Page 151

Chapter 9 Raster/Laser Control

Laser Framing Status Definition: Usage: Serial Cmd: Default: Options:

Sets the length of time the laser is ON during each scan so that only a selected portion of the scan width is effectively scanned. This is useful to avoid highly reflective objects, for filtering out unwanted signals and avoiding the wrong symbols. <K700,laser on/off status,laser framing status,laser off time,laser on time,laser power> Disabled 0 = Disabled 1 = Enabled

To adjust laser framing, click the Raster button, click the Laser Setup tab, and click the Laser On and Laser Off arrows up or down as shown in figure 9-1.

Scan beam direction

Laser Off Time

Figure 9-2 Laser Framing Pattern

When Laser Framing is not enabled, the scan beam sweeps across the full scan width. When enabled, the laser remains OFF during the Laser Off time (see figure 9-2) beginning at the start of the scan beam sweep and turns ON for the duration of the Laser On Time. Because scan widths are not always perfectly symmetrical, the most effective way to setup laser framing is to experiment with the Laser On and Laser Off commands until you get the best results.

MS-880 Industrial Long Range Scanner User’s Manual

9-7

9–Raster/Laser Con-

Figure 9-1 ESP Laser Framing Control

Laser On Time

Page 152

Laser Setup

Laser Off Time Definition:

Serial Cmd: Default: Options:

Duration of laser off time, in increments of 1/255th of the total scan width. Important: The combined values of Laser Off Time and Laser On Time cannot exceed 255 which is the duration of one scan. <K700,laser on/off status,laser framing status,laser off time,laser on time,laser power> 0 0 to 255

When setting up Laser Framing in ESP-MP, use the Off Time and On Time arrows to constrain or expand the laser framing area.

Laser On Time Definition:

Serial Cmd: Default: Options:

Duration of laser on time, in increments of 1/255th of the total scan width. Important: The combined values of Laser Off Time and Laser On Time cannot exceed 255 which is the duration of one scan. <K700,laser on/off status,laser framing status,laser off time,laser on time,laser power> 255 0 to 255

9–Raster/Laser Con-

Laser Power Definition: Usage: Serial Cmd: Default: Options:

9-8

The user can select the Low or High laser power. Applications that read past 45 inches typically use High. <K700,laser on/off status,laser framing status,laser off time,laser on time,laser power> Low 0 = Low (laser power = 2 mW) 1 = High (laser power = 3.25 mW)

MS-880 Industrial Long Range Scanner User’s Manual

Page 153

Chapter 9 Raster/Laser Control

Raster/Laser by Embedded Menu To access the Raster/Laser Framing menu, send <@> to open the Autocalibration Setup menu (figure 6-5 on page 6-8) and type the number 2.

Figure 9-3 Embedded Calibration Results Menu

Raster Settings

Sweep Rate, Raster Status, and Raster On/Off are explained more fully in “Raster Setup” on page 9-3.

Laser Settings Laser settings are also dynamically updated at the bottom of the menu. Full instructions and menus for laser operations can be found in “Laser Setup” on page 9-6. Press the enter key (CR) to return to the main calibration menu or Esc to leave the Auto Calibration Setup menu entirely.

MS-880 Industrial Long Range Scanner User’s Manual

9-9

9–Raster/Laser Con-

Type the letter T to increment the Top Offset and an R to decrement. Notice that the Top Offset reading at the bottom of the menu changes accordingly. Type the letter B to increment the Bottom Offset and V to decrement.

Page 154

9–Raster/Laser Con-

Raster/Laser by Embedded Menu

9-10

MS-880 Industrial Long Range Scanner User’s Manual

Page 155

Scanner Setup

10 Chapter Contents Scan Speed ...............................................................................................10-4 Focus.........................................................................................................10-5 Gain ...........................................................................................................10-6 Transition Sample Counter........................................................................10-7 Maximum Element.....................................................................................10-8

This section includes AGC and factory settings such as Focus, Gain, Tracking, and Scan Speed. Note: The characters NULL <> and , can only be entered through embedded menus, not through ESP or serial commands.

MS-880 Industrial Long Range Scanner User’s Manual

10-1

10–Scanner Setup

Chapter

Page 156

10–Scanner Setup

Scanner Setup by ESP

Click on this button to bring up the Scanner Setup menu. To change a setting, double-click on the setting and use your curser to scroll through the options.

Scanner Setup by Serial Command old comds new cmds

Scan Speed Gain/Tracking

KE KD

K500, K504,

Transition Counter

KH

K505,

Focus Scanner Setup Status Request

K.

K508,

<K500,scan speed> <K504,gain,tracking> <K505,transition counter status,transition counter threshold> <K508, focus position,zero position>

KU?

K699?

<K699?>

Command Title

Format

Note: To save Factory Settings changes to NOVRAM, send <Zp>, or if in ESP, select Save to Scanner/Advanced/Send and Save, Including Factory. (See “Save Active Settings including Factory for Power-on” on page -26.)

10-2

MS-880 Industrial Long Range Scanner User’s Manual

Page 157

Chapter 10 Scanner Setup

For information on accessing the embedded menus, see “Embedded Menus” on page 23. From the Main menu, scroll down through the options and select the following screen:

MS-880 Industrial Long Range Scanner User’s Manual

10-3

10–Scanner Setup

Scanner Setup by Embedded Menu

Page 158

10–Scanner Setup

Scan Speed

Scan Speed (factory setting) Definition:

Allows the user to set the number of scans per second by controlling the spinning mirror motor speed.

Usage:

Typically, to ensure a minimum number of scans, faster scan speeds are used for fast moving labels and/or longer labels with larger bar widths. Note: Scan beams will be moving faster across labels further out in the scan range since the moving beam is being projected from a spinning mirror.

Serial Cmd:

<KE500,scan speed>

Default:

400

Options:

400 to 1000

10-4

MS-880 Industrial Long Range Scanner User’s Manual

Page 159

Chapter 10 Scanner Setup

Focus Position Definition:

Focus is the act of narrowing the focal spot to its smallest size and sharpest edges at the desired distance from the scanner by moving an internal lens in discrete movement with a stepper motor. Focus Position is the length of the internal lens.

Usage:

Focus is generally done by the Autocalibration routine (see Chapter 6, “Autocalibration”). However, in certain applications involving specific focal lengths, manual control of focus might be appropriate.

Serial Cmd:

<K508,focus position,zero position>

Default:

10 (inside focal length)

Options:

0 to 40 (nearest to furthermost focal length)

Zero Position (factory setting) Definition:

This field displays the current factory-set zero focus position.

Usage:

This field is READ ONLY, and is set by a qualified technician during factory setup.

Serial Cmd:

<K508,focus position,zero position>

MS-880 Industrial Long Range Scanner User’s Manual

10-5

10–Scanner Setup

Focus

Page 160

10–Scanner Setup

Gain Adjustment

Gain Adjustment Definition:

Sets a voltage amplitude value for the analog circuitry.

Usage:

Can be useful for fine-tuning gain or when portions of analog signals spike as in the case of specular reflection or extraneous noise. Caution: Changes to Gain are typically done by qualified technicians in Autocalibration. See Chapter 6, “Autocalibration.”

Serial Cmd:

<K504,gain,tracking>

Default:

Default is 40 (nominal). Gain is optimized at the factory before shipment; the default value may not be 40.

Options:

0 to 255

Tracking Definition:

Tracks peak signals and selects an amplitude reference point to sample the analog signals for converting to digital.

Usage:

Useful in fine-tuning tracking or when conditions of poor contrast or blurred bar edges exist. Caution: Changes to Tracking are typically done in Autocalibration. See Chapter 6, “Autocalibration.”

Serial Cmd:

<K504,gain,tracking>

Default:

Tracking is optimized at the factory before shipment; the default value may not be 40.

Options:

0 to 127

10-6

MS-880 Industrial Long Range Scanner User’s Manual