Chapter 2
Click OK.
The new sensor will then appear in the External Sensor list in the Setup dialog box.
151
Page 152
Chapter 2
OaktonLog Plus
2.5.5. Sensor Calibration A new OaktonLog comes fully calibrated. After a period of long use, however, you may want to re-calibrate the humidity or the temperature sensors. The humidity calibration affects both OaktonLog and OaktonLog plus and should be carried out while OaktonLog is connected to the PC. The temperature calibration affects only the OaktonLog Plus software and can be carried out regardless of whether OaktonLog is connected to the PC. 1. Calibration password To prevent accidental change of the calibration, the calibration procedure is protected by a password. The default password is: 1234. To change the password:
152
Page 153
OaktonLog Plus 1. 2. 3. 4. 5. 6.
Chapter 2
Click Logger on the menu toolbar, then click Calibration. Click Change Password to open the Change Password dialog box. Enter the current password in the Current Password edit box. Enter the new password in the New Password edit box. Enter the new password a second time in the Confirm New Password edit box to confirm your new password. Click OK.
Note: The password must include at least 4 characters and is case sensitive.
153
Page 154
Chapter 2
OaktonLog Plus
2. Humidity calibration To calibrate the humidity sensor you will need a humidity chamber. 1. Connect the OaktonLog to the PC 2. Set up the OaktonLog to record every 10 seconds (see page 137). 3. Click Logger on the menu bar, then click Calibration. 4. Enter the calibration password, then click OK. 5. Select Humidity in the Choose sensor drop list
154
Page 155
6.
OaktonLog Plus
Calibration
Choose sensor : 7 Calibration Values : Reference MicroLog value value First value : 33.00 QO Second value : [76.00 fj
Default Calibrate Close
Click Default to restore the original values
Chapter 2
155
Page 156
Chapter 2 7. 8. 9. 10. 11. 12. 13. 14.
156
OaktonLog Plus
Disconnect OaktonLog from the PC and insert it into the humidity chamber. Set the humidity chamber to the first reference value. Wait until the humidity level is stabilized and write down the OaktonLog’s reading. Repeat the last step with the second reference value. Connect the OaktonLog to the PC. Enter the two OaktonLog values into the OaktonLog Value edit boxes. Enter the two reference values into the Reference Value edit boxes. Click Calibrate. To restore default calibration, click Default.
Page 157
OaktonLog Plus 3.
Chapter 2
Temperature calibration 1. Click Logger on the menu bar, then click Calibration. 2. Enter the calibration password, then click OK. 3. Select Temperature or Ext. Temperature in the Choose sensor drop list. 4. Click the Default button to restore the original values. 5. Enter the two OaktonLog values into the OaktonLog Value edit boxes. 6. Enter the two reference values into the Reference Value edit boxes. 7. Click Calibrate. 8. To restore default calibration, click Default.
157
Page 158
Chapter 2
OaktonLog Plus
2.5.6. Communication Setup Communication between a PC and an OaktonLog take place automatically whenever you send a command to OaktonLog. However, the Communication Setup dialog box can be used for more advanced communication options. Click on the Logger menu and then select Com setup to open the Communication Setup dialog box:
158
Page 159
OaktonLog Plus
Chapter 2
The checked COM ports are available for communication. Click Try to connect to establish communication. Click Work off-line to work with saved files.
159
Page 160
Chapter 2
OaktonLog Plus
2.6. Toolbar Buttons 2.6.1. Main (Upper) Toolbar Opens saved files (enabled only in Open Offline View)
160
Print…
Opens the Print Option dialog box (enabled only in Offline View)
Export
Exports the displayed data to an Excel spreadsheet. (enabled only in Offline View)