Page 121
Chapter 6 Autocalibration
Note: Autocalibration by Serial Command is useful for quick calibration; however, to do autocalibration and save your settings to the Configuration Database, or to make changes to related parameters, see “Autocalibration by ESP” on page 6-4 or “Autocalibration by Embedded Menu” on page 6-8. For a quick calibration of your label, send the <@CAL> serial command (see “Autocalibrate Scanner” on page 1-13 for details) and observe the results. For example, an attempt to autocalibrate a bar code symbol yielded the following results:
It was well within the recommended range for that size label (see “Position Scanner and Symbol” on page 1-12) yet it failed to decode. To find out if a code type other than Code 39 was being used, a <P> command was sent to enable all codes and the <@CAL> command was repeated. This time it passed:
Note: A user can abort autocalibration at any time by pressing the ESC key on the terminal. No other keystroke has any effect while in Autocalibration.
Saving for Power-up To save focus, tracking, and gain settings changed in Autocalibration for power-up, send <Zp>.
MS-880 Industrial Long Range Scanner User’s Manual
6-7
6–Autocalibration
Autocalibration by Serial Command
Page 122
6–Autocalibration
Autocalibration by Embedded Menu
Autocalibration by Embedded Menu To calibrate via the embedded Autocalibration menu, do the following: 1. Place the sample label at a fixed position in the center of the read range being tested. 2. Enter Autocalibration menu by sending an <@> command from a terminal.
Figure 6-5 Embedded Autocalibration Setup Menu 3. If the application requires a change in scan rate, do the motor speed adjustment now with the M or N commands. Note: All single letter commands must be entered in upper case. 4. Start Autocalibration by typing an upper case A from your terminal while in the calibration menu. Note: A user can abort autocalibration at any time by pressing the ESC key on the terminal. No other keystroke has any effect while in Autocalibration. 5. Wait for Autocalibration to cycle through its settings. An “Autocalibration in Progress” message will be displayed. A new text line will appear below the menu displaying the progress of the autocalibration. 6. As with the serial command, repeat the routine at other positions in your scan range and width and save each position to the Configuration Database. 7. If not successful an “Autocalibration Failed” message will be displayed. If successful an “Autocalibration Passed” message as shown in the above menu will be displayed along with other settings.
6-8
MS-880 Industrial Long Range Scanner User’s Manual
Page 123
Chapter 6 Autocalibration
Scanner Settings Motor Speed, Gain Setting, and Tracking adjustments affect the read performance. Focus controls a stepper motor that physically slides a lens back and forth to control focal length. These settings are adjusted automatically in autocalibration. The status of these parameters is displayed on the bottom line of the menu under Motor, Gain, Track, and Focus (see “Overview of Optoelectrics” on page 8-4. These settings are incremented or decremented by pressing the indicated key shown on the menu. PDF-417 Decode, AGC Status, Laser Power, and Raster Status use single commands to toggle between states.
AGC Status Typing S toggles AGC between Enable/Disable, as displayed on a the status line near the bottom of the menu. With AGC disabled this value is simply the Gain setting. However, with AGC enabled, this value represents the gain the scanner has to set to maintain the desired signal amplitude. If a label is in close, the AGC level should be a small value because the scanner does not need much gain to achieve the desired signal amplitude. However, as a label gets farther away from the scanner, this value will increase to maintain the signal amplitude.
Laser Power Typing L toggles between High and Low laser settings. See “Laser Power” on page 98 for details.
PDF417 Decode Test Typing D toggles the PDF417 Decode Test between Enable/Disable, as displayed on a the status line near the bottom of the menu. Once set to Enabled, PDF417 only is enabled and a special test mode is enabled that displays the decodes per second of a full PDF417 label. While autocalibration is not performed in this mode, you can still adjust settings manually. This special read rate will remain enabled until the Autocalibration menu is exited or PDF417 is disabled with another D command. Note: This mode averages decode and capture time over a series of readings, thus assuring for a smooth read-rate output, but at the cost of delayed reaction to label movement and user setting changes.
MS-880 Industrial Long Range Scanner User’s Manual
6-9
6–Autocalibration
Note: Focus positions Inside, Outside and Set posted near the bottom of the above menu are explained more fully in “Calibration Results Embedded Menu” on page 6-11. Note: If you are using the 9-pin configuration port and doing autocalibration by serial command or embedded menus, you will be able to view scanned bar code data in the menu, but upon exiting the autocalibration menu, you will not be able to see bar code data without sending the following command: <K146,,1>. If using the ESP, this is not necessary.
Page 124
6–Autocalibration
Autocalibration by Embedded Menu
Autocalibration Menu Options START Autocalibration Typing A initiates an Autocalibration sequence. Once entered, no other keystrokes will be accepted with the exception of the ESC key, which terminates the routine.
RECALL Initial Settings Typing B restores the original settings that were present upon entering the Autocalibration Setup Menu. These settings include Motor Speed, Gain, Tracking, and Focus.
SAVE Settings to Database Typing C saves the current Gain, Tracking, and Focus settings to the Configuration Database. The database is made up of indexes 1 through 7. Current settings will be saved to the first open index. See Chapter 8, “Optoelectrics.”
EXIT Calibration Setup menu To exit the Autocalibration Setup menu, press the ESC key. The following prompt will be displayed: “Do you want to save calibration settings for power on? (Y=yes N=no).” Typing Y will cause the current settings in the Calibration Setup menu to be saved to Flash Memory for recall when the scanner is powered up at a later date. These settings include Motor Speed, Gain, Tracking, Focus, AGC Status, Laser Power, and Raster Status. Entering N will cause the scanner to be reset without saving the settings to Flash memory.
Menu Selections Typing 1 will take you to the Calibration Results menu (page 6-11). Typing 2 will take you to the Raster/Laser Framing Setup menu (page 9-9). Typing 3 will take you to the Configuration Database menu (page 7-6). Typing 4 will take you to the Advanced Setup menu (page 6-12).
6-10
MS-880 Industrial Long Range Scanner User’s Manual
Page 125
Chapter 6 Autocalibration
To access Calibration Results menu from within the Autocalibration Setup menu, (figure 6-5 on page 6-8) type the number 1.
Figure 6-6 Embedded Calibration Results Menu The Calibration Results menu (an example shown here) shows the results of the autocalibration process for Read Rate, Gain, and Tracking both the first and second passes. Note that the first pass has only one reading for each parameter.
Focus Positions In the example above, the focus position selected from the first calibration pass (position 26) represents the optimal focus position for Read Rate, Gain, and Tracking. The second “fine tune” pass commences 3 focus positions in from this and steps out until the read rate drops off to zero.
Focal Range The algorithm selects the best focus position (SET) as the midpoint between those positions (IN and OUT) at which read rates drop off by no more than 6-2/3 percent of the maximum read rate. For example, in the above menu, the read rate for focus position 24 is 73. Since 73 is more than 6-2/3% below the maximum read rate of 100% shown elsewhere, the IN range falls under the next focus position, 25. A mirrored calculation is performed to locate the OUT focal range. Press the enter key (CR) to return to the main calibration menu or Esc to leave the Auto Calibration Setup menu entirely.
MS-880 Industrial Long Range Scanner User’s Manual
6-11
6–Autocalibration
Calibration Results Embedded Menu
Page 126
6–Autocalibration
Autocalibration by Embedded Menu
Advanced Autocalibration Setup Embedded Menu The Advanced Autocalibration Setup menu allows you to change the starting focus position and the number of focus steps during the first pass. For example, if you are preparing to do autocalibration on labels whose minimum distances from the scanner is 40 inches, there is no reason for autocalibration to cycle through all of the focal positions inside 40 inches. Figure 6-7 includes a table that associates several starting focus positions with minimum label distances from the scanner. Use this table to estimate your starting focus position. To increase Starting Focus Position, type T. To decrease it, type R. To increase the First Pass Focus Step, type G. To decrease it, type F. Increasing First Pass Focus Step will speed up the autofocus process, but can also cause autofocus to skip over a bar code symbol and not decode it. Note: All single letter commands must be entered in upper case.
Figure 6-7 Advanced Autocalibration Menu Auto Range is disabled by default. To enable, press A. See page 6-6 for an explanation of auto range.
6-12
MS-880 Industrial Long Range Scanner User’s Manual
Page 127
Calibration Database
Chapter
7
Overview of Calibration Database ...............................................................7-2 Calibration Database by ESP ......................................................................7-3 Calibration Database by Serial Command ..................................................7-4 Calibration Database by Embedded Menu..................................................7-6
This section explains the database setup for storing the optical settings derived from autocalibration. Note: The characters NULL <> and , can only be entered through embedded menus, not through ESP or serial commands.
Calibration Database by Serial Command Command Title Optoelectric Calibration Database old cmds new cmds
Format
K227,
<K227,database index,focus position, gain,tracking>
MS-880 Industrial Long Range Scanner User’s Manual
7-1
7–Calibration Data-
Chapter Contents
Page 128
Overview of Calibration Database
7–Calibration Data-
Overview of Calibration Database Definition:
The configuration database allows the user to save up to seven opticalelectrical settings. The database can be used simply to save most commonly used settings to make the scanner easily adaptable to changing applications.
Usage:
Useful in applications where a variety of optical-electrical settings are required to read and decode more than one label type, position, etc.
Serial Cmd:
<K227,database index,focus position,gain,tracking>
Indexes of optical settings can be entered either as individual optical settings or as a index of settings obtained from the autocalibration procedure. They can be entered in three ways: • by ESP in the calibration database menu • by serial commands • by Embedded menu commands
Calibration with Multilabel With Multilabel enabled for more than one label, the scanner may not locate labels at varying ranges. The reason for this is that once a label is located from the optical database, the scanner will continue to search for labels that meet that index of focus, tracking, and gain settings.
7-2
MS-880 Industrial Long Range Scanner User’s Manual
Page 129
Chapter 7 Calibration Database
Calibration Database by ESP To Load the current settings obtained from autocalibration into the database: Perform the autocalibration procedures described in “Autocalibration by ESP” on page 6-4. and the Database tab as shown in
2. Move the Database Index Indicator to the Index of your choice by clicking on the arrows in the Database Index Selector. 3. Click on the Load Current to Database button. This will put the Focus, Gain, and Tracking settings that reside in the Main tab into the database index. 4. Repeat this procedure for other indexes.
Database Index Indicator
Database Index Selector
Figure 7-1 Database Menu You can also enter in individual values in any of the seven indexes by double-clicking anywhere on the index row and making changes in the popup window, or by returning to the Main tab in Autocalibration and typing in the values of your choice. Database index settings can be applied later in operation. See Chapter 8, “Optoelectrics.”
MS-880 Industrial Long Range Scanner User’s Manual
7-3
7–Calibration Data-
1. Click on autocalibration menu figure 7-1.
Page 130
Calibration Database by Serial Command
Calibration Database by Serial Command Note: The Configuration Database is best developed from the ESP Calibration menu.
7–Calibration Data-
Database Index Definition:
Stores up to 7 user defined database records that can be entered or set from autocalibration and retrieved dynamically. See "Overview of Optoelectrics" on page 8-4 for instructions on setting up the database.
Serial Cmd:
<K227,database index,focus position,gain,tracking>
Options:
1 to 7
Focus Position Definition:
Associates a specific focus position with a database index.
Serial Cmd:
<K227,database index,focus position,gain,tracking>
Options:
0 to 40
Gain Definition:
Associates a specific gain setting with a database index.
Serial Cmd:
<K227,database index,focus position,gain,tracking>
Options:
0 to 255
Tracking Definition:
Associates a specific tracking setting with a database index.
Serial Cmd:
<K227,database index,focus position,gain,tracking>
Options:
0 to 127
7-4
MS-880 Industrial Long Range Scanner User’s Manual