Page 141

OaktonLog Plus

Chapter 2

The sensors’ present readings will appear in the labels next to the sensor name, once the OaktonLog begins logging for the first time. Temperature unit To set the temperature unit in the OaktonLog display, select the option you want (Celsius or Fahrenheit). Interval Determines the logging interval, or the time interval between successive data recordings. The time format is hh:mm:ss. Set the time setting to select a recording time interval from between 10 seconds and 2 hours. For example, to set a time interval of one hour, five minutes

141

Page 142

Chapter 2

OaktonLog Plus and thirty seconds, click the hours (hh) and type 1 or use the arrows to select 01, then click the minutes (mm) and type 5 or use the arrows, and finally, click the seconds (ss) and type 30 or select 30 using the arrows.

Timer run Click the Timer run check box if you want OaktonLog to start recording at a predetermined time. This option is convenient if you are using several OaktonLogs at once and want them to all begin logging at the same time. Use the time and date selectors to set the start time. Cyclic run In Cyclic run mode, OaktonLog overwrites the old measurements (starting with the oldest one)

142

Page 143

OaktonLog Plus

Chapter 2 once the OaktonLog’s memory is full. Click the Cyclic run check box to operate in this mode. Use this mode if you intend to operate the system continuously. If the check box is clear, OaktonLog will operate in Normal run mode and will stop recording when the memory is full. Alarm levels Type in the desired minimum and maximum alarm levels. If OaktonLog records a reading that exceeds either of these levels, the software's alarm will sound, the OaktonLog’s LCD will start to flash and the cradle’s alarm will sound.

143

Page 144

Chapter 2

OaktonLog Plus

To stop the LCD’s flashing, press either of its two keys. To stop the cradle’s alarm, press both the OaktonLog’s keys simultaneously. The default alarm levels are the lower and upper ends of the sensors. Click Cancel Alarm to restore the default levels. Cradle time settings Workday hours Use the up and down arrow buttons to set the daily period when you want the Alarm and/or the Daily Download to be active, or type the desired period in manually.

144

Page 145

OaktonLog Plus

Chapter 2

Alarm delay Click the drop list to select the time delay between the time OaktonLog records a reading that exceeds the alarm levels and the time the alarm will sound. Alarm duration Click the drop list to select the time for which the alarm will operate before it stops automatically. Complete the setup Click Send setup to send the new settings to OaktonLog, which will complete the setup. Click Cancel if you don’t want to change the setup.

145

Page 146

Chapter 2

OaktonLog Plus

Note: The Send setup command erases all existing data in the OaktonLog, and it will automatically begin recording. If you set it to Timer run mode, OaktonLog will wait in standby mode displaying time.

, and will begin recording at the specified

2.5.2. Auto setup When setting-up multiple Cradles use the Auto Setup tool. This tool will automatically set the minimum logging interval, depending on the number of units you are using, and set the start time for each unit. To perform auto Setup:

146

Page 147

OaktonLog Plus 1. 2. 3. 4.

Chapter 2

Connect the first Cradle to the PC Click Logger on the menu bar, then click Auto Setup In the Number of units drop list select the number of Cradles you are setting up, then click Start Set up the first unit as you would normally

Note: The minimum logging interval is already set in the Interval box and you cannot select a shorter time interval 5. 6.

Click OK Click Send

After setup is completed the Auto Setup tool is ready to setup the second unit (notice that the title bar on the Auto Setup dialog has changed to ‘Unit 2’)

147

Page 148

Chapter 2 1. 2. 3. 4.

OaktonLog Plus

Connect another Cradle to the PC Change the comment (optional). Click Send Repeat this procedure with every cradle

To change the settings of a specific unit click Edit on the Auto Setup dialog and select the new settings before executing the Send command. These settings will be saved until the next time you change them. Use the forward arrow and backward arrow buttons on the Auto Setup dialog to navigate to a specific unit. Notice that this will change the automatic sequence

148

Page 149

OaktonLog Plus

Chapter 2

2.5.3. Saving setup When setting up multiple Cradles use the Save Setup option in the Setup dialog. 1. After you have finished the settings selection of the first Cradle and before sending the setup command, click Save setup. This will save all the setup settings. 2. 3. 4. 5.

on Connect another Cradle to the PC, click Setup the main toolbar, then click Load Setup to load your setup settings. Change the Cradle ID Change the comment (optional). Click Send Setup.

149

Page 150

Chapter 2

OaktonLog Plus

2.5.4. Defining a custom sensor You can use OaktonLog with any sensor that has a 0 – 20mA current output or a 0 – 10V voltage output. 1. Click the Logger menu and then select Define New Sensors to open the Define New Sensor dialog box. 2. Click Add to add a new sensor to the list. 3. In the Based On drop list, select an external sensor that matches your sensor’s output 4. Type the sensor’s name in the Sensor Name edit box. 5. Type the sensor’s unit in the Sensor Unit edit box. 6. In the Calibration Values section, enter two values of your sensor that correspond to the base sensor values. For example, see the definition for a 0 – 100mbar pressure sensor whose output is 0 – 20mA: