Page 191
External Sensors
Chapter 6
Probe OD 3.2 mm Max. Temperature 150°C
6.2. Temperature -10°C to 10°C This is a high-precision temperature sensor designed for refrigeration monitoring Specifications: Range: -10°C to 10°C Resolution: 0.2°C Accuracy: ±0.2°C Probe Length: 50 mm Probe OD: 4 mm
191
Page 192
Chapter 6
External Sensors
Operating Temperature: -10°C to 50°C
6.3. Temperature -100°C to 120°C PT100 Adapter This is a wide range temperature adapter designed to process PT100 output Specifications: Range: -100°C to 120°C Resolution: 1°C Accuracy: ±1°C Adapter operating Temperature: -10°C to 50°C Screw terminals for PT100 connection
192
Page 193
External Sensors
Chapter 6
6.4. Voltage Sensor This is a general sensor that will measure any device or transmitter that produces a linear analog output of 0-10V. The voltage can easily be converted to the correct measured units with the help of the MicroLab software. Specifications: Range: 0-10V Resolution: 0.05V Accuracy: ±3% before calibration Input Impedance: 3MΩ Calibration: Two point calibration OV protection: +30V
193
Page 194
Chapter 6
External Sensors
6.5. Current Sensor This OaktonLog external sensor can sample any device or transmitter producing a linear current of 0-20mA. The 0-20mA can be converted to the correct measured units using the MicroLab software’s Calibration option. Specifications: Range: 0-20mA Resolution: ±0.1mA Accuracy: ±3 before calibration Calibration: Two point calibration OC protection: 55mA
194
Page 195
External Sensors
Chapter 6
6.6. pH Sensor This external sensor monitors the pH level of liquids. Specifications: Range: 1-14pH Resolution: 0.116pH Accuracy: 5% of reading Calibration: Single point, done with a small trimmer on the sensor.
6.7. Contact Adapter This OaktonLog sensor monitors Reed Relay contacts and switch status (open/closed) to identify the correlation between phenomena such as temperature change and door status.
195
Page 196
Chapter 6
External Sensors
Specifications: Range: Open/Close Connector: Screw Terminal Cable Length: 2.5m Internal Pull-Up Resistor: No need for external power source
196
Page 197
Specifications Specifications OaktonLog Models: EC600 temperature and external sensor data logger EC650 temperature, relative humidity and external sensor data logger Inputs: Two built-in sensors: Temperature: -30°C - 50°C (resolution 0.5°C, accuracy ±0.6°C)
197
Page 198
Specifications Relative Humidity: 0-100% (resolution 0.5%, accuracy ±3%) External sensor: Voltage 0 to 10V Current 4 to 20mA Temperature -50°C to 100°C pH 0 to 14pH Contact Open/Closed Outputs: Two digit 7-segment LCD IRDA interface to HP-printer and host computer RS232 serial communication at 19,200 bps
198
Page 199
Specifications Memory Capacity: 16,000 recording samples Power Supply: Internal Lithium Battery - 3.6V 1.2AH 1/2AA Battery Life - approximately two years, replaceable (Lifespan may vary with use of external sensors) Sampling Rate: User defined: from once every 10 seconds to every 2 hours Dimensions: Diameter - 72mm Thickness - 22.9mm
199
Page 200
Specifications Weight - 55gr Standards Water and dust proof IP65 standard, compliance for EC600 model CE and FCC standard compliance
OaktonLog Cradle Audible alarm. Serial communication channels: RS232 at 19.2Kbps USB at 1.5Mbps