Page 331

Camera Setup

Capture and Decode After calibration, click the Capture and Decode button.

Capture and Decode button

Symbol data appears here

Capture and Decode View Notice that following a capture and successful decode: • 1D symbols have a green line through them. • 2D symbols are surrounded by green boxes. The new settings are uploaded to ESP and displayed in the Symbol Information box below the Video screen.

Test (Video Capture) 1. With your calibrated symbol in the recommended position, click the Test button. Notice that the Number Of Symbols decoded is posted and Percent/Decode is dynamically updated. You can also click the Decodes per Second checkbox which shifts the output from Percent Decode to Decodes/Second. Un-clicking it returns the output to Percent Decode. 2. Click Stop to end the read rate test.

Quadrus Verifier User’s Manual

11-5

Page 332

Evaluation

Evaluation In Evaluation, you can view images currently in the Verifier, capture and decode a symbol, save it as a digital file, and perform histogram and line scan evaluations. When you click on the Evaluation tab, you will see the images that are currently stored in the Verifier. Click Receive to refresh this view. Click Capture/Decode to display the current stored image. Only one capture/ decode event will occur, regardless of read cycle settings. Click Read to trigger a read cycle. If there is enough time in the read cycle, up to 32 good reads (or 6 full-scale images) can be captured and displayed, depending on the size of the images, and depending on the number enabled in Number of Captures under Capture Mode in the Camera Setup menu tree to the left of the tabs. If you click Save, the current image will be saved to the location of your choice. Note: An image can only be saved in the format in which it was uploaded to ESP. JPEG images will be saved as .jpg, and bitmaps will be saved as .bmp. When you click the JPEG Image box, notice that the Line Scan button is grayed out. The JPEG option allows faster captures and transfers, but since the JPEG standard compresses image data, it is not suitable for the more rigorous demands of line scan evaluation. JPEG also allows you to adjust the image quality (resolution) by adjusting the sliding tab between 1 and 100, 1 being the lowest quality and 100 being the highest. When possible, use the highest quality; when image transfer speed must be increased, use a lower image quality setting. Adjustments for this setting will depend on your specific hardware and software limitations.

11-6

Quadrus Verifier User’s Manual

Page 333

Camera Setup

Histogram Usage:

Useful in determining quality and contrast of symbols.

Definition:

A histogram is a graphical presentation of the numeric count of the occurrence of each intensity (gray level) in an image. The horizontal axis is the values of gray levels and the vertical axis is the number of pixels for each gray level. Note: Since histograms are performed in the Verifier, the results will be save regardless of whether the image was captured as a BMP or JPEG. 1. From the Evaluation window, click the Histogram button. The current image is transferred into the histogram operation. This may take a moment since all relevant pixels are being evaluated intensively. 2. When the Histogram window opens, you might need to expand the window and/or adjust the scroll bars in order to bring the image into view. 3. To generate a histogram, click and drag your mouse pointer diagonally across the symbol or a portion of it. An area of interest box is a dashed blue line that can be selected and moved by placing a cursor inside the box, or resized by clicking and dragging the handles.

Histogram

Quadrus Verifier User’s Manual

11-7

Page 334

Evaluation

Threshold (Histogram) Enable Autothreshold is checked by default. To manually adjust the threshold, 1. Uncheck the Enable Autothreshold checkbox. 2. Move the threshold marker (vertical green dashed line) in the Histogram chart. Hint: This should be midway between the maximum and minimum curves. 3. Click on the Send Threshold button to adopt the new threshold position.

Histogram Evaluation In the histogram shown below, the first peak (from the left) is the “Low Peak”. Its highest point occurs at the 20 in the dark to light range. The next peak is the “High Peak”, which occurs at 141. The minimum (dark) and maximum (light) represent the entire range of pixels that are derived from the capture. Threshold is the mean setting (81 on this graph) and represented by the vertical dashed green line. In practical terms, the threshold represents the point at which all pixels on the left are regarded as dark and all pixels on the right are regarded as light. You can adjust the threshold by unchecking the Enable Autothreshold checkbox, moving the threshold, and clicking Send Threshold to save the new setting.

11-8

Quadrus Verifier User’s Manual

Page 335

Camera Setup

Line Scan 1. From the Evaluation tab, click the Line Scan button. The window shown in the figure below appears. Hint: When the Histogram window opens, you might need to expand the window and/ or adjust the scroll bars in order to bring the image into view. 2. Drag your mouse pointer across the image, horizontally. It will create a dashed horizontal line. Pixel information and a visual representation of the dark and light pixels will be displayed below the image. When you center your cursor over the dashed line, it becomes a double arrow cross. This will allow you to move this line anywhere in the window. You can also use your keyboard arrows to move this line incrementally in any direction. As with the histogram, the line scan compares light and dark pixels, but in a spacial distribution. On the Y axis of the graph, 0 is black and 255 is white; the X axis represents the horizontal axis of the symbol as described by the line scan. When you click anywhere on the graph, a vertical red line appears at that point and its position and value (in terms of black to white) are updated in the Pixel Info table to the left of the graph, in this case 237. In addition, a horizontal, dashed red line is displayed that indicates the average value.

Line Scan

Quadrus Verifier User’s Manual

11-9

Page 336

Region of Interest (ROI)

Region of Interest (ROI) You can narrow the active image area (field of view) by defining a specific area or region of interest. This is especially useful for increasing decode speed, improving threshold, and selecting the desired symbol from multiple symbols the FOV. Notes: 1. When using video output, the Region of Interest image will only be shown in triggered video mode. When in live video mode, the entire image from the image sensor will be shown. 2. If external camera is in use, the command will range check for the camera and will allow a larger row and column size then the sensor has. The overall operation of the Window of Interest will not be effected if a larger row or column size is selected. 3. The Region of Interest can be disabled by setting all values to 0 or setting the Region of Interest area to equal the image sensor area.

Caution: Region of Interest will shrink the field of view and therefore could cause symbols to be missed in dynamic applications.

11-10

Quadrus Verifier User’s Manual

Page 337

Camera Setup

Region of Interest by ESP 1. From the Camera menu, click the ROI tab to bring up Region of Interest. If you haven’t already captured an image, click the Capture and Decode button to decode the present image. If successful, the Good Read indicator on the ROI tab will be green and the symbol will appear in the pane below. Note: You can resize the image by clicking and dragging the ESP window from the lower right corner. This is useful where very small symbols are being read. 2. Click and drag your cursor over the symbol that you want to isolate for reading. Notice that the surrounding area goes black. You can use the handles on the image area that you have just drawn to resize the region of interest. You can also click on the center of the region of interest and move it about. 3. Test the new settings in Read Rate Mode. Note: to remove the region of interest, click the Reset button or just click anywhere in the ROI pane. Notice that all pixels not in the ROI are defined as black. Because the Quadrus Verifier has much less processing to do in a smaller window, read rates typically increase dramatically. One possible downside is that the chance of missing a symbol increases with the smaller window. Always verify that your ROI will be large enough to allow for any random movement of symbols in your FOV.

Region of Interest View

Quadrus Verifier User’s Manual

11-11

Page 338

Region of Interest (ROI)

Region of Interest by Serial Command The exact size and position of the ROI within the image area can be defined numerically in terms of pixels. Region of Interest Parameters shows where to locate the start position of the row and column pointers and how to measure the column depth and row width dimensions.

Top (Row Pointer) Definition:

Defines the row position of the upper-left starting point of the window.

Serial Cmd:

<K516,top,left,height,width>

Default:

0

Options:

0 to 496

Left (Column Pointer) Definition:

Defines the column position of the upper-left starting point of the window.

Serial Cmd:

<K516,top,left,height,width>

Default:

0

Options:

0 to 656

Height (Row Depth) Definition:

Defines the size, in rows, of the window. Maximum value is defined as the maximum row size of image sensor minus the Top value.

Serial Cmd:

<K516,top,left,height,width>

Default:

0

Options:

0 to 496

Width (Column Width) Definition:

Defines the size, in columns, of the window. Maximum value is defined as the maximum column size of Image sensor minus the Left value.

Serial Cmd:

<K516,top,left,height,width>

Default:

0

Options:

0 to 656

11-12

Quadrus Verifier User’s Manual

Page 339

Camera Setup

IP Database The Image Processing Database allows users to save multiple collections of camera and IP settings, and to apply them sequentially during a read cycle. (See Chapter 12, IP Database.)

Quadrus Verifier User’s Manual

11-13

Page 340

Dynamic Setup

Dynamic Setup You can visually determine where moving symbols will appear in the FOV during a read cycle by adjusting the delay time before capture. Note: If not already in Edge and Rapid modes, the reader will automatically change to those settings when you open the Dynamic window. 1. First set the Capture #. 2. Next set the Number of Captures and Number of Symbols that will be read during the read cycle. 3. Set Delay time. If Number of Captures is set to 1, Delay time will be the time between the start of a read cycle and the first capture. If set to any number other than 1, Delay time will be the time between the capture set in Capture # and the previous capture. 4. Start the moving application and trigger the read cycle. 5. Adjust the Delay setting so that the symbol appears near the center of the window. For example, if Number of Symbols is set to 3 and Capture # is set to 1, the Delay time will represent the time before the capture. You can change the number by manually sliding the bar tab back and forth or by typing in a number. Numbers in Dynamic Setup View the Delay box are rounded to the nearest value in ±32µS increments. Note: If a read cycle trigger occurs before ESP has finished processing/receiving an image from the previous read cycle, the trigger will be ignored. This is done to ensure that ESP remains in sync with the reader.

11-14

Quadrus Verifier User’s Manual