For the Omron V520

Page 28

Section 6 - Scanner Servicing and Maintenance

V520-LHA7127

Section 6 Scanner Servicing and Maintenance The V520-LHA7127 scanner contains no user adjustable or serviceable parts in the interior of the scanner. All product service must be performed by the Omron Service Department. Opening the scanner will void the warranty and could expose the operator to LED light. The LED’s are classified as a Class 1 LED Product per IEC 825-1 (1997). !Caution Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous LED exposure. !Caution Class 1 LED power up to 15 mW in a 0.1ms pulse at 635-670 nm could be accessible in the interior. The V520-LHA7127 is warranted for 5 years including parts and workmanship. If you need warranty or out-ofwarranty repair first call (847) 843-7900 to obtain a Returned Material Authorization (RMA) number. You will be provided a number and shipping instructions. There is no scheduled maintenance required for the V520-LHA7127. The scanner can be cleaned using a water dampened, lint free or lens cloth. Be careful to avoid excessive moisture that would penetrate the housing or obscure the window. While use of cleaning fluids other than water are not recommended, a neutral detergent or ehtanol would be preferred if necessary. Do not use bleach at full or diluted strength as damage to the painted case and/or window may result.

19

Page 29 read from a scan by OCR; it may hold reading errors

i 20-LHA7127 Section 6 - Scanner Servicing and Maintenance V5

20

Page 30

Appendix A - Programming Menus & Commands

V520-LHA7127

Appendix A Programming Menus & Commands Index 1. Global Default & Scanner Configurations ............................................................................ 22 2. Symbology Selection............................................................................................................ 26 3. Add/Activate Specific Bar Code Symbologies...................................................................... 27 4. Delete/Deactivate Specific Bar Code Symbologies ............................................................. 29 5. Options for UPC ................................................................................................................... 30 6. Options for Code 39 ............................................................................................................. 31 7. Options for Codabar (NW-7) ................................................................................................ 32 8. Options for 2 of 5.................................................................................................................. 33 9. Options for IATA ................................................................................................................... 34 10. Fixing the Number of Digits.................................................................................................. 35 11. Creating a Prefix and/or Suffix ............................................................................................. 36 12. Setting Prefixes (identifying the symbology) ........................................................................ 37 13. Setting Suffixes (identifying the symbology)......................................................................... 38 14. Direct Input of Numeric Characters...................................................................................... 40 15. Direct Input of Alpha Characters .......................................................................................... 41 16. Direct Input of Control Characters........................................................................................ 42 17. Enable Number of Characters/Digits Transmitted ................................................................ 42 18. Disable Number of Characters/Digits Transmitted ............................................................... 44 19. Communication Parameters................................................................................................. 45 20. Handshaking ........................................................................................................................ 47 21. Buzzer Operation ................................................................................................................. 48 22. Positive & Negative Bar Codes ............................................................................................ 50 23. Setting the Trigger Function ................................................................................................. 50 24. Selecting the Read Mode..................................................................................................... 51 25. Scanner Timeout (trigger enabled mode only)..................................................................... 53 26. Redundant Decoding ........................................................................................................... 54 27. Transmitting Error Indications (trigger enabled only)............................................................ 55 28. Print Quality Adjustments..................................................................................................... 55 29. LED Operation ..................................................................................................................... 56 30. Sequencer Controls ............................................................................................................. 57

21

Page 31

Appendix A - Programming Menus & Commands

V520-LHA7127

1. Global Default & Scanner Configurations Computer Command

Function/Description

Bar CodeCommand

(Z7)

START/END Programming Menu

1Y

Clear all prefixes

1Z

Clear all suffixes

P6

Disable Graph Mode

P9

Enable Graph Mode

U0

Scanner Test Mode

U1

Return all parameters to Default settings

U3

Straight Across Scan Mode

U4

Label Verification Mode

U5

Label Registration Mode

U8

Read Rate Mode

U9

Graph Mode (TBD)

Z1

Reset all values set by command

Z2

Save all values set by command (not lost upon power off)

Z3

Display software settings and version number

Z4

Display prefix/suffix value length (expressed in

-[8-2Z-2[-Q7-Q:-V1-V2-V!-V5-V6-V9-V:-[2-[3-[!-[5-

Note Where computer commands appear in parenthesis, bar code commands must be used to program the scanner. Scanner default settings are indicated by a pointing hand ( ) symbol throughout the menus to follow.

%

Use Z2 (Save all values set by command) after computer command to store settings in memory prior to power off.

22

Page 32

Appendix A - Programming Menus & Commands

V520-LHA7127

U1 Command (U1) The U1 command resets all parameters for an established set of defaults defined as follows: • Auto-discriminates 4 (not all) symbologies: WPC (UPC, EAN, JAN), Interleaved 2 of 5, NW-7 (Codabar) and Code 39. • No fixed number of digits defined (32 digits maximum). • Normal scan, multiple scan. • Trigger enabled. • Right contrast. • Small and large. • No verification, no “Bad Read (BR).” • Buzzer 3KHz and 2KHz, volume maximum. • Buzzer after decoding (200 ms). • Good LED after decoding (green 200 ms). • RS-232C (9600 bps, 8, 1, np, RTS/CTS, no control sequence). Preprogrammed Unit The unit is preprogrammed out-of-the-box with additional commands beyond those specified in the “U1” default. To restore the operation of the unit to its “out-of-the-box” configuration, should it be necessary, the following chain of commands is recommended: <ESC> U1 <CR>

(Establishes U1 defaults)

<ESC> P0 <CR>

(No handshake)

<ESC> S7 <CR>

(LED ON - Trigger disabled)

<ESC> B5 <CR>

(Enable Code 93)

<ESC> B6 <CR>

(Enable Code 128)

<ESC> Z2 <CR>

(Save values in non-volatile memory)

23

Page 33

Appendix A - Programming Menus & Commands

V520-LHA7127

Scanner Test Mode (U0) This mode resets all unit parameters to a predefined set of defaults for the purpose of putting the scanner into a test mode which reads barcodes continuously. The predefined defaults are the following: • Reads all barcodes by symbology • No fixed number of digits defined (32 digits minimum). • Normal scan, continuous scan. • Large only. • No verification, no “Bad Read (BR).” • Buzzer 3KHz, volume maximum. • Buzzer after decoding (10 ms). • Good LED after decoding (green, 100 ms). • RS-232C (9600 baud, 8, 1, no parity, no handshaking). Straight Across Scan Mode (U3) This mode resets all unit parameters to predefined defaults for the purpose of allowing straight across scanning of up to 4 labels. The predefined defaults are the following: • Reads all barcodes by symbology. • No fixed number of digits defined (32 digits max.) • Normal scan, multiple scan. • Trigger disable. • Small and large. • No verification, no “Bad Read (BR).” • Buzzer 3KHz and 2KHz, volume max. • Buzzer after decoding (200 ms). • Good LED after decoding (green, 200 ms). • RS-232C (9600 baud, 8, 1, no parity, np, no handshaking). Label Verification Mode (U4) This mode resets all unit parameters to a predefined set of defaults for the purpose of verifying barcodes against a set of pre-registered labels (up to 5 labels).With the trigger set to use the red LED lights when there is a bad read and stays on for 200 ms. The unit determines good or bad read based upon the registration of labels input during the use of the “U5” command mode defined below. It reads only the registered labels as normal. When it encounters an unregistered label, the red LED lights and sounds a “bad” buzzer. The predefined defaults are the following: • Reads all barcodes by symbology. • No fixed number of digits defined (32 digits max.). • Normal scan, multiple scan. • Trigger enabled. Read timing is coupled with trigger. • Small and large. • No verification, no “Bad Read.” • Buzzer and 3KHz and 2 KHz, volume max. • Buzzer after decoding (200 ms). • Good LED after decoding (green, 200 ms). • RS-232C (9600 bps, 8, 1, np, no handshaking).

24

Page 34

Appendix A - Programming Menus & Commands

V520-LHA7127

Label Registration Mode (U5) This mode resets all unit parameters to a predefined set of defaults for the purpose of registering up to 5 labels prior to using the verification mode defined previously. When this “U5” command is activated, an intermittent (fast) buzzer sounds. Scan the labels you want to register (up to 5) and when done scan the Z7 label. The predefined defaults are the following: • Reads all barcodes by symbology. • No fixed number of digits defined (32 digits max.). • Mass scan, multiple scan. Trigger disabled. Read timing is coupled with trigger. • Small and large. • No verification. No “Bad Read (BR).” • Buzzer 8 KHz and 2KHz, volume maximum. • Buzzer after decoding (200 ms). • RS-232C (9600 bps, 8, 1, np, no handshaking). Read Rate Mode (U8) This mode resets all unit parameters to a predefined set of defaults for the purpose of checking the scanner read rate. When activated, the test result is sent via RS-232C and indicates the ratio of the number of successful reads to the total number of attempts. The test result is reported in the following format: 700 d OK: 93.5% *CODE-39 TEST* Where: • 700d indicates the number of times the decoder ran while scanning at 700 scans per second (the number can be lower the 700 with poor quality barcodes). • 93.5% indicates the percentage of good reads to total attempts. • The control character to be output is CR (0DE carriage return line fed), “ESC[3A” (cursor 3 line UP).

The predefined defaults are the following: • Reads all barcodes by symbology. • No fixed number of digits defined (32 digits max.). • Normal scan, multiple scan. • Trigger disabled. • Small and large. • Buzzer 3 KHz, volume max. • Buzzer after decoding (10 ms.) • Good LED after decoding (green 100 ms) • RS-232C (9600 bps, 8, 1, np, no handshaking).

25

Page 35

Appendix A - Programming Menus & Commands

V520-LHA7127

2. Symbology Selection Computer Command

Function/Description

Bar Code Command

(Z7)

START/END Programming Menu

A0

Enable all bar code symbologies

A1

Only WPC (including Addon)

A2

Only Code 39

A3

Only Codabar (NW-7)

A4

Only 2 of 5 (Industrial or Interleaved)

A5

Only Code 93

A6

Only Code 128

A7

Only MSI/Plessey

A8

Only IATA

J1

Only UPC (A/E)

J2

Only UPC (A/E)+2

J3

Only UPC (A/E)+5

J4

Only EAN

J5

Only EAN +2

-[8-B1-B2-B3-B!-B5-B6-B7-B8-B9-K2-K3-K!-K5-K6Continued on next page.....

26

Page 36

Appendix A - Programming Menus & Commands

V520-LHA7127

Symbology Selection (cont.) Computer Command

Function/Description

Bar Code Command

(Z7)

START/END Programming Menu

J6

Only EAN +5

J7

Only Industrial 2 of 5

J8

Only Interleaved 2 of 5

Z9

Remote Menu

-[8-K7-K8-K9-[:-

3. Add/Activate Specific Bar Code Symbologies Computer Command

Function/Description

Bar Code Command

(Z7)

START/END Programming Menu

B1

Enable all WPC (including Addon)

B2

Enable Code 39

B3

Enable Codabar (NW-7)

B4

Enable 2 of 5 (Industrial/Interleaved)

B5

Enable Code 93 (Factory Set)

B6

Enable Code 128 (Factory Set)

-[8-C2-C3-C!-C5-C6-C7Continued on next page.....

27

Page 37

Appendix A - Programming Menus & Commands

V520-LHA7127

Add/Activate Specific Bar Code Symbologies (cont.)

28

Computer Command

Function/Description

(Z7)

START/END Programming Menu

B7

Enable MSI/Plessey

B8

Enable IATA

R1

Enable UPC (A/E)

R2

Enable UPC (A/E) +2

R3

Enable UPC (A/E) +5

R4

Enable EAN (13/8)

R5

Enable EAN (13/8) +2

R6

Enable EAN (13/8) +5

R7

Enable Industrial 2 of 5

R8

Enable Interleaved 2 of 5

Bar Code Command