Page 71

Utilities

Counters You can request counter numbers or reset counters via ESP by clicking Utilities and Counters from the pull-down menus or from the settings window clicking the Utilities button and the Counters tab as in figure 4-7.

Figure 4-7 Counter Controls

The Xs in all counter commands denote a numeric value from 00000 to 65,535. After reaching the maximum numeric limit of 65,535, an error message will be displayed and the counter will automatically rollover and start counting again at 00000. To obtain the cumulative total of counts after the rollover has occurred, add 65,536 per each rollover (the scanner does not keep track of the number of rollovers) to the current count. Note: All counter values will be lost if you can cycle power to the scanner or send an <A>, <Ad>, <Z>, or <Zp> command.

Noread Counter Clicking Request or sending serial command <N> displays the total number of noreads that have occurred since power-on or the last Noread Counter Reset command.

Noread Counter Reset Clicking Reset or sending serial command <O> sets Noread Counter to 00000.

VS-310 Scanner User’s Manual

4-7

4–Utilities

Clicking Reset for any counter will reset that counter to 0. Clicking Request for any counter will return the current number for that counter.

Page 72

Chapter 4

Utilities Trigger Counter Clicking Request or sending serial command <T> displays the total number of triggers since power-on or the last Trigger Counter Reset command.

Trigger Counter Reset Clicking Reset or serial command <U> sets the trigger counter to 00000.

Match Counter (or Good Read Counter) Clicking Request or sending serial command <V> displays the total number of good reads matching the master label or, if Match Code is not enabled, the number of good reads since power-on or the last Match Counter Reset command. This counter is always enabled, but will only work as a match count when Match Code is enabled. If Match Code is not enabled, this counter records the number of good reads. This count can be requested at any time.

Match Counter Reset Clicking Reset or sending serial command <W> sets the Match Counter to 00000.

Mismatch Counter Clicking Request or sending serial command <X> displays the message the number of labels successfully read that do not match the master label since power-on or the last Mismatch Counter command.

Mismatch Counter Reset

4–Utilities

Clicking Reset or sending serial command <Y> sets the Mismatch Counter to zero.

4-8

VS-310 Scanner User’s Manual

Page 73

Utilities

Master Label Note: Match code operations are intended for use when the scanner is in a triggered mode, that is with an external or serial trigger enabled. If Match Code is enabled while the scanner is in Continuous Read, the scanner defaults to Continuous Read 1 Output and label data must change before the scanner will output data again, unless a timeout, if enabled, occurs. When a master label has been defined, scanned labels are compared with the master label and under certain conditions relay driver signals may be output and internal counters advanced. You can enable Match Code in the Operations menu after clicking the MENU button on the settings window or by sending the serial command <E>. This allows you to subsequently enter a master label and begin comparing labels against the master label.1 You can access master label functions in ESP by clicking Utilities and Master Label from the pull-down menus, or from the settings window, clicking the Utilities button and the Master Label tab as shown in figure 4-8.

4–Utilities

Figure 4-8 Master Label Commands

After enabling Match Code, a master label can be downloaded to the scanner by entering the master label data by keystroke in the SEND box or by clicking NEXT LABEL/MASTER to instruct the scanner to load the next label scanned as the master label. Note: You will still need to trigger a read cycle (see “Trigger,” on page 4-5) before the new master label will appear upon clicking REQUEST. 1. If no master label is entered following the Enable Master Label, every subsequent decoded label will be a “mismatch” and will increment the mismatch counter by one. VS-310 Scanner User’s Manual

4-9

Page 74

Chapter 4

Utilities Send Master Label Information Typing in data after Send or sending a serial command <)XXXX)> downloads master label information from the host or a terminal. Master label information can be downloaded at any time and saved in nonvolatile memory with a <Z> command. A stored master label will not affect standard operations unless Match Code is enabled. The Xs denote alphanumeric data, from 1 to 31 characters.

Request Master Label Information Clicking Request or sending serial command <)> immediately sends the master label information to the host. To prevent conflicts with outputting label data, first send the <I> command (Disable Laser Scanning (Laser Off)). Note: If the master label information has previously been stored in nonvolatile RAM (by a <Z> command), cycling the power will restore that information.

Store Next Label Scanned as Master Label Clicking Next Label/Master or sending serial command <G> causes the scanner to read the next bar code label scanned as the master label if Match Code has been enabled. All subsequently decoded labels are compared against the master label information stored in RAM. (See “Match Code” on page 2-15.)

Delete Master Label Information

4–Utilities

Clicking Delete Master Label or sending serial command <))> deletes master label information that has previously been loaded by either <)XXXX)> Download Master Label Information Command or <G> Store Next Label as Master Label command.

4-10

VS-310 Scanner User’s Manual

Page 75

Utilities

Reset You can send reset commands by clicking Utilities and Reset from the pulldown menus or, from the status window, clicking the Utilities button and clicking the Reset tab as in figure 4-9.

Figure 4-9 Reset/Default Menu Commands

Reset Clicking Reset or sending serial command <A> resets all configuration commands in RAM and resets all counters and operating parameters (figure 4-10).

<A>

RAM Saved NOVRAM Settings

Note: A reset will cause the numeric counters in use to loose their count; record all data that you wish to save prior to sending this command.

Clicking Default or sending serial command <Ad> restores the ROM default status to RAM, resets the scanner (figure 4-11), and writes to NOVRAM.1

Default ROM Settings

Figure 4-10 Reset

<Ad>

Note: If you are unable to communi- cate with the scanner, you may need to perform a hardware reset 2

4–Utilities

Default

Reset

Reset RAM

Saved NOVRAM Settings

Default ROM Settings

Figure 4-11 Default

1. See NOVRAM limitation note at the end of the Reset section. VS-310 Scanner User’s Manual

4-11

Page 76

Chapter 4

Utilities Reset/Save Clicking Reset/Save or sending serial command <Z> saves all current settings for power-on and resets the scanner (figure 4-12). (See note at end of this section.) The values of numeric counters are not saved by this command.

Reset/Restore Saved Settings Clicking Reset/Restore Saved Settings or sending serial command <An> copies saved settings to RAM and resets the scanner (figure 4-13).

<Z>

Reset RAM

Saved NOVRAM Settings

Default ROM Settings

Figure 4-12 Reset/Save

<An

Reset RAM

Saved NOVRAM Settings

Save Scanner Type

Standard ROM Settings

Figure 4-13 Reset/Restore

Note: This command is not available in ESP software. <Zp> Save Scanner Type, for Power-on. Allows you to save the current settings for Scanner Type1 to NOVRAM. Once you send this command, default values can only be recalled if you reenter them individually and then save them again with this command.

<Zp

RAM

NOVRAM

ROM

Figure 4-14 Save Scanner Type

4–Utilities

Note: The total NOVRAM saves (the combined number of Default <Ad>,

Reset/Save <Z>, and <Zp> commands executed) can be at least 10,000. In normal usage this will exceed the life of the scanner. If frequent changes to the operating parameters are required, it is recommended that the <Z> command be used only when the current configuration has been changed and the changes will be permanent.

2. A hardware reset can be done by pressing an indented switch on an IB-105 (if attached) two times within three seconds or shorting pin 7 (default pin) to pin 4 (ground) of the scanner’s host connector two times within 3 seconds. 1. Scanner Type should not be changed from its factory setting.

4-12

VS-310 Scanner User’s Manual

Page 77

Utilities

Checksum/Part Number Commands Clicking Part # or sending <#> displays software part number. Clicking Checksum or sending <!> a four-digit hex number (corresponding to a given software version) used to verify a scanner’s EPROM.

Figure 4-15 Checksum/Part Number

Other Operational Commands Note: The following serial commands are also enabled from “Code Types

Menu” in chapter 2 as menu commands. <P> Autodiscriminate All Codes. Enables the scanner to decode all available bar code types without changing scanner configuration settings. Note: For maximum scanning speed, enable only those bar code symbologies used in the application. <Q> Enable Code 39 Only. Allows only Code 39 labels to be read. <S> Enable I 2 of 5 Only. Allows only Interleaved 2 of 5 labels to be read.

VS-310 Scanner User’s Manual

4-13

4–Utilities

<R> Enable Codabar Only. Allows only Codabar labels to be read.

Page 78

Chapter 4

Utilities

Terminal Mode You can enter serial commands and read scanned data from the terminal window. Clicking Terminal in the settings window brings up an active terminal window and displays it alongside the settings window as shown in figure 4-16.

Figure 4-16 Terminal Mode

4–Utilities

Serial commands to the scanner and label and read rate data from the scanner are displayed dynamically. Pressing the Stop button ends the terminal session and returns ESP to menu configuration mode. Clicking the Go button on the toolbar re-activates the terminal window.

4-14

VS-310 Scanner User’s Manual

Page 79

A

Appendices

Contents Appendix A, Scanner Specifications .......................................................A-2 Appendix B, Mating Connector............................................................ A-4 Appendix C, Accessory Cables ........................................................... A-5 Appendix D, Connectivity with the IB-105 ........................................... A-6 Appendix E, ASCII Table.................................................................... A-12 Appendix F, Multidrop Communications............................................ A-13 Appendix G, Orientation .................................................................... A-17 Appendix H, Grounding and Shielding .............................................. A-19 Appendix I, Bar Code Symbology...................................................... A-20 Appendix J, Glossary of Terms ......................................................... A-21

Appendices

VS-310 Scanner User’s Manual

A-1

Page 80

Appendices

Appendix A, Scanner Specifications MECHANICAL:

ENVIRONMENT: Weight: Approximately 3 oz. (85 g) Housing: IP54 Operating Temperature: 32° to 104°F (0° to 40°C) Humidity: Up to 90% (non-condensing) storage/ operating humidity

2.61" 6.63 cm

LASER LIGHT: Type: Semiconductor visible laser

0.70" 1.78 cm

Front

Laser beam height

2.29" 5.82 cm 2.15" 5.46 cm 1.10" 2.79 cm

M4 diode (670 nm nominal) Safety class: Designed for CDRH Class II

SCANNING PARAMETERS: Type: 10-sided, single line or raster, spinning mirror Scan rate: 60 scans per second (low density), 100 scans per second (high density) Scan width: 4" (10.6 cm) @ 2.75" (6.99 cm) from window Pitch: ±50°; Skew: ±40° Label contrast: 25% min. absolute dark to light differential at 670 nm wavelength

COMMUNICATIONS : Interface: RS-232 or RS-422/485

0.3"

CONNECTORS: A single 9-pin mini Din connects the

Bottom 0.42" 1.07 cm

0.84" 2.13 cm

VS-310 with included Mating Connector kit 98-200021-01 (page A-4). It also connects to accessory cables (page A-5) which in turn connect the VS-310 to the IB-105 interface box (page A-6).

CODE TYPES: Code 39, Codabar, Interleaved 2 of 5, Code 93, Code 128

PROTOCOLS: Point-to-Point, Point-to-Point w/XON/

XOF, Polling Mode D, Multidrop, User Defined, and User-defined Multidrop

Back

ELECTRICAL: Power Supply requirement: +5 VDC ±4% @ 200 ma

STATUS LIGHTS: P

D

TM

2.29" 5.82 cm

1.10" 2.79 cm

Color

Condition

Yellow

On power-on

Green

Normal operations: illuminates whenever a bar code label is decoded and remains on until a new trigger occurs. Continuous Read or Continuous Read 1 Output: flashes for each good read.

Status LED

M4

Appendices

LED Power LED

CERTIFICATIONS: Designed for: CDRH, FCC, TüV, UL/cUL, CE

Top

Figure A-1 VS-310 Dimensions

A-2

VS-310 Scanner User’s Manual