be edited.
Entering a New Data Set Creating a New Data Column Before you can enter data, an editable data column must exist. To create one, press F3 to open the Edit menu and select New Data Column. The keypad shortcut for creating a new data column is sequence.
F3
, 1 , pressed in pqrs
The new editable column takes the place of the first blank column in the Table. If there is no blank column, and there are three or fewer other columns, the new editable column is added on the right side of the table. If the Table is already displaying four columns, the new editable column replaces the fourth column.
Create an editable data column
F3
Page 39
X p l o r e r
G L X
U s e r s ’
G u i d e
Entering Numeric Data In a New Column In the new editable column, you will see a box around the first cell and a blinking cursor inside the cell. 1.
Type a number on the keypad (or attached keyboard), then press When editing data in the Table,
= has the same effect as
33
Box with blinking cursor
.
.
2.
The box and cursor will step to the next cell in the column. Enter another value and repeat.
3.
When you are finished, press Esc . (You must press value before you press Esc .)
to lock in the last
If you would like to change or add to the data after you have pressed refer to “Editing an Existing Data Set” on page 34.
Esc
,
Entering Text Data in a New Column In addition to numeric data, the Table allows you to create a column of text. 1.
Create a new data column as described on page 32.
2.
Press F3 to open the Edit menu. If Num Lock is turned on, select it to turn it off and enable multipress text input mode. (If you are using an attached keyboard, this step is unnecessary.)
Press 4 or click here to turn off Num Lock ghi
F3
For more information, see “Num Lock” on page 30 and “Multipress Text Input Mode” on page 90.
3.
Type a non-numeric string of text and press
.
The non-numeric text string can contain numerals, but it must have at least one letter or punctuation mark, so the GLX knows that it is supposed to be text. This limitation applies to only one cell in the column. After non-numeric text has been entered in one cell, the GLX will treat everything in that column as text, even if it looks like a number.
4.
A dialog box will appear, reading: “Treat this column as text data?” Press F1 to select OK. The column is now configured to contain text data. If you enter a numeric value it will be treated as text only. You can recognize a text-only data set by its left-justification (contents appear on the left side of the column) as opposed to numeric data, which is always right-justified.
5.
Continue entering text in subsequent cells. Press after you type in each cell to lock in that text string and step to the next cell. You can press on a blank cell to leave it empty.
6.
When you are finished entering data, press Esc . (You must press in the last value before you press Esc .) If you would like to change or add to the text data after you have pressed Esc , refer to “Editing an Existing Data Set” on page 34.
F1
F2 to lock Text data left-justified
Numeric data right-justified
Page 40
34
T a b l e
Naming Manually Entered Data and Adding Units By default, new editable data sets are named “Data 1,” “Data 2,” etc., and they do not have defined units. To change the name and add units, open the data source menu and select Data Properties.31
31To open the data source menu and select Data Properties For more information, see “Data Source” on page 28, and “Data Properties” on page 69.
Editing an Existing Data Set The above section, “Entering a New Data Set” starting on page 32, describes how to create and enter data in a new column. This section describes how to change and add to a manually entered data set that has previously been created.
Whenever you have one or more editable columns displayed in the Table, one of the editable cells is surrounded by a box. The box is either an Editing Box (solid lines) or a Navigating Box (dashed lines). Editing Box The Editing Box32 is drawn with solid lines, and always contains a blinking cursor. It may be on a cell that already contains data, or it may be on an empty cell.
Keypad 1. If the blinking cursor is visible, press Esc . 2. Press to light up the active fields. 3. Use the arrow keys to move the highlight to the data source field. 4. Press to open the menu. 5. Use the arrow keys to highlight Data Properties and press ; or press the number on the keypad corresponding to Data Properties. Mouse 1. Click the data source field to open the menu. 2. Click Data Properties.
32If you do not see the Editing Box, press
F2
.
To move the Editing Box to a different cell, refer to “Navigating Box” below.
If the Editing Box is on an empty cell, anything that you type on the keypad (or attached keyboard) will appear in that cell. When the Editing Box is on a cell that already contains data, what you type will either replace or amend the contents of the cell, depending on whether the contents are highlighted. •
•
If the cell contents are highlighted, they will be replaced by any number or text that you type. If you press Õ , the entire contents of the cell will be deleted.
Editing Box on an empty cell
If the cell contents are not highlighted, they will be amended at the location of the blinking cursor. If you press Õ , the character to the left of the cursor will be deleted.
By default, when the Editing Box appears on a cell, the contents (if any) are highlighted. To remove the highlight, press the left or right arrow key, then use the left or right arrow key to move the blinking cursor to the position where you would like to make an edit.
The existing contents will be replaced
The existing contents will be amended
After you have changed the cell contents, you must press to lock in the change. If you want to cancel the change, or leave the cell without entering data, press Esc instead of . If the contents of a cell are not highlighted, you can restore the highlight by pressing F2 . Navigating Box The Navigating Box33 is drawn with dashed lines. Use the up and down arrow keys to move it within an editable column. Use the left and right arrow keys to move it to other editable columns.
33If an editable column exists, but you do not see the Navigating Box, press
Esc
.
Page 41
X p l o r e r
G L X
U s e r s ’
G u i d e
35
Use the arrow keys to move the Navigating Box
Position the Navigating Box on a cell where you would like to enter or edit data, then press F2 to make the Editing Box appear. To restore the Navigating Box, press Esc . If you would like to add data to the bottom of a data set, position the Navigating Box on the first empty cell below the existing data and press F2 .
Press F2 to edit this cell
Press F2 to add to this data set
Table Scrolling with Manually Entered Data At any time, up to eight cells of data are visible in a manually entered data column. As with regular Table scrolling (see page 29), use the up or down arrow key to scroll the Table up or down and make other rows of data visible. To scroll the Table when manually entered data column is displayed, the dashed Navigating Box must be visible.34 To scroll up, press the up arrow key repeatedly (or press and hold it) to move the Navigating Box to the first visible cell, then continue to press (or hold) the key. To scroll down, press the down arrow key repeatedly (or press and hold it) to move the Navigating Box to the last visible cell, then continue to press (or hold) the key.
Press to scroll up
Press to scroll down
All columns scroll together. The Table will not scroll past the bottom of the column containing the Navigating Box. In the table picture below, the Navigating Box is at the end of the shorter data column. To scroll the Table down, press the left arrow key to place the Navigating Box on the longer manually entered column.
34If the Navigating Box is not visible, press Esc .
Page 42
36
T a b l e
Graphing Manually Entered Data Numeric Data After you have created an editable numeric data set in the Table, you can display it in the Graph. In the Graph, select the editable data from the horizontal or vertical data source menu (see page 15). Text Data A text data set can be displayed on the horizontal axis of the Graph. First select a numeric data as the vertical data source, then select the text data as the horizontal data source (see page 15). The text from each cell of the Table appears on the Graph as a label next to the corresponding data point.
Graph displaying manually entered text data on the horizontal axis
Page 43
X p l o r e r
G L X
U s e r s ’
G u i d e
37
Di gits D isp lay Use the Digits display to show data from up to eight data sources. This display monitors and shows live data even when the GLX is not recording. Data Source
Units
Digits display
To Open the Digits Display From the Home Screen, do one of the following: use the arrow keys to highlight the Digits icon, then press
; or click the Digits icon.
Choosing Data to Display Data Source Select the data source field35 at the top of each section to select a data source. A data source would usually be a sensor measurement or a calculation based on a sensor measurement. Sensor measurements or calculations based on sensor measurements can vary while data collection is in progress. If you select a static data source (such as manually entered data), the Digits display will show only the last data point.
For more information on selecting data from a data source menu, see “Data Source Menus” on page 89.
Also select the data source field to access the Data Properties of the displayed data. See “Data Properties” on page 69 for more information. Units Select the units field35 at the top of each display to select different units for the selected data source (if available).
Digits Function Keys Press F1 , F2 , F3 , or F4 to divide the display into two, four, six, or eight segments. The GLX automatically selects data sources, if available, for new segments.
The Digits icon on the Home Screen
35To select a data source field
Keypad 1. Press to light up the active fields. 2. Use the arrow keys to move the highlight to the data source or units field. 3. Press again to open the menu. 4. Use the arrow keys to highlight the desired menu option and press ; or press the number on the keypad corresponding to the desired menu option. Mouse 1. Click the data source or units field to open the menu. 2. Click the desired menu option.
Page 44
38
M e t e r
Mete r The Meter display simulates an analog meter with a needle that deflects in proportion to a sensor measurement or sensor-based calculation. This display monitors and shows live data even when the GLX is not recording. Data Source
Units
Meter display
To Open the Meter Display From the Home Screen, do one of the following: use the arrow keys to highlight the Meter icon, then press
; or click the Meter icon.
Choosing Data to Display Data Source Select the data source field36 at the top of the Meter display to select a data source. A data source would usually be a sensor measurement or a calculation based on a sensor measurement. Sensor measurements or calculations based on sensor measurements can vary while data collection is in progress. If you select a static data source (such as manually entered data), the Meter display will show only the last data point.
For more information on selecting data from the a data source menu, see “Data Source Menus” on page 89.
Units Select the units field36 at the top of the display to select different units for the selected data source (if available).
Auto Scale and Reset Press F1 to automatically scale the Meter. If the GLX is recording data, the lower and upper limits of the meter will be reset to equal the lowest and highest values measured thus far in the current or most recent data run. Press F2 to restore the default scale.
Print Press
F4 to print the Meter. See page 91 for more information.
The Meter icon on the Home Screen
36To select a data source or units field
Keypad 1. Press to light up the active fields. 2. Use the arrow keys to move the highlight to the desired field. 3. Press again to open the menu. 4. Use the arrow keys to highlight the desired menu option and press ; or press the number on the keypad corresponding to the desired menu option. Mouse 1. Click the desired field to open the menu. 2. Click the desired menu option.