For the Microscan MS-890

Page 221

Utilities Default/Reset/Save

Status Requests

<A> <Ard> <An> <Z> <Zrd> <?> <?1> <K?> <K??> <K?# <Knnn?> <Knnn??> <Knnn?#> <Knnn?*> <Knnnd>

Reset (does not save for Power-On) Reset and Restore Microscan Default Parameters Reset and Read Flash Parameters Save Current Settings for Power-On Restore Microscan Default Parameters and Save for Power-On Status Byte Extended Status All Commands Status Single Command Status All Command Descriptions Single Command Description All Command Ranges Single Command Range Single Command Wildcard Single Command Default

MS-890 Industrial Automation Scanner User’s Manual

12-3

Page 222

Read Rate

Read Rate You can access Read Rate from the Utilities menu. To start or end a read rate test, right-click Read Rate and select Start or Stop. To change from percent read rate to decodes per second, first end the read rate by selecting Stop, then right-click select Decodes/Sec.

Read Rate by Serial Command Enter Single Symbol Percent Test Sending <Cp> instructs the scanner to transmit the percentage of decodes and any scanned symbol data.

Enter Multisymbol Percent Test Sending <Cm> instructs the scanner to transmit the percentage of scans decoded and symbol data (if any) for multiple symbols.

Enter Single Symbol Read Rate Test Sending <C> instructs the scanner to transmit the decodes per second and symbol data (if any). The decode rate can vary dramatically due to the angle and location of the symbol in relation to the scan line. This test is very useful in aligning and positioning the scanning device during installation.

Enter Multisymbol Read Rate Test Sending <Cs> instructs the scanner to transmit the decode rate and symbol data (if any) for multiple symbols.

Enter Percent PDF Read Rate Test Sending <CPDF> instructs the scanner to transmit the number of decodes per second on the full PDF symbol. This test includes both capture and decode and translation time. Note that the output format of the data is done as a running average of 750ms windows. This means that you can remove the symbol from the view of the scanner and it will continue to transmit some history of the symbol for a few outputs.

Enable PDF Information Sending the <a1> check box will cause PDF417 symbology data to be prefaced with information consisting of error correction level, number of rows, number of columns, number of informative code words and the number of data characters. This feature can be disabled by using the exit read rate command <J>. 12-4

MS-890 Industrial Automation Scanner User’s Manual

Page 223

Utilities

Enable Extended Read Rate Test for Single and Multisymbol Sending <Ce> after <C> or <Cs> will generate a an extended mode that applies to both single and multiple symbols. Extended Mode instructs the scanner to output the decode rate, the symbol data, and the transition data. The example below shows output with Postamble enabled and defined as Carriage Return/Line Feed. Decode rate

Symbol data

Number of bar/space transitions

Note: The <Ce> command has no immediate effect if not preceded by <C> or <Cs>.

End Read Rate Test Sending <J> ends both the Percent test and the Decodes/Second test for both single and multisymbol.

MS-890 Industrial Automation Scanner User’s Manual

12-5

Page 224

Counters

Counters Counter commands can be a numeric value from 00000 to 65535. After reaching the maximum numeric limit of 65,535, the counter will automatically rollover and start counting again at 00000 and report any errors. To obtain the cumulative total of counts after the rollover has occurred, add 65,536 per each rollover (the scanner does not keep track of the number of rollovers) to the current count. Note: All counter values will be lost if power is recycled to the scanner or the scanner receives a reset or save command.

Counters by Serial Command No Read Counter Sending <N> displays the total number of No Reads that have occurred since the last reset.

No Read Counter Reset Sending <O> sets No Read Counter to 00000.

Trigger Counter Sending <T> displays the total number of triggers since the last reset.

Trigger Counter Reset Sending <U> sets the trigger counter to 00000.

12-6

MS-890 Industrial Automation Scanner User’s Manual

Page 225

Utilities

Good Read/Match Counter Sending <V> displays the total number of good reads matching the master symbol if Matchcode is enabled. This counter is always enabled, but will only work as a match count when Matchcode is enabled. If Matchcode is not enabled, this counter records the number of good reads. This count can be requested at any time.

Good Read/Match Counter Reset Sending <W> sets the match counter to 00000.

Mismatch Counter With Matchcode enabled, sending <X> displays the number of decoded symbols since the last reset that do not match the master symbol.

Mismatch Counter Reset Sending <Y> sets the mismatch counter to zero.

MS-890 Industrial Automation Scanner User’s Manual

12-7

Page 226

Device Control

Device Control Device Control is accessible from the Utilities menu.

Device Control by Serial Command Enable Laser Scanning Sending <H> will turn the laser on continuously. Note: Enable Laser Scanning does not relate to Laser On/Off command.

Disable Laser Scanning Sending <I> will turn the laser off and end the current read cycle. This feature is useful during extended periods of time when no symbols are being scanned or the scanner is being configured. Disabling laser scanning will not affect any downloaded commands to the scanner. Note: Disable Laser Scanning does not relate to Laser On/Off, which, when enabled, also turns off the laser but only between read cycles.

Motor On Sending <KE> turns the spinning mirror and raster mirror motors on (if not already running). Note: The spinning mirror reaches full speed after a short time delay.

Motor Off Sending <KF> turns the spinning mirror motor off. This command is useful for long idle periods. Note: Laser turns off whenever motor is off.

12-8

MS-890 Industrial Automation Scanner User’s Manual

Page 227

Utilities

Output 1 Sending <L1> activates the link between Output 1(+) and Output 1(–) of the host connector (regardless of Master Symbol or Output 1 status).

Output 2 Sending <L2> activates the link between Output 2(+) and Output 2(–) of the host connector (regardless of Master Symbol or Output 2 status).

Output 3 Sending <L3> activates the link between Output 3(+) and Output 3(–) of the host connector (regardless of Master Symbol or Output 3 status).

MS-890 Industrial Automation Scanner User’s Manual

12-9

Page 228

Differences from Default

Differences from Default Clicking the Differences from Default button will cause ESP to check all stored configuration settings and compare them to default settings. All settings that are different than default will appear in the left column (shown below), and descriptions of those settings will appear in the right column. Click this button for a list of ESP configuration settings that are different than default settings.

Click Generate Barcode to bring up the Bar Code Dialog. Then create configuration symbols containing the configuration commands of your choice.

Send configuration settings to the scanner without saving by clicking Send to Reader.

Click Save As to save the report as plain text or a tab-delimited text file.

Send configuration settings to the scanner and save in ESP by clicking Send and Save.

• To create a symbol containing any of the command settings in the table, click Generate Barcode. This will bring up the Bar Code Dialog. • To save the Differences from Default report, either as plain text or as a tab-delimited text file, click Save As. • Click Send and Save to send the settings to the scanner and save them, or Send to Reader to send the settings without saving them. Important: The use the Differences from Default feature, you must connect to the scanner and Receive Reader Settings via the Send/Recv button on the toolbar.

12-10

MS-890 Industrial Automation Scanner User’s Manual

Page 229

Utilities

Master Database Important: Master Database is used only for comparing entire symbols, when Sequential and Wild Card are NOT enabled, and Start Position is equal to 0.

Master Database Overview Usage:

Used where more than one master symbol is required, as in a Multisymbol setup, for matching and other matchcode operations.

Definition:

Allows you to define up to 10 master symbols as the master symbol database, which can be entered by keyboard, scanned in, displayed, or deleted by serial or ESP commands. 1. Click the Master Database tab. 2. Set the Master Symbol Database Size. 3. Select database index you want to enter the master symbol. 4. Do one of the following to enter master symbol data: • Double-click the index row to type data directly into index. (See Enter Master Symbol Data on page 12-13.) • Click Read Symbol into Selected Index to enter the next decoded symbol. Set Master Symbol Database Size here.

Loads the next symbol decoded into the selected index.

Double-click a row to enter data into popup dialog box. Loads the scanner’s saved master symbols into ESP.

MS-890 Industrial Automation Scanner User’s Manual

Saves the displayed database to the scanner.

12-11

Page 230

Master Database

Master Symbol Database Size Definition:

Master Symbol Database Size allows you to select 1 to 10 master symbols for the master symbol database.

ESP:

First set the number of master symbols you want to include by double-clicking Database Size and entering the number.

Serial Cmd:

<K224,master symbol database size>

Default:

1

Options:

1 to 10

Set Master Symbol Database Size here.

12-12

MS-890 Industrial Automation Scanner User’s Manual