The text of a manual · part 29 of 43

QX-870 Industrial Raster Scanner User Manual, pages 281 to 290

Pages 281 to 290 of 426

The original document is the PDF (7.0 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Microscan QX-870

Page 281

Utilities

Outline of Full Auto Calibration Process Auto Calibration runs a calibration cycle for the scanner to optimize basic symbol decoding parameters. Focus, Gain, Shutter Speed, and Symbol Type are all configured during Auto Calibration. To configure the operation of Auto Calibration see Calibration Options. The calibration cycle can be entered either through ESP, serial command, or the EZ Button. 1. If laser framing is enabled, set scan line to full. 2. If laser power calibration is enabled, set laser power to factory default. 3. If motor speed calibration is enabled, set motor speed to factory default. 4. If video calibration is enabled, set gain and tracking to default. 5. If symbology detection is enabled: initialize symbology detection array with current symbology status configuration; enable all symbologies (except PDF417 and Pharmacode); symbology detection continues to operate concurrently with all remaining calibration processes. 6. Set LEDs to display 20% LED on, all others off. 7. Pre-position the raster: As-Is: no change; Straight: set raster straight out; Search or Frame: sweep the raster through full sweep angle. If a symbol is found, place raster in the middle of the symbol. If no symbol is found, set raster straight out. 8. Set LEDs to display 40% LED on, all others off. 9. Perform one pass of symbology detection (100 scans are decoded). 10. Perform motor and laser calibration (if enabled). 11. If motor speed calibration is enabled, motor speed starts at 300 sps, and is increased by 50 sps until 500 sps. After 500 sps, motor speed is increased by 100 sps until 1,000 sps. Motor speed calibration is limited to a maximum of 1000 sps, or motor speed is not changed. 12. If laser power calibration is enabled, at each motor speed increment, laser power is cycled between low, medium, and high power settings, or laser power is not changed. 13. If video calibration is enabled, and AGC is enabled (not in Fixed Gain operation), several gain and tracking pairs are used at each motor speed and laser power combination, or video settings are unchanged. 14. A “best-ever” read rate value is initialized to 0 to start the calibration process. 15. At each motor speed/laser power/video combination, a new read rate is acquired. The settings used to acquire the new read rate are recorded as the new “best” if: • The new read rate is better than 95%. Go to the next motor speed. • The new read rate is greater than the best read rate. Go to the next laser power. • The new read rate is equal to the best read rate, and the motor speed is greater. Go to the next laser power. • The new read rate is equal to the best read rate, and the laser power is less. Go to the next laser power. 16. Set LEDs to display 60% LED on, all others off. 17. Restore user-configured video (gain and tracking) settings. QX-870 Industrial Raster Scanner User Manual

7-25

Page 282

Other Serial Utility Commands 18. Perform video calibration, (if enabled): take initial read rate at current settings; calibrate video, select new gain and tracking; take final read rate with new gain and tracking; if final read rate is < 2%, video calibration fails; if final read rate is > or = initial read rate, video calibration passes and new settings take effect; if final read rate is < initial read rate, video settings revert to pre-calibration values (this is not considered a “failure”; all other settings calibrated by other calibration processes like motor speed and laser power). 19. Perform raster framing if the raster parameter of the calibration command is set to “Frame”. 20. Perform laser framing, if enabled. 21. If symbology detection is enabled, update user NOVRAM with any new symbologies detected during calibration. Notes: If any step in the calibration process fails, of if the user cancels calibration, the remainder of calibration is canceled and all settings are restored to their previous state. This does not apply to symbology status when symbology detection is enabled. Any symbols that are decoded during calibration will be enabled, even if any other calibration processes fail or are canceled.

7-26

QX-870 Industrial Raster Scanner User Manual

Page 283

Utilities

Bar Code Configuration Mode Bar Code Configuration Mode is a way of programming the scanner by using symbols. Bar Code Configuration Mode can be entered two different ways: 1. By forcing the scanner into Bar Code Configuration Mode by serial command <BCCFG>. 2. By configuring one of the four EZ Button positions (Single Beep, Two Beeps, Three Beeps, Four Beeps) to Bar Code Configuration Mode. Once Bar Code Configuration Mode has been entered, symbols can be thought of as serial data. Configure the scanner by printing symbols in Microscan’s serial command format. Commands are processed as if the data were streamed in through the serial port. The scanner will acknowledge the symbol with a beep, green flash, and echo the serial data to the host. If the command causes the scanner to produce more serial output, such as serial verification or counter requests, the data will be routed to the host. Bar Code Configuration Mode can be exited by any reset <A> or <Z> command as well as a <J> or a quick press and release of the EZ Button. The command to exit Bar Code Configuration Mode can be included as part of the symbol. For example, try encoding <K200,4><K220,1><J> into a symbol. This configures the scanner to enable Serial Trigger Mode, to program a new trigger to end the read cycle, and to exit Bar Code Configuration Mode with <J>. To end all EZ Button functions, press the EZ Button once and quickly release.

QX-870 Industrial Raster Scanner User Manual

7-27

Page 284

Embedded Menus

Embedded Menus The QX-870 contains Embedded Menus to provide easy configuration in situations where a ESP Software may not be available or applicable. The Embedded Menus consist of a Main Menu and a sub-menu for every group of commands. The following example shows how a user can operate the QX-870’s Embedded Menus. Upon connection, send a <D> command to the reader to bring up the main menu.

Main Menu For the purpose of this example, assume the user wants to navigate to the Symbologies menu. The user would select C from the list. =========================[MAIN MENU]============================ A) COMMUNICATION B) SCANNER SETUP C) SYMBOLOGIES D) SCANNER I/O SETUP E) DIAGNOSTICS SETUP F) CONFIG IP DATABASE G) OUTPUT ORDER FILTER H) MATCHCODE -- PLEASE PRESS A-H TO EDIT, ESC TO EXIT |

Command Menu The user then selects B to navigate to the Code 128 menu. =======================[SYMBOLOGIES]============================ ---- Code Type --------------------- Status ---------A) CODE 39 ENABLE B) CODE 128 DISABLE C) CODE I 2 OF 5 DISABLE D) CODABAR DISABLE E) UPC DISABLE F) CODE 93 DISABLE G) PHARMA CODE DISABLE H) PDF417 DISABLE ------------- Global Parameters ------------I) Quiet Zone Status = DISABLE J) Symbology Identifier Status = DISABLE K) Background Color = White -- PLEASE PRESS A-K TO EDIT, 1=MAIN MENU, 2=PREVIOUS MENU, ESC TO EXIT

7-28

|

QX-870 Industrial Raster Scanner User Manual

Page 285

Utilities

Parameter Menu The user then selects B to configure the Fixed Symbol Length Status parameter. ===========================[CODE 128]============================ A) Code 128 Status = DISABLE B) Fixed Symbol Length Status = DISABLE C) Symbol Length = 10 D) EAN-128 Status = DISABLE E) Output Format = Standard F) Application Record Separator Status = DISABLE G) Application Record Separator Character = , H) Application Record Brackets = DISABLE I) Application Record Padding = DISABLE -- PLEASE PRESS A-I TO EDIT, 1=MAIN MENU, 2=PREVIOUS MENU, ESC TO EXIT

|

Parameter Options Menu At this level, the current setting and system default are both shown. The user can select the desired option and go back to the previous menu, or press Enter or ESC to return to the previous menu without any change to the parameter. ==================[Fixed Symbol Length Status]=================== A) DISABLE [CURRENT] [DEFAULT] B) ENABLE -- PLEASE PRESS A-B TO CHANGE THE PARAMENTER, ESC TO CANCEL |

Another example of a Parameter Options-level menu is the Symbol Length, shown below. In this menu, the user is prompted to enter a desired value in a range from 1 to 128, and the menu shows a system default of 10. After a desired value has been entered, the user must press Enter to confirm that the value is valid and return to the previous menu, or press ESC to return to the previous menu without any change to the parameter. The user could also return to the previous menu by pressing Enter when no value has been entered. =======================[Symbol Length]=========================== A) Symbol Length = 10 Rang: 1 - 128 System Default: 10 -- PLEASE TYPE THE NEW VALUE, THEN ENTER, ESC TO CANCEL -- NEW VALUE = |

QX-870 Industrial Raster Scanner User Manual

7-29

Page 286

Embedded Menus

7-30

QX-870 Industrial Raster Scanner User Manual

Page 287

Appendices Contents Appendix A General Specifications .............................................................................................. A-2 Appendix B Electrical Specifications ............................................................................................ A-6 Appendix C Ground and Shield Considerations......................................................................... A-15 Appendix D Flying Lead Cordset Pinouts................................................................................... A-17 Appendix E Updated Pin Assignment of QX Connector T ......................................................... A-18 Appendix F Serial Commands.................................................................................................... A-20 Appendix G Protocol Commands ............................................................................................... A-85 Appendix H ASCII Table ............................................................................................................ A-94 Appendix I Configuring Ethernet TCP/IP.................................................................................... A-95 Appendix J Using EtherNet/IP.................................................................................................... A-98 Appendix K Operating EtherNet/IP Serial Gateway Data Fields .............................................. A-107 Appendix L Allen-Bradley Version 16 PLC Setup .................................................................... A-109 Appendix M Allen-Bradley Version 20 PLC Setup ................................................................... A-123 Appendix N EIPScan Setup ..................................................................................................... A-130 Appendix O Interface Standards .............................................................................................. A-136 Appendix P Glossary of Terms................................................................................................. A-137

QX-870 Industrial Raster Scanner User Manual

A-1

Page 288

General Specifications

Appendix A, General Specifications Mechanical Height: 4.59" (109 mm) Width: 3.74" (95 mm) Depth: 1.76" (45 mm) Weight: 16 oz. (453 g)

Environmental Enclosure: Die-cast aluminum Operating Temperature: 0° to 50° C (32° to 122° F) Storage Temperature: –50° to 75° C (–63° to 167° F) Humidity: up to 90% (non-condensing)

Emissions Heavy Industrial: EN 61000-6-2:2005 Radiated Emissions: EN 55022:2006 Class A 30-1000 MHz Conducted Emissions: EN 55022:2006 Class A .15-30 MHz

Note: Nominal dimensions shown. Typical tolerances apply.

QX-870 Dimensions

Communication Interface Standard Interfaces: RS-232, RS-422, RS-485, Ethernet

Symbologies Standard Offering: Code 39, Codabar, Code 93, Interleaved 2 of 5, Code 128, PDF417, MicroPDF417, Pharmacode, UPC, GS1 DataBar Applications Standards: EAN-128, AIAG

Laser Light Type: Laser diode Output Wavelength: 655 nm Beam Divergence: 0.4mrad (typ.) Pulse Duration: 40~186µs Maximum Power: 1.75mW Operating Life: 50,000 hours @ 25° C Safety Class: Visible laser: Class 2, 655 nm

Scanning Parameters Mirror Type: Rotating, 10-faceted Scan Rate: Adjustable from 300 to 1,400 scans per second; Default = 500 Scan Width Angle: Typically 60° Pitch: ±50° max. Skew: ±40° max. Symbol Contrast: 25% min. absolute dark-to-light differential at 655 nm wavelength Raster Mirror Performance: 1° - 10°: 80 sweeps per second; 11° - 20°: 60 sweeps per second; 21° - 34° (max.): 40 sweeps per second; 35° - 36° (max.): 20 sweeps per second

Protocols Point-to-Point, Point-to-Point with RTS/CTS, Point-to-Point with XON/XOFF, Point-to-Point with RTS/CTS and XON/XOFF, Multidrop, Daisy Chain, User-Defined Multidrop, Ethernet TCP/IP, EtherNet/IP

Discrete I/O Input 1 (Trigger / New Master): Optoisolated, 4.5–28V rated (13 mA at 24VDC); New Master is (–) to signal ground Outputs (1, 2, 3): Optoisolated, 1–28V rated (ICE < 100 mA at 24VDC, current limited by user)

A-2

QX-870 Industrial Raster Scanner User Manual

Page 289

Appendices

QX-1 Interface Device Mechanical Height: .75" (19.1 mm) Width: 2.50" (83.5 mm) Depth: 3.15" (80 mm) Weight: 7 oz. (200 g)

QX-1 Dimensions

QX-870 Industrial Raster Scanner User Manual

A-3

Page 290

General Specifications Read Ranges

Low Density

Low Density

Medium Density

High Density

Medium Density

High Density

Narrow Bar Width

Read Range

.0075" (.191 mm)

10” to 12" (254 mm to 305 mm)

.010" (.254 mm)

7” to 16" (178 mm to 406 mm)

.015" (.381 mm)

6” to 19" (152 mm to 483 mm)

.020” (.508 mm)

5” to 22” (127 mm to 559 mm)

.040” (1.02 mm)

4” to 30” (102 mm to 762 mm)

.0075" (.191 mm)

2.5” to 5.5" (64 mm to 140 mm)

.010" (.254 mm)

1.5” to 7" (38 mm to 178 mm)

.015" (.381 mm)

1.5” to 8.5" (38 mm to 216 mm)

.020” (.508 mm)

1.5” to 11” (38 mm to 279 mm)

.030” (.762 mm)

1” to 12” (25 mm to 305 mm)

.0033” (.084 mm)

Call Microscan

.005” (.127 mm)

4” to 5” (102 mm to 127 mm)

.0075" (.191 mm)

3.5” to 6.75" (89 mm to 171 mm)

.010" (.254 mm)

3.25” to 8" (83 mm to 203 mm)

.015" (.381 mm)

3.25” to 9" (83 mm to 229 mm)

Read Ranges based on a Grade A Code 39 label running at 500 scans per second.

A-4

QX-870 Industrial Raster Scanner User Manual