The text of a manual · part 29 of 43
QX-830 Compact Industrial Scanner User Manual, pages 281 to 290
Pages 281 to 290 of 422
The original document is the PDF (7.7 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Microscan QX-830
Page 281
Utilities
Other Serial Utility Commands Calibration without Menu Sending the <@CAL> command from ESP’s Terminal initiates auto-calibration without entering the auto-calibration menu. The calibrated parameters are determined by the settings of the Calibration Options command.
Calibration with Menu Sending the <@> command from ESP’s Terminal enters the auto-calibration menu. The initial state of the menu is determined by the settings of the Calibration Options command.
When the Auto Calibration menu is entered, or when calibration is beginning, all NOVRAM settings that will be calibrated, manipulated during the calibration process, or configured by the calibration menu are saved to an “initial” settings buffer: • AGC Mode • Gain • Tracking • Motor Speed • Laser On Position • Laser Off Position • Laser Framing Status • Laser Power • Laser On/Off Status Note: Calibration Options and symbology status are not backed up. Any changes made to Calibration Options, or any new symbologies detected, can only be restored by cycling power or sending a “Recall” command: <Arp>, <Arc>, <Zrd>, or <Zrc>. QX-830 Compact Industrial Scanner User Manual
7-23
Page 282
Other Serial Utility Commands
Calibration Progress Indicators The calibration process occurs in five phases: Quick Focus, Search Pass, Focus Pass, Medium Pass, and Fine-Tune Pass. The scanner quickly locates the configuration setup, defines a calibration range for the parameters, and then tunes the parameters for optimal performance. During the calibration process, the scanner has a number of ways to communicate its progress to the user. The following three types of feedback are the primary ways that the scanner communicates calibration progress. LEDs 20% 20%, 40% 20%, 40%, 60% 20%, 40%, 60%, 80% 20%, 40%, 60%, 80%, 100%
Calibration Progress Search Pass in progress Locate Center Focus in progress Medium Pass in progress Fine-Tune Pass in progress Calibrated Successfully (holds for green flash duration)
Beeper/Green Flash The beeper and green LED flash are used to indicate calibration status as shown in the table below: Indicator 2 quick beeps 5 quick beeps 1 long beep 1 sec. Green Flash
7-24
Status Calibration has started Calibration has failed Calibration has been completed successfully Calibration has been completed successfully
QX-830 Compact Industrial Scanner User Manual
Page 283
Utilities
Outline of Full Auto Calibration Process Auto Calibration runs a calibration cycle for the scanner to optimize basic symbol decoding parameters. Focus, Gain, Shutter Speed, and Symbol Type are all configured during Auto Calibration. To configure the operation of Auto Calibration see Calibration Options. The calibration cycle can be entered either through ESP, serial command, or the EZ Button. 1. If laser framing is enabled, set scan line to full. 2. If laser power calibration is enabled, set laser power to factory default. 3. If motor speed calibration is enabled, set motor speed to factory default. 4. If video calibration is enabled, set gain and tracking to default. 5. If symbology detection is enabled, initialize symbology detection array with current symbology status configuration. 6. Enable all symbologies (except PDF417 and Pharmacode). 7. Symbology detection continues to operate concurrently with all remaining calibration processes. 8. Set LEDs to display 20% LED on, all others off. 9. Set LEDs to display 40% LED on, all others off. 10. Perform one pass of symbology detection (100 scans are decoded). 11. Perform motor and laser calibration (if enabled). 12. If motor speed calibration is enabled, motor speed starts at 300 sps, and is increased by 50 sps until 500 sps. After 500 sps, motor speed is increased by 100 sps until 1,000 sps. Motor speed calibration is limited to a maximum of 1000 sps, or motor speed is not changed. 13. If laser power calibration is enabled, at each motor speed increment, laser power is cycled between low, medium, and high power settings, or laser power is not changed. 14. If video calibration is enabled, and AGC is enabled (not in Fixed Gain operation), several gain and tracking pairs are used at each motor speed and laser power combination, or video settings are unchanged. 15. A “best-ever” read rate value is initialized to 0 to start the calibration process. 16. At each motor speed/laser power/video combination, a new read rate is acquired. The settings used to acquire the new read rate are recorded as the new “best” if: • The new read rate is better than 95%. Go to the next motor speed. • The new read rate is greater than the best read rate. Go to the next laser power. • The new read rate is equal to the best read rate, and the motor speed is greater. Go to the next laser power. • The new read rate is equal to the best read rate, and the laser power is less. Go to the next laser power. 17. Set LEDs to display 60% LED on, all others off. 18. Restore user-configured video (gain and tracking) settings. 19. Perform video calibration, (if enabled). QX-830 Compact Industrial Scanner User Manual
7-25
Page 284
Other Serial Utility Commands
Outline of Full Auto Calibration Process (continued) 20. Take initial read rate at current settings. 21. Calibrate video, select new gain and tracking settings. 22. Take final read rate with new gain and tracking settings. (If the final read rate is < 2%, video calibration fails; if final read rate is > or = the initial read rate, video calibration passes and new settings take effect; if final read rate is < initial read rate, video settings revert to pre-calibration values. This is not a considered a “failure” - all other settings calibrated by other calibration processes (motor speed, laser power, etc) will still take effect. 23. Perform laser framing, (if enabled). 24. If symbology detection is enabled, update user NOVRAM with any new symbologies detected during calibration. Note: If any step in the calibration process fails, or the user cancels calibration, the remainder of calibration is canceled, and all user settings are restored to their pre-calibration state. This does not apply to symbology status when symbology detection is enabled. Any symbologies that are decoded during calibration will be enabled, even if any other calibration processes fail or are canceled.
7-26
QX-830 Compact Industrial Scanner User Manual
Page 285
Utilities
Bar Code Configuration Mode Bar Code Configuration Mode is a way of programming the scanner by using symbols. Bar Code Configuration Mode can be entered two different ways: 1. By forcing the scanner into Bar Code Configuration Mode by serial command <BCCFG>. 2. By configuring one of the four EZ Button positions (Single Beep, Two Beeps, Three Beeps, Four Beeps) to Bar Code Configuration Mode. Once Bar Code Configuration Mode has been entered, symbols can be thought of as serial data. Configure the scanner by printing symbols in Microscan’s serial command format. Commands are processed as if the data were streamed in through the serial port. The scanner will acknowledge the symbol with a beep, green flash, and echo the serial data to the host. If the command causes the scanner to produce more serial output, such as serial verification or counter requests, the data will be routed to the host. Bar Code Configuration Mode can be exited by any reset <A> or <Z> command as well as a <J> or a quick press and release of the EZ Button. The command to exit Bar Code Configuration Mode can be included as part of the symbol. For example, try encoding <K200,4><K220,1><J> into a symbol. This configures the scanner to enable Serial Trigger Mode, to program a new trigger to end the read cycle, and to exit Bar Code Configuration Mode with <J>. To end all EZ Button functions, press the EZ Button once and quickly release.
QX-830 Compact Industrial Scanner User Manual
7-27
Page 286
Embedded Menus
Embedded Menus The QX-830 contains Embedded Menus to provide easy configuration in situations where ESP Software may not be available or applicable. The Embedded Menus consist of a Main Menu and a sub-menu for every group of commands. The following example shows how a user can operate the QX-830’s Embedded Menus. Upon connection, send a <D> command to the reader to bring up the main menu.
Main Menu For the purpose of this example, assume the user wants to navigate to the Symbologies menu. The user would select C from the list. =========================[MAIN MENU]============================ A) COMMUNICATION B) SCANNER SETUP C) SYMBOLOGIES D) SCANNER I/O SETUP E) DIAGNOSTICS SETUP F) CONFIG IP DATABASE G) OUTPUT ORDER FILTER H) MATCHCODE -- PLEASE PRESS A-H TO EDIT, ESC TO EXIT |
Command Menu The user then selects B to navigate to the Code 128 menu. =======================[SYMBOLOGIES]============================ ---- Code Type --------------------- Status ---------A) CODE 39 ENABLE B) CODE 128 DISABLE C) CODE I 2 OF 5 DISABLE D) CODABAR DISABLE E) UPC DISABLE F) CODE 93 DISABLE G) PHARMA CODE DISABLE H) PDF417 DISABLE ------------- Global Parameters ------------I) Quiet Zone Status = DISABLE J) Symbology Identifier Status = DISABLE K) Background Color = White -- PLEASE PRESS A-K TO EDIT, 1=MAIN MENU, 2=PREVIOUS MENU, ESC TO EXIT
7-28
|
QX-830 Compact Industrial Scanner User Manual
Page 287
Utilities
Parameter Menu The user then selects B to configure the Fixed Symbol Length Status parameter. ===========================[CODE 128]============================ A) Code 128 Status = DISABLE B) Fixed Symbol Length Status = DISABLE C) Symbol Length = 10 D) EAN-128 Status = DISABLE E) Output Format = Standard F) Application Record Separator Status = DISABLE G) Application Record Separator Character = , H) Application Record Brackets = DISABLE I) Application Record Padding = DISABLE -- PLEASE PRESS A-I TO EDIT, 1=MAIN MENU, 2=PREVIOUS MENU, ESC TO EXIT
|
Parameter Options Menu At this level, the current setting and system default are both shown. The user can select the desired option and go back to the previous menu, or press Enter or ESC to return to the previous menu without any change to the parameter. ==================[Fixed Symbol Length Status]=================== A) DISABLE [CURRENT] [DEFAULT] B) ENABLE -- PLEASE PRESS A-B TO CHANGE THE PARAMENTER, ESC TO CANCEL |
Another example of a Parameter Options-level menu is the Symbol Length, shown below. In this menu, the user is prompted to enter a desired value in a range from 1 to 128, and the menu shows a system default of 10. After a desired value has been entered, the user must press Enter to confirm that the value is valid and return to the previous menu, or press ESC to return to the previous menu without any change to the parameter. The user could also return to the previous menu by pressing Enter when no value has been entered. =======================[Symbol Length]=========================== A) Symbol Length = 10 Rang: 1 - 128 System Default: 10 -- PLEASE TYPE THE NEW VALUE, THEN ENTER, ESC TO CANCEL -- NEW VALUE = |
QX-830 Compact Industrial Scanner User Manual
7-29
Page 288
Embedded Menus
7-30
QX-830 Compact Industrial Scanner User Manual
Page 289
Appendices Contents Appendix A General Specifications .............................................................................................. A-2 Appendix B Electrical Specifications ............................................................................................ A-6 Appendix C Ground and Shield Considerations......................................................................... A-14 Appendix D Flying Lead Cordset Pinouts................................................................................... A-16 Appendix E Serial Commands ................................................................................................... A-17 Appendix F Protocol Commands................................................................................................ A-80 Appendix G ASCII Table ............................................................................................................ A-89 Appendix H Configuring Ethernet TCP/IP .................................................................................. A-90 Appendix I Using EtherNet/IP..................................................................................................... A-93 Appendix J Operating EtherNet/IP Serial Gateway Data Fields............................................... A-102 Appendix K Allen-Bradley Version 16 PLC Setup .................................................................... A-104 Appendix L Allen-Bradley Version 20 PLC Setup .................................................................... A-118 Appendix M EIPScan Setup ..................................................................................................... A-125 Appendix N Interface Standards .............................................................................................. A-131 Appendix O Glossary of Terms ................................................................................................ A-132
QX-830 Compact Industrial Scanner User Manual
A-1
Page 290
General Specifications
Appendix A, General Specifications Mechanical Height: 2.59" (66 mm) Width: 3.47" (88 mm) Depth: 1.38" (35 mm) Weight: 7.5oz. (212 g)
Environmental Enclosure: Die-cast aluminum, IP65 rated Operating Temperature: 0° to 50° C (32° to 122° F) Storage Temperature: –29° to 70° C (–20° to 158° F) Humidity: Up to 90% (non-condensing)
Emissions Heavy Industrial: EN 61000-6-2:2005 Radiated Emissions: EN 55022:2006 Class A 30-1000 MHz Conducted Emissions: EN 55022:2006 Class A .15-30 MHz
Communication Interface Standard Interfaces: RS-232, RS-422, RS-485, Ethernet 1.20” 30.5 mm
Symbologies Standard Offering: Code 39, Codabar, Code 93, Interleaved 2 of 5, Code 128, PDF417, MicroPDF417, Pharmacode, UPC, GS1 DataBar Applications Standards: EAN-128, AIAG
Laser Light
.77” 19.6 mm 1.13” 28.7 mm
Type: Laser diode Output Wavelength: 655 nm Beam Divergence: 0.4mrad (typ.) Pulse Duration: 40~186µs Maximum Power: 1.75mW Operating Life: 50,000 hours @ 25° C Safety Class: Visible laser: Class 2, 655 nm Note: Nominal dimensions shown. Typical tolerances apply.
QX-830 Dimensions Scanning Parameters Mirror Type: Rotating, 10-faceted Optional Raster Mirror Image: 10 raster scan lines over a 2° arc, or 0.500” raster height at 8” (203 mm) Scan Rate: Adjustable from 300 to 1,400 scans per second; Default = 500 Scan Width Angle: Typically 60° Pitch: ±50° max. Skew: ±40° max. Symbol Contrast: 25% min. absolute dark-to-light differential at 655 nm wavelength
Protocols Point-to-Point, Point-to-Point with RTS/CTS, Point-to-Point with XON/XOFF, Point-to-Point with RTS/CTS and XON/XOFF, Multidrop, Daisy Chain, User-Defined Multidrop, Ethernet TCP/IP, EtherNet/IP
Discrete I/O Input 1 (Trigger / New Master): Optoisolated, 4.5–28V rated (13 mA at 24VDC); New Master is (–) to signal ground Outputs (1, 2, 3): Optoisolated, 1–28V rated (ICE < 100 mA at 24VDC, current limited by user)
A-2
QX-830 Compact Industrial Scanner User Manual